Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Reference Electrode Test

Since there is no universal junction which is compatible with or will perform well in all types of samples, one possible solution to difficulties may require the use of another reference electrode with a different type of junction. A general indication of preferred junctions for different types of samples was shown in Table 3.3. Most often, a sleeve junction will provide good performance and should be selected if the disadvantages of the flow rate and maintenance discussed in Chapter 3 are acceptable. Chapter 3 also discusses altering the reference filling solution to make the junction more compatible with the sample. [Pg.143]


Solid-state reference electrodes for potentiometric sensors are currently under research. The main problem to be faced in developing this type of electrode lies in connecting the ionic conducting (usually aqueous) solution with an electronic conductor. Since the reference electrode has to maintain a defined potential, the electrochemical reaction with components of the electrolyte has to be avoided. Oxides, mixed oxides, and polyoxometalate salts of transition elements can be proposed for preparing solid-state reference electrodes. Tested compounds include tungsten and molybdenum oxides (Guth et al., 2009). [Pg.204]

Dross, R. and Maynard, B. 2007. In-situ reference electrode testing for cathode carbon corrosion. ECS Transactions 11 1059-1068. [Pg.68]

Two methods are used to measure pH electrometric and chemical indicator (1 7). The most common is electrometric and uses the commercial pH meter with a glass electrode. This procedure is based on the measurement of the difference between the pH of an unknown or test solution and that of a standard solution. The instmment measures the emf developed between the glass electrode and a reference electrode of constant potential. The difference in emf when the electrodes are removed from the standard solution and placed in the test solution is converted to a difference in pH. Electrodes based on metal—metal oxides, eg, antimony—antimony oxide (see Antimony AND ANTIMONY ALLOYS Antimony COMPOUNDS), have also found use as pH sensors (8), especially for industrial appHcations where superior mechanical stabiUty is needed (see Sensors). However, because of the presence of the metallic element, these electrodes suffer from interferences by oxidation—reduction systems in the test solution. [Pg.464]

Reference Electrodes and Liquid Junctions. The electrical cincuit of the pH ceU is completed through a salt bridge that usually consists of a concentrated solution of potassium chloride [7447-40-7]. The solution makes contact at one end with the test solution and at the other with a reference electrode of constant potential. The Hquid junction is formed at the area of contact between the salt bridge and the test solution. The mercury—mercurous chloride electrode, the calomel electrode, provides a highly reproducible potential in the potassium chloride bridge solution and is the most widely used reference electrode. However, mercurous chloride is converted readily into mercuric ion and mercury when in contact with concentrated potassium chloride solutions above 80°C. This disproportionation reaction causes an unstable potential with calomel electrodes. Therefore, the silver—silver chloride electrode and the thallium amalgam—thallous chloride electrode often are preferred for measurements above 80°C. However, because silver chloride is relatively soluble in concentrated solutions of potassium chloride, the solution in the electrode chamber must be saturated with silver chloride. [Pg.466]

Electrodes. Test and reference electrodes and, in some cases, an auxiliary electrode. [Pg.2430]

Point (a) only concerns simple metal electrodes and needs to be tested for each case. Point (b) is important for the measuring instrument being used. In this respect, polarization of the reference electrode leads to less error than an ohmic voltage drop at the diaphragm. Point (c) has to be tested for every system and can result in the exclusion of certain electrode systems for certain media and require special measures to be taken. [Pg.87]

The pipe/soil potential is conventionally measured at a test point with a reference electrode Bn. [Pg.135]

According to Fig. 3-24 the reference electrode is remotely positioned, outside the voltage cones of the pipeline holidays. Thus it may also be positioned above a section of the pipeline where there are no defects in the coating, which can be tested by a conventional Cg measurement. If now the electrode Bq lies on such a section (0gQ = 0g ), but electrode B, already lies in the voltage cone of a defect in the pipe coating, Eq. (3-61) applies for the position, a, ... [Pg.136]

Figure 19-1 shows the experimental setup with the position of the steel test pieces and the anodes. The anodes were oxide-coated titanium wires and polymer cable anodes (see Sections 7.2.3 and 7.2.4). The mixed-metal experimental details are given in Table 19-1. The experiments were carried out galvanostatically with reference electrodes equipped to measure the potential once a day. Thus, contamination of the concrete by the electrolytes of the reference electrodes was excluded. The potentials of the protected steel test pieces are shown in Table 19-1. The potentials of the anodes were between U(2u-cuso4 = -1-15 and -1.35 V. [Pg.429]

Reference electrodes at the test points may only be needed part of the time, depending on the mode of operation of the protective systems (e.g., for monitoring or for permanent control of potential-controlled protection current equipment). Potentiostatic control is always preferred to galvanostatic systems where operational parameters are changing. [Pg.465]

Pipeline connection with reference electrode and pipe current test point, including material and earthworks DM 4500... [Pg.494]

The variation in potential between the sheath of a lead-covered cable lying parallel to a d.c. railway system and a reference electrode placed close to the cable is shown in Fig. 10.36. The shut-down potential of the cable in this test was -570 mV [Cu/CuS04(sat.)]. There is a marked positive change in... [Pg.230]

Various types of reference electrodes have been considered in Section 20.3, and the potentials of these electrodes and their variation with the activity of the electrolyte are listed in Table 21.7, Chapter 21. It is appropriate, however to point out here that the saturated calomel electrode (S.C.E.), the silver-silver chloride electrode and the copper-copper sulphate electrode are the most widely used in corrosion testing and monitoring. [Pg.1010]

One way of overcoming the liquid junction potential problem is to replace the reference electrode by an electrode composed of a solution containing the same cation as in the solution under test, but at a known concentration, together with a rod of the same metal as that used in the indicator electrode in other words we set up a concentration cell (Section 2.29). The activity of the metal ion in the solution under test is given by... [Pg.549]

Some commercial electrodes are supplied with a double junction. In such arrangements, the electrode depicted in Fig. 15.1(h) is mounted in a wider vessel of similar shape which also carries a porous disc at the lower end. This outer vessel may be filled with the same solution (e.g. saturated potassium chloride solution) as is contained in the electrode vessel in this case the main function of the double junction is to prevent the ingress of ions from the test solution which may interfere with the electrode. Alternatively, the outer vessel may contain a different solution from that involved in the electrode (e.g. 3M potassium nitrate or 3M ammonium nitrate solution), thus preventing chloride ions from the electrode entering the test solution. This last arrangement has the disadvantage that a second liquid junction potential is introduced into the system, and on the whole it is preferable wherever possible to choose a reference electrode which will not introduce interferences. [Pg.553]

So-called combination electrodes may be purchased in which the glass electrode and the saturated calomel reference electrode are combined into a single unit, thus giving a more robust piece of equipment, and the convenience of having to insert and support a single probe in the test solution instead of the two separate components. [Pg.557]

The pressed disc (or pellet) type of crystalline membrane electrode is illustrated by silver sulphide, in which substance silver ions can migrate. The pellet is sealed into the base of a plastic container as in the case of the lanthanum fluoride electrode, and contact is made by means of a silver wire with its lower end embedded in the pellet this wire establishes equilibrium with silver ions in the pellet and thus functions as an internal reference electrode. Placed in a solution containing silver ions the electrode acquires a potential which is dictated by the activity of the silver ions in the test solution. Placed in a solution containing sulphide ions, the electrode acquires a potential which is governed by the silver ion activity in the solution, and this is itself dictated by the activity of the sulphide ions in the test solution and the solubility product of silver sulphide — i.e. it is an electrode of the second kind (Section 15.1). [Pg.560]

As an alternative to plotting a calibration curve, the method of standard addition may be used. The appropriate ion-selective electrode is first set up, together with a suitable reference electrode in a known volume (Ft) of the test solution, and then the resultant e.m.f. ( t) is measured. Applying the usual Nernst equation, we can say... [Pg.571]

Reference electrodes are usually a calomel or a silver-silver chloride electrode. It is advisable that these be of the double-junction pattern so that potassium chloride solution from the electrode does not contaminate the test solution. Thus, for example, in titrations involving glacial acetic acid as solvent, the outer vessel of the double junction calomel electrode may be filled with glacial acetic acid containing a little lithium perchlorate to improve the conductance. [Pg.589]

The H-type cell devised by Lingane and Laitinen and shown in Fig. 16.9 will be found satisfactory for many purposes a particular feature is the built-in reference electrode. Usually a saturated calomel electrode is employed, but if the presence of chloride ion is harmful a mercury(I) sulphate electrode (Hg/Hg2 S04 in potassium sulphate solution potential ca + 0.40 volts vs S.C.E.) may be used. It is usually designed to contain 10-50 mL of the sample solution in the left-hand compartment, but it can be constructed to accommodate a smaller volume down to 1 -2 mL. To avoid polarisation of the reference electrode the latter should be made of tubing at least 20 mm in diameter, but the dimensions of the solution compartment can be varied over wide limits. The compartments are separated by a cross-member filled with a 4 per cent agar-saturated potassium chloride gel, which is held in position by a medium-porosity sintered Pyrex glass disc (diameter at least 10 mm) placed as near the solution compartment as possible in order to facilitate de-aeration of the test solution. By clamping the cell so that the cross-member is vertical, the molten... [Pg.609]

Logarithmic scale for expressing acidity or alkalinity of water (7.0 to 0 indicates increasing acidity 7.0 to 14 indicates increasing alkalinity). Measured by means of a glass electrode/reference electrode pair immersed in the water sample under test. The potential difference depends upon the pH which is then displayed on a pH meter (high input impedance, millivoltmeter). [Pg.749]


See other pages where Reference Electrode Test is mentioned: [Pg.306]    [Pg.1506]    [Pg.106]    [Pg.100]    [Pg.142]    [Pg.143]    [Pg.306]    [Pg.1506]    [Pg.106]    [Pg.100]    [Pg.142]    [Pg.143]    [Pg.942]    [Pg.466]    [Pg.467]    [Pg.110]    [Pg.292]    [Pg.312]    [Pg.479]    [Pg.230]    [Pg.1008]    [Pg.1020]    [Pg.1120]    [Pg.533]    [Pg.535]    [Pg.550]    [Pg.560]    [Pg.563]    [Pg.583]    [Pg.610]    [Pg.190]    [Pg.340]   


SEARCH



Reference electrodes

Reference tests

Test electrode

© 2024 chempedia.info