Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration and analysis

Anionic and neutral polymers are usually analyzed successfully on Syn-Chropak GPC columns because they have minimal interaction with the appropriate mobile-phase selection however, cationic polymers adsorb to these columns, often irreversibly. Mobile-phase selection for hydrophilic polymers is similar to that for proteins but the solubilities are of primary importance. Organic solvents can be added to the mobile phase to increase solubility. In polymer analysis, ionic strength and pH can change the shape of the solute from mostly linear to globular therefore, it is very important to use the same conditions during calibration and analysis of unknowns (8). Many mobile phases have been used, but 0.05-0.2 M sodium sulfate or sodium nitrate is common. [Pg.316]

Beebe, K.R., Kowalski, B.R., "An Introduction to Multivariate Calibration and Analysis", Anal. Chem. 1987 (59) 1007A-1017A. [Pg.191]

K, Beebe and B.R. Kowalski, An introduction to multivariate calibration and analysis. Anal. [Pg.56]

Quantitative analysis in this fashion is dependent on the physical parameters of the sample being prqperly reproduced in prepared standards. With water samples that are essentially clean, that is, drinking water, an appropriate standard can readily be prepared in distilled water. However, if the sample of concern is questionable in salt content and likely deviates significantly from distilled water, it would then be appropriate to make an overwhelming addition of salt to both sample and standard prior to calibration and analysis. [Pg.409]

Solids. Conventional near-infrared reflectance analyzers use a variety of methods to position the sample into the incident collimated beam and collect reproducibly the diffusely reflected radiation to measure the absorption which takes place in the body of the sample traversed. Solid samples are ground with care to achieve reproducible and reasonably uniform granulation for calibration and analysis measurements. The overall scattering characteristics of the sample which shift the reflectance baseline and control the depth of sample penetration and opportunity for absorption become a part of the method and the empirical analytical equation (6,9). Solid sampling is summarized on the following table. [Pg.279]

TABLE 13-6. Calibration and Analysis Data for Caffeine in a Beverage"... [Pg.401]

The concentration of iron in the standard has to correspond with that of the sample solution. In the case of alloy steel of varying iron concentration, pure-iron solution is added to the stock solution and to the sample solution in excess. Thus, the difference in the concentration of iron in the sample compared with that in the standard is negligible. Interferences due to the matrix then remain constant during calibration and analysis. Calibration with model solutions is carried out by following the same pattern as with standard solutions. [Pg.243]

Capillary gas chromatography (GC) was used to determine the aminopropyldisiloxane concentration. An 11-m column (0.2-mm internal diameter) coated with a dime thy Isiloxane stationary phase was used. Temperature-programming techniques and a flame ionization detector were used. Tetradecane was used as an internal standard. Details of the chromatographic conditions have been reported earlier (8). A Varian Vista 402 data station simplified the calibration and analysis for both the HPLC and GC methods. [Pg.148]

The availability of secondary standards (i.e., sample matrix-based standards) has the advantage that the calibration and analysis can be performed in a similar matrix, however, without complete elimination of the matrix effects [133]. For many analytes that are structurally heterogeneous (e.g., some hormones and tumour markers), even the development of reference methods constitutes a serious problem [133]. [Pg.618]

The internal standard method is based on the comparison of signals of the substance to be determined and another substance with similar spectral properties mixed to the calibration and analysis solution in the same concentration. This procedure was introduced to TLC by Klaus in 1972 82) but was not widely adopted. Internal standard techniques lead to more precise results 87,88) by using electronic integrators83) or automated systems 84,85,86). The calculation is based on Eq. (37) where the internal standard factor is calculated from the results in peakheight or area from the calibration track according to Eq. (36)i... [Pg.90]

It is clear from Figure 28-19 that the part of a flame to be used in an analysis must vary from element to element and that the position of the flame with respect to the source must be reproduced closely during calibration and analysis. Generally, the flame position is adjusted to yield a maximum absorbance reading. [Pg.863]

Bro R, Heimdal H, Enzymatic browning of vegetables. Calibration and analysis of variance by multiway methods, Chemometrics and Intelligent Laboratory Systems, 1996,34, 85-102. [Pg.353]

This will improve analytical confidence and traceability and reduce, or eliminate, analytical errors due to sensing element placement and/or sample conditioning variability between calibration and analysis steps. [Pg.347]

AMETEK s ProMaxion process mass spectrometer can monitor up to 32 components. Automated sample switching at the multiport allows unattended analysis of process and calibration gases. Different calibration and analysis methods can be assigned to each sample port. A membrane inlet system is incorporated for ambient gas sampling. The available mass range is 1-200 amu. The detection range is 10 ppm to 100 % with a Faraday cup detector lower LODs are possible with an electron multiplier. [Pg.241]

The authors thank Susan Reynolds, David Shafer, Russ Bone, Dennis Wegener, James Hiekey, Ben Powell, Riek Tharp, Susan Singletary Gordon and Leslie Deal for their assistance in data acquisition or interpretation. The authors especially thank Paul Walker for assistance with Rock-Eval 6 calibration and analysis. The manuscript was improved by reviews by S. Larter and K. Weissenburger. [Pg.86]

X-ray detector electronic settings (Fig. 2). For elements where narrow voltage windows are required to filter out interferences from higher-order peaks, calibration and analysis current should be varied so as to (a) minimize changes in X-ray count rate, and (b) minimize voltage pulse distribution shift. Detector settings are EMP-specific. [Pg.360]

Total X-ray counts from the detector are simply accumulated and output using a programmable logic controller (PLC). The total counts are expressed as count rate (total counts divided by analysis count time) and sent to the software for use in calibration and analysis algorithms. [Pg.112]

E10.05 E0181-93el Standard Test Methods for Detector Calibration and Analysis of Radionuclides... [Pg.415]

Another disadvantage of the A -matrix approach results from the fact that calibration and analysis are connected to the inversion of a matrix. Although this is not a problem from the point... [Pg.244]

Of the three modes of calibration used in instrumental analysis, external, internal, and standard addition, the latter provides for the most accurate analytical results for samples which exhibit a matrix interference. The external mode of calibration is used to convert instrumental response to concentration when matrix interferences are not considered a factor. A series of standard solutions which contain the metal of interest are prepared from precise dilution of a certified standard stock solution. The calibration curve is obtained and a least squares regression is performed on the x, y data points. The best-fit line is used to establish the calibration curve. Samples which contain the metal at an unknown concentration level in a sample matrix nearly identical to that used to prepare the serial standards can be run and the data interpolated to give the concentration. In contrast, the standard addition mode of calibration requires that calibration and analysis... [Pg.530]


See other pages where Calibration and analysis is mentioned: [Pg.411]    [Pg.233]    [Pg.96]    [Pg.102]    [Pg.6]    [Pg.248]    [Pg.233]    [Pg.398]    [Pg.414]    [Pg.88]    [Pg.78]    [Pg.266]    [Pg.270]    [Pg.299]    [Pg.317]    [Pg.148]    [Pg.346]    [Pg.377]    [Pg.316]    [Pg.742]    [Pg.233]    [Pg.269]   


SEARCH



Calibration analysis

© 2024 chempedia.info