Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential thermal analysis temperature calibration

If much well-ordered kaolinite is present, the assymmetric peaks are not prominent in the patterns from random samples, and the basal reflections are sharper and much enhanced in intensities in patterns from oriented samples. If much disordered kaolinite is present, the assymmetric peaks are prominent in the first patterns, and the basal reflections are much enhanced in the second. Chemical pretreatments prior to X-ray diffraction, such as those proposed by Wada [1965] and Alexiades and Jackson [1965], are sometimes useful in determining amounts of kaolinite and halloysite. Where the halloysite is tubular, it is easily detected in electron micrographs, although the amount can seldom be determined. Amounts of hydrated halloysite can be determined if allophane is not present in differential thermal analysis by calibrating and measuring the low-temperature endothermic peak. [Pg.150]

Differential Thermal Analysis. High temperature differential thermal analyses were obtained with a Dupont Model 1200 instrument. Samples were heated from room temperature to 950° C at a rate of 20°C/min in a slow stream of hydrogen. Molybdenum cups were used to hold the sample and alumina reference. The instrument was calibrated with sodium chloride (mp 800° C). [Pg.383]

Differential thermal analysis was performed with the DuPont 900 differential thermal analyzer the heating rate was usually 10°C. per minute. To determine heats of reaction, the calorimeter attachment to the Du Pont instrument was employed. Planimeter determinations of peak areas were converted to heat values by using standard calibration curves. For the infrared spectra either a Beckman IR5A instrument or a Perkin Elmer 521 spectrophotometer with a Barnes Engineering temperature-controlled chamber, maintained dry, was used. Specimens for infrared were examined, respectively, as Nujol mulls on a NaCl prism or as finely divided powders, sandwiched between two AgCl plates. For x-ray diffraction studies, the acid-soap samples were enclosed in a fine capillary. Exposures were 1.5 hours in standard Norelco equipment with Cu Ko radiation. For powder patterns the specimen-to-film distance was 57.3 mm. and, for long-spacing determinations, 156 mm. [Pg.76]

Frey s variant of the silvered vessel test has been in use in the Germany. In its variant, different amounts of heat are supplied to the electric heating elements mounted inside the Dewar flask, and the temperature differences between the interior of the Dewar vessel and the furnace are measured by thermocouples. A calibration curve is plotted from the values thus obtained, and the heat of decomposition of the propellant is read off the curve. In this way, the decomposition temperature at a constant storage temperature can be determined as a function of the storage time, and the heat of decomposition of the propellants can thus be compared with each other. If the measurements are performed at different storage temperatures, the temperature coefficient of the decomposition rate can be calculated. (-< also Differential Thermal Analysis.)... [Pg.284]

Using differential scanning calorimetry (DSC) (or, less directly, differential thermal analysis (DTA)) (see Section 2.8.5., above) it is possible to measure several of the thermodynamic properties of solids and of solid state reactions. The DSC response is directly proportional to the heat capacity, Cp, of the sample, so that by use of a calibrant it is possible to obtain values of this fundamental thermodynamic property, at a particular temperature, or as an average over a specified temperature range. Other thermodynamic properties are readily derived from such measurements ... [Pg.66]

The decomposition temperatures of hydrates were measured by means of differential thermal analysis (DTA) under the conditions of excess gas in a stainless steel flask that was developed specially for the investigation of hydrate formation with a gaseous guest at high hydrostatic pressure. The hydrate decomposition temperature was measured with a chromel-alumel thermocouple to the accuracy of 0.3 K. The thermocouple was calibrated with the use of temperature standards. Pressure was measured with a Bourdon-tube pressure gauge. The error of the pressure measurements did not exceed 0.5 %. This procedure was described in more detail previously.The gases used in the investigation... [Pg.554]

There are several different types of instrument covered under the term DSC, which have evolved from differential thermal analysis (DTA) and measure the temperature difference between sample and reference pans located in the same furnace. This is then converted to heat flow using a calibration factor. A detailed analysis of DSC requires consideration of the various sources of heat loss, and these are generally captured in the calibration routine for the instrument. Absolute temperature calibration is achieved through the use of pure indium (156.6 °C) and tin (231.9 °C) melting-point standards. A comprehensive analysis of the theory of DSC contrasted with DTA may be found in several reference works (Richardson, 1989, Gallagher, 1997). [Pg.197]

As with many other analytical techniques, the temperature axis used in differential thermal analysis (and DSC) must be calibrated with materials having known transition temperatures. The International Confederation of Thermal Analysis (ICTA) has been very active in developing a set of standard materials for this purpose (19) and has worked with the U.S. National Bureau of Standards to have these materials made commercially available (20). The U.S. National Bureau of Standards GM 754-GM 760 DTA temperature standards are listed in Table 6.2. They cover the temperature range from —83 to 925 C. The results of an ICTA round-robin study with 24 cooperating laboratories have been reported by Menis and Sterling (20). [Pg.309]

Thermal gravimetric analysis and differential thermal analysis (TGA/DTA) can be performed by using a SDT 2960 Simultaneous Differential Thermal Analyzer (TA Instruments, Inc., New Castle, DE). The instrument was calibrated with gold supplied by Perkin-Ehner. Samples (70 mg) of as-prepared powders were hand-pressed in a 3 mm dual action die and placed inside Pt sample cups and heated at the rates of 10 K/ min from ambient temperature to 1400°C. The reference material was used as a pellet of a-alumina. A flow of synthetic air at 50mL/min was maintained during the experiments. [Pg.494]

Test method for glass transition temperatures by differential scanning calorimetry or differential thermal analysis Standard practice for calibration of temperature scale for thermogravimetry... [Pg.106]

Differential Scanning Calorimetry. Differential scanning calorimetry is in some respects a modification of differential thermal analysis, and the schematic diagram of a differential scanning calorimeter (dsc) is similar to that of the dta device shown in Figure 12. The essential difference, however, is that the measured quantity here is the differential power supplied to the two wells, rather than the temperature difference. In other words, the dsc device maintains the same programmed temperature in each well and records the power required to achieve this. If a transition takes place in the sample, a characteristic excursion in the measured differential power is observed. The nature of these excursions can be related to the transitional properties of the sample. Furthermore, by proper calibration with a reference material of known thermal properties, the specific heat capacity of the sample may be obtained. [Pg.1172]

Standard Test Method for Assessing the Thermal Stability of Chemicals by Method of Differential Thermal analysis. (Specimens weight varies from 20 to 30 mg. The reference was an empty pan. The Heat Rate used was 10 °C/min) Standard Test Method for Temperature Calibration of Differential scanning Calorimeters and Differential Thermal Analyzers... [Pg.208]

Barshad, L, 1952. Temperature and heat of reaction calibration of the differential thermal analysis apparatus. Am, Mineral 37 667-694. [Pg.523]

Thermal Properties. The glass transition temperature (Tg) and the decomposition temperature (Td) were measured with a DuPont 910 Differential Scanning Calorimeter (DSC) calibrated with indium. The standard heating rate for all polymers was 10 °C/min. Thermogravimetric analysis (TGA) was performed on a DuPont 951 Thermogravimetric Analyzer at a heating rate of 20 °C/min. [Pg.157]

Instrumental. All Differential Scanning Calorimetry (DSC) runs were conducted on a Perkin-Elmer DSC-7 attached through a TAC-7 Thermal Analysis Controller to a DEC computer station 325c. All runs were conducted at 10°C/min. in N2 unless otherwise stated. Thermogravimetry was run on a Perkin Elmer TGA-7 attached through the same system as the DSC. Isothermal aging of samples in sealed capillaries were conducted in the DSC cell of a DuPont 9(X) Thermal Analyzer after temperature calibration. [Pg.170]

Differential scanning calorimetry (DSC) of PET and its nanocomposites was performed on a Perkin-Elmer DSC 7 thermal analysis system on typically 7 mg of material at a scanning speed of 10°C/min from room temperature to the melting point of the PET. Before evaluation, the thermal runs were subtracted similar runs of an empty pan. The DSC equipment was calibrated using indium as a standard. [Pg.1876]

The ability of the complementary Py-GC-MS technique to differentiate between linear and branched polysiloxanes has been investigated. Erzin and co-workers [56] investigated the ability of Py-GC-MS to detect trace levels of silicone polymer in recycled paper, and to differentiate between linear and branched polysiloxanes. The mass spectrometer had a mass range of 100-700 m/z and was operated by single ion monitoring and by selected ion data collection to enhance resolution and detection. A pyrolysis temperature of 750 °C was used. Silicone polymer scraped from the backing sheet was used for calibration. Thermal desorption GC-MS at 225 °C was used for the analysis of volatile components. It was concluded that concentrations of parts per million, and possibly as low as parts per billion could be measured. [Pg.207]


See other pages where Differential thermal analysis temperature calibration is mentioned: [Pg.551]    [Pg.141]    [Pg.211]    [Pg.45]    [Pg.612]    [Pg.534]    [Pg.218]    [Pg.682]    [Pg.193]    [Pg.143]    [Pg.4]    [Pg.181]    [Pg.418]    [Pg.101]    [Pg.92]    [Pg.3243]    [Pg.3011]    [Pg.278]    [Pg.91]    [Pg.346]    [Pg.418]    [Pg.829]    [Pg.159]    [Pg.198]    [Pg.697]    [Pg.455]    [Pg.386]    [Pg.533]    [Pg.712]    [Pg.129]    [Pg.116]   
See also in sourсe #XX -- [ Pg.309 ]

See also in sourсe #XX -- [ Pg.8 ]




SEARCH



Analysis temperature

Calibration analysis

Differential analysis

Differential temperature analysis

Differential thermal analysis temperature

Temperature calibration

Temperature differential

Thermal analysis temperature)

Thermal temperature

Thermalization temperature

© 2024 chempedia.info