Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector drift

They studied the effect of the mass detectors drift tube temperature on the low-molecular-mass TGs. Solutions of 10 mg/ml of tributyrin, tricaproin, tricaprylin, tricaprin, and trilaurin were injected twice at each of the following drift tube temperatures 20,25, 30,45, and 60°C. Five replications of the HPLC analysis were performed for one sample of ewe s milk fat to determine the reproducibility of the HPLC method. The TG composition was estimated in accordance with the method based on the calculation of the equivalent carbon numbers (ECNs) of the HPLC chromatographic peaks and in the molar composition in fatty acids, analyzed by GLC, collected at the HPLC chromatograph outlet. The HPLC fractions were collected every 40 s at the outlet of the column after 14 min there were no peaks before that time. [Pg.238]

The error terms might be correlated with each other or with an external parameter (i.e., time, injection sequence). It is always a good idea to plot the residuals vs. time and injection sequence to evaluate possible correlation effects. Detector drift, carry-over or other elfects can be easily detected with these plots. For a correct evaluation of the independence of the error, however, the calibration solutions should be injected in a randomized manner and correction for heteroskedasticity and/or curvilinearity should be applied on the calibration model. [Pg.146]

Two sets of factors that work in opposite directions are not considered in Eq. (2.19). The equation overestimates the FWHM by -/f, where F is the empirically observed Fano factor. The existence of this factor is attributed to the circumstance that the generated electrons do not necessarily act independently in producing the ionization pulse, and so the peak is narrower than when attributed to random events. On the other hand, detector drift, noise, and incomplete carrier collection each contributes to widening the FWHM (Knoll 1989). [Pg.36]

In summary, a gas chromatograph functions as follows. An inert carrier gas (like helium) flows continuously from a large gas cylinder through the injection port, the column, and the detector. The flow rate of the carrier gas is carefully controlled to ensure reproducible retention times and to minimize detector drift and noise. The sample is injected (usually with a microsyringe) into the heated injection port where it is vaporized and carried into the column, typically a capillary column 15 to 30 m long, coated on the inside with a thin (0.2 fim) film of high boiling liquid (the stationary phase). The sample partitions between the mobile and stationary phases, and is separated into individual components based on relative solubility in the liquid phase and relative vapor pressures. [Pg.14]

As all spectroscopists know and have observed, spectrometers do not always collect data with an ideal baseline. Due to a variety of problems (detector drift, changing environmental conditions such as temperature, spectrometer purge, sampling accessories, etc.), the baseline of a given spec-... [Pg.152]

A double-beam spectrometer is illustrated in Figure 16-9. The mirrors directing the interferometer beam through the sample and reference cells arc oscillated rapidly compared to the movement of the interferometer mirror so that sample and reference information can be obtained at each mirror position. The double-beam design compensates for source and detector drifts. [Pg.229]

Table 3. The Effect of Temperature Stabilisation on Detector Drift and Noise... Table 3. The Effect of Temperature Stabilisation on Detector Drift and Noise...
If the distance from the ion source to the detector is d, then the time (t) taken for an ion to traverse the drift tube is given by Equation 26.3. [Pg.190]

In this mode, ions are formed continuously in the ion source (a), but the electrostatic accelerating potential is applied in pulses (b). Thus, a sample of ions is drawn into the drift region (c) with more ions formed in the source. As shown in Figure 26.1, the ions separate according to m/z values (d) and arrive at the detector (e), the ions of largest m/z arriving last. [Pg.194]

Upon acceleration through an electric potential of V volts, ions of unknown m/z value reach a velocity v = f2zeV/m]" ). The ions continue at this velocity (drift) until they reach the detector. Since the start (to) and end (r) times are known, as is the length d of the drift region, the velocity can be calculated, and hence the m/z value can be calculated. In practice, an accurate measure of the distance d is not needed because it can be found by using ions of known m/z value to calibrate the system. Accurate measurement of the ion drift time is crucial. [Pg.220]

In a time-of-flight (TOF) mass spectrometer, ions formed in an ion source are extracted and accelerated to a high velocity by an electric field in an analyzer consisting of a long, straight drift tube. The ions pass along the tube until they reach a detector. [Pg.406]

In effect, the ions race each other along the drift tube, but the winners are always the ions of smallest m/z value, since these have the shortest flight times. The last to arrive at the detector are always those of greatest mass, which have the longest flight times. However, as in a race, for there to be a separation at the finish line (the detector), the ions must all start from the ion source at the same time (no handicapping allowed ). [Pg.406]

After acceleration through an electric field, ions pass (drift) along a straight length of analyzer under vacuum and reach a detector after a time that depends on the square root of their m/z values. The mass spectrum is a record of the abundances of ions and the times (converted to m/z) they have taken to traverse the analyzer. TOP mass spectrometry is valuable for its fast response time, especially for substances of high mass that have been ionized or selected in pulses. [Pg.407]

In time-of-flight (TOF) mass spectrometers, a pulse of ions is accelerated electrically at zero time. Having attained a maximum velocity, the ions drift along the flight tube of the analyzer. The times of arrival of ions at a detector are noted. [Pg.410]

X-rays are collected and analy2ed in ema in one of two ways. In wds, x-rays are dispersed by Bragg diffraction at a crystal and refocused onto a detector sitting on a Rowland circle. This arrangement is similar to the production of monochromati2ed x-rays for xps described above. In the other approach, edx, x-rays are all collected at the same time in a detector whose output scales with the energy of the x-ray (and hence, Z of the material which produces the x-ray.) Detectors used for ema today are almost exclusively Li-drifted Si soHd-state detectors. [Pg.285]

Germanium metal is also used in specially prepared Ge single crystals for y-ray detectors (54). Both the older hthium-drifted detectors and the newer intrinsic detectors, which do not have to be stored in hquid nitrogen, do an exceUent job of spectral analysis of y-radiation and are important analytical tools. Even more sensitive Ge detectors have been made using isotopicahy enriched Ge crystals. Most of these have been made from enriched Ge and have been used in neutrino studies (55—57). [Pg.281]

Development of a benchtop energy dispersive analyser BRA-18 is carrying out which is based on Si-drift detector and x-ray tube with side window range of the elements to be determined is extended from Mg to U. The distinctive feature of the device is that a specimen to be analysed is placed in the open air. [Pg.76]

The results of simulation have been confirmed by determination of Fe traces in quai tz sand, Cu and Mo in flotation tails and Ag in waste fixing waters on BRA-17-02 analyzer based on X-ray gas-filled electroluminescent detector and on BRA-18 analyzer based on Si-drift detector. The results of the simulation conform satisfactory with the experimental data in the mentioned cases the optimum filtration results in 2 to 5 times lowering of the detection limit. [Pg.134]

Fast concentration and sample injection are considered with the use of a theory of vibrational relaxation. A possibility to reduce a detection limit for trinitrotoluene to 10 g/cnf in less than 1 min is shown. Such a detection limit can by obtained using selective ionization combined with ion drift spectrometry. The time of detection in this case is 1- 3 s. A detection technique based on fluorescent reinforcing polymers, when the target molecules strongly quench fluorescence, holds much promise for developing fast detectors. [Pg.165]


See other pages where Detector drift is mentioned: [Pg.55]    [Pg.55]    [Pg.180]    [Pg.182]    [Pg.695]    [Pg.308]    [Pg.180]    [Pg.182]    [Pg.74]    [Pg.73]    [Pg.66]    [Pg.537]    [Pg.275]    [Pg.55]    [Pg.55]    [Pg.180]    [Pg.182]    [Pg.695]    [Pg.308]    [Pg.180]    [Pg.182]    [Pg.74]    [Pg.73]    [Pg.66]    [Pg.537]    [Pg.275]    [Pg.800]    [Pg.1353]    [Pg.1354]    [Pg.1781]    [Pg.1786]    [Pg.2084]    [Pg.189]    [Pg.190]    [Pg.193]    [Pg.291]    [Pg.194]    [Pg.57]    [Pg.368]    [Pg.382]    [Pg.11]    [Pg.123]    [Pg.125]    [Pg.358]    [Pg.552]   
See also in sourсe #XX -- [ Pg.116 , Pg.117 ]




SEARCH



Detector, lithium-drifted

Detectors lithium drifted silicon

Detectors noise and drift

Detectors, HPLC drift

Drift

Drift, HPLC detector response

Drifting

Lithium drifted germanium detectors

Silicon-drift-detectors

© 2024 chempedia.info