Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Constant heat flow, Calorimeter with

Calorimeters with constant heat flow. Constant heat flow calorimeters are characterized by a constant temperature difference between the calorimetric vessel and the cover. To this group of calorimeters also belong the high-speed calorimeters for the measurement of heat capacities and the heats of modification transformation of substances, which are electrical conductors or semiconductors, where the heating is provided by their electrical resistance. [Pg.235]

As already indicated, Tian s equation supposes (1) that the temperature of the external boundary of the thermoelectric element 8e, and consequently of the heat sink, remains constant and (2) that the temperature Oi of the inner cell is uniform at all times. The first condition is reasonably well satisfied when the heat capacity of the heat sink is large and when the rate of the heat flux is small enough to avoid the accumulation of heat at the external boundary. The second condition, however, is physically impossible to satisfy since any heat evolution necessarily produces heat flows and temperature gradients. It is only in the case of slow thermal phenomena that the second condition underlying Tian s equation is approximately valid, i.e., that temperature gradients within the inner cell are low enough to be neglected. The evolution of many thermal phenomena is indeed slow with respect to the time constant of heat-flow calorimeters (Table II) and, in numerous cases, it has been shown that the Tian equation is valid (16). [Pg.210]

The home-made heat-flow calorimeter used consisted of a high vacuum line for adsorption measurements applying the volumetric method. This equipment comprised of a Pyrex glass, vacuum system including a sample holder, a dead volume, a dose volume, a U-tube manometer, and a thermostat (Figure 6.3). In the sample holder, the adsorbent (thermostated with 0.1% of temperature fluctuation) is in contact with a chromel-alumel thermocouple included in an amplifier circuit (amplification factor 10), and connected with an x-y plotter [3,31,34,49], The calibration of the calorimeter, that is, the determination of the constant, k, was performed using the data reported in the literature for the adsorption of NH3 at 300 K in a Na-X zeolite [51]. [Pg.286]

In a heat flow calorimeter, a feedback controller is used to maintain a constant desired reactor temperature by adjusting the jacket temperature. From (1), with a constant calibration probe heat flow, at steady state (dT/dt = 0), the overall heat transfer coefficient can be found from... [Pg.144]

RC measurements can be classified either as devices using jacketed vessels with control of the jacket temperature (heat balance calorimeters, heat flow calorimeters and temperature oscillation calorimeters) or as devices using a constant surrounding temperature, e.g., jacketed vessels with a constant jacket temperature, (isoperibolic calorimeters and power compensation calorimeters) such instruments may also feature single or double cells. [Pg.89]

Here designates the heat transferred from the bomb. Next, let us assume that the heat of reaction is determined in steady-state flow calorimeter with 1 = entering fluid and 2 = exiting fluid, and with = 0, AF = 0, and AK = 0. Then if the process takes place at constant pressure the general energy balance reduces to... [Pg.449]

Calorimeter with constant heat flow dQ/dt = constant, Tc = Tg = constant... [Pg.232]

Controlling the pressure and flow of a fuel gas alone cannot dehver a constant heat flow if the gas composition varies. The control problems presented by gas composition variations can be overcome by adding a hi-speed calorimeter to monitor heat value and by using a conventional flow orifice to measure flow to an existing burner control system. These two measurements, heat value and flow rate, in combination with the conventional control and feedback loop based on loan, provide the basis for effective, continuous feed forward control of the burning process. [Pg.136]

The value of AH can be determined experimentally by measuring the heat flow accompanying a reaction at constant pressure. When heat flows into or out of a substance, tire temperature of the substance changes. Experimentally, we can determine tire heat flow associated with a chemical reaction by measuring tire temperature change it produces. The measurement of heat flow is calorimetry an apparatus used to measure heat flow is a calorimeter. [Pg.169]

Scanning condition Tp = Tp(t) or Tm = TmW with Tp = constant Calorimeters involving the measurement of a temperature difference (heat flow calorimeters) or with a compensation of the thermal effect by thermoelectric effects (power compensation calorimeters). [Pg.92]

It is pertinent to note here again that in these heat flow calorimeters, the sample crucibles and their supports (or vessels) must govern the thermal behavior of the instrument in other words, their heat capacity and thermal resistance must be large compared with that of the sample and reference substance. This naturally affects the sensitivity of the calorimeter this is a factor to be considered in selecting the thermal resistance and thereby the time constant of the instrument. [Pg.193]

Thermochemistry is concerned with the study of thermal effects associated with phase changes, formation of chemical compouncls or solutions, and chemical reactions in general. The amount of heat (Q) liberated (or absorbed) is usually measured either in a batch-type bomb calorimeter at fixed volume or in a steady-flow calorimeter at constant pressure. Under these operating conditions, Q= Q, = AU (net change in the internal energy of the system) for the bomb calorimeter, while Q Qp = AH (net change in the enthalpy of the system) for the flow calorimeter. For a pure substance. [Pg.351]

As noted earlier, for a reaction at constant pressure, such as that taking place in an open coffee-cup calorimeter, the heat flow is equal to the change in enthalpy. If a reaction is carried out at constant volume (as is the case in a sealed bomb calorimeter) and there is no mechanical or electrical work involved, no work is done. Under these conditions, with w = 0, the heat flow is equal to the change in energy, AE. Hence we have... [Pg.216]

Figure 6-17 illustrates a constant-volume calorimeter of a type that is often used to measure q for combustion reactions. A sample of the substance to be burned is placed inside the sealed calorimeter in the presence of excess oxygen gas. When the sample bums, energy flows from the chemicals to the calorimeter. As in a constant-pressure calorimeter, the calorimeter is well insulated from its surroundings, so all the heat released by the chemicals is absorbed by the calorimeter. The temperature change of the calorimeter, with the calorimeter s heat capacity, gives the amount of heat released in the reaction. [Pg.393]

The measurement of an enthalpy change is based either on the law of conservation of energy or on the Newton and Stefan-Boltzmann laws for the rate of heat transfer. In the latter case, the heat flow between a sample and a heat sink maintained at isothermal conditions is measured. Most of these isoperibol heat flux calorimeters are of the twin type with two sample chambers, each surrounded by a thermopile linking it to a constant temperature metal block or another type of heat reservoir. A reaction is initiated in one sample chamber after obtaining a stable stationary state defining the baseline from the thermopiles. The other sample chamber acts as a reference. As the reaction proceeds, the thermopile measures the temperature difference between the sample chamber and the reference cell. The rate of heat flow between the calorimeter and its surroundings is proportional to the temperature difference between the sample and the heat sink and the total heat effect is proportional to the integrated area under the calorimetric peak. A calibration is thus... [Pg.313]

FIGURE 8.8 A calorimeter for measuring the heat flow in a reaction at constant pressure (AH). The reaction takes place inside an insulated vessel outfitted with a loose-fitting top, a thermometer, and a stirrer. Measuring the temperature change that accompanies the reaction makes it possible to calculate AH. [Pg.310]

In these calorimeters, the temperature of the surroundings is controlled and maintained constant or scanned. So the temperature of the sample is allowed to vary as well as the heat flow to the surroundings. Hence the results are more difficult to evaluate than with the techniques described above. Therefore these instruments are often semi-quantitative. [Pg.85]

The basic operation of the gaseous flow calorimeters is essentially identical to that of the flow-through solution-phase calorimeters with an external gas/vapour source that is passed, through a single calorimetric cell, across the solid of interest and the resulting heat change measured. For these instruments, the detectors are thermistors in direct contact with the solid under study. The form of the returned data is volts as a function of time. The signal can be converted to J s via a calibration constant. [Pg.113]

Isoperibolic calorimeters also comprise calorimeters based on the measurement of heat flow, since they fulfill the condition Tg = constant, changes. With these calorimeters, the temperature difference (Tc - Tg) is not measured but directly the heat flow between the calorimetric vessel and the cover. [Pg.233]

Solution calorimetry involves the measurement of heat flow when a compotmd dissolves into a solvent. There are two types of solution calorimeters, that is, isoperibol and isothermal. In the isoperibol technique, the heat change caused by the dissolution of the solute gives rise to a change in the temperature of the solution. This results in a temperature-time plot from which the heat of the solution is calculated. In contrast, in isothermal solution calorimetry (where, by definition, the temperature is maintained constant), any heat change is compensated by an equal, but opposite, energy change, which is then the heat of solution. The latest microsolution calorimeter can be used with 3-5 mg of compound. Experimentally, the sample is introduced into the equilibrated solvent system, and the heat flow is measured using a heat conduction calorimeter. [Pg.221]

The open-system calorimeter allows a constant flow to pass through a reaction tube surrounded by a water bath (Figure A5.2). The reaction is initiated and the apparatus is run until a steady state is reached, with the temperature in the centre of the reaction tube held equal to the inlet temperature by a control system (not. shown). The temperature drop across the reaction tube will thus be rendered negligible. The heat flow marked on the diagram is from the water bath to the reaction tube, in accordance once more with the sign convention of equation (13.34), although an exothermic reaction will cause the heat to flow out of the reaction tube and into the water. [Pg.349]

The heat-flow rate of the sample calorimeter, consisting of a pan and the sample, and the reference calorimeter, consisting usually of an empty pan, is governed by the rate of temperature change, q (in K min ), and the heat capacity, Cp (in 1K ). The heat capacity measured at constant pressure, p, and composition, n, can then be represented by Cp = (8H/8T)p , with H being the enthalpy and T, the temperature. The overall heat capacity of the sample calorimeter is written as = (mCp + Q), where m... [Pg.332]


See other pages where Constant heat flow, Calorimeter with is mentioned: [Pg.171]    [Pg.215]    [Pg.215]    [Pg.223]    [Pg.84]    [Pg.154]    [Pg.169]    [Pg.337]    [Pg.337]    [Pg.273]    [Pg.170]    [Pg.274]    [Pg.232]    [Pg.574]    [Pg.333]    [Pg.177]    [Pg.287]    [Pg.44]    [Pg.56]    [Pg.270]    [Pg.172]    [Pg.60]    [Pg.273]    [Pg.50]    [Pg.276]    [Pg.48]    [Pg.334]   
See also in sourсe #XX -- [ Pg.232 ]




SEARCH



Calorimeter constant

Calorimeter, flow

Calorimeters

Constants with

© 2024 chempedia.info