Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mobile-phase temperature and

This expresses tR as a function of the fundamental column parameters t0 and k tR can vary between t0 (for k = 0) and any larger value (for k > 0). Since to varies inversely with solvent velocity u, so does tR. For a given column, mobile phase, temperature, and sample component X, k is normally constant for sufficiently small samples. Thus, tR is defined for a given compound X by the chromatographic system, and tR can be used to identify a compound tentatively by comparison with a tR value of a known compound. [Pg.498]

Figure 10.3. Examples of chromatograms. (a) The separation of 20 essential amino acids (in derivatized form) by RPLC using 1 mm inside diameter by 150 mm long microbore column packed with 4 ft m silica support (with C18 CBP) particles and a water/ acetonitrile mobile phase gradient. (6) The fractionation of gasoline using SFC with C02 mobile phase (temperature and pressure programmed) in 0.25 mm x 50 cm packed column with 5 fim polymeric support particles. (Courtesy of Frank J. Yang.)... Figure 10.3. Examples of chromatograms. (a) The separation of 20 essential amino acids (in derivatized form) by RPLC using 1 mm inside diameter by 150 mm long microbore column packed with 4 ft m silica support (with C18 CBP) particles and a water/ acetonitrile mobile phase gradient. (6) The fractionation of gasoline using SFC with C02 mobile phase (temperature and pressure programmed) in 0.25 mm x 50 cm packed column with 5 fim polymeric support particles. (Courtesy of Frank J. Yang.)...
For a given column type, mobile phase, temperature, and detection system, both N and R are constant and / depends only on the H/v ratio. We have looked at H versus v curves before (Figures 12.2 and 12.4) we note here that H/v is simply the slope of the straight line on such a plot connecting the origin with the point (the specific H and v values) represent-... [Pg.286]

It should be noted that Eqs. (1.7) and (1.8) are valid only if the migration velocity of a sample zone is constant during the elution, which means that the plate number can be determined only from isocratic chromatograms obtained at a constant composition of the mobile phase, temperature and flow rate. Plate number values evaluated from a gradient-elution chromatogram are subject to gross errors and have no real meaning. [Pg.23]

Factors which affect selectivity, mobile phase, temperature and nature of solute can be shown by the following selective coefficient equation ... [Pg.220]

The separation factor (Figure 2.4) is dependent on a number of factors, such as the nature of the stationary phase, the mobile phase, temperature, and the compounds of interest. [Pg.11]

The three types of gradients that have been used the most in TLC are mobile phase, temperature, and stationary phase gradients. Planned mobile phase gradients must be distinguished from the natural, uncontrolled gradients that result from solvent demixing during development. [Pg.30]

When a solute elutes from the column, the thermal conductivity of the mobile phase decreases and the temperature of the wire filament, and thus its resistance, increases. A reference cell, through which only the mobile phase passes, corrects for any time-dependent variations in flow rate, pressure, or electrical power, ah of which may lead to a change in the filament s resistance. [Pg.569]

Consider the heat balance over a time (dt) during which there is a flow of mobile phase (dv) and a temperature change (d0), then... [Pg.211]

Figure 4.21 demonstrates the effect of temperature on the resolution of PEOs on a TSK-GEL G6000PWxi. and G3000PWxl in series. Increased temperature will decrease mobile phase viscosity and improve diffusion, which will improve resolution. [Pg.114]

Today, the use of CHIRBASE as a tool in aiding the chemist in the identification of appropriate CSPs has produced impressive and valuable results. Although recent developments diminish the need for domain expertise, today the user must possess a certain level of knowledge of analytical chemistry and chiral chromatography. Nevertheless, further refinements will notably reduce this required level of expertise. Part of this effort will include the design of an expert system which will provide rule sets for each CSP in a given sample search context. The expert system will also be able to query the user about the specific requisites for each sample (scale, solubility, etc.) and generate rules which will indicate a ranked list of CSPs as well their most suitable experimental conditions (mobile phase, temperature, pH, etc.). [Pg.122]

Although cSFC shows relatively poor figures of merit (speed, sensitivity, detection dynamic range and sample capacity) as well as a limited application area, its applications tend to be unique. These include solutes that can be solvated with pure SCCO2 and quantified with FID. Linear density programs typical in cSFC are ideal for homologous series found in surfactants, many prepolymers, etc. Selectivity in cSFC, which can be achieved by mobile phase density and temperature programming, relies on selective interactions with the stationary phase. Quantitative analysis in cSFC may be rendered difficult by small injected volumes the use of internal standards is recommended. [Pg.207]

Parameters that should be tested in HPLC method development are flow rate, column temperature, batch and supplier of the column, injection volume, mobile phase composition and buffer pH, and detection wavelength [2], For GC/GLC methods, one should investigate the effects of column temperature, mobile phase flow rate, and column lots or suppliers [38], For capillary electrophoresis, changes in temperature, buffer pH, ionic strength, buffer concentrations, detector wavelength, rinse times, and capillaries lots and supplier should be studied [35, 36], Typical variation such as extraction time, and stability of the analytical solution should be also evaluated [37],... [Pg.256]

The quantity N is approximately constant for different bands or peaks in a chromatogram for a given set of operating conditions (a particular column and mobile phase, with fixed mobile-phase velocity, and temperature). Hence N is a useful measure of column efficiency the relative ability of a given column to provide narrow bands (small values of tw) and improved separations. [Pg.500]

Since the separation process in CEC has a number of attributes similar to those of HPLC, the most important variables affecting the separation are the same for both of these techniques. However, in HPLC mobile phase, flow and separation are independent variables. Therefore, the most important operational variables are the analyte-sorbent interactions that can be modulated by the chemistry of the packing, composition of the mobile phase, and temperature. In contrast, the CEC column has a dual role as it serves as both (i) a flow driving device and (ii) separation unit at the same time. Although the set of variables typical of HPLC is also effective in CEC, their changes may affect in one way or another both column functions. Therefore, optimization of the separation process in CEC is more complex than in HPLC. [Pg.35]

Procedure The HPLC is carried out using (a) a Vydac C18 column, for proteins and peptides, maintained at 40 °C, (b) as the mobile phase at a flow rate of 1 ml per minute, a mixture of 48 volumes of mobile phase A and 52 volumes of mobile phase B prepared and maintained at a temperature of not less than 20 °C, and (c) a detection wavelength of 214 nm. [Pg.473]

McCalley [82,83] showed that the retention of some bases at neutral pH increased with temperature over the range ambient to 60" C, in contrast to the usual effect of decrease in retention for neutral compounds. This observation accounted for some of the marked selectivity differences that can be observed in the separation of mixtures containing different classes of compound with increasing temperature. Pronounced increases in efficiency with increasing temperature were demonstrated for basic compounds at a mobile phase pH of 7. These increases were over and above any expected due to decreased mobile phase viscosity and increased solute diffusivity, which were shown for the same compounds at pH 3. It was later demonstrated that the increase... [Pg.338]

A firm understanding of the chromatographic system is expected of the analyst who developed an analytical procedure. However, when this procedure is transferred to another analyst, problems may occur. The problems may stem from differences in the chromatographic system, column variance, temperature fluctuations, mobile phase variability, and other factors. The standard means to ensure that the procedure transfers (technology transfer) successfully is through the use of system suitability... [Pg.373]


See other pages where Mobile-phase temperature and is mentioned: [Pg.337]    [Pg.77]    [Pg.184]    [Pg.20]    [Pg.337]    [Pg.77]    [Pg.184]    [Pg.20]    [Pg.62]    [Pg.284]    [Pg.342]    [Pg.1148]    [Pg.289]    [Pg.290]    [Pg.558]    [Pg.699]    [Pg.775]    [Pg.206]    [Pg.210]    [Pg.241]    [Pg.392]    [Pg.90]    [Pg.131]    [Pg.257]    [Pg.62]    [Pg.539]    [Pg.568]    [Pg.366]    [Pg.372]    [Pg.435]    [Pg.152]    [Pg.321]    [Pg.480]    [Pg.267]    [Pg.150]   
See also in sourсe #XX -- [ Pg.61 ]




SEARCH



Mobile phases temperature

Mobility and

Temperature mobilities

© 2024 chempedia.info