Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Follower current

In the voltammograms in figures 11.33a and 11.33b, the current is monitored as a function of the applied potential. Alternatively, the change in current following... [Pg.513]

Ion Channels. The excitable cell maintains an asymmetric distribution across both the plasma membrane, defining the extracellular and intracellular environments, as well as the intracellular membranes which define the cellular organelles. This maintained a symmetric distribution of ions serves two principal objectives. It contributes to the generation and maintenance of a potential gradient and the subsequent generation of electrical currents following appropriate stimulation. Moreover, it permits the ions themselves to serve as cellular messengers to link membrane excitation and cellular... [Pg.279]

This defines the maximum permissible composite error at the rated accuracy limit primary current, followed by letter P for protection. The standard prescribed accuracy... [Pg.477]

Two platinum electrodes are immersed in sulphuric acid of suitable concentration containing the nitrate ion to be determined and a potential of about 100 millivolts is applied. Upon titration with 0.4M ammonium iron(II) sulphate solution there is an initial rise in current followed by a gradual fall, with a marked increase at the end point the latter is easily determined from a plot of current against volume of iron solution added. The concentration of water should not be allowed to rise above 25 per cent (w/w). The temperature of the solution should not exceed 40 °C. [Pg.636]

The mass spectrometric currents follow largely, but not completely the faradaic current signals. The contributions to the respective faradaic currents resulting from complete oxidation to CO2, which are calculated using the calibration constant K (see Section 13.2), are plotted as dashed lines in the top panels in Fig. 13.3. For the calculations of the partial reaction currents, we assumed six electrons per CO2 molecule formation and considered the shift in the potential scale caused by the time... [Pg.427]

The electrochemical responses of the 100 electrodes in the array were measured simultaneously using a potentiostat, two 64-channel current followers, and two data... [Pg.574]

Figure 16.4 Electrochemical screening instrumentation, consisting of a wave generator, a potentiostat, a current follower, and a PC with a data acquisition card [Guerin et al., 2004]. Figure 16.4 Electrochemical screening instrumentation, consisting of a wave generator, a potentiostat, a current follower, and a PC with a data acquisition card [Guerin et al., 2004].
The quantity k3 may be considered as an instrumental constant to be determined in a blank experiment—that is, without added solute. In this case, the current is given by I(t)/I(0) = (1 - vt/d) exp( - k3 t), from which k can be determined. With the solute added, the current initially decays exponentially (fast decay) from which is determined h + k2 + k3, while the ratio of the initial plateau to the initial current gives k2/(k] + k2 + k ). The detachment rate k2 is now obtained from the last two numbers, and then the attachment rate fe, is also obtained since k3 is already predetermined. In short, both attachment (kj and detachment (k2) rates are obtainable from the time dependence of the cell current following a brief pulse of ionizing radiation. [Pg.349]

The third set of models is for understanding the actual currents, and the pathways that the currents follow through molecular transport junctions. This is to some... [Pg.11]

Various other methods devised for measuring frequency are all based on the fact that the bubble creates a disturbance in a continuously occurring phenomenon. This distrubance can be utilized as an obstruction to the passage of light, x-rays, or y-rays or as a disturbance in the noise pattern or in the flow of current. Following are some of the methods which make use of the above ideas. [Pg.263]

For times t> t, where the reoxidation of the previously generated species Red takes place, the diagram represents the double potential step response. Under these conditions, the anodic current follows the equation ... [Pg.125]

ET much slower than transport (ET control). The current follows the But-ler-Volmer equation (2)... [Pg.5]

Fig. 5.43 Voltage control circuit with the aid of an operational amplifier (a) and a circuit of potentiostat (b). VF voltage follower CF current follower. Fig. 5.43 Voltage control circuit with the aid of an operational amplifier (a) and a circuit of potentiostat (b). VF voltage follower CF current follower.
Fig. 30. Classical operational amplifier circuits, (a) A current follower (b) a current integrator. Fig. 30. Classical operational amplifier circuits, (a) A current follower (b) a current integrator.
The signal Vu, for compensation of iRu, must be proportional to this quantity and identical in phase. A convenient source of Vu is the output of the current follower, OA3, of Figure 7.1 A. Adjustment of the magnitude is provided by a potentiometer so that some fraction F of iRu is compensated. Because of the wide range of resistor values that may be used in the circuit, Vu will not generally be equal to F(iRu), but will be scaled according to the equation... [Pg.218]

Accepting these results for an idealized system confirms the intuitive notion that the use of positive feedback from the current follower is identical to physically moving the reference electrode closer to the working electrode. There are no additional concepts to cope with in relating response to physical parameters. It will become evident that this situation allows a rather simple assessment of loop gain and a clear view of corrective measures that will lead to stability with Ru compensation. [Pg.221]

Figure 8.3 Current-follower circuit for resistance measurement (a) e0 = -ERr/Ru (b) e0 = -ERU/Rr. Figure 8.3 Current-follower circuit for resistance measurement (a) e0 = -ERr/Ru (b) e0 = -ERU/Rr.
The pulses are provided by a precision bipolar voltage source, which is switched into the input of the pulsing amplifier by the switch at point A in the circuit. A very accurate crystal-controlled timing circuit (not shown) drives the switch to ensure that the pulses are symmetrical. The pulsing amplifier inverts the signal as shown by waveform B and supplies current to the cell. The cell current is amplified by the current follower, the output of which is illustrated by waveform C. [Pg.261]

The small area of a microelectrode, with its proportionately low capacitance, allows its use at very short time scales compared to the time scale used with a classical voltammetric electrode. As we have seen earlier in this chapter, when microelectrodes are used at short time scales, the current follows the behavior expected for diffusion in one dimension. Thus, the development of high-speed voltammetric methods with microelectrodes was a logical step, and has greatly expanded the scope and capabilities of electrochemical techniques [41]. Rapid electrochemical methods allow evaluation of the larger rate constants of rapid heterogeneous and/or homogeneous reactions. For example, theories of hetero-... [Pg.381]


See other pages where Follower current is mentioned: [Pg.91]    [Pg.938]    [Pg.105]    [Pg.592]    [Pg.575]    [Pg.576]    [Pg.178]    [Pg.432]    [Pg.288]    [Pg.132]    [Pg.101]    [Pg.167]    [Pg.526]    [Pg.136]    [Pg.265]    [Pg.21]    [Pg.108]    [Pg.159]    [Pg.160]    [Pg.161]    [Pg.158]    [Pg.164]    [Pg.240]    [Pg.1272]    [Pg.175]    [Pg.176]    [Pg.184]    [Pg.245]    [Pg.256]    [Pg.259]   
See also in sourсe #XX -- [ Pg.134 ]

See also in sourсe #XX -- [ Pg.145 ]

See also in sourсe #XX -- [ Pg.62 , Pg.63 , Pg.66 , Pg.67 ]

See also in sourсe #XX -- [ Pg.65 ]

See also in sourсe #XX -- [ Pg.101 ]




SEARCH



© 2024 chempedia.info