Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Constant calibration

Simultaneous Cr O " and MnOj" detemiination (Ewing, 1985) is a multivariate spectrophotometric analysis that requires detemiination of a matrix of four calibration constants, one for each unknown at each of two wavelengths. [Pg.83]

The ellipsometer user will always get data but unfortunately may not always know when the data or the results of analysis are correct. Improper optical alignment, bad calibration constants, reflection from the back surface of partially trans-... [Pg.409]

A variation of the Shepard Cane incorporates both electrodes on a single rod. A circuit diagram of this type of instrument is shown in Fig. 10.51. The single rod has a calibration constant, determined by the area of the positive electrode. At balance the resistivity is given by the resistance multiplied by the rod constant. This instrument is portable and simple to operate, but it cannot be recommended when accurate results are required. [Pg.253]

R.L. Bohon, AnalChem 35 (12), 1845-52 (1963) CA 60,1527 (1964) Approx heats of expin, Qv were detd on mg amounts of propints and expls by differential thermal analysis (DTA). Small-screw-cap metal cupsi sealed with a Cu washer served as constant vol sample containers the initial cup pressure could be controlled from 0 to approximately lOOOpsia. The calibration constant was calcd for each run from the total heat capacity of the cup and the relaxation curve, thereby compensating for equipment variations. [Pg.945]

When 0.113 g of benzene, C6H6, burns in excess oxygen in a calibrated constant-pressure calorimeter with a heat capacity of 551 J-(°C) I, the temperature of the calorimeter rises by 8.60°C. Write the thermochemical equation for... [Pg.361]

If a solution tends to be independent of shear, then the measurement of viscosity (q) is based on Poiseuille s law can be made easy by grouping all those terms related to a specific viscometer as a calibration constant A. [Pg.88]

An added benefit of the direct SEC-[n] calibration approach is that a new independent way of determining K and a values, using only broad MW standards, has also resulted. As few as three standards (or four, if all are narrow MWD) are needed to obtain both MW and [x]] calibration curves for a particular polymer-solvent system by using the broad-standard, linear calibration approach. From the experimental calibration constants of the two calibration curves, one can calculate K and a directly as described later. [Pg.92]

Upon applying the calibration constants obtained from the data of one sample to evaluate the SEC data of the other samples, the calculated Mn and values correspond fairly well (within 10-20%) with the absolute MW parameters of the samples. This agreement also suggests that the samples probably have similar chain structures. The distribution functions for samples PN-1 and IL-22 are plotted in Figures 4 and 5. The molecular weight distributions of both polymers are similar to distribution curves reported for derivatized poly(organo)phosphazenes (4-10). [Pg.252]

The CO2 signal was calibrated by comparing against the well-known amount of CO2 formation produced during oxidation of a saturated CO adlayer ( CO stripping ) [Jusys et al., 2001], determining the calibration constant K between the two signals via the relation... [Pg.416]

The mass spectrometric currents follow largely, but not completely the faradaic current signals. The contributions to the respective faradaic currents resulting from complete oxidation to CO2, which are calculated using the calibration constant K (see Section 13.2), are plotted as dashed lines in the top panels in Fig. 13.3. For the calculations of the partial reaction currents, we assumed six electrons per CO2 molecule formation and considered the shift in the potential scale caused by the time... [Pg.427]

The entity ot is the so-called isomer shift calibration constant, c is the speed of light, Co is the electric constant, and Eq is the nuclear transition energy. (The Coulomb constant k = l/(47rco), which was dropped in (4.1), is re-inserted here.) A comprehensive derivation of this expression is found in [8, 9]. [Pg.80]

A calibration of the popular B3LYP and BP86 density functionals for the prediction of Fe isomer shifts from DFT calculations [16], using a large number of complexes with a wide range of iron oxidation states and a span of about 2 mm s for the isomer shifts, yielded a value for the calibration constant a = —0.3666 mm s a.u. (see Chap. 5). Note the negative sign, which indicates that a positive isomer shift of a certain compound relative to a reference material reveals a lower electron density at the nuclei in that compound as compared to nuclei in the reference material. [Pg.81]

Based on the results of Trautwein and coworkers [67], the calibration constant carries a factor of 1.3 for compensating for the neglect of relativistic effects... [Pg.159]

Instrument calibration is done during the analysis of samples by interspersing standards among the samples. Following completion of the samples and standards, a linear calibration curve is estimated from the response of the standards using standard linear regression techniques. The calibration constants obtained from each run are used only for the samples quantitated in that run. Drastic changes or lack of linearity may indicate a problem with the detector. [Pg.359]

The formal transformation of the critical value into the sample domain is necessary to estimate correctly the limit of detection. If the sensitivity is known and without of any uncertainty (e.g. in case of error-free calibration constants fi), then the analytical value at CV is calculated by... [Pg.229]

In principle, AHvap can be obtained from other sources and (M) can be calculated from Equation (33) without any a priori assumptions. In reality there are always thermal leaks and the instruments are calibrated with a known standard. If AV oc AT is the measured voltage difference and K the instrument calibration constant, (M)n is obtained as [20]... [Pg.217]

The so-called Lupolen standard 25 is a well-known secondary standard in the field of SAXS. In conjunction with the Kratky camera it is easily used, because its slit-smeared intensity J(s) /V is constant over a fairly wide range, and this level is chosen as the calibration constant. In point-focus setups the SAXS of the Lupolen standard neither shows a constant intensity region, nor is the reported calibration constant of any use. [Pg.106]

A proper calibration constant for any beamline geometry is the invariant Q. Thus, the Lupolen standard or any other semicrystalline polymer that previously has been calibrated in the Kratky camera can be made a secondary standard for a point-focus setup, after its invariant Q has been computed in absolute units - based on a measurement of its SAXS in the Kratky camera. [Pg.106]

It may be required to have known quantities to which to compare the sample. These may be solutions with which the analyte reacts, or instrument readings or calibration constants obtained through known quantities. The analyte may also need to be physically or chemically separated from the sample matrix. [Pg.6]

It is required to have known quantities to which to compare the sample. This is the instrument readings or calibration constants obtained through known quantities. [Pg.153]

We also indicated in Section 6.2 that the calibration of an instrument for quantitative analysis can utilize a single standard, resulting in a calibration constant, or a series of standards, resulting in a standard curve. If a single standard is used, the value of K is the calibration constant. It is found by determining the instrument response for a standard solution of the analyte and then calculating K ... [Pg.159]

In the above equations, Rs is the readout for the standard solution, Cs is the concentration of the analyte in the standard solution, R(J is the readout for the unknown sample solution, C, is the concentration of the analyte in the unknown solution, and the proportionality constant (K) is the calibration constant. [Pg.159]


See other pages where Constant calibration is mentioned: [Pg.240]    [Pg.1395]    [Pg.1397]    [Pg.1411]    [Pg.684]    [Pg.309]    [Pg.99]    [Pg.103]    [Pg.104]    [Pg.1]    [Pg.80]    [Pg.160]    [Pg.354]    [Pg.544]    [Pg.544]    [Pg.545]    [Pg.545]    [Pg.428]    [Pg.435]    [Pg.504]    [Pg.282]    [Pg.173]    [Pg.273]    [Pg.58]    [Pg.303]    [Pg.109]    [Pg.152]   
See also in sourсe #XX -- [ Pg.51 , Pg.102 , Pg.171 , Pg.172 ]

See also in sourсe #XX -- [ Pg.54 , Pg.66 ]

See also in sourсe #XX -- [ Pg.551 ]

See also in sourсe #XX -- [ Pg.213 ]




SEARCH



Calibration constant determination

Calibration of the Spring Constant

Gauge calibration constant

Isomer shift calibration constant

© 2024 chempedia.info