Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Isotope ratio measurements, mass

Isotope Dilution Analysis. Isotope Ratio Measurements. Mass Spectrometry Overview. [Pg.247]

The combination of gas chromatography and an isotope ratio measurement mass spectrometer (ir/n-GC-MS) was used to study the detoxification and degradation pathways of tropane alkaloids involving iV-demethylation to the respective nor-compounds. Molinie et al. developed protocols for sample preparation, including liquid-hquid and SPE extraction, and analysis of the kinetic isotope effects (KlEs) with irm-GC-MS and ir/M-elemental analysis (EA)-MS [100]. Degradation experiments were performed with a Pseudomonas strain that uses tropane alkaloids as sole source of carbon and nitrogen. The authors found that liquid-liquid extraction from the aqueous medium in combination with ir/n-GC-MS was the most convenient method for measurement of the ratio. Later, Kosieradzka... [Pg.1030]

Since detailed chemical structure information is not usually required from isotope ratio measurements, it is possible to vaporize samples by simply pyrolyzing them. For this purpose, the sample can be placed on a tungsten, rhenium, or platinum wire and heated strongly in vacuum by passing an electric current through the wire. This is thermal or surface ionization (TI). Alternatively, a small electric furnace can be used when removal of solvent from a dilute solution is desirable before vaporization of residual solute. Again, a wide variety of mass analyzers can be used to measure m/z values of atomic ions and their relative abundances. [Pg.285]

Because variations in accurate isotope ratio measurements typically concern only a few parts per 1000 by mass and there are no universal absolute ratios, it is necessary to define some standards. For this purpose, samples of standard substances are produced and made available at two major centers IAEA (International Atomic Energy Authority, U.K.) and NIST (National Institute for Standards and Technology, U.S.). Standards from other sources are also available. These primary standards can be used as such, or alternative standards can be employed if the primary ones are not available. However, any alternative standards need to be related accurately to the primary ones (see formulae below). For example, the material PDB (PeeDee belemnite), used particularly as a standard for the ratio of isotopes, is no longer readily available, and a new standard, VPDB,... [Pg.354]

For example, if a carbonaceous sample (S) is examined mass spectrometrically, the ratio of abundances for the carbon isotopes C, in the sample is Rg. This ratio by itself is of little significance and needs to be related to a reference standard of some sort. The same isotope ratio measured for a reference sample is then R. The reference ratio also serves to check the performance of the mass spectrometer. If two ratios are measured, it is natural to assess them against each other as, for example, the sample versus the reference material. This assessment is defined by another ratio, a (the fractionation factor Figure 48.2). [Pg.354]

Almost any type of analyzer could be used to separate isotopes, so their ratios of abundances can be measured. In practice, the type of analyzer employed will depend on the resolution needed to differentiate among a range of isotopes. When the isotopes are locked into multielement ions, it becomes difficult to separate all of the possible isotopes. For example, an ion of composition CgHijOj will actually consist of many compositions if all of the isotopes ( C, C, H, H, 0, O, and 0) are considered. To resolve all of these isotopic compositions before measurement of their abundances is difficult. For low-molecular-mass ions (HjO, COj) or for atomic ions (Ca, Cl), the problems are not so severe. Therefore, most accurate isotope ratio measurements are made on low-molecular-mass species, and resolution of these even with simple analyzers is not difficult. The most widely used analyzers are based on magnets, quadrupoles, ion traps, and time-of-flight instruments. [Pg.365]

Accurate, precise isotope ratio measurements are important in a wide variety of applications, including dating, examination of environmental samples, and studies on drug metabolism. The degree of accuracy and precision required necessitates the use of special isotope mass spectrometers, which mostly use thermal ionization or inductively coupled plasma ionization, often together with multiple ion collectors. [Pg.369]

This is the basic process in an inductively coupled plasma discharge (ICP). The excited ions can be examined by observing the emitted light or by mass spectrometry. Since the molecules have been broken down into their constituent atoms (as ions) including isotopes, these can be identified and quantified by mass spectrometry, as happens with isotope ratio measurements. [Pg.388]

Accurate, precise isotope ratio measurements are used in a variety of applications including dating of artifacts or rocks, studies on drug metabolism, and investigations of environmental issues. Special mass spectrometers are needed for such accuracy and precision. [Pg.426]

Cochran JK, Masque P (2003) Short-lived U/Th-series radionuclides in the ocean tracers for scavenging rates, export fluxes and particle dynamics. Rev Mineral Geochem 52 461-492 Cohen AS, O Nions RK (1991) Precise determination of femtogram quantities of radium by thermal ionization mass spectrometry. Anal Chem 63 2705-2708 Cohen AS, Belshaw NS, O Nions RK (1992) High precision uranium, thorium, and radium isotope ratio measurements by high dynamic range thermal ionization mass spectrometry. Inti J Mass Spectrom Ion Processes 116 71-81... [Pg.56]

Neder H, Heusser G, Laubenstein M (2000) Low-level y-ray germanium-spectrometer to measure veiy low primordial radionuclide concentrations. ApplRadiat Isot 53 191-195 Palacz ZA, Freedman PA, Walder AJ (1992) Thorium isotope ratio measurements at high abundance sensitivity using a VG 54-30, an energy-filtered thermal ionization mass spectrometer. Chem Geol 101 157-165... [Pg.58]

Walder AJ, Freedman PA (1992) Isotopic ratio measurement using a double focusing magnetic sector mass analyser with an inductively coupled plasma as an ion somce. J Anal At Spectrom 7 571-575... [Pg.59]

Walder AJ, Roller D, Reed NM, Hutton RC, Freedman PA (1993) Isotope ratio measurement by inductively coupled plasma multiple collector mass spectrometry incorporating a high efficiency nebulization system. J Anal At Spectrom 8 1037-1041... [Pg.59]

Table 1 lists MORE studies to date and Figures 2 and 3 present data from different areas of ridge. The first study of U-series disequilibria in MORE was the pioneering work of Condomines et al. (1981) (Fig. 2A). These workers analyzed samples having a relatively wide range in composition (Mg 72 to 57) from the FAMOUS region of the Mid-Atlantic Ridge (MAR 37°N) by combined alpha spectrometry (for U and Th isotopic ratios) and mass spectrometry (isotope dilution measurements for U and Th... [Pg.179]

Applications Sector instruments are applied for niche applications such as high-resolution measurements and fundamental ion chemistry studies. Magnetic sector mass spectrometers remain the instrument of choice in areas of target compound trace analysis, accurate mass measurement and isotope ratio measurement. [Pg.388]

Flegal and Stukas [406] described the special sampling and processing techniques necessary for the prevention of lead contamination of seawater samples, prior to stable lead isotopic ratio measurements by thermal ionisation mass spectrometry. Techniques are also required to compensate for the absence of an internal standard and the presence of refractory organic compounds. The precision of the analyses is 0.1 -0.4% and a detection limit of 0.02 ng/kg allows the tracing of lead inputs and biogeochemical cycles. [Pg.191]

TI is a very precise and accurate method in stable isotope ratio measurements and quantification of inorganic elements, for example, by isotope dilution mass spectrometry [8]. Because TI is a continuous ion source, it could be coupled to any analyzer that is suitable for such sources. However, because the strength of TI lies in the quantitative precision and accuracy, sector analyzers are preferred to ensure maximum quality. [Pg.19]

T. Walczyk. Iron Isotope Ratio Measurements by Negative Thermal Ionisation Mass Spectrometry using FeF Molecular Ions. Int. J. Mass Spectrom. Ion Proc., 161(1997) 217-227. [Pg.71]

Walder, A.J. and Furuta, N. (1993). High-precision lead isotope ratio measurements by inductively coupled plasma multiple collector mass spectrometry. Analytical Sciences 9 675-680. [Pg.74]

Secondary isotope effects are small. In fact, most of the secondary deuterium KIEs that have been reported are less than 20% and many of them are only a few per cent. In spite of the small size, the same techniques that are used for other kinetic measurements are usually satisfactory for measuring these KIEs. Both competitive methods where both isotopic compounds are present in the same reaction mixture (Westaway and Ali, 1979) and absolute rate measurements, i.e. the separate determination of the rate constant for the single isotopic species (Fang and Westaway, 1991), are employed (Parkin, 1991). Most competitive methods (Melander and Saunders, 1980e) utilize isotope ratio measurements based on mass spectrometry (Shine et al., 1984) or radioactivity measurements by liquid scintillation (Ando et al., 1984 Axelsson et al., 1991). However, some special methods, which are particularly useful for the accurate determination of secondary KIEs, have been developed. These newer methods, which are based on polarimetry, nmr spectroscopy, chromatographic isotopic separation and liquid scintillation, respectively, are described in this section. The accurate measurement of small heavy-atom KIEs is discussed in a recent review by Paneth (1992). [Pg.234]

B is the magnetic field intensity, r is the radius of the ion path, and V is the accelerating potential. In analytical mass spectrometers either B or V is varied systematically so that ions of different m/z are sequentially focused on the collector and the spectrum recorded. Such whole molecule mass spectrometers have been employed occasionally for isotope ratio measurements but their use is restricted to samples... [Pg.215]

Because there is so little mass bias in the mass analyzer, a discussion of ion transfer optics and collectors is not presented. The ion transfer optics of the magnetic sector mass analyzer, and the collectors used for isotope ratio measurements are critical design elements in all isotope ratio mass spectrometers and recent reviews of these items can be found in Habfast... [Pg.114]

As the name implies, the sample is introduced into the mass spectrometer as a gas (Nier 1940). There are two types of sources, the classic viscous flow source and the continuous flow source. The viscous flow source typically consists of two identical inlet systems that are coupled to the mass spectrometer by a change-over valve, which allows rapid switching for comparison of isotope ratios measured for sample and standard gases. In the continuous flow source, samples gas is introduced as a bubble in a non-reactive carrier gas stream. [Pg.114]

Rigorous correction for instrumental mass bias is required if the precision of an isotope ratio measurement needs to be greater than l%o per mass unit. This concept is well illustrated by the definitive Ca isotope work of Russell et al. (1978), which used a double-spike approach. Prior to the Ca isotope investigation of Russell et al. (1978), natural mass-dependent Ca... [Pg.117]

The MC-ICP-MS consists of four main parts 1) a sample introduction system that inlets the sample into the instrument as either a liquid (most common), gas, or solid (e.g., laser ablation), 2) an inductively coupled Ar plasma in which the sample is evaporated, vaporized, atomized, and ionized, 3) an ion transfer mechanism (the mass spectrometer interface) that separates the atmospheric pressure of the plasma from the vacuum of the analyzer, and 4) a mass analyzer that deals with the ion kinetic energy spread and produces a mass spectrum with flat topped peaks suitable for isotope ratio measurements. [Pg.118]

Hirata T, Hayano Y, Ohno T (2003) Improvements in precision of isotopic ratio measurements using laser ablation-multiple collector-ICP-mass spectrometry reduction of changes in measured isotopic ratios. J Anal At Spectrom 18 1283-1288... [Pg.148]

WaIczyk T (1997) Iron isotope ratio measurements by negative thermal ionization mass spectrometry using FeF4" molecular ions. Int J Mass Spectrom Ion Proc 161 217-227... [Pg.150]

Using the three measured ratios, Ca/ Ca, Ca/ " Ca and Ca/ " Ca, three unknowns can be solved for the tracer/sample ratio, the mass discrimination, and the sample Ca/ Ca ratio (see also Johnson and Beard 1999 Heuser et al. 2002). Solution of the equations is done iteratively. It is assumed that the isotopic composition of the Ca- Ca tracer is known perfectly, based on a separate measurement of the pure spike solution. Initially it is also assumed that the sample calcium has a normal Ca isotopic composition (equivalent to the isotope ratios listed in Table 1). The Ca/ Ca ratio of the tracer is determined based on the results of the mass spectrometry on the tracer-sample mixture, by calculating the effect of removing the sample Ca. This yields a Ca/ Ca ratio for the tracer, which is in general different from that previously determined for the tracer. This difference is attributed to mass discrimination in the spectrometer ion source and is used to calculate a first approximation to the parameter p which describes the instrumental mass discrimination (see below). The first-approximation p is used to correct the measured isotope ratios for mass discrimination, and then a first-approximation tracer/sample ratio and a first-approximation sample CeJ Ca... [Pg.259]


See other pages where Isotope ratio measurements, mass is mentioned: [Pg.219]    [Pg.513]    [Pg.513]    [Pg.219]    [Pg.513]    [Pg.513]    [Pg.369]    [Pg.56]    [Pg.660]    [Pg.52]    [Pg.114]    [Pg.21]    [Pg.70]    [Pg.168]    [Pg.195]    [Pg.199]    [Pg.113]    [Pg.118]    [Pg.123]    [Pg.148]    [Pg.156]    [Pg.158]    [Pg.229]    [Pg.231]   


SEARCH



Isotope Ratio Measurements by Gas Source Mass Spectrometry

Isotope measurement

Isotope ratio measurements

Isotope ratios

Isotopes masses

Isotopes, measuring

Isotopic masses

Isotopic measurement

MASS RATIO

Mass measurements

Measuring Mass

© 2024 chempedia.info