Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Signal intensities

The signal intensity of spectral features contains quantitative information but it is highly sensitive to changes in experimental conditions. Therefore, special precautions need to be taken to extract quantitative information from mass spectra. Discussion of quantitative data treatment is also included in Chapter 8. In this section, we will focus on the basic features of mass spectra that carry information on ion abundances. [Pg.242]

The signal intensity of a spectral feature depends on multiple factors. The first determining factor is the concentration of analyte before ionization, which is also the amount of the molecule of interest. However, signal intensity depends even more on ionization efficiency - a parameter that is very hard to predict and control. Signal intensity also relies on the transmission efficiency of ions from ion sources to detectors, and the detection efficiency of detectors towards the mass and kinetic energy of incoming ions [11,12]. Since the influence of every parameter on signal intensity is unknown, mass spectra only reveal relative abundances of ions. These relative abundances can be used to estimate absolute abundances and concentrations of analytes in the analyzed samples. [Pg.242]

Before going any further, it is necessary to define various terms relating to ion abundance. Amplitude, peak area, and the S/N ratio all link to ion abundances. However, they have [Pg.242]

In an IR spectrum, some signals will be very strong in comparison with other signals on the same spectrum  [Pg.9]

That is, some bonds absorb IR radiation very efficiently, while other bonds are less efficient at absorbing IR radiation. The efficiency of a bond at absorbing IR radiation depends on the strength of the dipole moment for that bond. For example, compare the following two highlighted bonds  [Pg.9]

Since the carbonyl group has a larger dipole moment, the carbonyl group is more efficient at absorbing IR radiation, producing a stronger signal  [Pg.10]

Carbonyl groups often produce the strongest signals in an IR spectrum, while C=C bonds often produce fairly weak signals. In fact, some alkenes do not even produce any signal at all. For example, consider the IR spectrum of 2,3-dimethyl-2-butene  [Pg.10]

This aUcene is symmetrical. That is, both vinylic positions are electronically identical, and the bond has no dipole moment at all. As such, this C=C bond is completely inefficient at absorbing IR radiation, and no signal is observed. The same is hue for symmetrical C C bonds. [Pg.10]

Symmetrical application of magnetic field gradients around the refocusing pulse in a spin echo will refocus the signal from static molecules. However, microscopic molecular motion (Brownian motion) will cause dephasing of individual magnetic moments to a degree which is dependent on the freedom of the molecular motion of the water in cells. Freedom of the molecular motion [Pg.748]

Diffusion can be measured separately in various directions by application of diffusion gradients in different directions during individual measurements. In order to obtain an approximation of the average diffusion of water in an asymmetrical cell, measurement of diffusion in three orthogonal directions (for example, along X, Y, and Z dinections) are obtained and averaged. However, this is an approximation and a true value can only be obtained by determination of the diffusion tensor. This requires a minimum of six individual measurements. This can be understood once the student is familiar with the definition of a tensor. [Pg.748]

The diffusion tensor can than be diagonalized to generate a diffusion tensor of the form  [Pg.749]

FA ranges from zero for a spherical cell to one for an infinitely long cell with zero cross sectional area. [Pg.749]

T2 ofprotons onmolecules are dependent on molecular motion. Large molecules such as membranes and proteins with slow molecular motion have a very short T2 ( 1 ms). As a result, the linewidth of these molecules is very broad compared to free water. The relationship betw een linewidth (full width at half maximum) and the T2 of molecules is given by  [Pg.750]


It should be noted that this technique is not without some disadvantages. The blackbody emission background in the near IR limits the upper temperature of the sample to about 200°C [43]. Then there is the dependence of the Raman cross-section ( equation (B 1.3.16) and equation ( B1.3.20)-equation ( B 1.3.21)) which calls for an order of magnitude greater excitation intensity when exciting in the near-IR rather than in the visible to produce the same signal intensity [39]. [Pg.1200]

Experimentally, this phenomenon is difficult to observe (Ihrs sr however again electronic resonance enliancement is seen to greatly increase the signal intensity [14 ]. [Pg.1214]

For phase encoding the phase twist is most connnonly varied by incrementing in a series of subsequent transients as tiiis results in a constant transverse relaxation attenuation of the signal at the measurement position. The signal intensity as a fiinction of G is... [Pg.1525]

ELDOR is tlie acronym for electron-electron double resonance. In an ELDOR experiment [28] one observes a rednction in the EPR signal intensity of one hyperfme transition that results from the saturation of another EPR transition within the spin system. ELDOR measurements are still relatively rare bnt the experiment is fimily established in the EPR repertoire. [Pg.1571]

Figure Bl.15.16. Two-pulse ESE signal intensity of the chemically reduced ubiqumone-10 cofactor in photosynthetic bacterial reaction centres at 115 K. MW frequency is 95.1 GHz. One dimension is the magnetic field value Bq, the other dimension is the pulse separation x. The echo decay fiinction is anisotropic with respect to the spectral position. Figure Bl.15.16. Two-pulse ESE signal intensity of the chemically reduced ubiqumone-10 cofactor in photosynthetic bacterial reaction centres at 115 K. MW frequency is 95.1 GHz. One dimension is the magnetic field value Bq, the other dimension is the pulse separation x. The echo decay fiinction is anisotropic with respect to the spectral position.
Section 13 16 NMR spectra are rarely integrated because the pulse FT technique distorts the signal intensities... [Pg.577]

The ratio of the signals intensity to the average intensity of the surrounding... [Pg.379]

Figure 3.13 Change of signal intensity I(S) with retardation 3... [Pg.56]

The majority of infrared spectra are obtained by an absorption rather than an emission process and, as a result, the change of signal intensity 1(5) with retardation 5 appears very different from that in Figure 3.13. [Pg.57]

The spin-lattice relaxation time, T/, is the time constant for spin-lattice relaxation which is specific for every nuclear spin. In FT NMR spectroscopy the spin-lattice relaxation must keep pace with the exciting pulses. If the sequence of pulses is too rapid, e.g. faster than BT/max of the slowest C atom of a moleeule in carbon-13 resonance, a decrease in signal intensity is observed for the slow C atom due to the spin-lattice relaxation getting out of step. For this reason, quaternary C atoms can be recognised in carbon-13 NMR spectra by their weak signals. [Pg.10]

Thus, in the series of Ti measurements of 2-octanol (42, Fig. 2.27) for the methyl group at the hydrophobic end of the molecule, the signal intensity passes through zero at Tq = 3.8 s. From this, using equation 10, a spin-lattice relaxation time of Ti = 5.5 s can be calculated. A complete relaxation of this methyl C atom requires about five times longer (more than 30 s) than is shown in the last experiment of the series (Fig. 2.27) Tj itself is the time constant for an exponential increase, in other words, after T/ the difference between the observed signal intensity and its final value is still 1/e of the final amplitude. [Pg.64]

NOE Nuclear Overhauser effect, change of signal intensities (integrals) dining decoupling experiments decreasing with spatial distance of nuclei... [Pg.267]


See other pages where Signal intensities is mentioned: [Pg.37]    [Pg.15]    [Pg.877]    [Pg.1199]    [Pg.1206]    [Pg.1215]    [Pg.1308]    [Pg.1309]    [Pg.1510]    [Pg.1525]    [Pg.1529]    [Pg.1531]    [Pg.1538]    [Pg.1539]    [Pg.1566]    [Pg.1570]    [Pg.1570]    [Pg.1570]    [Pg.1571]    [Pg.1641]    [Pg.1781]    [Pg.1786]    [Pg.1787]    [Pg.1788]    [Pg.2061]    [Pg.551]    [Pg.55]    [Pg.401]    [Pg.48]    [Pg.442]    [Pg.467]    [Pg.8]    [Pg.10]    [Pg.30]    [Pg.48]    [Pg.49]    [Pg.51]    [Pg.213]    [Pg.216]   
See also in sourсe #XX -- [ Pg.28 , Pg.50 , Pg.51 , Pg.60 , Pg.61 ]

See also in sourсe #XX -- [ Pg.4 ]

See also in sourсe #XX -- [ Pg.430 ]

See also in sourсe #XX -- [ Pg.28 , Pg.50 , Pg.51 , Pg.60 , Pg.61 ]

See also in sourсe #XX -- [ Pg.14 ]

See also in sourсe #XX -- [ Pg.133 ]

See also in sourсe #XX -- [ Pg.141 , Pg.142 , Pg.155 , Pg.350 ]

See also in sourсe #XX -- [ Pg.67 , Pg.165 , Pg.242 ]

See also in sourсe #XX -- [ Pg.64 ]

See also in sourсe #XX -- [ Pg.13 , Pg.536 ]

See also in sourсe #XX -- [ Pg.129 ]




SEARCH



Absolute signal intensities

Chemical signals typical intensity

Diffusion-weighted signal intensity

ESR signal intensity

Electron signal intensity

General principles signal intensity

H NMR Intensity of Signals

High intensity transient signal

Infrared signals intensity

Intensity of signals

Intensity of the Signals

Intensive signals

Intensive signals

Low-signal-intensity

Mass spectrometry signal intensity

Microarray signal intensities

Normalized signal intensity

Normalized signal intensity calculation

Protein signal intensity

Raman spectrometry signal intensity

Relative signal intensities

SH signal intensity

Showing increase in signal intensity

Signal , generally intensity

Signal Characteristics Intensity

Signal intensities dephasing

Signal intensity spectra)

Signal intensity spin-lattice relaxation time

Signal intensity versus time curves

Signal intensity, fluorescence

Signal intensity, nuclear magnetic resonance

Spin lattice signal intensity

TERS signal intensities

Transient grating signal intensity

Upconverted signal intensity

© 2024 chempedia.info