Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector issues

Because testing CWAs in the open air is not possible, potential interfering substances are tested in the laboratory at controlled exposure levels. The detector s ability to detect the CWA vapor is tested in combination with potential interfering vapors at the 0.1% and 1% headspace saturation concentration level of interferent vapor, providing that the interferent does not cause the detector to alarm. Such testing reveals whether the detector issues false positive or false negative results. [Pg.54]

Proceedings of the Workshop on Physics and Detector Issues for a High-Lurninosity Asymmetric B Factory at SLAC, D. Hitlin, ed., SLAC-373 (1991). [Pg.25]

S. Shapiro R. Morrison Pixel Detectors, General Vertex Detector Issues Oct 20,1989... [Pg.205]

Two six-month-long workshops, each with four meetings, were then held in 1990-91 to refine the physics objectives of the program and to identify the R D required to build a new detector capable of matching the physics requirements. These workshops were very well attended by a cross section of the worldwide community that has expressed interest in CP violation physics. Indeed, the Workshop report, Workshop on Physics and Detector Issues for a High-Luminosity Asymmetric B Factory at -SLAC (SLAC-373) [17], has become the Bible for those developing asymmetric B Factory detectors. [Pg.214]

Detection of cantilever displacement is another important issue in force microscope design. The first AFM instrument used an STM to monitor the movement of the cantilever—an extremely sensitive method. STM detection suffers from the disadvantage, however, that tip or cantilever contamination can affect the instrument s sensitivity, and that the topography of the cantilever may be incorporated into the data. The most coimnon methods in use today are optical, and are based either on the deflection of a laser beam [80], which has been bounced off the rear of the cantilever onto a position-sensitive detector (figme B 1.19.18), or on an interferometric principle [81]. [Pg.1693]

D. S. Simons. Int. J. Mass Spectrom. Ion Process. 55,15,1983. General discussion of the OMS technique and its applications. Contains a discussion of detector dynamic range issues. [Pg.597]

Several practical issues of the scatterometer must be considered in the case of characterizing nominally smooth surfaces. The incident laser beam may be collimated, but more commonly it is brought to a focus at a distance defined by the arc in which the detector rotates. In addition, a deflection mirror or an optical fiber might be used to direct light to the detector element. These features permit measurements close to the specular and transmitted beams, and this is critical to folly characterize the scattered light. This is especially significant since the scattered light intensity... [Pg.718]

Direct instrument control (or the lack of it) was an important issue for the earlier version of CDS. The scheme of connecting the detector channels through A/Ds to CDS worked well in analytical laboratories across the pharmaceutical industry. The scheme provided enough flexibility so that the CDS could collect data from a variety of instruments, including GC, HPLC, IC, SFC, and CE. It was equally important that the CDS could be connected to instruments that were manufactured by different vendors. It was not uncommon to find a variety of instruments from different vendors in a global pharmaceutical research company. The disadvantage of this scheme was that the instrument metadata could not be linked to the result file of each sample analyzed. It could not be guaranteed that the proper instrument parameters were used in sample analysis. Another need came from the increased use of... [Pg.56]

Absorbance detectors are also commonly used in combination with postcolumn reactors. Here, most issues of detector linearity and detection limit have to do with optimization of the performance of the reactor. In a typical application, organophosphorus compounds with weak optical absorbances have been separated, photolyzed to orthophosphate, and reacted with molybdic acid, with measurement being performed by optical absorbance.58... [Pg.18]

Concentration assays are often the least demanding, since usually the component to be measured is abundant and minor components scarce. Even if resolution is poor or there is detector noise, accurate measurements of concentration can still be obtained. In concentration assays, the principal requirements are stringency in the precision of sample dilution and measurement of column losses of the major component. Detector calibration, another important issue in concentration assays, has been discussed above. [Pg.27]

Detection is also frequently a key issue in polymer analysis, so much so that a section below is devoted to detectors. Only two detectors, the ultra-violet-visible spectrophotometer (UV-VIS) and the differential refractive index (DRI), are commonly in use as concentration-sensitive detectors in GPC. Many of the common polymer solvents absorb in the UV, so UV detection is the exception rather than the rule. Refractive index detectors have improved markedly in the last decade, but the limit of detection remains a common problem. Also, it is quite common that one component may have a positive RI response, while a second has a zero or negative response. This can be particularly problematic in co-polymer analysis. Although such problems can often be solved by changing or blending solvents, a third detector, the evaporative light-scattering detector, has found some favor. [Pg.333]

The detection of the migrating sample boundary in CE can be accomplished by UV, fluorescent, electrochemical, radiochemical, conductivity, and mass spectrometry (MS) means. The use of high-sensitivity detection systems is always a key issue in CE applications. The sensitivity of HPCE detectors may be at least 2 to 3 orders of magnitude better than that of HPLC detectors. Since the detection cell volume is very small, the concentration sensitivity... [Pg.397]


See other pages where Detector issues is mentioned: [Pg.35]    [Pg.179]    [Pg.63]    [Pg.180]    [Pg.457]    [Pg.35]    [Pg.179]    [Pg.63]    [Pg.180]    [Pg.457]    [Pg.364]    [Pg.1264]    [Pg.1632]    [Pg.231]    [Pg.212]    [Pg.420]    [Pg.92]    [Pg.337]    [Pg.255]    [Pg.294]    [Pg.17]    [Pg.18]    [Pg.19]    [Pg.59]    [Pg.68]    [Pg.205]    [Pg.17]    [Pg.828]    [Pg.21]    [Pg.33]    [Pg.45]    [Pg.74]    [Pg.333]    [Pg.375]    [Pg.376]   
See also in sourсe #XX -- [ Pg.62 , Pg.63 , Pg.64 , Pg.65 ]




SEARCH



Isotope detector issues

Single detector issues

© 2024 chempedia.info