Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration set

The PLS calibration set was built mixing in an agate mortar different amounts of Mancozeb standard with kaolin, a coadjuvant usually formulated in agrochemicals. Cluster analysis was employed for sample classification and to select the adequate PLS model acording with the characteristics of the sample matrix and the presence of other components. [Pg.93]

In order to evaluate possible classes among samples considered, a clustering analysis was carried out before PFS treatment for selecting properly a reduced but well representative calibration set. [Pg.142]

The standard requires the supplier to safeguard inspection, measuring, and test facilities including both test hardware and test software from adjustments which would invalidate the calibration setting. [Pg.420]

How do you safeguard test facilities from adjustments that would invalidate the calibration setting ... [Pg.424]

The primary reason for die popularity of the Pople and DH style basis sets is the extensive calibration available. There have been so many calculations reported with these basis sets that it is possible to get a fairly good idea of the level of accuracy that can be attained with a given basis. This is of course a self-sustaining procedure, the more calculations that are reported with a given basis, tlie more popular it becomes, since the calibration set becomes larger and larger. [Pg.161]

The CHI can be obtained without preliminary method development direcdy from a single fast-gradient run with a cycle time less than 15 min with a 150-mm column [40] or 5 min with 50-mm column [42]. In this case, the obtained retention time, tr, is expressed within an organic phase concentration (< o) scale using a calibration set of compounds. CHI value can be obtained from ... [Pg.342]

In case of fast gradient (below 15 min), S could be considered constant for all the investigated molecules and wiU only have a small influence on the retention time of the compounds. Thus, the gradient retention times, of a calibration set of compounds are linearly related to the ( )o values [39]. Moreover, Valko et al. also demonstrated that the faster the gradient was, the better the correlation between t, and < )o [40]. Once the regression model was established for the calibration standards, Eq. 8 allowed the conversion of gradient retention times to CHI values for any compound in the same gradient system. Results are then suitable for interlaboratory comparison and database construction. The CH I scale (between 0 and 100) can be used as an independent measure of lipophilicity or also easily converted to a log P scale. [Pg.342]

In 1978, Ho et al. [33] published an algorithm for rank annihilation factor analysis. The procedure requires two bilinear data sets, a calibration standard set Xj and a sample set X . The calibration set is obtained by measuring a standard mixture which contains known amounts of the analytes of interest. The sample set contains the measurements of the sample in which the analytes have to be quantified. Let us assume that we are only interested in one analyte. By a PCA we obtain the rank R of the data matrix X which is theoretically equal to 1 + n, where rt is the number of interfering compounds. Because the calibration set contains only one compound, its rank R is equal to one. [Pg.298]

From now on, we adopt a notation that reflects the chemical nature of the data, rather than the statistical nature. Let us assume one attempts to analyze a solution containing p components using UV-VIS transmission spectroscopy. There are n calibration samples ( standards ), hence n spectra. The spectra are recorded at q wavelengths ( sensors ), digitized and collected in an nx.q matrix S. The information on the known concentrations of the chemical constituents in the calibration set is stored in an nxp matrix C. Each column of C contains the concentrations of one of the p analytes, each row the concentrations of the analytes for a particular calibration standard. [Pg.353]

The CLS method hinges on accurately modelling the calibration spectra as a weighted sum of the spectral contributions of the individual analytes. For this to work the concentrations of all the constituents in the calibration set have to be known. The implication is that constituents not of direct interest should be modelled as well and their concentrations should be under control in the calibration experiment. Unexpected constituents, physical interferents, non-linearities of the spectral responses or interaction between the various components all invalidate the simple additive, linear model underlying controlled calibration and classical least squares estimation. [Pg.356]

We chose the number of PCs in the PCR calibration model rather casually. It is, however, one of the most consequential decisions to be made during modelling. One should take great care not to overfit, i.e. using too many PCs. When all PCs are used one can fit exactly all measured X-contents in the calibration set. Perfect as it may look, it is disastrous for future prediction. All random errors in the calibration set and all interfering phenomena have been described exactly for the calibration set and have become part of the predictive model. However, all one needs is a description of the systematic variation in the calibration data, not the... [Pg.363]

Van der Voet [21] advocates the use of a randomization test (cf. Section 12.3) to choose among different models. Under the hypothesis of equivalent prediction performance of two models, A and B, the errors obtained with these two models come from one and the same distribution. It is then allowed to exchange the observed errors, and c,b, for the ith sample that are associated with the two models. In the randomization test this is actually done in half of the cases. For each object i the two residuals are swapped or not, each with a probability 0.5. Thus, for all objects in the calibration set about half will retain the original residuals, for the other half they are exchanged. One now computes the error sum of squares for each of the two sets of residuals, and from that the ratio F = SSE/JSSE. Repeating the process some 100-2(K) times yields a distribution of such F-ratios, which serves as a reference distribution for the actually observed F-ratio. When for instance the observed ratio lies in the extreme higher tail of the simulated distribution one may... [Pg.370]

There are two points of view to take into account when setting up a trmning set for developing a predictive multivariate calibration model. One viewpoint is that the calibration set should be representative for the population for which future predictions are to be made. This will generally lead to a distribution of objects in experimental space that has a higher density towards the center, tailing out to the boundaries. Another consideration is that it is better to spread the samples more or... [Pg.371]

Several approaches have been investigated recently to achieve this multivariate calibration transfer. All of these require that a small set of transfer samples is measured on all instruments involved. Usually, this is a small subset of the larger calibration set that has been measured on the parent instrument A. Let Z indicate the set of spectra for the transfer set, X the full set of spectra measured on the parent instrument and a suffix Aor B the instrument on which the spectra were obtained. The oldest approach to the calibration transfer problem is to apply the calibration model, b, developed for the parent instrument A using a large calibration set (X ), to the spectra of the transfer set obtained on each instrument, i.e. and Zg. One then regresses the predictions (=Z b ) obtained for the parent instrument on those for the child instrument yg (=Z b ), giving... [Pg.376]

For the extraction of rubber and rubber compounds a wide variety of solvents (ethyl acetate, acetone, toluene, chloroform, carbon tetrachloride, hexane) have been used [149]. Soxtec extraction has also been used for HDPE/(Tinuvin 770, Chimassorb 944) [114] and has been compared to ultrasonic extraction, room temperature diffusion, dissolution/precipitation and reflux extraction. The relatively poor performance of the Soxtec extraction (50% after 4h in DCM) as compared with the reflux extraction (95% after 2-4 h in toluene at 60 °C) was described to the large difference in temperature between the boiling solvents. Soxtec was also used to extract oil finish from synthetic polymer yam (calibration set range of 0.18-0.33 %, standard error 0.015 %) as reference data for NIRS method development [150]. [Pg.72]

Especially in the case of biochemical and environmental systems and generally in ultra trace analysis, SAM is frequently applied. By addition of standard solutions to the sample a similar behaviour of the calibration set and the sample is created provided that the analyte is added in form of... [Pg.172]

A very limited version of this question, does in fact, sometimes appear, when the question arises of how many samples from a given calibration set to keep in reserve for... [Pg.135]

The other obvious extension, which is more useful for the case where you may still have to measure samples with the same characteristics as the old ones, is to simply keep adding to and expanding the calibration set as new samples become available. The new samples then not only allow you to test for robustness, but inclusion of such samples will actually make the calibration more robust. I think we all know this intuitively, but I have also been able to prove this mathematically. [Pg.137]

The technique is not optimal the instrument response (Y) is a predictor of analyte values (X). The limitation for modeling is in the representation of calibration set chemistry, sample presentation, and unknown variations of instrument and operator during measurement. [Pg.381]

In this chapter as a continuation of Chapters 58 and 59 [1, 2], the confidence limits for the correlation coefficient are calculated for a user-selected confidence level. The user selects the test correlation coefficient, the number of samples in the calibration set, and the confidence level. A MathCad Worksheet ( MathSoft Engineering Education, Inc., 101 Main Street, Cambridge, MA 02142-1521) is used to calculate the z-statistic for the lower and upper limits and computes the appropriate correlation for the z-statistic. The upper and lower confidence limits are displayed. The Worksheet also contains the tabular calculations for any set of correlation coefficients (given as p). A graphic showing the general case entered for the table is also displayed. [Pg.393]

Figure 68-1 (a) Artificial data representing a linear relationship between the two variables. This data represents a linear, one-variable calibration, (b) The same artificial data extended in a linear manner. The extrapolated calibration line (broken line) can predict the data beyond the range of the original calibration set with equivalent accuracy, as long as the data itself is linear. [Pg.466]

Again, however, if there is perfect linearity in the relationship of the absorbance at both wavelengths with respect to the concentrations of the components, one should equally well be able to extrapolate the model beyond the range of either or both components in the calibration set, just as in the univariate case. [Pg.467]

The Mahalanobis Distance statistic provides a useful indication of the first type of extrapolation. For the calibration set, one sample will have a maximum Mahalanobis Distance, Z) ax. This is the most extreme sample in the calibration set, in that, it is the farthest from the center of the space defined by the spectral variables. If the Mahalanobis Distance for an unknown sample is greater than ZTax, then the estimate for the sample clearly represents an extrapolation of the model. Provided that outliers have been eliminated during the calibration, the distribution of Mahalanobis Distances should be representative of the calibration model, and ZEax can be used as an indication of extrapolation. [Pg.499]

Ensure that the responses from samples are close to the mean response, y, of the calibration set. This will decrease the error contribution from the least-squares estimate of the regression line. [Pg.89]


See other pages where Calibration set is mentioned: [Pg.607]    [Pg.62]    [Pg.172]    [Pg.189]    [Pg.342]    [Pg.238]    [Pg.349]    [Pg.350]    [Pg.357]    [Pg.363]    [Pg.371]    [Pg.372]    [Pg.374]    [Pg.374]    [Pg.376]    [Pg.378]    [Pg.52]    [Pg.704]    [Pg.383]    [Pg.382]    [Pg.316]    [Pg.179]   
See also in sourсe #XX -- [ Pg.140 , Pg.153 , Pg.179 , Pg.201 , Pg.211 , Pg.223 , Pg.273 , Pg.274 , Pg.317 , Pg.369 , Pg.378 , Pg.404 , Pg.409 ]

See also in sourсe #XX -- [ Pg.265 ]




SEARCH



Calibration data set

General Guidelines for Calibration Data Sets

Number of samples in the calibration set

© 2024 chempedia.info