Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Spectrometer requirements

The narrowest PFI bands in the present study are 3 cm-1 FWHM, using a 0.5 V/cm extraction field with the lasers attenuated to minimize effects of space charge. We measure band positions at the intensity maxima. These are reproducible to better than 1 cm-1. The bandwidth is limited by the rotational contour and also by the ionization process. A major advantage of ZEKE-PFI over more traditional photoelectron techniques is that the energy calibration is that of the tunable dye lasers, which are quite stable from day to day. In contrast, both electrostatic analyzers and time-of-flight photoelectron spectrometers require frequent calibration. [Pg.163]

Mass spectrometry is another widely used technology for hydrogen sensing in the industry. Process mass spectrometers such as gas chromatographs must be safety certified to operate in the plant environment. Mass spectrometers require special air-conditioned... [Pg.501]

Any mass spectrometer requires mass calibration before use. However, the procedures to perform it properly and the number of calibration points needed may largely differ between different types of mass analyzers. Typically, several peaks of well-known m/z values evenly distributed over the mass range of interest are necessary. These are supplied from a well-known mass calibration compound or mass reference compound. Calibration is then performed by recording a mass spectrum of the calibration compound and subsequent correlation of experimental m/z values to the mass reference list. Usually, this conversion of the mass reference list to a calibration is accomplished by the mass spectrometer s data system. Thereby, the mass spectrum is recalibrated by interpolation of the m/z scale between the assigned calibration peaks to obtain the best match. The mass calibration obtained may then be stored in a calibration file and used for future measurements without the presence of a calibration compound. This procedure is termed external mass calibration. [Pg.99]

Despite these caveats, many analyses are done on land either because the sample can be stored without changing the concentrations of the analytes it contains or because the apparatus required for the analysis cannot be operated on board a ship. For instance, some radionuclides are measured on land for both reasons. Samples analyzed on land can be spiked immediately with another (artificial) isotope of the same element to fill the adsorption sites on container walls and to serve as an internal standard. The mass spectrometers required for isotope ratio analyses are often too sensitive to vibration and motion for shipboard use. Analytes present at greater than trace levels, or which can be stabilized with some pretreatment, may also be analyzed successfully on shore. [Pg.39]

Fluorine NMR data can be collected readily on most spectrometers, requiring only minor adjustments to instrumentation used to run proton samples The fluonne-19 nucleus is easily detected (relative abundance, 100%, spin, 1/2) and generates a wealth of spectral information to assist in structure elucidation To take full advantage of all the spectral evidence available, lH, 13C, and 19F chemical shifts and coupling constants should be acquired and correlated... [Pg.1040]

Data Acquisition, Storage, and Display. Fourier-transform spectrometers require access to an on-line computer in order to permit rapid transformation of interferograms to spectra, and, in fact, the development of commercial instruments was in large... [Pg.23]

Shortly afterwards, this work was extended by the incorporation of a mass spectrometer into the system, thus enabling on-line NMR and MS data to be obtained with on-line collection of the eluent for off-line FT-IR spectroscopy [22]. The incorporation of the mass spectrometer required the addition of a small proportion of ammonium acetate, dissolved in methanol, to the deuterated chloroform used as the eluent in order to promote the ionisation of the analytes. The inclusion of methanol and ammonium acetate to the solvent obviously introduced new signals into the NMR spectra, and in addition resulted in the loss of exchangeable protons from the analytes which had been observable when chloroform alone was used as the solvent. This work demonstrated the feasibility of multiple hyphenation ( hypernation ) but the off-line nature of the FT-IR data acquisition, with the inevitable delay inherent in offline analysis, represents a slight disadvantage. In addition, volatile components may well be lost as the solvent is evaporated. This can be a problem that, together with analyte instability, is exacerbated with such interfaces when reversed-phase eluents are used since these require heating in order to ensure removal of the solvent. [Pg.52]

A true double beam spectrometer requires separate optical paths for sample and reference, and involves simultaneous comparison of the two. It is possible to construct a pseudo-double beam spectrometer, employing only one detector and optical path, taking advantage of the capabilities of a computer system attached to the detector. [Pg.137]

This pseudo-double beam spectrometer requires that the source be stable over the measurement of the three required arrays. It also consumes a large amount of computational time forbidding the very rapid acquisition of spectra. [Pg.140]

A Fourier transform is the mathematical technique used to compute the spectrum from the free induction decay, and this technique of using pulses and collecting transients is called Fourier transform spectroscopy. A Fourier transform spectrometer requires sophisticated electronics capable of generating precise pulses and accurately receiving the complicated transients. A good 13C NMR instrument usually has the capability to do H NMR spectra as well. When used with proton spectroscopy, the Fourier transform technique produces good spectra with very small amounts (less than a milligram) of sample. [Pg.603]

However, introducing a sample into a mass spectrometer requires the transfer of the sample at atmospheric pressure into a region of high vacuum without compromising the latter. In the same way, producing efficient ion-molecule collisions requires the mean free path to be reduced to around 0.1 mm, implying at least a 60 Pa pressure in a region of the... [Pg.10]

Another interesting characteristic of the TOF analyser lies in its easy mass calibration with only two reference points. As in all the mass spectrometers, the TOF mass spectrometer requires a calibration equation to relate and convert the physical property that is measured to a mass value. For the TOF spectrometer, the physical property that is measured during an analysis is the flight time of the ions. As already mentioned, the flight time of an ion is related to its mass by the following equation ... [Pg.128]

For all applications in spectroscopy it is necessary to adapt the optical conductance of the fiber to the other elements of the instrument. Laser radiation can be transported easily by one fiber, but its connection to a spectrometer principally needs a bundle of fibers, because spectrometers have an optical conductance which usually exceeds that of one fiber. Besides, irradiation of the entrance slit of a grating spectrometer requires the fibers to be arranged in a row, while an interferometer requires a circular arrangement at an image of the Jacquinot stop. The optical conductance of one fiber may be approximated by... [Pg.105]

You should only operate an FAAS system iiiuli t iliiecl supervision. The instrument should be located under a fume extraction hood. The spectrometer requires approximately 20 min to warm up before switching the gases on and using the instrument. [Pg.177]

As the measurement is carried out by a process spectrometer requiring a clean gas stream containing the furfural to be determined, the first stage of the measurement process consists of an apparatus producing such a stream. It is shown schematically in Figure 42. [Pg.86]


See other pages where Spectrometer requirements is mentioned: [Pg.577]    [Pg.338]    [Pg.1011]    [Pg.492]    [Pg.502]    [Pg.238]    [Pg.704]    [Pg.39]    [Pg.277]    [Pg.97]    [Pg.235]    [Pg.382]    [Pg.145]    [Pg.395]    [Pg.469]    [Pg.486]    [Pg.101]    [Pg.33]    [Pg.149]    [Pg.361]    [Pg.450]    [Pg.155]    [Pg.66]    [Pg.130]    [Pg.166]    [Pg.97]    [Pg.24]    [Pg.421]    [Pg.312]    [Pg.126]    [Pg.251]    [Pg.145]    [Pg.478]    [Pg.1583]    [Pg.986]   
See also in sourсe #XX -- [ Pg.245 ]




SEARCH



© 2024 chempedia.info