Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Resolution preamplifier

Fig. 12.4. Single-tube STM. The tube piezo scanner is adhered inside a sturdy metal cylinder, which sits on three screws on the base plate. The two front screws make the coarse approaching. The rear screw makes fine approaching by using the two front screws as the pivot axis. The rear screw is actuated by a stepping motor for automatic approaching. The preamplifier (not shown) is mounted directly on top of the metal cylinder to eliminate the microphone effect of the coaxial cable between the tip and the input of the preamplifier. The entire unit is rigid enough that a mediocre vibration isolation device can provide atomic resolution. (After Hansma et al., 1988.)... Fig. 12.4. Single-tube STM. The tube piezo scanner is adhered inside a sturdy metal cylinder, which sits on three screws on the base plate. The two front screws make the coarse approaching. The rear screw makes fine approaching by using the two front screws as the pivot axis. The rear screw is actuated by a stepping motor for automatic approaching. The preamplifier (not shown) is mounted directly on top of the metal cylinder to eliminate the microphone effect of the coaxial cable between the tip and the input of the preamplifier. The entire unit is rigid enough that a mediocre vibration isolation device can provide atomic resolution. (After Hansma et al., 1988.)...
Schematic diagrams for radio-frequency (r.f.) electronics are shown in Fig. 5.2. The circuit in Fig. 5.2(a) is a simple heterodyne circuit. The pulse length is defined by the switch SI. The speed of this switch determines the minimum pulse length, and hence the minimum lens focal length, and hence the highest frequency of the microscope thus the limit to the resolution is ultimately determined by the highest useful speed of this switch. S2 is the single-pole-double-throw (s.p.d.t.) switch, and Al the low-noise preamplifier. The mixer is a device that takes two inputs, the radio-frequency signal and the... Schematic diagrams for radio-frequency (r.f.) electronics are shown in Fig. 5.2. The circuit in Fig. 5.2(a) is a simple heterodyne circuit. The pulse length is defined by the switch SI. The speed of this switch determines the minimum pulse length, and hence the minimum lens focal length, and hence the highest frequency of the microscope thus the limit to the resolution is ultimately determined by the highest useful speed of this switch. S2 is the single-pole-double-throw (s.p.d.t.) switch, and Al the low-noise preamplifier. The mixer is a device that takes two inputs, the radio-frequency signal and the...
Inside the vacuum chamber a cooled first-stage field-effect-transistor (FET) preamplifier adjacent to the Ge(Li) detector ensures maximum resolution. [Pg.211]

Ge(Li) Detector Characteristics. Resolution measurements for the 18-cm.8 Ge(Li) detector were made with the anticoincidence shield in the inoperative mode, with a normal operating bias of 1700 volts, and with a preamplifier designed in our Laboratory (3, 4), and operated in conjunction with a Tennelec TC-200 linear amplifier. Resolution at 1.33 M.e.v. was 2.62 k.e.v., FWHM (Figure 4). The electronic pulser resolution for the amplifier system at a slightly lower energy was 1.86 k.e.v., the total capacitance of the detector was 28 pF, the noise slope was 0.035 k.e.v./pF, and the leakage current at 1700 volts was 0.5 X 10"9 amp. [Pg.216]

The temporal resolution of both methods is limited by the risetime of the IR detectors and preamplifiers, rather than the delay generators (for CS work) or transient recorders (SS) used to acquire the data, and is typically a few hundred nanoseconds. For experiments at low total pressure the time between gas-kinetic collisions is considerably longer, for example, approximately 8 /is for self-collisions of HF at lOmTorr. Nascent rotational and vibrational distributions of excited fragments following photodissociation can thus be obtained from spectra taken at several microseconds delay, subject to adequate SNR at the low pressures used. For products of chemical reactions, the risetime of the IR emission will depend upon the rate constant, and even for a reaction that proceeds at the gas-kinetic rate the intensity may not reach its maximum for tens of microseconds. Although the products may only have suffered one or two collisions, and the vibrational distribution is still the initial one, rotational distributions may be partially relaxed. [Pg.29]

In general, the spatial resolution of PSD s is a function of the signal to noise ratio after the preamplifiers. Since the noise amplitude is proportional to the bandwidth of the amplifier stages, it is absolutely necessary to select this bandwidth such that the maximum information, timing or amplitude, is extracted from the signal, and that the minimum amount of noise is passed. [Pg.68]

X-ray spectrometry is generally carried out with Si(Li) detectors. The set-up is similar to that applied to y-ray spectrometry with i-Ge or Ge(Li) detectors cooling of the detector in a cryostat, operation in combination with a preamplifier, an amplifier and a multichannel analyser. The energy resolution is very good, as already mentioned in section 7.6, and makes it possible to distinguish the characteristic X rays of neighbouring elements. Some X-ray emitters that may be used for calibration purposes are listed in Table 7.6. [Pg.115]

The electron-hole pairs, as the electric charge, are swept from the detector diode. A preamplifier collects the charge to produce an output of electrical pulse whose voltage amplitude is proportional to the X-ray photon energy. The energy resolution of the detector (R) in eV can be estimated. [Pg.183]

E is the energy of characteristic X-ray line and Fisa constant called the Fano factor, which has a value of 0.12 for Si(Li). electronic noise factor, plays an important role in the resolution. Reduction of the electronic noise will improve the resolution of the EDS detector. Thus, the Si(Li) diode and the preamplifier are mounted in a cylindrical column (the cryostat) so that they can operate at the temperature of liquid nitrogen (-196°C) in order to reduce the electrical noise and increase the signal-to-noise ratio. [Pg.184]

Figure II. Diagram of a small portion of a charge-injection device array showing row and column FLT selection and video preamplifier. Individual pixel (resolution) elements are comprised of a pair ofmetal oxide insulated capacitor plates one plate on a row and one plate on a column. (Reproduced with permission from Ref. 26. Copyright 1981, J. Figure II. Diagram of a small portion of a charge-injection device array showing row and column FLT selection and video preamplifier. Individual pixel (resolution) elements are comprised of a pair ofmetal oxide insulated capacitor plates one plate on a row and one plate on a column. (Reproduced with permission from Ref. 26. Copyright 1981, J.
The most important elements affecting the energy resolution of a radiation detection system are the three statistical factors mentioned above in relation to the width F. It is worth repeating that in energy measurements it is the energy resolution of the counting system (detector-preamplifier-amplifier) that is the important quantity and not the energy resolution of just the detector. [Pg.301]


See other pages where Resolution preamplifier is mentioned: [Pg.106]    [Pg.106]    [Pg.1425]    [Pg.1426]    [Pg.3029]    [Pg.58]    [Pg.72]    [Pg.212]    [Pg.143]    [Pg.37]    [Pg.225]    [Pg.227]    [Pg.98]    [Pg.125]    [Pg.53]    [Pg.226]    [Pg.125]    [Pg.718]    [Pg.139]    [Pg.317]    [Pg.329]    [Pg.79]    [Pg.6051]    [Pg.37]    [Pg.3458]    [Pg.517]    [Pg.212]    [Pg.213]    [Pg.61]    [Pg.266]    [Pg.266]    [Pg.272]    [Pg.317]    [Pg.32]    [Pg.1425]    [Pg.1426]    [Pg.3029]    [Pg.292]    [Pg.353]    [Pg.339]   
See also in sourсe #XX -- [ Pg.283 ]




SEARCH



Preamplifiers

© 2024 chempedia.info