Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

GC-MS

FIGURE 13 44 Diagram of a gas chromatograph When connected to a mass spectrometer as in GC/MS the effluent is split into two streams as it leaves the column One stream goes to the detector the other to the mass spectrometer (Adapted with permission from H D Durst and G W Gokel Experimental Organic Chemistry Inti eti McGraw Hill New York 1987)... [Pg.572]

Ring Currents Aromatic and Antiaromatic Magnetic Resonance Imaging Spectra by the Thousands Gas Chromatography GC/MS and MS/MS... [Pg.1329]

Yang, M. J. Orton, M. L. Pawliszyn, J. Quantitative Determination of Caffeine in Beverages Using a Combined SPME-GC/MS Method, /. Chem. Educ. 1997, 74,... [Pg.226]

In GC-MS effluent from the column is introduced directly into the mass spectrometer s ionization chamber in a manner that eliminates the majority of the carrier gas. In the ionization chamber all molecules (remaining carrier gas, solvent, and solutes) are ionized, and the ions are separated by their mass-to-charge ratio. Because each solute undergoes a characteristic fragmentation into smaller ions, its mass spectrum of ion intensity as a function of mass-to-charge ratio provides qualitative information that can be used to identify the solute. [Pg.571]

For mixture.s the picture is different. Unless the mixture is to be examined by MS/MS methods, usually it will be necessary to separate it into its individual components. This separation is most often done by gas or liquid chromatography. In the latter, small quantities of emerging mixture components dissolved in elution solvent would be laborious to deal with if each component had to be first isolated by evaporation of solvent before its introduction into the mass spectrometer. In such circumstances, the direct introduction, removal of solvent, and ionization provided by electrospray is a boon and puts LC/MS on a level with GC/MS for mixture analysis. Further, GC is normally concerned with volatile, relatively low-molecular-weight compounds and is of little or no use for the many polar, water soluble, high-molecular-mass substances such as the peptides, proteins, carbohydrates, nucleotides, and similar substances found in biological systems. LC/MS with an electrospray interface is frequently used in biochemical research and medical analysis. [Pg.59]

Specialized detectors and inlet systems for GC/MS and LC/MS are described in Chapters 36 and 37, respectively. [Pg.250]

As described above, the mobile phase carrying mixture components along a gas chromatographic column is a gas, usually nitrogen or helium. This gas flows at or near atmospheric pressure at a rate generally about 0,5 to 3.0 ml/min and evenmally flows out of the end of the capillary column into the ion source of the mass spectrometer. The ion sources in GC/MS systems normally operate at about 10 mbar for electron ionization to about 10 mbar for chemical ionization. This large pressure... [Pg.254]

In a GC/MS combination, passage of the separated components (A, B, C, D) successively into the mass spectrometer yields their individual spectra. [Pg.256]

Once a mass spectrum from an eluting component has been acquired, the next step is to try to identify the component either through the skill of the mass spectroscopist or by resorting to a library search. Most modem GC/MS systems with an attached data station include a large library of spectra from known compounds (e.g., the NIST library). There may be as many as 50,000 to 60,000 stored spectra covering most of the known simple volatile compounds likely to be met in analytical work. Using special search routines under the control of the computer, one can examine... [Pg.257]

By connecting a gas chromatograph to a suitable mass spectrometer and including a data system, the combined method of GC/MS can be used routinely to separate complex mixtures into theii individual components, identify the components, and estimate their amounts. The technique is widely used. [Pg.260]

The EPA Contract Laboratory Program (CLP) has responsibility for managing the analysis programs required under the U.S. Comprehensive Environmental Response, Compensation, and Liability Act (CERCLA). The approved analytical methods are designed to analyze water, soil, and sediment from potentially hazardous waste sites to determine the presence or absence of certain environmentally harmful organic compounds. The methods described here all require the use of GC/MS. [Pg.295]

To satisfy the Resource Conservation and Recovery Act (1977) and its amendment for hazardous and solid waste (1984), the 80(K) Series Methods have been designed to analyze solid waste, soUs, and groundwater. In particular, methods 8240/8260 require the use of a purge-and-trap device in conjunction with packed or capillary GC/MS, respectively, for the analysis of purgeable organic compounds. Methods 8250/8270 concern analyses for the less-volatile bases, neutrals, and acids by GC/MS after extraction from the matrix by an organic solvent. [Pg.296]

The next step is to show that the response for the analysis of any target compound is linear. This step is known as the initial calibration and is achieved by the analysis of standards for a series of specified concentrations to produce a five-point calibration curve (Figure 41.2a, b). On subsequent days, a continuing calibration must be performed on calibration check compounds to evaluate the calibration precision of the GC/MS system. [Pg.299]

Before sample preparation, surrogate compounds must be added to the matrix. These are used to evaluate the efficiency of recovery of sample for any analytical method. Surrogate standards are often brominated, fluorinated, or isotopically labeled compounds that are not expected to be present in environmental media. If the surrogates are detected by GC/MS within the specified range, it is... [Pg.299]

At such a rate of scanning, it is even possible to examine eluants from capillary GC (gas chromatography) columns during GC/MS operations. [Pg.413]

A detector is needed to sense when the separated substances are emerging from the end of the column. A mass spectrometer (MS) makes a very good, sensitive detector and can be coupled to either GC or LC to give the combined techniques of GC/MS or LC/MS, respectively. [Pg.414]

The coupled methods, GC/MS and LC/MS, form very powerful combinations for simultaneous separation and identification of components of mixtures. Hence, these techniques have been used in such widely disparate enterprises as looking for evidence of life forms on Mars and for testing racehorses or athletes for the presence of banned drugs. [Pg.414]

Gas chromatography/ma.ss spectrometry (GC/MS) is an analytical technique combining the advantages of a GC instrument with those of a mass spectrometer. [Pg.414]

Because a GC and an MS both operate in the gas phase, it is a simple matter to connect the two so that separated components of a mixture are passed sequentially from the GC into the MS, where their mass spectra are obtained. This combined GC/MS is a very powerful analytical technique, the two instruments complementing each other perfectly. [Pg.415]

The combined GC/MS system provides more information than is obvious from the simple sum of the two separate instruments. [Pg.415]


See other pages where GC-MS is mentioned: [Pg.572]    [Pg.573]    [Pg.586]    [Pg.616]    [Pg.621]    [Pg.4]    [Pg.4]    [Pg.77]    [Pg.185]    [Pg.242]    [Pg.245]    [Pg.253]    [Pg.254]    [Pg.255]    [Pg.255]    [Pg.256]    [Pg.257]    [Pg.257]    [Pg.258]    [Pg.277]    [Pg.278]    [Pg.279]    [Pg.283]    [Pg.296]    [Pg.296]    [Pg.323]    [Pg.324]    [Pg.414]   
See also in sourсe #XX -- [ Pg.97 ]

See also in sourсe #XX -- [ Pg.100 , Pg.266 ]

See also in sourсe #XX -- [ Pg.174 , Pg.473 , Pg.473 ]

See also in sourсe #XX -- [ Pg.611 ]




SEARCH



Chromatography GC/MS)

Classical detectors, for GC-MS systems

Combined techniques - GC-MS

Coupling of Purge and Trap with GC-MS Systems

Developing a GC-MS Method

Evaluation of GC-MS Analyses

Fast GC-MS

GC and MS

GC-EI-MS-analysis

GC-ICP-MS

GC-MS (gas chromatograph-mass

GC-MS (gas chromatography and

GC-MS Operation

GC-MS Technology

GC-MS analysis

GC-MS analytical procedure

GC-MS and Isotope Ratio Mass Spectrometry

GC-MS calibration

GC-MS combination

GC-MS interfacing

GC-MS of Volatile Organic Compounds Using Purge and Trap Extraction

GC-MS procedure

GC-MS-selected ion monitoring

GC-MS-selective ion monitoring

GC/MS (gas chromatography with mass

GC/MS experiments

GC/MS system

GC/MS tuning

GC—MS jet separator

GC—MS studies

GC—MS technique

Gas chromatograph mass spectrometry GC-MS)

Gas chromatography with mass spectrometry GC/MS)

Gas chromatography-coupled mass spectrometry GC-MS)

Gas chromatography-mass spectrometer GC-MS)

Gas chromatography-mass spectrometry GC-MS)

Gas chromatography-mass spectroscopy GC-MS)

General comments on GC-MS techniques in water analysis

Interface GC-MS)

Isotope Ratio Monitoring GC-MS

Notations in irm-GC-MS

Organic acids observed and quantified prior to GC-MS

Organic acids observed in blood and urine using GC-MS

Pyrolysis-gas chromatography-mass spectrometry Py GC MS)

Quantification by GC-MS

Recent GC-MS Studies of Different Elements

© 2024 chempedia.info