Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector preamplifier amplifier

The gamma ray scintillation spectrometer (Fig. 1) consisted of two single channel analyzers coupled to a common sodium iodide well detector, preamplifier, amplifier and scalers. By setting each analyzer for the appropriate energy the two isotopes were determined. [Pg.29]

The most important elements affecting the energy resolution of a radiation detection system are the three statistical factors mentioned above in relation to the width F. It is worth repeating that in energy measurements it is the energy resolution of the counting system (detector-preamplifier-amplifier) that is the important quantity and not the energy resolution of just the detector. [Pg.301]

Reactor Gold sample Indium sample BF3 detector Preamplifier Amplifier MPC counter... [Pg.486]

Figure Bl.10.2. Schematic diagram of a counting experiment. The detector intercepts signals from the source. The output of the detector is amplified by a preamplifier and then shaped and amplified friitlier by an amplifier. The discriminator has variable lower and upper level tliresholds. If a signal from the amplifier exceeds tlie lower tlireshold while remaming below the upper tlireshold, a pulse is produced that can be registered by a preprogrammed counter. The contents of the counter can be periodically transferred to an online storage device for fiirther processing and analysis. The pulse shapes produced by each of the devices are shown schematically above tlieni. Figure Bl.10.2. Schematic diagram of a counting experiment. The detector intercepts signals from the source. The output of the detector is amplified by a preamplifier and then shaped and amplified friitlier by an amplifier. The discriminator has variable lower and upper level tliresholds. If a signal from the amplifier exceeds tlie lower tlireshold while remaming below the upper tlireshold, a pulse is produced that can be registered by a preprogrammed counter. The contents of the counter can be periodically transferred to an online storage device for fiirther processing and analysis. The pulse shapes produced by each of the devices are shown schematically above tlieni.
The pulses received from preamplifier have wide variation in energies of the particles. The gain in the amplifier is of the order of 8000 such that a 1 mV signal is amplified to approximately 8 V while still maintaining the proportionality of the energy delivered by the particle (more often y-Ray) to the detectors. The amplified pulse is then delivered to pulse height analyzer. [Pg.3088]

Dead time, or resolving time, of a counting system is defined as the minimum time that can elapse between the arrival of two successive particles at the detector and the recording of two distinct pulses. The components of dead time consist of the time it takes for the formation of the pulse in the detector itself and for the processing of the detector signal through the preamplifier-amplifi-er-discriminator-scaler (or preamplifier-amplifier-MCA). With modern electronics, the longest component of dead time is that of the detector, and for this... [Pg.73]

The detector described above is of the continuous type. Position-sensitive detectors of the discrete type have also been developed. They consist of individual semiconductor elements all placed on the same base material, with each element connected to its own preamplifier-amplifier system. Two-dimensional detectors of the continuous type as well as of the discrete type have... [Pg.461]

A basic counting system for activation analysis consists of a detector [Ge or Si(Li)], electronics (i.e., preamplifier, amplifier), and a multichannel analyzer (MCA). Modern MCAs do much more than record the data. They are minicomputers or are connected to computers that store and analyze the recorded data. Examples are the ADCAM architecture offered by EG G ORTEC and the Genie-ESP VAX-based Data Acquistion and Analysis System offered by Canberra. [Pg.532]

Time resolution is limited by the response speed of the detector and amplifier, and the speed of taking signals into the AD converter. If the pulse width of the external stimulus is short enough at a time resolution of about 100 ns, an absorbance change of smaller than 10 " is detectable [1]. In measurements using a photovoltaic MCT detector capable of producing output in both AC and direct current (DC) and a high-speed preamplifier [1,9], a time resolution of about 10 ns has been achieved. [Pg.293]

In optical domain, preamplifier is no more an utopia and is in actual use in fiber communication. However quantum physics prohibits the noiseless cloning of photons an amplifier must have a spectral density of noise greater than 1 photon/spatial mode (a "spatial mode" corresponds to a geometrical extent of A /4). Most likely, an optical heterodyne detector will be limited by the photon noise of the local oscillator and optical preamplifier will not increase the detectivity of the system. [Pg.368]

Adjustable Workbench (PAW) instrument assembly. The SH shown in Figs. 3.15 and 3.16 contains the electromechanical transducer (mounted in the center), the main and reference Co/Rh sources, multilayered radiation shields, detectors and their preamplifiers and main (linear) amplifiers, and a contact plate and sensor. The contact plate and contact sensor are used in conjunction with the IDD to apply a small preload when it places the SH holding it firmly against the target. The electronics board contains power supplies/conditioners, the dedicated CPU, different kinds of memory, firmware, and associated circuitry for instrument control and data processing. The SH of the miniaturized Mossbauer spectrometer MIMOS II has the dimensions (5 x 5.5 x 9.5) cm and weighs only ca. 400 g. Both 14.4 keV y-rays and 6.4 keV Fe X-rays are detected simultaneously by four Si-PIN diodes. The mass of the electronics board is about 90 g [36],... [Pg.55]

Radiation detector output signals are usually weak and require amplification before they can be used. In radiation detection circuits, the nature of the input pulse and discriminator determines the characteristics that the preamplifier and amplifier must have. Two stages of amplification are used in most detection circuits to increase the signal-to-noise ratio. [Pg.81]

The concentration of 222Rn in air was determined with a radon measurement detector. The detector allows realizing continuous radon monitoring. It consists of an electronic unit and a scintillation cell. The electronic unit contains power supply, amplifier, discriminator, timer, counter, and indicator. The scintillation cell contains the zinc sulfide scintillator, photomultiplier, preamplifier, high voltage power supply and chamber with a volume of 200 mL over the scintillator. This chamber is filled with the gas to be analyzed. The air is either pumped or diffuses into the scintillation cell. The scintillation count is processed by electronics, and radon concentrations for predetermined intervals are stored in the memory of the device. [Pg.221]

The detector output signal is generated by a current preamplifier for photovoltaic detectors, such as InSb, and by a simple detector bias circuit shown in Fig. 4 for photoconductive detectors, such as PbS and Hg Ge. The voltage signal derived from the bias circuit is normally preamplified and forwarded to a phase-sensitive synchronous detector usually embodied in a lock-in amplifier (Stewart, 1970 Blass, 1976b). [Pg.166]


See other pages where Detector preamplifier amplifier is mentioned: [Pg.1420]    [Pg.1421]    [Pg.187]    [Pg.1420]    [Pg.1421]    [Pg.200]    [Pg.189]    [Pg.65]    [Pg.1420]    [Pg.1421]    [Pg.187]    [Pg.1420]    [Pg.1421]    [Pg.200]    [Pg.189]    [Pg.65]    [Pg.58]    [Pg.59]    [Pg.66]    [Pg.298]    [Pg.14]    [Pg.74]    [Pg.196]    [Pg.138]    [Pg.116]    [Pg.15]    [Pg.4191]    [Pg.1573]    [Pg.136]    [Pg.132]    [Pg.441]    [Pg.1302]    [Pg.468]    [Pg.486]    [Pg.70]    [Pg.123]    [Pg.673]    [Pg.112]    [Pg.26]    [Pg.82]    [Pg.210]    [Pg.45]    [Pg.23]   


SEARCH



Amplifiers

Preamplifiers

© 2024 chempedia.info