Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detectors closed-cell

For closed-cell detectors, including the photoionization detector, the electron-capture detector and the TCD, extracolumn band broadening can be excessive unless specially designed devices with small cell volume are used or the detector is operated at subambient pressure. At reduced column outlet pressure, the carrier-gas velocity in the detector is increased, and the cell is swept out more quickly. Extra gas, called makeup gas, can be introduced into the detector cell to sweep the cell more rapidly and reduce peak broadening and distortion. [Pg.247]

Coded arrays were originally conceived for applications in X-ray imaging (Mertz and Young 1961 and Dicke 1968). A coded array is defined to be a pattern on a periodic two-dimensional lattice which associates with each lattice point a 0 or a 1 indicating whether the lattice point is open or closed . In coded-aperture imaging, the open and closed lattice points become open and closed cells in an opaque mask which casts a shadow of the photon source on a position-sensitive detector. For a review see Caroli et al. (1987). A uniformly redundant array, or URA (Fenimore and Cannon 1978), is a particular form of coded array. For a URA, each possible vector displacement between pairs of inequivalent open lattice points occurs a uniform number of times. (Equivalent lattice points are separated by a period of the array.)... [Pg.221]

The experimental set-up is given in figure 6. A closed reactor-detector system was used to enable detection of small mole fluxes. The stirred cell reactor is 0.10 m in diameter and was filled before each experiment with 120 ml of charged 2.0 M DIPA solution. The gas phase in the system was circulated by means of a flexible tube pump over a flow-through cell in a Perkin Elmer model 257 Infrared Grating Spectrophotometer for CO2 detection. Although spectrophotometers are not exceptionally well-suited for quantitative measurements, we preferred this type of analysis compared to gas chromatography for example because it does not influence the gas phase. [Pg.364]

As the average of the intensity of a modulated light beam is nonzero, the heat energy in the illuminated volume will rise continuously. Therefore, the temperature will slowly increase and the density will decrease until the heat deposition rate is equal to the loss rate due to heat conduction. This process is also governed by the heat-diffusion equation. For a closed cell the average density is constant therefore a pressure rise will occur. This DC component of the released heat power density changes the thermodynamic state of the material, in particular in very small PA detectors ( cm ). [Pg.654]

A radioactive source was transferred from one container to another by remote operation in a shielded cell. A radiation detector, interlocked with the cell door, prevented anyone from opening the cell door when radiation could be detected inside it. To make sure the interlock was working, an operator tried to open the cell door, by remote control, during a transfer. He found he could open it. He then found that the closing mechanism would not work. Fortunately he had not opened the door very far. [Pg.275]

FTIR spectra were collected with a Nicolet 740 spectrometer and a custom built in situ gas flow cell. The spectrometer was equipped with a MCT-B detector cooled by liquid nitrogen. Approximately 15 mg of the MgO catalyst sample was pressed into a self-supported disc and placed in a sample holder located at the center of the cell. The temperature in the cell was measured with a thermocouple placed close to the catalyst sample. Transmission spectra were collected in a single beam mode with a resolution of 2 cm 1. Prior to introduction... [Pg.390]

Therefore, the sensitivity of the measurement is greatly improved by decreasing the distance of the flow cell from the detector. Practically, the flow cell should be positioned as close as possible to the photosensitive area of the detector. [Pg.340]

Due to close proximity of the electrodes in the Hewlett-Packard HP 1049A Electrochemical Detector cell, little conductance is already sufficient for proper functioning of the electrode system. [Pg.49]

Figure 3.8 Amperometric detectors (a) measure the current that flows between the working electrode, usually a glassy carbon electrode, and a reference electrode, at a fixed voltage, usually close to the discharge potential for the compound. Coulometric detectors (b) are less common and are designed with a porous carbon flow cell so that all the analyte reacts in the cell, the amount of current consumed during the process being proportional to the amount of the substance. Figure 3.8 Amperometric detectors (a) measure the current that flows between the working electrode, usually a glassy carbon electrode, and a reference electrode, at a fixed voltage, usually close to the discharge potential for the compound. Coulometric detectors (b) are less common and are designed with a porous carbon flow cell so that all the analyte reacts in the cell, the amount of current consumed during the process being proportional to the amount of the substance.
Figure 16.1 Data recorded on a Marresearch (Marresearch GmbH, Hans Bockler-Ring 17, 22851 Norderstedt, Germany) amorphous Se detector from a crystal of BTV, shown in close up, to illustrate the low point spread and excellent separation of spots from the large cell (795 X 822 X 753 A ). X-ray source ESRFID14 EH1, wavelength 0.933 A. Figure 16.1 Data recorded on a Marresearch (Marresearch GmbH, Hans Bockler-Ring 17, 22851 Norderstedt, Germany) amorphous Se detector from a crystal of BTV, shown in close up, to illustrate the low point spread and excellent separation of spots from the large cell (795 X 822 X 753 A ). X-ray source ESRFID14 EH1, wavelength 0.933 A.
Figure 3.8 — (A) Biosensors used in different FI manifolds to perform reaction-rate measurements (I) stopped-flow manifold (II) iterative flow-reversal system (III) open-closed configuration S sample B buffer P pump IV injection valve PC personal computer IMEC immobilized enzyme cell D detector W waste SV switching valve. (B) Types of recordings obtained by using the three types of biosensors and measurements to be performed on them in order to develop reaction-rate methods. (Reproduced from [50] with permission of Elsevier Science Publishers). Figure 3.8 — (A) Biosensors used in different FI manifolds to perform reaction-rate measurements (I) stopped-flow manifold (II) iterative flow-reversal system (III) open-closed configuration S sample B buffer P pump IV injection valve PC personal computer IMEC immobilized enzyme cell D detector W waste SV switching valve. (B) Types of recordings obtained by using the three types of biosensors and measurements to be performed on them in order to develop reaction-rate methods. (Reproduced from [50] with permission of Elsevier Science Publishers).
The equipment required to develop this type of sensor is very simple and resembles closely that used to implement ordinary liquid-solid separations in FI manifolds. The only difference lies in the replacement of the packed reactor located in the transport-reaction zone with a packed (usually photometric or fluorimetric) flow-cell accommodated in the detector. Whether the packing material is inert or active, it should meet the following requirements (a) its particle diameter should be large enough (< 80-100 fim) to avoid overpressure (b) it should be made of a material compatible with the nature of the integrated detection system e.g. almost transparent for absorbance measurements) and, (c) the retention/elution process should be fast enough to avoid kinetic problems. [Pg.214]


See other pages where Detectors closed-cell is mentioned: [Pg.229]    [Pg.247]    [Pg.229]    [Pg.247]    [Pg.320]    [Pg.12]    [Pg.35]    [Pg.160]    [Pg.247]    [Pg.656]    [Pg.426]    [Pg.404]    [Pg.506]    [Pg.85]    [Pg.171]    [Pg.16]    [Pg.155]    [Pg.238]    [Pg.154]    [Pg.290]    [Pg.838]    [Pg.1011]    [Pg.240]    [Pg.296]    [Pg.500]    [Pg.49]    [Pg.53]    [Pg.196]    [Pg.337]    [Pg.452]    [Pg.175]    [Pg.176]    [Pg.520]    [Pg.66]    [Pg.105]    [Pg.158]    [Pg.390]    [Pg.409]    [Pg.447]   
See also in sourсe #XX -- [ Pg.247 ]




SEARCH



Closed cell

Detector cell

© 2024 chempedia.info