Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Linear calibration function calculation

The calibration function calculation uses multiple linear regression to obtain hydrodynamic volume as a polynomial function of elution volume for a given column set. [Pg.133]

The a coefficients have to be calculated using regression procedures. In the case of a calibration function of a higher degree, the calibration graph is curved. The latter also can be approximated by a polygone, where different linear calibration functions are used for well-defined concentration ranges. [Pg.85]

The coefficients a are calculated by regression procedures. Alternatively, different linear calibration functions may be used for different concentration ranges. [Pg.659]

For the practical determination of the detection criteria, the standard deviation Syo of the blank value Jq or the standard deviation of the intercept a of the linear calibration function is used, where the standard deviation should be calculated from at least = 6 measuring values. The detection criteria is then regarded as the upper limit of the blank value scatter... [Pg.962]

In some special analyses it may be necessary to extend the absorbance measurements beyond the linear range (i.e., beyond the validity of Beer-Lamberts s law). In these cases a non-linear calibration curve has to be generated by measuring concentrations versus spectrophotometer readings. The sample concentrations are then calculated according to the non-linear calibration function or corresponding correction terms applied to the linearly calculated sample concentrations. [Pg.225]

The sum of squared differences of each standard to the middle of the operating range (Qxx) and the squared standards (Sxx), both given on the concentration scale, are the only parameters that were additionally required to proceed with comparing linear calibration functions directly [5]. All other input refers to common calibration characteristics. On the important left-hand-side of the spreadsheet, one can see the calculation proceeding step-by-step in the way which is commonly known in chemical laboratories for validation purposes [7], i.e. explicitly comparing precision data and mean values applied to two measured series rmder repeatable conditions (F-test, t-test). [Pg.136]

If the mass ratio between the two matched peak maxima differed at any time during the evaporation of the sample by more than 40 ppm (due to drift of the instrument, sample contamination, or electrical interference), the sample was disregarded. The calibration functions calculated by linear regression for 12 calibration samples containing known picomole quantities of each compound had the correlation coefficients 0.9920, 0.9914 and 0.9865, for cadaverine, monoacetylcadaverine and monopropionylcadaverine, respectively. The reported concentrations were measured as quantities more than three times higher than their blanks. They were not corrected for losses during extraction and thin-layer chromatography. [Pg.206]

In most cases, a linear calibration function can be used, as discussed earlier. It can be calculated by a linear regression with a number of values (c , Xi) where c is the concentration of a standard sample and Xi is the radiation intensity, absorption or ion current obtained for the element to be determined. It has the form ... [Pg.89]

Consequently, the proof of calibration should never be limited to the presentation of a calibration graph and confirmed by the calculation of the correlation coefficient. When raw calibration data are not presented in such a situation, most often a validation study cannot be evaluated. Once again it should be noted that nonlinearity is not a problem. It is not necessary to work within the linear range only. Any other calibration function can be accepted if it is a continuous function. [Pg.104]

If the relationship between the signal and the concentration is not linear, we may apply a second order calibration function. For details about this slightly more difficult calculation see ISO 8466-2. [Pg.189]

If it is not clear, the Mandel-test may be applied for linearity check. We calculate the linear and the 2" order calibration function and the respective residual standard deviations. If the F-test (as described in the shde) delivers a significant difference between the residual standard deviations, this shows that the 2 order calibration function significantly better describes the calibration. So this function should be preferred. If it is not significantly better, we should use the hnear function... [Pg.190]

The most common calibration model or function in use in analytical laboratories assumes that the analytical response is a linear function of the analyte concentration. Most chromatographic and spectrophotometric methods use this approach. Indeed, many instruments and software packages have linear calibration (regression) functions built into them. The main type of calculation adopted is the method of least squares whereby the sums of the squares of the deviations from the predicted line are minimised. It is assumed that all the errors are contained in the response variable, T, and the concentration variable, X, is error free. Commonly the models available are Y = bX and Y = bX + a, where b is the slope of the calibration line and a is the intercept. These values are the least squares estimates of the true values. The following discussions are only... [Pg.48]

Multi point calibration will be recommended if minimum uncertainty and maximum consistency are required over a wide range of pH(X) values [21, 22]. The calibration function of the electrode is then calculated by linear regression of the difference in cell voltage results from the standard pH values. This calibration procedure is also recommended for characterising the performance of electrode systems. [Pg.210]

In relation to Methods IV and V, the instantaneous MMDs were simulated with the aim of calculating the (noise-free) calibrations log M V) and log M (V) (Fig. IB). These functions were obtained from the (noise-free or non-truncated) mass chromatogram in order to illustrate their true shapes in the complete range of the measured chromatogram. The resulting ad hoc calibrations are non-linear, generally less steep than log M(V), and close to the effective linear calibration log Af(V)liv (Fig. IB). [Pg.152]

The EPA MDL determination specifies a minimum of seven (k = n = 7) replicate spikes prepared at a single appropriately low concentration (generally one to five times the expected MDL). The spiked matrix typically is reagent water, or clean sand or sodium sulfate for sediment samples, or a well characterized natural material that does not contain the substance for tissue samples the point here is to avoid matrix interferences, a condition that will apply to real world analytical samples only if appropriate clean-up procedures are employed. These spiked blank matrices are processed through the entire analytical method, preferably completed within a short period (a few days). It is assumed that the calibration function is linear in both concentration-response and in the fitting parameters (i.e.. Equation [8.19a] is valid) at least in the low concentration region, and that the frequency distribution of measured concentrations in the low concentration spiked blank has a normal distribution (Eigure 8.11(a)). Another important assumption of the EPA MDL calculation is that the variance Vy,i... [Pg.421]


See other pages where Linear calibration function calculation is mentioned: [Pg.102]    [Pg.193]    [Pg.343]    [Pg.58]    [Pg.47]    [Pg.948]    [Pg.952]    [Pg.561]    [Pg.565]    [Pg.112]    [Pg.116]    [Pg.78]    [Pg.83]    [Pg.84]    [Pg.84]    [Pg.130]    [Pg.265]    [Pg.869]    [Pg.425]    [Pg.3415]    [Pg.298]    [Pg.513]    [Pg.517]    [Pg.84]    [Pg.156]    [Pg.618]    [Pg.57]    [Pg.152]    [Pg.418]    [Pg.84]    [Pg.658]    [Pg.155]   
See also in sourсe #XX -- [ Pg.282 ]




SEARCH



Calibration linear

Linear calibration function

Linear functional

Linear functionals

Linear functions

Linearity calculation

© 2024 chempedia.info