Liquid junction measurement

In fact, some care is needed with regard to this type of concentration cell, since the assumption implicit in the derivation of A2.4.126 that the potential in the solution is constant between the two electrodes, caimot be entirely correct. At the phase boundary between the two solutions, which is here a semi-pemieable membrane pemiitting the passage of water molecules but not ions between the two solutions, there will be a potential jump. This so-called liquid-junction potential will increase or decrease the measured EMF of the cell depending on its sign. Potential jumps at liquid-liquid junctions are in general rather small compared to nomial cell voltages, and can be minimized fiirther by suitable experimental modifications to the cell. [Pg.602]

As a result of a variable liquid-junction potential, the measured pH may be expected to differ seriously from the determined from cells without a liquid junction in solutions of high acidity or high alkalinity. Merely to affirm the proper functioning of the glass electrode at the extreme ends of the pH scale, two secondary standards are included in Table 8.14. In addition, values for a 0.1 m solution of HCl are given to extend the pH scale up to 275°C [see R. S. Greeley, Anal. Chem. 32 1717 (I960)] ... [Pg.931]

When the potential of an electrochemical cell is measured, the contribution of the liquid junction potential must be included. Thus, equation 11.1 is rewritten as... [Pg.471]

Electrochemical methods covered in this chapter include poten-tiometry, coulometry, and voltammetry. Potentiometric methods are based on the measurement of an electrochemical cell s potential when only a negligible current is allowed to flow, fn principle the Nernst equation can be used to calculate the concentration of species in the electrochemical cell by measuring its potential and solving the Nernst equation the presence of liquid junction potentials, however, necessitates the use of an external standardization or the use of standard additions. [Pg.532]

Reference Electrodes and Liquid Junctions. The electrical cincuit of the pH ceU is completed through a salt bridge that usually consists of a concentrated solution of potassium chloride [7447-40-7]. The solution makes contact at one end with the test solution and at the other with a reference electrode of constant potential. The Hquid junction is formed at the area of contact between the salt bridge and the test solution. The mercury—mercurous chloride electrode, the calomel electrode, provides a highly reproducible potential in the potassium chloride bridge solution and is the most widely used reference electrode. However, mercurous chloride is converted readily into mercuric ion and mercury when in contact with concentrated potassium chloride solutions above 80°C. This disproportionation reaction causes an unstable potential with calomel electrodes. Therefore, the silver—silver chloride electrode and the thallium amalgam—thallous chloride electrode often are preferred for measurements above 80°C. However, because silver chloride is relatively soluble in concentrated solutions of potassium chloride, the solution in the electrode chamber must be saturated with silver chloride. [Pg.466]

Standard potentials Ee are evaluated with full regard to activity effects and with all ions present in simple form they are really limiting or ideal values and are rarely observed in a potentiometric measurement. In practice, the solutions may be quite concentrated and frequently contain other electrolytes under these conditions the activities of the pertinent species are much smaller than the concentrations, and consequently the use of the latter may lead to unreliable conclusions. Also, the actual active species present (see example below) may differ from those to which the ideal standard potentials apply. For these reasons formal potentials have been proposed to supplement standard potentials. The formal potential is the potential observed experimentally in a solution containing one mole each of the oxidised and reduced substances together with other specified substances at specified concentrations. It is found that formal potentials vary appreciably, for example, with the nature and concentration of the acid that is present. The formal potential incorporates in one value the effects resulting from variation of activity coefficients with ionic strength, acid-base dissociation, complexation, liquid-junction potentials, etc., and thus has a real practical value. Formal potentials do not have the theoretical significance of standard potentials, but they are observed values in actual potentiometric measurements. In dilute solutions they usually obey the Nernst equation fairly closely in the form ... [Pg.363]

An element of uncertainty is introduced into the e.m.f. measurement by the liquid junction potential which is established at the interface between the two solutions, one pertaining to the reference electrode and the other to the indicator electrode. This liquid junction potential can be largely eliminated, however, if one solution contains a high concentration of potassium chloride or of ammonium nitrate, electrolytes in which the ionic conductivities of the cation and the anion have very similar values. [Pg.549]

In view of the problems referred to above in connection with direct potentiometry, much attention has been directed to the procedure of potentio-metric titration as an analytical method. As the name implies, it is a titrimetric procedure in which potentiometric measurements are carried out in order to fix the end point. In this procedure we are concerned with changes in electrode potential rather than in an accurate value for the electrode potential with a given solution, and under these circumstances the effect of the liquid junction potential may be ignored. In such a titration, the change in cell e.m.f. occurs most rapidly in the neighbourhood of the end point, and as will be explained later (Section 15.18), various methods can be used to ascertain the point at which the rate of potential change is at a maximum this is at the end point of the titration. [Pg.549]

A typical set of experimental data290a,290b is shown in Fig. 11. All measurements converge to the value measured by Grahame.286 At present, the of Hg in water can be confidently indicated5 as -0.433 0.001 V (SCE), i.e., -0.192 0.001 V (SHE). The residual uncertainty is related to the unknown liquid junction potential at the boundary with the SCE, which is customarily used as a reference electrode. The temperature coefficient of of the Hg/H20 interface has been measured and its significance discussed.7,106,1 8,291... [Pg.57]

Japaridze et al.m 323 have studied the interface between Hg and a number of vicinal and nonvicinal diols such as 1,2-, 1,3-, 2,3- and 1,4-butanediol (BD), ethanediol (ED), and 1,3-propanediol. KF and LiC104 were used as surface-inactive electrolytes. The potential of zero charge was measured by the capacitance method against an SCE in water without correction for the liquid junction potential at the solvent/H20 contact (such a potential drop is estimated to be in the range of 20 to 30 mV). The potential of the capacitance minimum was found to be independent of the electrolyte concentration while capacitance decreased with dilution. Therefore, Emin was taken to measure E . These values are reported in Table 4. [Pg.59]

Capacitance and interfacial tension measurements were used to study the interface between Hg and mixtures of acetone + nitromethane.330 The potential was measured against an SCEin H20 and corrected for the liquid junction potential by measuring the half-wave potential of the ferrocene-... [Pg.61]

Following the same procedure, the kinetic constants have been determined for very different electrochemical conditions. When n-WSe2 electrodes are compared in contact with different redox systems it is, for example, found9 that no PMC peak is measured in the presence of 0.1 M KI, but a clear peak occurs in presence of 0.1 M K4[Fe(CN)6], which is known to be a less efficient electron donor for this electrode in liquid junction solar cells. When K4[Fe(CN)6] is replaced by K3[Fe(CN)6], its oxidized form, a large shoulder is found, indicating that minority carriers cannot react efficiently at the semiconductor/electrolyte junction (Fig. 31). [Pg.487]

At present, the microwave electrochemical technique is still in its infancy and only exploits a portion of the experimental research possibilities that are provided by microwave technology. Much experience still has to be gained with the improvement of experimental cells for microwave studies and in the adjustment of the parameters that determine the sensitivity and reliability of microwave measurements. Many research possibilities are still unexplored, especially in the field of transient PMC measurements at semiconductor electrodes and in the application of phase-sensitive microwave conductivity measurements, which may be successfully combined with electrochemical impedance measurements for a more detailed exploration of surface states and representative electrical circuits of semiconductor liquid junctions. [Pg.519]

The possibility of measuring a nonequilibrium liquid junction (diffusion) potential by using voltaic cells has been checked also ... [Pg.35]

A representative ISE is shown schematically in Fig. 1. The electrode consists of a membrane, an internal reference electrolyte of fixed activity, (ai)i , ai and an internal reference electrode. The ISE is immersed in sample solution that contains analyte of some activity, (ajXampie and into which an external reference electrode is also immersed. The potential measured by the pH/mV meter (Eoe,) is equal to the difference in potential between the internal (Eraf.int) and external (Eref.ext) reference electrodes, plus the membrane potential (E emb), plus the liquid junction potential... [Pg.4]

Under potentiostatic conditions, the photocurrent dynamics is not only determined by faradaic elements, but also by double layer relaxation. A simplified equivalent circuit for the liquid-liquid junction under illumination at a constant DC potential is shown in Fig. 18. The difference between this case and the one shown in Fig. 7 arises from the type of perturbation introduced to the interface. For impedance measurements, a modulated potential is superimposed on the DC polarization, which induces periodic responses in connection with the ET reaction as well as transfer of the supporting electrolyte. In principle, periodic light intensity perturbations at constant potential do not affect the transfer behavior of the supporting electrolyte, therefore this element does not contribute to the frequency-dependent photocurrent. As further clarified later, the photoinduced ET... [Pg.220]

In initial ET rate measurements, both the NB and aqueous phases contained 0.1 M TEAP, enabling measurements to be made with a constant Galvani potential difference across the liquid junction. In these early studies, the concentration of Fc used in the organic phase (phase 2) was at least 50 times the concentration of the electroactive mediator in the aqueous phase which contained the probe UME (phase 1). This ensured that the interfacial process was not limited by mass transport in the organic phase, and that the simple constant-composition model, described briefly in Section IV, could be used. [Pg.314]

Oxygen electrode. In principle, a classical oxygen electrode in a liquid electrolyte would be possible if an electrode material were known on the surface of which the redox system 02/0H is electrochemically reversible however, Luther26 measured its standard potential from the following cell without a liquid junction ... [Pg.59]

Thermodynamics describes the behaviour of systems in terms of quantities and functions of state, but cannot express these quantities in terms of model concepts and assumptions on the structure of the system, inter-molecular forces, etc. This is also true of the activity coefficients thermodynamics defines these quantities and gives their dependence on the temperature, pressure and composition, but cannot interpret them from the point of view of intermolecular interactions. Every theoretical expression of the activity coefficients as a function of the composition of the solution is necessarily based on extrathermodynamic, mainly statistical concepts. This approach makes it possible to elaborate quantitatively the theory of individual activity coefficients. Their values are of paramount importance, for example, for operational definition of the pH and its potentiometric determination (Section 3.3.2), for potentiometric measurement with ion-selective electrodes (Section 6.3), in general for all the systems where liquid junctions appear (Section 2.5.3), etc. [Pg.39]

Two types of methods are used to measure activity coefficients. Potentiometric methods that measure the mean activity coefficient of the dissolved electrolyte directly will be described in Section 3.3.3. However, in galvanic cells with liquid junctions the electrodes respond to individual ion activities (Section 3.2). This is particularly true for pH measurement (Sections 3.3.2 and 6.3). In these cases, extrathermodynamical procedures defining individual ion activities must be employed. [Pg.55]

In potentiometric measurements the simplest approach to the liquid-junction problem is to use a reference electrode containing a saturated solution of potassium chloride, for example the saturated calomel electrode (p. 177). The effect of the diffusion potential is completely suppressed if the solutions in contact contain the same indifferent electrolyte in a sufficient... [Pg.125]

It would appear from Eq. (3.2.8) that the pH, i.e. the activity of a single type of ion, can be measured exactly. This is not, in reality, true even if the liquid junction potential is eliminated the value of Eref must be known. This value is always determined by assuming that the activity coefficients depend only on the overall ionic strength and not on the ionic species. Thus the mean activities and mean activity coefficients of the electrolyte must be employed. The use of this assumption in the determination of the value of Eref will, of course, also affect the pH value found from Eq. (3.2.8). Thus, the potentiometric determination of the pH is more difficult than would appear at first glance and will be considered in the special Section 3.3.2. [Pg.184]

The potentiometric measurement of physicochemical quantities such as dissociation constants, activity coefficients and thus also pH is accompanied by a basic problem, leading to complications that can be solved only if certain assumptions are accepted. Potentiometric measurements in cells without liquid junctions lead to mean activity or mean activity coefficient values (of an electrolyte), rather than the individual ionic values. [Pg.203]

For current practice, the described method of pH measurement is too tedious. Moreover, not hydrogen but glass electrodes are used for routine pH measurements (see Section 6.3). Then the expression for the EMF of the cell consisting of the glass and reference electrodes contains a constant term from Eq. (6.3.10), in addition to the terms present in Eq. (3.3.3) this term must be obtained by calibration. Further, a term describing the liquid junction potential between the reference electrode and the measured solution must also be included. [Pg.204]


See other pages where Liquid junction measurement is mentioned: [Pg.942]    [Pg.471]    [Pg.494]    [Pg.568]    [Pg.33]    [Pg.519]    [Pg.652]    [Pg.225]    [Pg.226]    [Pg.227]    [Pg.230]    [Pg.231]    [Pg.262]    [Pg.287]    [Pg.28]    [Pg.57]    [Pg.43]    [Pg.193]    [Pg.207]    [Pg.208]    [Pg.211]    [Pg.212]    [Pg.629]    [Pg.630]    [Pg.125]    [Pg.164]   
See also in sourсe #XX -- [ Pg.209 , Pg.216 ]