Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Frequency analysis method

Now, let us suppose that the system is in a black-box, and that all we can know of it is the observable Q, sampled up to a finite time interval. This is the typical situation occurring in Physics, where one obtains information on some system on the basis of the output of an experiment. The amplitudes and frequencies (6) can be numerically computed from g(t), for example, by means of the frequency analysis method (Laskar 1990, Laskar et al. 1992). However, if we are interested mainly in recognizing the quasi-periodic nature of the solution, it is not necessary to use a refined frequency analysis, but it is sufficient to compute the Fast Fourier transform of time interval [—T,T], where >( ) is a suitable analytic window on [—T, T] (see Section 4 for details). Figure 1 shows an example of such an analysis. Within the precision of our computation (a line is identified with an error of about 10-5 in frequency) we can easily recognize that the spectrum of g(t) is a line spectrum. Now, we consider the more interesting quasi-integrable Hamiltonian ... [Pg.169]

Feng Z, Liang M, Chu F (2013) Recent advances in time-frequency analysis methods for machinery fault diagnosis a review with application examples. Mech Syst Signal Pr 38(1) 165-205... [Pg.1849]

Figure B2.1.8 Dynamic absorption trace obtained with the dye IR144 in methanol, showing oscillations arising from coherent wavepacket motion (a) transient observed at 775 mn (b) frequency analysis of the oscillations obtained using a linear prediction, smgular-value-decomposition method. Figure B2.1.8 Dynamic absorption trace obtained with the dye IR144 in methanol, showing oscillations arising from coherent wavepacket motion (a) transient observed at 775 mn (b) frequency analysis of the oscillations obtained using a linear prediction, smgular-value-decomposition method.
Risk-Based Inspection. Inspection programs developed using risk analysis methods are becoming increasingly popular (15,16) (see Hazard ANALYSIS AND RISK ASSESSMENT). In this approach, the frequency and type of in-service inspection (IS I) is determined by the probabiUstic risk assessment (PRA) of the inspection results. Here, the results might be a false acceptance of a part that will fail as well as the false rejection of a part that will not fail. Whether a plant or a consumer product, false acceptance of a defective part could lead to catastrophic failure and considerable cost. Also, the false rejection of parts may lead to unjustified, and sometimes exorbitant, costs of operation (2). Risk is defined as follows ... [Pg.123]

The frequency analysis step involves estimating the likelihood of occurrence of each of the undesired situations defined in the hazard identification step. Sometimes you can do this through direct comparison with experience or extrapolation from historical accident data. While this method may be of great assistance in determining accident frequencies, most accidents analyzed by QRA are so rare that the frequencies must be synthesized using frequency estimation methods and models. [Pg.36]

A sound is generally not a pure tone, as the latter is only emitted from particular sources. It can be demonstrated that a sound can be divided into different pure tones (superposition method). The waves at different frequencies give the spectrum of the sound, which also describes its energy distribution. In frequency analysis, the spectrum is divided into octave bands. An octave band is defined as the frequency range with its upper boundary twice the frequency of its lower boundary. For every octave band, a central band frequency ( f. ) is defined as follows ... [Pg.793]

The most important non-faradaic methods are conductometric analysis and (normal) potentiometric analysis in the former we have to deal essentially with the ionics and in the latter mainly with the electrodics. Strictly, one should assign a separate position to high-frequency analysis, where not so much the ionic conductance but rather the dielectric and/or diamagnetic properties of the solution are playing a role. Nevertheless, we shall still consider this techniques as a special form of conductometry, because the capacitive and inductive properties of the solution show up versus high-frequency as a kind of AC resistance (impedance) and, therefore, as far as its reciprocal is concerned, as a kind of AC conductance. [Pg.28]

At this point, we should disclose a little secret. Just from the terminology, we may gather that control analysis involves quite a bit of mathematics, especially when we go over stability and frequency response methods. That is one reason why we delay introducing these topics. Nonetheless, we have to accept the prospect of working with mathematics. We would be lying if we say that one can be good in process control without sound mathematical skills. [Pg.8]

The second chapter by Peter Verveer and Quentin Hanley describes frequency domain FLIM and global analysis. While the frequency domain technique for fluorescence lifetime measurement is sometimes counterintuitive, the majority of the 10 most cited papers using FLIM have taken advantage of the frequency domain method as stated by these authors. The global analysis of lifetime data in the frequency domain, resolving both E and /d has contributed significantly to this advantage. [Pg.11]

K.Kulpa, A.Wojtkiewicz, M.Nalecz, J.Misiurewicz, The simple analysis method of nonlinear frequency distortions in FMCW radar , Journal of Telecommunications and Information Technology, No. 4, 2001, pp. 26-29. [Pg.241]

Another analysis method was based on the local wave vector estimation (LFE) approach applied on a field of coupled harmonic oscillators.39 Propagating media were assumed to be homogeneous and incompressible. MRE images of an agar gel with two different stiffnesses excited at 200 Hz were successfully simulated and compared very well to the experimental data. Shear stiffnesses of 19.5 and 1.2 kPa were found for the two parts of the gel. LFE-derived wave patterns in two dimensions were also calculated on a simulated brain phantom bearing a tumour-like zone and virtually excited at 100-400 Hz. Shear-stiffnesses ranging from 5.8 to 16 kPa were assumed. The tumour was better detected from the reconstructed elasticity images for an input excitation frequency of 0.4 kHz. [Pg.229]

As indicated in Section 4.1 (and as should be apparent from the discussion thus far in this chapter), titrimetric analysis methods heavily utilize solution chemistry, and therefore volumes of solutions are prepared, measured, transferred, and analyzed with some degree of frequency in this type of analysis. It should not be surprising that analytical laboratory workers need to be well versed in the selection and proper use of the glassware and devices used for precise volume measurement. [Pg.79]

Particle-Size Analysis Methods for particle-size analysis are shown in Fig. 17-34, and examples of size-analysis methods are given in Table 17-1. More detailed information may be found in Lapple, Chem. Eng., 75( 11), 140 (1968) Lapple, Particle-Size Analysis, in Encyclopedia of Science and Technology, 5th ed., McGraw-Hill, New York, 1982 Cadle, The Measurement of Airborne Particles, Wiley, New York, 1975 Lowell, Introduction to Powder Surface Area, 2d ed., Wiley, New York, 1993 and Allen, Particle Size Measurement, 4th ed, Chapman and Hall, London, 1990. Particle-size distribution may be presented on either a frequency or a cumulative basis the various methods are discussed in... [Pg.24]

For this purpose, we cleaned the lower than 1000 cm and higher than 1500 cm" areas by subtracting the Raman spectrum of the solvent (see Analysis Methods) in order to make some possible modifications of the sugar-phosphate part of DNA upon complexation conspicuous. Furthermore, we could expect to refine the nature of the weak interactions of SOAz with DNA b s by observing the low frequencies characteristic of these bases. [Pg.67]

The increase in time resolution of advanced sorption uptake methods and the joint use of sorption and radio-spectroscopic techniques allow for a more detailed analysis of the so-called "non-Fickian" behaviour of sorbing species in the intracrystalline bulk phase [18,28,29,76]. Correspondingly, information on molecular dynamics has been obtained for n-butane and 2-but ne in NFI zeolites by means of the single step frequency response method and C n.m.r. line-shape analysis [29]. As can be seen from Figures 4 and 5, the ad- / desorption for both sorbates proceeds very quickly, but with a... [Pg.204]

This method is by far the most widely used around the world and generally relies on the direct analytical determination of a key inhibitor component species, such as C1O4, Zn, Mo, PO4, SiC>3, or phosphonate. The degree of control is a function of the frequency and method of analysis, plus the variability in cooling system operation. If inhibitor and other testing is only carried out infrequently (say every 2 to 3 weeks) and only simple field test kits are employed for all analyses, it is unlikely that good long-term control can be satisfactorily achieved. [Pg.376]

Automatic analyzers have been used for many years to provide on-line analytical data in power houses, oil refineries, and other major industries. These take periodic water samples from a bypass line at a frequency ranging from every 2 minutes to almost continuously. These analyzers use conventional wet chemistry, colorimetric analysis methods and provide for... [Pg.377]

These results clearly indicate that the multi-frequency dynamic analysis method allows us to estimate the contribution of different relaxation mechanisms during curing of elastomers, and the changes in chemical and physical networks densities can be studied separately. [Pg.105]

Stomatal density and index show great potential for paleoelevation reconstructions with low error margins, if additional error sources such as the presence of sun and shade morphotypes and especially uncertainty in sea-level C02 concentrations can be well constrained. Unlike other paleobotanical methods, stomatal frequency analysis is not restricted to angiosperm dominated floras, and has no requirements for a minimum amount of taxa present. The method will be most reliable when applied to fossil taxa that are closely related to extant species, and suitable taxa are most likely to be found for periods when C02 concentrations were not much higher than ambient (380 ppmV). [Pg.235]

O EIEs of 1. 026-1.030 were computed from three experimentally determined frequencies that represent the most isotopically sensitive vibrational modes for side-on peroxide structures. This approach may be superior to the full frequency analysis specifically for the third row metals where the DFT method is less reliable. [Pg.439]


See other pages where Frequency analysis method is mentioned: [Pg.46]    [Pg.26]    [Pg.46]    [Pg.26]    [Pg.1582]    [Pg.39]    [Pg.226]    [Pg.13]    [Pg.241]    [Pg.75]    [Pg.179]    [Pg.250]    [Pg.1049]    [Pg.222]    [Pg.147]    [Pg.56]    [Pg.58]    [Pg.703]    [Pg.271]    [Pg.458]    [Pg.1105]    [Pg.1204]    [Pg.376]    [Pg.1468]    [Pg.379]    [Pg.389]    [Pg.365]    [Pg.265]    [Pg.112]   
See also in sourсe #XX -- [ Pg.25 , Pg.168 ]




SEARCH



© 2024 chempedia.info