Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Peak height variables

The analysis can be performed for several values of the relative peak height n, using the appropriate values of f(n) taken from Table I. Thus several estimates of Ed are obtained and either an average of them is calculated with its standard deviation, or provided a dependence of Ed on n is encountered, conclusions on the variability of Ed with coverage are drawn. As an alternative, only the half-widths of the peak are treated, as long as the experimental data are not distorted by some adjacent peak. Obviously, more information is obtained in such a case. The ratios of the half-widths taken at various values of n and compared with those given in Table I represent a criterion for the fit of the value of the desorption order. Since the estimates of Ed are free of contributions of fcd, the Tm relations can be used to estimate grossly the value of fcd, similarly as in Section V.C.2.b. [Pg.380]

Results The raw data consisted of peak height ratios of signal internal standard, see data files VALIDl.dat (primary validation m - 0 repeats at every concentration), VALID2.dat (between-day variability), and VALID3. dat (combination of a single-day calibration with several repeats at 35 and 350 [ng/mlj in preparation of placing QC-sample concentration near these values). Fig. 4.29 shows the results of the back-calculation for all three files, for both the lin/lin and the log/log evaluations. Fig. 4.30 shows the pooled data from file VALID2.dat. [Pg.256]

Successful use of modern liquid chromatography in the clinical laboratory requires an appreciation of the method s analytical characteristics. The quantitative reproducibility with respect to peak height or peak area is quite good. With a sample loop injector relative standard deviations better than 1% are to be expected. The variability of syringe injection (3-4% relative standard deviation) requires the use of an internal standard to reach the 1% level (2,27). [Pg.236]

Both absolute quantitation and relative quantitation of species in mixtures is of interest in some circumstances. Quantitation in a 5-minute analysis can be achieved by addition of an internal standard, ideally the target microorganism grown in special media to incorporate heavy isotopes92-95 and determination of the relative peak heights of pairs of proteins from the analyte and the standard. Isotope-labeled proteins or peptides, selected to match proteins or peptides characteristic of target microorganisms, can also serve as internal standards for isotope ratio measurement. The addition of unmatched proteins or peptides is less reliable for either ESI or MALDI measurements because of unpredictable suppression in the variable mixture. [Pg.269]

Even a cursory glance at a spectrum will show quite how wide a typical UV-visible spectroscopic band actually is. For example, the band for the methyl viologen dication shown in Figure 9.4 is about 100 nm wide at half peak height. There is clearly an additional variable to consider. [Pg.453]

Before analysis. It Is necessary to arrange the relevant data In a data matrix which consists of n objects (laboratories, samples, methods, etc.) arranged In rows with p columns of variables (concentrations, peak heights, etc.). The objects are designated with a subscript 1, and the variables are designated with a k. An element In the matrix, Xi , represents the value of variable k for object 1. Columns show the values of the particular variable k over all n objects, and.rows show the values of all p variables for a particular object 1. [Pg.106]

Fig. 4.6. Method for determining the concentration of residual impurities, [Impjj jjj., in a reaction mixture if the impurity is a catalyst or co-catalyst. The observed variable x can be peak-height h for a GLC method, absorbance A for spectroscopy, conductivity k for conductimetry, current i for polarography, or rate constant k for kinetics, etc. Fig. 4.6. Method for determining the concentration of residual impurities, [Impjj jjj., in a reaction mixture if the impurity is a catalyst or co-catalyst. The observed variable x can be peak-height h for a GLC method, absorbance A for spectroscopy, conductivity k for conductimetry, current i for polarography, or rate constant k for kinetics, etc.
The method CLASSY attempts to bring together the appealing ideas of SIMCA and the Kernel density estimation of ALLOC (CLassification by ALLOC and SIMCA SYnergy) It has been applied to the classification of French wines (Bourgogne and Bordeaux) by classical chemica( and physical variables and by peak height of head-space chromatography. [Pg.125]

Reading mode - peak height,single pen recorder with variable 1-10 mv input... [Pg.113]

Precision. The ability of the injector to draw the same amount of sample in replicate injections is crucial to the precision and accuracy for peak-area or peak-height comparison for external standard quantitation [10,11]. If the variability of the sample and standard being injected into the column is not controlled tightly, the basic principle of external standard quantitation is seriously compromised. No meaningful comparison between the responses of the sample and the standard can be made. The absolute accuracy of the injection volume is not critical as long as the same amount of standard and sample is injected. [Pg.177]

Fig. 6.1.9A-C Electrochemical detector (a) and florescence chromatograms (b), the latter generated following post-column oxidation. A Standards - 50 nM. 1 BH4 (retention time = 5.12 min) 2 dihydroneop-terin (retention time = 4.17 min), lox oxidized BH4 generated by electrochemical detector oxidation. N.B. This peak is variable in height/area and is not used for quantification ... Fig. 6.1.9A-C Electrochemical detector (a) and florescence chromatograms (b), the latter generated following post-column oxidation. A Standards - 50 nM. 1 BH4 (retention time = 5.12 min) 2 dihydroneop-terin (retention time = 4.17 min), lox oxidized BH4 generated by electrochemical detector oxidation. N.B. This peak is variable in height/area and is not used for quantification ...
These parameters are largely self-explanatory, and generally refer to changes in electrode behavior with time or repeated surface preparation. Since the DME is reproducible to 1 % or better, it serves as a standard to approach with solid electrodes. The variables in question generally include background current, analytical signal (e.g., voltammetric peak height), electron transfer rate, and electrode area. [Pg.298]

Most chemometricians prefer inverse methods, but most traditional analytical chemistry texts introduce the classical approach to calibration. It is important to recognise that there are substantial differences in terminology in the literature, the most common problem being the distinction between V and y variables. In many areas of analytical chemistry, concentration is denoted by V, the response (such as a spectroscopic peak height) by y However, most workers in the area of multivariate calibration have first been introduced to regression methods via spectroscopy or chromatography whereby the experimental data matrix is denoted as 6X , and the concentrations or predicted variables by y In this paper we indicate the experimentally observed responses by V such as spectroscopic absorbances of chromatographic peak areas, but do not use 6y in order to avoid confusion. [Pg.5]

The 70 p-liter size of the flow cell contributed to the broadening of the chromatographic peaks observed in the fluorescence mode and could account for some loss in selectivity. The second variable, detector design, could account for the unexpected ratio in peak heights (compared with data in Table 2) obtained in Figure 4 for diphenyl and naphthalene the glass envelop that houses the photomultipler tube may absorb some of the fluorescence of biphenyl (emission maximum 305 nm) while the fluorescence of naphthalene (emission maximum 322 nm) passes through. [Pg.123]

Figure 4.6 Configurations of the reconstructed spectral parameters in the FPT+ Panel (i) the absolute values d I of the amplitudes at the corresponding chemical shifts, Re(v ). Panel (ii) the ratios dj /lm(i ) that are proportional to the peak heights. Panel (iii) distributions of poles via the harmonic variable z in the complex z+—plane. Panel (vi) distributions of the fundamental complex frequencies in the complex v+—plane. Figure 4.6 Configurations of the reconstructed spectral parameters in the FPT+ Panel (i) the absolute values d I of the amplitudes at the corresponding chemical shifts, Re(v ). Panel (ii) the ratios dj /lm(i ) that are proportional to the peak heights. Panel (iii) distributions of poles via the harmonic variable z in the complex z+—plane. Panel (vi) distributions of the fundamental complex frequencies in the complex v+—plane.
In order to compensate for variations during sample analysis (e.g. thermal instabilities, variability in flow rate and also electronic instability in the mass analyzer), samples are usually analyzed together with an internal standard, which is always added to the sample in the same amount. All measured peak areas or peak heights can be normalized on the signal of the internal standard, which helps to eliminate fluctuations during the individual measurement. [Pg.608]

In practice, relaxation parameters are measured as a series of HSQC-type spectra as a function of a variable time delay, and peak heights (or volumes) are used to fit relaxation rates.73 The next step is the determination of the overall rotational correlation time of the molecule, typically from the T /T2 ratio obtained for the residues. In such calculations, outliers are generally excluded, that is those residues for which Rex or ie is significant (residues with T, /T2 ratio well above and below the average, respectively, Ref. 71). The second step is to determine the best fit between the relaxation parameters Ri, R2 and heteronuclear NOE and the motional parameter sets. The correspondence between these can be expressed... [Pg.52]

Data Collection Devices Modem data stations receive and store detector output and print out chromatograms complete with peak heights, peak areas, sample identification, and method variables. They are also used to program the liquid chromatograph, controlling most variables and providing for long periods of unattended operation. [Pg.840]

If silica CEC columns are subject to some variability during preparation, a given column, once installed, allows highly repeatable separation in the short term. Over a series of 4-10 repeat injections on to an ODS-silica column, test mixture retention times varied [22] by only 0.1-0.2% RSD at voltages between 10 and 30 kV. In the same tests, detected peak heights and areas varied only by approximately 1% RSD. Excellent linearity of peak areas and heights as a function of concentration was also observed. [Pg.173]


See other pages where Peak height variables is mentioned: [Pg.366]    [Pg.8]    [Pg.569]    [Pg.533]    [Pg.332]    [Pg.200]    [Pg.167]    [Pg.276]    [Pg.255]    [Pg.231]    [Pg.237]    [Pg.405]    [Pg.404]    [Pg.426]    [Pg.734]    [Pg.16]    [Pg.306]    [Pg.153]    [Pg.6]    [Pg.332]    [Pg.186]    [Pg.106]    [Pg.57]    [Pg.45]    [Pg.244]    [Pg.22]    [Pg.55]    [Pg.56]    [Pg.329]    [Pg.399]    [Pg.389]   
See also in sourсe #XX -- [ Pg.92 ]




SEARCH



Peak height

© 2024 chempedia.info