Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Temperature measurement differential

Thus by measuring the small amount of heat 5Q which is evolved when the adsorption increases by the small amount 6n mole at constant temperature, the differential molar energy of adsorption can be evaluated calorimetri-... [Pg.15]

The solid-liquid transition temperatures of ionic liquids can (ideally) be below ambient and as low as -100 °C. The most efficient method for measuring the transition temperatures is differential scanning calorimetry (DSC). Other methods that have been used include cold-stage polarizing microscopy, NMR, and X-ray scattering. [Pg.43]

If the measured ripple current is confirmed to be within the rating, we can then take the case temperature measurement as the basis for applying the normal 10°C doubling rule, even if the heat is coming from adjacent sources. Again, that is only because the case to core temperature differential is actually within the capacitor s design expectations. [Pg.106]

A TMA analyser will need to measure accurately both the temperature of the sample, and very small movements of a probe in contact with the surface of the sample. A typical analyser, as illustrated in Figure 11.20(a) and (b), uses a quartz probe containing a thermocouple for temperature measurement, and is coupled to the core of a linear variable differential transformer (LVDT). Small movements at the sample surface are transmitted to the core of the LVDT and converted into an electrical signal. In this way samples ranging from a few microns to centimetre thicknesses may be studied with sensitivity to movements of a few microns. For studying different mechanical properties the detailed construction of the probe will vary as is illustrated in Figure 11.20(c). [Pg.494]

Laboratory operation of equipment with a fixed bed of granules is not highly satisfactory because of difficulty of temperature control and measurement in both radial and axial directions. A short packed bed with extensive recycle, however, can achieve substantially isothermal behavior and measurable differential conversion. [Pg.105]

Differential thermal analysis (DTA) is a technique in which the temperature difference between the sample tested and a reference material is measured while both are subjected to the controlled temperature program. Differential scanning calorimetry (DSC) is a technique in which the heat flow difference between the sample and reference material is monitored while both are subjected to the controlled temperature program. Thermogravimetric analysis (TGA) is a technique in which the weight of a sample is monitored during the controlled temperature program. [Pg.424]

The difference of temperature between the sample under estimation and a thermally-inert reference material is continuously recorded as a function of furnace temperature in differential thermal analysis (DTA). In actual practice both TGA and DTA are regarded as complementary techniques whereby information gathered by the usage of one approach is invariably supplemented and enhanced by the application of the other method. The range of phenomena measurable during a DTA-run is found to be much larger than in a TGA-run. [Pg.198]

Radiance temperature measurements. In the high-temperature range, devices based on radiance measurement can be used. A differential thermal analysis... [Pg.546]

Area 300 is controlled using a distributed control system (DCS). The DCS monitors and controls all aspects of the SCWO process, including the ignition system, the reactor pressure, the pressure drop across the transpiring wall, the reactor axial temperature profile, the effluent system, and the evaporation/crystallization system. Each of these control functions is accomplished using a network of pressure, flow, temperature, and analytical sensors linked to control valves through DCS control loops. The measurements of reactor pressure and the pressure differential across the reactor liner are especially important since they determine when shutdowns are needed. Reactor pressure and temperature measurements are important because they can indicate unstable operation that causes incomplete reaction. [Pg.115]

Microcalorimetry has gained importance as one of the most reliable method for the study of gas-solid interactions due to the development of commercial instrumentation able to measure small heat quantities and also the adsorbed amounts. There are basically three types of calorimeters sensitive enough (i.e., microcalorimeters) to measure differential heats of adsorption of simple gas molecules on powdered solids isoperibol calorimeters [131,132], constant temperature calorimeters [133], and heat-flow calorimeters [134,135]. During the early days of adsorption calorimetry, the most widely used calorimeters were of the isoperibol type [136-138] and their use in heterogeneous catalysis has been discussed in [134]. Many of these calorimeters consist of an inner vessel that is imperfectly insulated from its surroundings, the latter usually maintained at a constant temperature. These calorimeters usually do not have high resolution or accuracy. [Pg.212]

An Integrated-Circuit Temperature Sensor for Calorimetry and Differential Temperature Measurement 193... [Pg.136]

Therefore, any result that follows from considerations of the form of Fick s second law applies to evolution of heat as well as concentration. However, the thermal and mass diffusion equations differ physically. The mass diffusion equation, dc/dt = V DVc, is a partial-differential equation for the density of an extensive quantity, and in the thermal case, dT/dt = V kVT is a partial-differential equation for an intensive quantity. The difference arises because for mass diffusion, the driving force is converted from a gradient in a potential V/u to a gradient in concentration Vc, which is easier to measure. For thermal diffusion, the time-dependent temperature arises because the enthalpy density is inferred from a temperature measurement. [Pg.79]

Thermal properties are measured and evaluated by some of the methods also mentioned in Chapter 2. For identification of transition temperatures, measurements of heats of fusion, and so on, differential thermal analysis (DTA) and differential scanning calorimetry (DSC) are much used. Thermal stability is measured by thermogravimetric analysis (TGA), although this technique can give overly optimistic results unless used with great care. [Pg.161]

Schawe, J. E. K. (1998). A description of the glass transition measured by temperature modulated differential scanning calorimetry. Colloid Polym. Sci. 276(7), 565-569. [Pg.834]

Signal processing. The signal from a sensor is usually related in a nonlinear fashion to the process variable of interest. For the output of the measurement device to be linear with respect to the process variable of interest, linearization is required. Furthermore, the signal from the sensor might be affected by variables other than the process variable. In this case, additional variables must be sensed, and the signal from the sensor compensated to account for the other variables. For example, reference junction compensation is required for thermocouples (except when used for differential temperature measurements). [Pg.54]

The time constants of pressure and differential pressure measurements are on the order of 0.1 seconds. Temperature measurement time constants are usually between 1 and 10 seconds. Composition measurements (analyzers) are even slower, varying from 5 seconds to 10 minutes. [Pg.177]

By diluting a reactive sample powder with an inert powder (e.g. AI2O3), the heat capacity of the sample can be made to match more closely that of the reference, hence baseline float can be dampened. This is a useful technique for reactions of significant thermal effect, e.g. combustion, since sample dilution diminishes the intensity of the differential temperature signal. Since the diluent adds a thermal resistance between the reaction zones and the temperature measuring device, the onset of reactions will shift to higher temperatures. [Pg.75]

For the examinations three different mono- and multifilament PET-yams were used. As seen by the effective temperature two of the fibers (220 dtex multifil and 360 monofil) were heat setted in air at 160°C. The experiments in air and supercritical C02 were carried out in a 400 ml autoclave, the DSC measurements (Differential Scanning Calorimetry) under pressure in a home-made apparatus with an integrated TA-Instruments calorimeter. [Pg.581]


See other pages where Temperature measurement differential is mentioned: [Pg.94]    [Pg.94]    [Pg.393]    [Pg.211]    [Pg.270]    [Pg.405]    [Pg.747]    [Pg.225]    [Pg.243]    [Pg.296]    [Pg.283]    [Pg.33]    [Pg.228]    [Pg.405]    [Pg.232]    [Pg.416]    [Pg.25]    [Pg.141]    [Pg.108]    [Pg.48]    [Pg.217]    [Pg.91]    [Pg.168]    [Pg.100]    [Pg.365]    [Pg.188]    [Pg.27]    [Pg.154]    [Pg.602]   
See also in sourсe #XX -- [ Pg.230 ]




SEARCH



Differential measurement

Differential scanning calorimetry transition temperature measurement

Temperature differential

Temperature measurement

Temperature measurement double-differential

Temperature measurement, differential scanning

Temperature measurement, differential scanning calorimetry

© 2024 chempedia.info