Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Stopped-flow instruments

Schematic diagram of a stopped-flow instrument whose operation is described in the text. Schematic diagram of a stopped-flow instrument whose operation is described in the text.
Volumes of activation for fast reactions are determined from the effects of high pressure on rate constants, as presented in Chapter 7. Several versions of stopped-flow instruments suitable for high-pressure experiments have been described.7-10... [Pg.256]

When a larger than tenfold excess of superoxide is applied to the solution of [Fe Por)(DMSO)2l, the reduced [Fe (Por)] complex is formed during the dead time of the stopped-flow instrument, and the second reaction step (Scheme 12), i.e., formation of the peroxo [Fe (Por)(02 )] complex is observed (Fig. 10). This second... [Pg.90]

Stopped-Flow Instrumentation and Design 92 Application of Stopped-Flow Techniques to Soil Constituent Reactions 93 Electric Field Methods 95 Introduction 95... [Pg.61]

For those primarily interested in the design and construction of stopped-flow instruments, the reader should consult contributions by a number of authors (Caldin, 1964 Gibson, 1969 Kustin, 1969 Chance, 1974). Commercial stopped-flow instruments are readily available and are manufactured in the United States, England, France, and Japan. Two of the current commercial units are discussed below. The major advance in stopped-flow instruments since 1970 has been in adapting stopped-flow units to a variety of detection systems, which has increased the kinds of reactions that can be studied. There is also stopped-flow instrumentation that enables reaction rates to be studied as a function of pressure. [Pg.92]

Bio-Logic Instrument and Laboratories (Meylan, France) manufactures an SFM-3 stopped-flow instrument (Fig. 4.17) that consists of three independent drive syringes driven by stepping motors, two mixers and a delay line, three observation windows, replaceable cuvettes, no stop-syringes, and efficient temperature regulation. At maximum flow rate, the minimum dead times range from 1.0 to 4.9 ms for fluorescence detection and 1.3 ms for transmittance. Currently, the Bio-Logic MOS-IOOO optical system employs fluorescence or absorbance detection, which is not suitable... [Pg.92]

By control experiments with a standard commercial stopped-flow instrument, it was checked that true kinetic information was derived, i.e. that there are no mixing effects any more, which limit the applicability of the device [133],... [Pg.151]

Kinetic measurements were carried out using a Hewlett-Packard diode-array spectrophotometer and/or a Hi-Tech small-volume single-mixing stopped-flow instrument (fastest mixing-time 5ms) in absorbance (turbidity) mode. [Pg.687]

Reactions completed within the dead time of stopped-flow instruments cannot be measured dead times are on the order of milliseconds for conventional instruments but can be somewhat shorter if special mixers are used. [Pg.6311]

Today a stopped-flow instrument consists of oifly the unit itself combined with a diode array detector and a computer allowing fast kinetic measurements of time-resolved UV-vis spectra under anaerobic, high pressure and/or low temperature conditions. Improvements have been made as well, for example, syringes are installed vertically instead of horizontally (to avoid problems with gas bubbles) and polyetheretherketone (PEEK) is used instead of Teflon for valves and flow tubes to improve the anaerobic capabilities of the instrument. Further, the syringe drives setup was optimized. Application of rapid-scan devices (usually, but not exclusively, diode arrays) allows complete spectra to be collected at very short time intervals during a reaction. [Pg.6311]

Because stopped-flow techniques are widely used with optical detection, samples should be prepared in solution and produce detectable signal changes after mixing into the cell. In some situations, if the reaction of some samples is very rapid and complete within the dead time of the stopped-flow instruments, the majority and indeed the entire kinetic time course may be lost. Selected adjustment of concentration, solution conditions, temperature, and so on, may be able to slow the reaction into an accessible time range, but this is not always possible or desirable. Such systems are not amenable to the stopped-flow technique. In general, other techniques will have to be used, and these will be... [Pg.6313]

It is good practice to check from time to time that the stability of the instrument is within the manufacturer s specification. It is necessary to test the reliability of the stopped-flow instrument using control experiments that test a range of parameters such as the dead time, mixing efficiency and signal output. In general, these tests will be the same for the instrument in the configuration for fluorescence studies as that for absorbance studies. [Pg.6314]

Kinetic data were obtained by recording time-resolved UV-vis spectra using a stopped-flow instrument combined with a thermostat and equipped with a highspeed diode array spectrometer. At least 10 kinetic runs were recorded under all conditions, and the reported rate constants represent the mean values. All kinetic measurements were carried out under pseudo-first-order conditions that is. the ligand concentration was in large excess (complex concentration 5 x 10 M). [Pg.6314]

Figure 4 Time-resolved spectra (800 spectra are acquired in the 390-705 nm range per shot. Only 19 spectra are present for clear observation.) of 6.0 x 10 M perphenazine mixed with 2.0 x 10 M K2S2O8 in 0.3 M H2SO4 buffer at 25.0 °C The reaction was monitored by photodiode array detector (PDA) attached to the stopped-flow instrument... Figure 4 Time-resolved spectra (800 spectra are acquired in the 390-705 nm range per shot. Only 19 spectra are present for clear observation.) of 6.0 x 10 M perphenazine mixed with 2.0 x 10 M K2S2O8 in 0.3 M H2SO4 buffer at 25.0 °C The reaction was monitored by photodiode array detector (PDA) attached to the stopped-flow instrument...
Zuberbiihler etal. reported stopped-flow kinetic investigations of the oxygenation reactions of a series of tefradentate copper(I) complexes in propionitrile, tetrahydro-furan, and acetone. The formation of [(L °)Cu (02 )]+ can only be followed below 203 K because at a higher temperature it occurs faster than the stopped-flow instrumental limit. On the basis of experimental results, the authors proposed a reaction mechanism, which involved four steps, at least three intermediates. On the basis of the kinetics analysis, spectra for species (intermediates and product) could be calculated. [Pg.6320]

The anaerobic hybrid mixture is mixed with an oxygenated solution of Hp in a stopped-flow instrument. The absorbance is monitored for 20 seconds, and the resulting observed rate constant is measured at 0.20 0.02 s. This results in an assembly free energy, AGasm. of —9.1 O.lkcal/mol when combined with the consensus on constant in Equation 6 (9). This value is in excellent agreement with the results of the equilibrium low-temperature isoelectric focusing experiment. [Pg.688]

Transient-kinetic techniques most often rely on the rapid mixing of reactants with enzyme to initiate the reaction. This mixing is essential so that all enzyme molecules start reaction in synchrony with one another therefore, the time dependence of the observable reactions dehnes the kinetics of interconversion of enzyme intermediate states. Because mixing requires a hnite amount of time, conventional methods are limited in their ability to measure very fast reactions. For example, a typical value for the dead time of a stopped-flow instrument is approximately 1 ms, which is because of the time it takes to hll the observation cell. Thus, reactions with a half-life of less than 1 ms (rate > 700 s ) are difficult to observe depending on the signal to noise... [Pg.1886]

Figure 4. Schematics of a sequential stopped-flow instrument. Figure 4. Schematics of a sequential stopped-flow instrument.

See other pages where Stopped-flow instruments is mentioned: [Pg.87]    [Pg.888]    [Pg.84]    [Pg.505]    [Pg.156]    [Pg.183]    [Pg.740]    [Pg.510]    [Pg.92]    [Pg.257]    [Pg.278]    [Pg.88]    [Pg.459]    [Pg.355]    [Pg.1258]    [Pg.6312]    [Pg.6313]    [Pg.6313]    [Pg.6313]    [Pg.6315]    [Pg.6315]    [Pg.6318]    [Pg.6318]    [Pg.6325]    [Pg.6326]    [Pg.6382]    [Pg.6563]    [Pg.158]    [Pg.175]    [Pg.305]    [Pg.477]    [Pg.218]   
See also in sourсe #XX -- [ Pg.7 ]




SEARCH



Flow instrumentation

Rapid scanning stopped-flow instrumentation

Stop-flow

Stopped flow

Stopped-flow circular dichrois instrumentation

Stopped-flow instrumentation

Stopped-flow instrumentation

© 2024 chempedia.info