Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Length capillary

Limits of detection become a problem in capillary electrophoresis because the amounts of analyte that can be loaded into a capillary are extremely small. In a 20 p.m capillary, for example, there is 0.03 P-L/cm capillary length. This is 1/100 to 1/1000 of the volume typically loaded onto polyacrylamide or agarose gels. For trace analysis, a very small number of molecules may actually exist in the capillary after loading. To detect these small amounts of components, some on-line detectors have been developed which use conductivity, laser Doppler effects, or narrowly focused lasers (qv) to detect either absorbance or duorescence (47,48). The conductivity detector claims detection limits down to lO molecules. The laser absorbance detector has been used to measure some of the components in a single human cell (see Trace AND RESIDUE ANALYSIS). [Pg.183]

The physics underlying Eqs. (74-76) is quite simple. A solidifying front releases latent heat which diffuses away as expressed by Eq. (74) the need for heat conservation at the interface gives Eq. (75) Eq. (76) is the local equilibrium condition at the interface which takes into account the Gibbs-Thomson correction (see Eq. (54)) K is the two-dimensional curvature and d Q) is the so-called anisotropic capillary length with an assumed fourfold symmetry. [Pg.889]

Here 4) = 7 T Cp/L is a capillary length proportional to the isotropic part of the surface energy 7 0 is the angle between the normal n to the interface and some fixed crystallographic direction at which d Q) is minimal e is the strength of the anisotropy. [Pg.889]

The experimental setup for high-speed CZE can be seen in Figure 9.8. Highspeed CZE, or fast CZE (FCZE), yielded 70 000 to 90 000 theoretical plates for the separation of amino acid mixtures. Complete separation was achieved in under 11s, using a capillary length of 4 cm (24). [Pg.208]

Fig. 10.13 Liquid velocity and the location of the meniscus versus capillary length (J = 10 m, g = 9.8 m/s, = 200 W/m ). Reprinted from Yarin et al. (2002) with permission... Fig. 10.13 Liquid velocity and the location of the meniscus versus capillary length (J = 10 m, g = 9.8 m/s, = 200 W/m ). Reprinted from Yarin et al. (2002) with permission...
Large-scale numerical simulation for samples that are many times os large as the critical wavelength is perhaps the only way to develop a quantitative understanding of the dynamics of solidification systems. Even for shallow cells, such calculations will be costly, because of the fine discretizations needed to be sure the dynamics associated with the small capillary length scales are adequately approximated. Such calculations may be feasible with the next generation of supercomputers. [Pg.329]

In order to simplify the analysis, we will consider the capillary flow of a hquid L in a horizontal small tube (diameter much smaller than the capillary length), in order to avoid complications due to gravity effects (Fig. 14). [Pg.310]

OS 33] ]R 16h] [P 25] For the nitration of a single-ring aromatic in a capillary-flow micro reactor, experiments were performed at two temperature levels, 60 and 120 °C [94]. Owing to the assumed increase in conversion rate with higher temperature, attempts were made to compensate for this by decreasing the capillary length at otherwise constant dimensions. For the 60 °C experiment, a very low... [Pg.455]

The shear history simulators operate at a single shear rate during an experiment and do not run shear ramps. For these Instruments, apparent viscosity at a single shear rate Is determined by the relationship of differential pressure (AP), capillary length (L) and radius (r), and volumetric flow rates (Q), as follows. [Pg.110]

Groendes and Mesler (1982) studied the saturated film boiling impacts of a 4.7 mm water droplet on a quartz surface of 460 °C. The fluctuation of the surface temperature was detected using a fast-response thermometer. The maximal temperature drop of the solid surface during a droplet impact was reported to be about 20 °C. Considering the lower thermal diffusivity of quartz, this temperature drop implies a low heat-transfer rate on the surface. Biance et al. (2003) studied the steady-state evaporation of the water droplet on a superheated surface and found that for the nonwetting contact condition, the droplet size cannot exceed the capillary length. [Pg.29]

The migration time increases with increasing capillary length and reduces with increasing mobilities of the compounds and applied voltages. [Pg.583]

An electrophoretic method was developed for the simultaneous determination of artificial sweeteners, preservatives and colours in soft drinks. The samples were degassed by sonication, filtered and used for analysis without any other pretreatment. Measurements were realized in uncoated fused-silica capillaries, the internal diameter being 50 ptm. Capillary lengths were 48.5 cm (40 cm to the detector) and 65.4 cm (56 cm to the detector). Capillaries were conditioned by washing them with (1 M sodium hydroxide (10 min), followed by 0.1 M sodium hydroxide (5 min) and water (5 min). Samples were injected hydrodinamically (250 mbar) at the anodic end. Analyses were performed at a voltage of 20 kV and the capillary temperature was 25°C. Analytes having ionizable substructure... [Pg.516]

Selection of capillary length is dictated by the type of capillary used and the required resolution. When using coated capillaries with insignificant EOF, separations can be achieved with relatively short capillaries of 20 to 30 cm effective length... [Pg.182]

Capillary length is one of the parameters commonly used to improve resolution, but in our experience, more dramatic effects can be accomplished by changing the mobility of the sample components through manipulation of pH or the use of buffer additives. In fact, we usually develop a method in the shortest capillary possible, and only when the separation is adequate do we increase the capillary length for final optimization. This practice saves time in methods development because more data can be collected if the analysis time is short. [Pg.183]

A,/V,/V, /V -tctramethylethylcncdiarninc (TEMED) can be used to block the distal section of the capillary. As a rule of thumb, the ratio of TEMED concentration (% v/v) to ampholyte concentration should be approximately equal to the ratio of the noneffective (or blind) capillary length to total length. [Pg.194]

The low level of EOF in coated capillaries permits separations to be carried out with very short effective capillary lengths. Earlier work using chemical mobilization was performed using capillaries as short as 11 cm with internal... [Pg.197]

The length of capillaries used for single-step CIEF is very important, especially when EOF is the driving force. The capillary length must be optimized according to the size of the injection and the velocity of EOF or flow due to hydraulic forces so that the sample will not reach the detection point before it has finished focusing. [Pg.198]

Some special approaches defined in the CE enantioseparation field can be found in References 37 and 38. In Reference 37, the authors tested four approaches to decrease the analysis time of a previously developed chiral separation method for amphetamine and its related compounds. The considered possibilities were (i) the short-end injection technique or (ii) increased electrical field combined with a capillary length reduction. [Pg.188]

Injection plug should be 1—2% of total capillary length Re-cut capillary, laser cuts are best. Follow procedure In reference 3... [Pg.56]


See other pages where Length capillary is mentioned: [Pg.12]    [Pg.363]    [Pg.379]    [Pg.754]    [Pg.756]    [Pg.891]    [Pg.901]    [Pg.442]    [Pg.111]    [Pg.304]    [Pg.303]    [Pg.451]    [Pg.456]    [Pg.460]    [Pg.460]    [Pg.476]    [Pg.477]    [Pg.104]    [Pg.620]    [Pg.75]    [Pg.193]    [Pg.198]    [Pg.216]    [Pg.62]    [Pg.332]    [Pg.22]    [Pg.39]    [Pg.45]    [Pg.46]    [Pg.47]    [Pg.48]   
See also in sourсe #XX -- [ Pg.12 , Pg.363 ]

See also in sourсe #XX -- [ Pg.14 , Pg.107 , Pg.118 ]

See also in sourсe #XX -- [ Pg.316 , Pg.329 ]

See also in sourсe #XX -- [ Pg.7 ]

See also in sourсe #XX -- [ Pg.14 ]

See also in sourсe #XX -- [ Pg.33 , Pg.77 ]

See also in sourсe #XX -- [ Pg.189 , Pg.270 , Pg.271 ]

See also in sourсe #XX -- [ Pg.198 , Pg.212 ]

See also in sourсe #XX -- [ Pg.23 , Pg.106 , Pg.122 , Pg.152 , Pg.169 , Pg.173 , Pg.181 ]

See also in sourсe #XX -- [ Pg.143 ]

See also in sourсe #XX -- [ Pg.140 ]

See also in sourсe #XX -- [ Pg.84 ]




SEARCH



Capillaries maximum allowed length

Capillary Length 1 the Role of Gravity

Capillary columns length

Extended Path-Length Capillaries

Generalized capillary length

The Capillary Length

© 2024 chempedia.info