Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrumentation injector

As regards LC-MS/MS, most compounds can be analyzed in the ESI (-I-) mode. There are several acidic compounds, however, that are more sensitive in the negative ESI (-) mode. To avoid precipitation of non-polar pesticides due to solubility shift, any dilution of extract with the aqueous component of the LC-eluent is better to be performed automatically in the instrument injector itself. [Pg.457]

A GPC system consists of various instruments. Injectors are used to introduce the polymer solution into the columns of separation. Pumps deliver the sample and solvent through the columns and the system. Detectors record the exit of fractions of the sample and count the number of molecules of a certain MW. The computer controls the test automatically, records the results, and calculates the different MW averages. The GPC system contains a number of different instruments that work together to provide the optimum system performance. Figure 17.3 shows a schematic of a gel permeation chromatograph with the basic components. [Pg.358]

The instrumentation of HdC, including a pump, an injector, a column (set), a detector, and a recorder or computer, is very similar to size exclusion chromatography SEC). The essence of this technique is the column. There are two types of HdC columns open microcapillary tubes and a nonporous gel-packed column. This chapter emphasizes column technology and selection and the applications of this technique on the molecular weight analysis of macromolecules. [Pg.597]

Although the OTHdC has several unique applications in polymer analysis, this technique has several limitations. First, it requires the instrumentation of capillary HPLC, especially the injector and detector, which is not as popular as packed column chromatography at this time. Second, as discussed previously, the separation range of a uniform capillary column is rather narrow. Third, it is difficult to couple capillary columns with different sizes together as SEC columns. [Pg.601]

When a problem is noted, it is generally best to look for solutions in easy-to-check-out areas. For example, check for leaks at all column connections and external fittings. Once the simple checks are completed, attempt to isolate the problem to the injector, column, detector (or transfer line), or instrument. [Pg.182]

Ensure that the actual instrument configuration conforms to what is written under Experimental supplier, models, modifications, consumables (HPLC or GC columns, gaskets, etc.), and software for the main instrument, peripherals (injectors, integrators, computers, printers, plotters, etc.), and ancillary equipment (vortexer, dispensers, balances, centrifuges, filters, tubing, etc.). [Pg.143]

The instrument used in this study was an in-house constructed HPGPC composed of a Waters Associates M-6000 solvent delivery system, Waters Associates U6K injector, Varian Instruments fixed... [Pg.207]

Figure A2.1 Waters ProMonix On-Line HPLC analyzer. The upper compartment door contains a keypad for programming and operation of the analyzer. The upper window allows viewing of indicator lights and a liquid crystal display that provides the operator with analyzer interface, programmed parameters, and instrument status results. The lower chamber contains the pumps, valves, injector, and detector(s) required for the chromatographic separation. The sample conditioning plate for online process monitoring is to the right of the analyzer. This is a typical process HPLC. (From Cotter, R.L. and Li, J.B., Lab Rob Autom., 1, 251,1989. With permission of VCH Publishers.)... Figure A2.1 Waters ProMonix On-Line HPLC analyzer. The upper compartment door contains a keypad for programming and operation of the analyzer. The upper window allows viewing of indicator lights and a liquid crystal display that provides the operator with analyzer interface, programmed parameters, and instrument status results. The lower chamber contains the pumps, valves, injector, and detector(s) required for the chromatographic separation. The sample conditioning plate for online process monitoring is to the right of the analyzer. This is a typical process HPLC. (From Cotter, R.L. and Li, J.B., Lab Rob Autom., 1, 251,1989. With permission of VCH Publishers.)...
GC-FTIR, GC-AED, GC-ICP-MS, cf. Chapter 7), fast GC separations (1996) and most recently the development of sophisticated injectors with temperatureprogramming capability and high-resolution systems (GC-ToFMS). As a result, modem GC systems are quite advanced (Scheme 4.3) and GC is one of the most widely applied instrumental techniques. [Pg.182]

With today s instruments the customer may specify an injector or an automatic reagent dispenser, and a choice of sample formats is available test tube, vial, microplate, Petri dish. The volumes of sample and reagent required to carry out a determination steadily decrease and are typically a few microliters. Sample temperature during the assays is generally controlled. [Pg.86]

The sample preparation step in Figure 3.9 in the simplest case would include only the aspiration, but often steps like dilution, pre-concentration, and other treatments will be performed before injection of the sample onto the separation column. This is easily performed on modem instruments that allow injector programming and the use of additional valves. This will be discussed further in the next section about parallelizing steps. If parallelizing is not possible and a purely serial analysis is required, the required treatment should be abandoned or performed off-line. [Pg.109]

FIA has also found wide application in pharmaceutical analysis.214,215 Direct UV detection of active ingredients is the most popular pharmaceutical analysis application of FIA. For single component analysis of samples with little matrix interference such as dissolution and content uniformity of conventional dosage forms, many pharmaceutical chemists simply replace a column with suitable tubing between the injector and the detector to run FIA on standard HPLC instrumentation. When direct UV detection offers inadequate selectivity, simple online reaction schemes with more specific reagents including chemical, photochemical, and enzymatic reactions of derivatization are applied for flow injection determination of pharmaceuticals.216... [Pg.269]

A simple system is comprised of an isocratic pump, a manual injector, a UV detector, and a strip-chart recorder. A schematic diagram of an HPLC instrument is shown in Fig. 15.4. This simple configuration is rarely used in most modern laboratories. A typical HPLC system is likely to consist of a multi-solvent pump, an autosampler, an on-line degasser, a column oven, and a UV/Vis or photodiode array detector all connected to and controlled by a data-handling workstation. Examples of modular and integrated systems are shown in Fig. 15.5. Some of the important instrumental requirements are summarized in Table 15.2. [Pg.503]


See other pages where Instrumentation injector is mentioned: [Pg.73]    [Pg.73]    [Pg.246]    [Pg.491]    [Pg.223]    [Pg.199]    [Pg.174]    [Pg.227]    [Pg.242]    [Pg.736]    [Pg.46]    [Pg.110]    [Pg.126]    [Pg.279]    [Pg.287]    [Pg.640]    [Pg.791]    [Pg.833]    [Pg.250]    [Pg.129]    [Pg.182]    [Pg.200]    [Pg.233]    [Pg.459]    [Pg.468]    [Pg.257]    [Pg.258]    [Pg.497]    [Pg.85]    [Pg.59]    [Pg.121]    [Pg.101]    [Pg.111]    [Pg.143]    [Pg.146]    [Pg.72]    [Pg.78]    [Pg.356]   
See also in sourсe #XX -- [ Pg.151 , Pg.151 , Pg.152 ]

See also in sourсe #XX -- [ Pg.5 , Pg.6 , Pg.7 ]




SEARCH



Injector

Instrumentation pumps, injectors and column design

© 2024 chempedia.info