Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detection calibration behavior

Cyclic voltammetry, square-wave voltammetry, and controlled potential electrolysis were used to study the electrochemical oxidation behavior of niclosamide at a glassy carbon electrode. The number of electrons transferred, the wave characteristics, the diffusion coefficient and reversibility of the reactions were investigated. Following optimization of voltammetric parameters, pH, and reproducibility, a linear calibration curve over the range 1 x 10 6 to 1 x 10 4 mol/dm3 niclosamide was achieved. The detection limit was found to be 8 x 10 7 mol/dm3. This voltammetric method was applied for the determination of niclosamide in tablets [33]. [Pg.83]

Once calibrated, the NIR analyzer was used to investigate a number of factors expected to affect the polymerization kinetics, including reaction temperature, initiator type, and initiator concentration (relative to monomer concentration). These experiments, in addition to improving process understanding, also mimicked the effects of inadequate process control during a reaction. Figure 15.1 shows the effect of reaction temperature on kinetics. The reaction rate nearly doubles when the temperature is raised from 65 to 75 °C, and the concentration of unreacted monomer after 85 minutes is reduced from 1.1 to 0.5%. In-hne NIR monitoring allows unusual behavior in either reaction rates or residual monomer levels to be detected and corrected immediately. [Pg.508]

Figure 2.3. Two experimental designs for modeling a potentially nonlinear system, (a) Two point calibration does not detect curvature (b) three point calibration can account for quadratic behavior. Figure 2.3. Two experimental designs for modeling a potentially nonlinear system, (a) Two point calibration does not detect curvature (b) three point calibration can account for quadratic behavior.
The ELS detector was previously also referred to as a mass detector, pointing to the fact that the response is (mainly) determined by the mass of the sample rather than by its chemical structure. Van der Meeren et al., though, demonstrated that the ELSD calibration curves of phospholipid classes were also dependent on the fatty acid composition (52). The dependence on the fatty acid composition is, however, completely different in nature and much less pronounced than for UV detection. The reason for this behavior is to be found in the partial resolution of molecular species, even during normal-phase chromatography. Thus, the peak shape depends not only on the chromatographic system but also on the fatty acid composition and molecular species distribution of the PL sample (47). Because it was shown before, based on both theoretical considerations and practical experiments, that the ELS detector response is generally inversely proportional to peak width (62,104), it follows that the molecular species distribution of the PL standards used should be similar to the sample components to be quantified. It was shown that up to 20% error may be induced if an inappropriate standard is used (52). [Pg.273]

Cao and Zeng [52] used of an oscillopolarographic method for the determination and the electrochemical behavior of omeprazole. Portions of standard omeprazole solution were treated with 1 ml 1 M ammonia/ ammonium chloride at pH 8.9 and the solution was diluted with water to 10 ml. The diluted solution was subjected to single sweep oscillopolaro-graphy with measurement of the derivative reduction peak at —1.105 V versus saturated calomel electrode. The calibration graph was linear from 0.5 to 10 /iM omeprazole with a detection limit of 0.2 fiM. The method was applied to the analysis of omeprazole in capsules with recoveries of 100-118.6% and RSD of 6.78%. The electrochemical behavior of omeprazole at the mercury electrode was also investigated. [Pg.213]

The electrochemical behavior of nimodipine was studied in ammonia buffer containing 10% (v/v) ethanol [8]. A single-sweep oscillopolaro-graphic method was then developed for nimodipine in tablets. The calibration graph (peak current at —0.73 V vs. concentration) was linear from 0.2 to 70 pM, and the detection limit was 10 pM. The same authors applied linear sweep voltammetry for the determination of nimodipine in tablets [9]. A reduction peak at —0.62V vs. the Ag/ACl reference... [Pg.360]

By comparing the size of the residuals of the test data with the residuals from the calibration samples, possible outlying behavior can be detected. For the three left-out samples, the sum-of-squared residuals are shown together with the sum-squared residuals obtained in the calibration model. For the calibration samples, the sum-of-squared residuals are given both as the fitted residuals and those obtained from leave-one-sample-out cross-validation. The fitted residuals are usually small because the model has been found specifically to minimize these. This makes them difficult to compare with residuals from new samples. By using residuals from leave-one-sample-out cross-validation, this overfit problem is minimized because the residuals from the calibration samples are obtained under similar conditions to the new samples. [Pg.270]

Figure 3. Net platelet retention behavior on polyethylene catheters in four goats. Thrombus platelet concentration catheter length was used to calibrate externally detected platelet retention curves. The repeatability of the initial slope ana the variable times to peak are shown. Key I, Goat 32 2, Goat 36 ... Figure 3. Net platelet retention behavior on polyethylene catheters in four goats. Thrombus platelet concentration catheter length was used to calibrate externally detected platelet retention curves. The repeatability of the initial slope ana the variable times to peak are shown. Key I, Goat 32 2, Goat 36 ...
The handbooks and operation manuals provided by tube suppliers usually give detailed information about the various reaction mechanisms [6-99]. Because of the chemical nature of detection mechanisms, with few exceptions they do not have absolute selectivity toward one single substance. In many cases, cross-sensitivities to chemically similar substances can be observed, which means that chemicals showing the same chemical reaction behavior will contribute to the indicated result As the calibration is performed only using the pure substance, the true concentration cannot be concluded from the indicated result if an interfering chemical is present in the air. Such kinds of cross-sensitivities are not easily foreseen. The true concentration of a contaminant is often lower than that indicated by the reading on the tube, which means the result is false positive. However, with... [Pg.271]

This chapter introduces the most important aspect of TEQA for the reader After the basics of what constitutes good laboratory practice are discussed, the concept of instrumental calibration is introduced and the mathematics used to establish such calibrations are developed. The uncertainty present in the interpolation of the calibration is then introduced. A comparison is made between the more conventional approach to determining instrument detection limits and the more contemporary approaches that have recently been discussed in the literature (1-6). These more contemporary approaches use least squares regression and incorporate relevant elements from statistics (7). Quality assurance/quality control principles are then introduced. The chapter ends with a comparison of the performance from two hypothetical labs. Every employer wants to hire an analyst who knows of and practices good laboratory behavior. [Pg.26]

Determination of molar masses and polydispersity of /polymers is often complicated because of the limited solubility particularly of highly fluorinated polymers. To avoid different conditions for the polymers and to ensure a certain comparison, SEC with a mixture of pentafluorophenol (PFP)/chloroform (1/3 vol/vol) as eluent (even if some of the samples were soluble in chloroform or tetrahydrofurane, THF), refractive index (RI) detection, and PMMA standards for calibration and determination of relative molar masses was employed. For some samples, molar masses could not be detected due to isorefractive behavior (no signal due to same RI of sample and eluent). This was observed in the series for P(MMA-co-sfMA-H10F10) at 20-25 mol% sfMA-HlOFlO and for P(MMA-co-sfMA-H2F8) at 20-30 mol% sfMA-H2F8. The chemical characterization of selected copolymers used for the present study is given in Table 11.2. [Pg.249]


See other pages where Detection calibration behavior is mentioned: [Pg.24]    [Pg.19]    [Pg.69]    [Pg.166]    [Pg.776]    [Pg.778]    [Pg.781]    [Pg.232]    [Pg.496]    [Pg.151]    [Pg.56]    [Pg.413]    [Pg.240]    [Pg.321]    [Pg.470]    [Pg.328]    [Pg.145]    [Pg.311]    [Pg.335]    [Pg.541]    [Pg.167]    [Pg.151]    [Pg.557]    [Pg.289]    [Pg.550]    [Pg.307]    [Pg.68]    [Pg.151]    [Pg.819]    [Pg.396]    [Pg.92]    [Pg.878]    [Pg.320]    [Pg.172]    [Pg.355]    [Pg.775]    [Pg.4345]    [Pg.589]   
See also in sourсe #XX -- [ Pg.2 , Pg.776 ]




SEARCH



Calibration Behavior in Charge Detection

© 2024 chempedia.info