Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standards calculation instructions

The top-down approach is often used when there are method validation data from properly conducted interlaboratory studies, and when the laboratory using reproducibility as the measurement uncertainty can demonstrate that such data are applicable to its operations. Chapter 5 describes these types of studies in greater detail. In assigning the reproducibility standard deviation, sR, to the measurement uncertainty from method validation of a standard method, it is assumed that usual laboratory variables (mass, volume, temperature, times, pH) are within normal limits (e.g., 2°C for temperature, 5% for timing of steps, 0.05 for pH). Clause 5.4.6.2 in ISO/ 17025 (ISO/IEC 2005) reads, In those cases where a well-recognized test method specifies limits to the values of the major sources of uncertainty of measurement and specifies the form of presentation of the calculated results, the laboratory is considered to have satisfied this clause by following the test method and reporting instructions. ... [Pg.171]

Calculation of the exact value of AGP is perhaps less instructive than the generalization we can make about actual free-energy changes in vivo, the energy released by ATP hydrolysis is greater than the standard free-energy change, AG °. [Pg.498]

Alternatively, the analysis may be performed by standard addition method (see Chapter 1.9) in which no calibration curve is required. Prepare an NH3 standard solution that is about 10 times as concentrated as the estimated concentration of NH3-N in the sample. Determine the electrode slope following the instruction manual. To 100 mL sample, add 1 mL 10 A NaOH and immerse the electrode and stir the solution. Record the millivolt value Ex when the reading is stable. Add 10 mL of standard solution into the sample. Mix thoroughly and record the stable millivolt reading E2. Calculate the millivolt difference AE as E2 - E and determine the concentration, Cx mg NH3-N/L, from the following expression ... [Pg.178]

Before using the pH electrode, it should be calibrated using two (or more) buffers of known pH. Many standard buffers are commercially available, with an accuracy of 0.01 pH unit. Calibration must be performed at the same temperature at which the measurement will be made care must be taken to match the temperature of samples and standards. The exact procedure depends on the model of pH meter used. Modern pH meters, such as the one shown in Figure 5.8, are microcomputer-controlled, and allow double-point calibration, slope calculation, temperature adjustment, and accuracy to +0.001 pH unit, all with few basic steps. The electrode must be stored in an aqueous solution when not in use, so that the hydrated gel layer of the glass does not dry out. A highly stable response can thus be obtained over long time periods. As with other ion-selective electrodes, the operator should consult the manufacturer s instructions for proper use. Commercial glass electrodes are remarkably... [Pg.176]

Procedure Use a Perkin-Elmer 403 atomic absorption spectrophotometer equipped with a deuterium arc background corrector, a digital readout device, and a burner head capable of handling 20% solids content. Blank the instrument with water following the manufacturer s operating instructions. Aspirate a portion of the Standard Preparation, and record the absorbance as As then aspirate a portion of the Sample Preparation, and record the absorbance as Av. Calculate the lead content, in milligrams per kilogram, of the sample taken by the formula... [Pg.168]

Procedure Using a suitably calibrated atomic absorption spectrophotometer and following the manufacturer s instructions for optimum operation of the spectrophotometer, determine the absorbance of the Standard Solution and of the Sample Solution at 589.0 nm. The absorbance of the Sample Solution is not greater than that of the Standard Solution. Viscosity of a 50% Solution Dissolve 200 g of sample, calculated on the dried basis and accurately weighed, in 200 mL of water contained in a 500-mL beaker. Equilibrate the solution at 25°, and measure its viscosity with a Brookfield viscometer A (model LVG, or equivalent), using a number 2 spindle at 20 rpm. [Pg.421]

The following experiment describes the release of gastric acid during the cephalic phase. The study involved human subjects who volunteered to have a tube inserted into a nostril and pushed down into the stomach, a standard procedure for collecting stomach juices. The subjects were instructed to be silent, to discuss sports, or to discuss food. Samples were removed at 15-minute inter als during the course of the experiment. The concentration of hydrogen ions (H" ) in the samples was used to calculate the rate of acid release. [Pg.67]

New Spreadsheet Calculations. Throughout the book we have introduced spreadsheets for problem solving, graphical analysis, and many other applications. Microsoft Excel has been adopted as the standard for these calculations, but the instructions could be readily adapted to other programs. Several chapters have tutorial discussions of how to enter values, formulas, and built-in functions. Many other examples worked in detail are presented in our companion book. Applications of Microsoft Excel in Analytical Chemistry. We have attempted to document each stand-alone spreadsheet with working formulas and entries. [Pg.1172]

An instructive two-dimensional calculation that reveals the stress magnifying effects of flaws is that of an elliptical hole in an elastic solid as depicted in fig. 2.12. The crucial idea is that, despite the fact that the specimen is remotely loaded with a stress uq which may be lower than the ideal strength needed to break bonds in a homogeneous fashion, locally (i.e. in the vicinity of the crack-like defect) the stresses at the termination of the major axis of the hole can be enhanced to values well in excess of the remote load. The exact solution to this problem can be found in any of the standard references on fracture and we will content ourselves with examining its key qualitative features. [Pg.60]

Activate cell D4, and make it carry the instruction = STDEV(B3 B5), which calculates the standard deviation according to eqn. (2.2-2). Again verify that STDEV indeed calculates correctly, then erase that test. [Pg.46]

As far as we are aware, no a priori calculations, have even been carried out to evaluate Pc from the molecular properties of the collision partners, its order of magnitude has only been found a posteriori for reaction systems. It is, therefore, impossible to make a direct numerical comparison between the frequency factors A found for the process of stepwise activation Eq. VII.36 and for the "all or nothing" kinetic theory activation Eq. VII.41. The following indirect comparison is, however, instructive. Measurements on the rate of activation of J2 at about 300°K in various inert gases such as He, Ne, A, Kr, and N2 have shown22 that the frequency factor A is of the order of about 5x 1016 cm3/mole /sec. Since the collision number Z is only about 1014 cm3/mole/sec at 300°K, even a value of Pc = 1, which corresponds to unit efficiency in direct collisional activation, could not raise the calculated A-value for the standard collision.theory (Eq. VII.41) to the observed one. This is, of course, one of the old and vexing problems in chemical kinetics,17... [Pg.390]

Many of the sulfur oxoacids and their salts are connected by oxidation-reduction equilibria some of the more important standard reduction potentials are summarized in Table 15.19 and displayed in graphic form as a volt-equivalent diagram (p. 435) in Fig. 15.28. By use of the couples in Table 15.19 data for many other oxidation-reduction equilibria can readily be calculated. (Indeed, it is an instructive exercise to check the derivation of the numerical data... [Pg.706]

The formal analysis of the mathematics required incorporating the linear combination of atomic orbitals molecular orbital approximation into the self-consistent field method was a major step in the development of modem Hartree-Fock-Slater theory. Independently, Hall (57) and Roothaan (58) worked out the appropriate equations in 1951. Then, Clement (8,9,63) applied the procedure to calculate the electronic structures of many of the atoms in the Periodic Table using linear combinations of Slater orbitals. Nowadays linear combinations of Gaussian functions are the standard approximations in modem LCAO-MO theory, but the Clement atomic calculations for helium are recognized to be very instructive examples to illustrate the fundamentals of this theory (67-69). [Pg.167]

Another extremely important aspect of an effective quality system is documentation. Accurate and full documentation of all activities is required to ensure the integrity of data generated in the laboratory. Standard operating procedures and working instructions should be prepared and used for the laboratory processes, and all such documents should be controlled. Any notes made, calculations, or changes to procedures should be recorded and, if necessary, explained. It is useful to adopt the phrase If you did not write it... [Pg.330]

Take all the means of the negative sera and calculate the mean and standard deviation (SD) of the negative population using the calculator. (Note Instruction should be given on the use of a calculator.) Noncourse users of the manual... [Pg.185]

In parentheses are the default values, sd is the default for s in DFIX and SADI, and for si in the SAME instruction for DANG twice the value of sd is applied, sf is the default standard uncertainty for CHIV and FLAT, su is the default values for si and s2 in DELU, and ss is the default value for s in SIMU. As mentioned above, the default values for st in SIMU and ISOR, as well as s2 in SAME are calculated from the respective s or si values (unless specified differently by the... [Pg.21]


See other pages where Standards calculation instructions is mentioned: [Pg.124]    [Pg.1439]    [Pg.6]    [Pg.67]    [Pg.68]    [Pg.342]    [Pg.37]    [Pg.100]    [Pg.120]    [Pg.15]    [Pg.51]    [Pg.134]    [Pg.1173]    [Pg.162]    [Pg.21]    [Pg.244]    [Pg.67]    [Pg.472]    [Pg.484]    [Pg.293]    [Pg.114]    [Pg.546]    [Pg.656]    [Pg.134]    [Pg.112]    [Pg.297]    [Pg.712]    [Pg.46]    [Pg.17]    [Pg.5]   
See also in sourсe #XX -- [ Pg.27 ]




SEARCH



Instructions

Standard calculating

© 2024 chempedia.info