Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sampled signals

In operation, a spark source is normally first flushed with argon to remove loose particulate matter from any previous analysis. The argon flow is then reduced, and the cathode is preheated or conditioned with a short bum time (about 20 sec). The argon flow is then reduced once more, and the source is ran for sufficient time to build a signal from the sample. The spark is then stopped, and the process is repeated as many times as necessary to obtain a consistent series of analyses. The arc source operates continuously, and sample signal can be taken over long periods of time. [Pg.114]

Taking Laplaee transforms of equation (7.1), whieh is the ideal sampled signal, gives... [Pg.202]

Quantitative analysis using FAB is not straightforward, as with all ionisation techniques that use a direct insertion probe. While the goal of the exercise is to determine the bulk concentration of the analyte in the FAB matrix, FAB is instead measuring the concentration of the analyte in the surface of the matrix. The analyte surface concentration is not only a function of bulk analyte concentration, but is also affected by such factors as temperature, pressure, ionic strength, pH, FAB matrix, and sample matrix. With FAB and FTB/LSIMS the sample signal often dies away when the matrix, rather than the sample, is consumed therefore, one cannot be sure that the ion signal obtained represents the entire sample. External standard FAB quantitation methods are of questionable accuracy, and even simple internal standard methods can be trusted only where the analyte is found in a well-controlled sample matrix or is separated from its sample matrix prior to FAB analysis. Therefore, labelled internal standards and isotope dilution methods have become the norm for FAB quantitation. [Pg.369]

Fig. 2.15. Changing of the true sample signal by the process of analytical measurement... [Pg.58]

The transition probability P is proportional to the square of the microwave power level. Equation (26) shows that if the product of the microwave power level and the relaxation time are sufficiently small so that 2PT 1, the rate of energy absorption in the sample (signal amplitude) will be proportional to the population difference and to the power level. If 2PTi 5>> 1, saturation occurs and the rate of energy absorption will no longer be proportional to the microwave power level. [Pg.280]

Figure 5.2 GALDI mass spectra of (a) fresh hydroxydammarenone (1, sodium adduct [M+Na]+ at m/z 465) and (b) an equimolar mixture of hydroxydammarenone (1 H, m/z 465), uvaol (5 U, m/z 465), oleanonic acid (3 On, m/z 477), oleanolic acid (2 Ol, m/z 479), andglycyrrhetinic acid (4 G, m/z 493) [39]. In (a), a small amount of (M+16) is visible at m/z 481 in the unaged sample. Signals marked with crosses at m/z 413, 469, 483, and 507are contaminants in the spectrometer... Figure 5.2 GALDI mass spectra of (a) fresh hydroxydammarenone (1, sodium adduct [M+Na]+ at m/z 465) and (b) an equimolar mixture of hydroxydammarenone (1 H, m/z 465), uvaol (5 U, m/z 465), oleanonic acid (3 On, m/z 477), oleanolic acid (2 Ol, m/z 479), andglycyrrhetinic acid (4 G, m/z 493) [39]. In (a), a small amount of (M+16) is visible at m/z 481 in the unaged sample. Signals marked with crosses at m/z 413, 469, 483, and 507are contaminants in the spectrometer...
In all three methods a blank is obtained. To ascertain the blank, excess sodium thiosulfate is added to the potassium iodide reagent at a concentration of 4.0 x 10 14 mol/1. Samples were re-analysed and the appropriate blank subtracted from the sample signal. [Pg.79]

Figure 7.8, Different type of optical fiber probes, hi is the incident light source, R is the dye chamber and D is the detector location, (a) The dye is entrapped around a single optical fiber, (b) The same optical fiber is used to route the incident light and the nuorescem signal. (Additional optical components are used to isolate the sample s signal.) (c) One optica] fiber is used to carry the incident light and the second fiber is use to carry the sample signal. Figure 7.8, Different type of optical fiber probes, hi is the incident light source, R is the dye chamber and D is the detector location, (a) The dye is entrapped around a single optical fiber, (b) The same optical fiber is used to route the incident light and the nuorescem signal. (Additional optical components are used to isolate the sample s signal.) (c) One optica] fiber is used to carry the incident light and the second fiber is use to carry the sample signal.
ENABLE COUNTERS The counters that generate /e and fs are armed simultaneously at the beginning of each run to maintain phase coherence between the excitation and sampling signals. The enabling of the counters causes the pulsed excitation to be turned on. [Pg.286]

Figure 3.21 (a) The signal (noise plus sample signals) obtained when a material is illuminated... [Pg.102]

Sequences of impulses, such as the output of an impulse sampler, can be z-transformed. For a specified sampling period 7, the z transformation of an impulse-sampled signal/,, is defined by the equation... [Pg.626]

Keep in mind the concept that we always take z transforms of impulse-sampled signals, not continuous functions. We will also use the notation... [Pg.626]

Calibration curve with confidence bands around the curve sample signal, and predicted concentration. (Reproduced with permission from D. G. Mitchell, W. N. Mills, J. S. Garden, and M. Zdeb,... [Pg.120]

Infrared spectra of zeolitic samples can be measured in several different modes. These include transmission, diffuse reflectance, attenuated total internal reflection (ATR) and emission. Transmission and diffuse reflectance are by far the most widely used of these techniques. In the transmission mode, the sample is placed directly in the infrared beam of the instrument and the light passing through or transmitted is measured by the detector. This transmitted signal (T) is ratioed to the open beam (no sample) signal (To) to get the transmission spectrum of the sample. The transmission spectrum is converted to an absorbance spectrum ... [Pg.112]

Figure 3.12 Generic illustration of incorrect valve designs for the case of horizontal flow. In these examples only part of the cross section of the stream is sampled all of the time intermittent sampling/signal acquisition is in no way better, see Figure 3.11. The sampling valves shown there offer no counteraction for gravitational segregation and/or flow differentiation effects. All solutions shown are structurally incorrect. A specific valve opening design alone does not eliminate IDE. Figure 3.12 Generic illustration of incorrect valve designs for the case of horizontal flow. In these examples only part of the cross section of the stream is sampled all of the time intermittent sampling/signal acquisition is in no way better, see Figure 3.11. The sampling valves shown there offer no counteraction for gravitational segregation and/or flow differentiation effects. All solutions shown are structurally incorrect. A specific valve opening design alone does not eliminate IDE.
Identify and quantify Apply some signal processing procedure or algorithm to infer the in situ concentration of explosive molecules in the samples Signal processor... [Pg.10]

Equation (4.20) was proposed by Hoskuldsson [65] many years ago and has been adopted by the American Society for Testing and Materials (ASTM) [59]. It generalises the univariate expression to the multivariate context and concisely describes the error propagated from three uncertainty sources to the standard error of the predicted concentration calibration concentration errors, errors in calibration instrumental signals and errors in test sample signals. Equations (4.19) and (4.20) assume that calibrations standards are representative of the test or future samples. However, if the test or future (real) sample presents uncalibrated components or spectral artefacts, the residuals will be abnormally large. In this case, the sample should be classified as an outlier and the analyte concentration cannot be predicted by the current model. This constitutes the basis of the excellent outlier detection capabilities of first-order multivariate methodologies. [Pg.228]

Figure 10.18—Photoacoustic detection device. The spectrum is obtained in the photoacoustic units PAS = 100/s//rcf (where 7S is the sample signal intensity and /ref is the reference intensity). The reference consists of activated charcoal. The spectrum resembles an absorption spectrum. Figure 10.18—Photoacoustic detection device. The spectrum is obtained in the photoacoustic units PAS = 100/s//rcf (where 7S is the sample signal intensity and /ref is the reference intensity). The reference consists of activated charcoal. The spectrum resembles an absorption spectrum.
We prefer calibration procedures with a linear response, in which the corrected analytical signal (= signal from sample — signal from blank) is proportional to the quantity of analyte. Although we try to work in the linear range, you can obtain valid results beyond the... [Pg.70]

A calibration curve shows the response of a chemical analysis to known quantities (standard solutions) of analyte. When there is a linear response, the corrected analytical signal (= signal from sample — signal from blank) is proportional to the quantity of analyte. Blank solutions are prepared from the same reagents and solvents used to prepare standards and unknowns, but blanks have no intentionally added analyte. The blank tells us the response of the procedure to impurities or interfering species in the reagents. The blank value is subtracted from measured values of standards prior... [Pg.73]

When the different samples were exposed to oxygen, signal 1 disappeared, but signal 2 remained unchanged. In contrast, when hydrogen was allowed to react at room temperature on the different sample, signal 2 disappeared while signal 1 increased in intensity. [Pg.270]

Finally, the ions (lowing down ihc quadrupole strike the Faraday plate defector. In some cases. Ihc signal is amplified further by an electron multiplier. Thus, there is obtained a spectrum of signal intensity versus m/e value. Each molecule has a unique fragmentation pattern so that a spectrum can he used as a fingerprint lor compound idenlilication. In addition, it is possible lo quantitate the amount of a particular compound by comparing sample signal intensity with the intensity produced hy a known amount of ihe compound. [Pg.973]


See other pages where Sampled signals is mentioned: [Pg.1019]    [Pg.769]    [Pg.201]    [Pg.201]    [Pg.34]    [Pg.53]    [Pg.288]    [Pg.30]    [Pg.231]    [Pg.251]    [Pg.274]    [Pg.77]    [Pg.71]    [Pg.558]    [Pg.192]    [Pg.102]    [Pg.102]    [Pg.103]    [Pg.616]    [Pg.624]    [Pg.624]    [Pg.625]    [Pg.142]    [Pg.420]    [Pg.70]    [Pg.111]    [Pg.705]    [Pg.385]    [Pg.70]   
See also in sourсe #XX -- [ Pg.65 ]




SEARCH



Digital signals sample rate/frequency

Sample Modulation and Signal Detection

Sample signal losses

Sampling Continuous Signals

Sampling signals

Sampling signals

Signal sampling rate

Surface spectroscopy, sample preparation Raman signal

© 2024 chempedia.info