Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standard Experimental Apparatus

For fraction collection the gas chromatographic (GC) collection tube is connected directly to the exit port of the GC detector through a stainless steel standard taper adapter. The collected sample is then transferred to a 0.1-mL conical vial for storage. The system is conveniently employed in the resolution and isolation of two-component mixtures (see Fig. 3.6). [Pg.21]


Standard Experimental Apparatus Moisture-protected Claisen head with... [Pg.467]

Standard Experimental Apparatus Moistiore Protected Reaction Apparatus... [Pg.705]

FIGURE 21.2 Experimental apparatus used to measure the standard reduction potential of the indicated redox couples (a) the acetaldehyde/ethanol couple, (b) the fumarate/succi-nate couple, (c) the Fe /Fe" couple. [Pg.675]

In 1996 the present writer offered the students a study of the 1-butanol/1-pentanol/cyclohexanol mixture using Scheffe s plan (mixture plan) which used a standard experimental matrix offered by LUMIERE software. The substance choice takes into account the limited possibilities of the Setaflash apparatus, the low toxicity of the chosen substances and their reciprocal chemical inertia. These had been... [Pg.69]

The two principal experimental apparatuses used to determine the density of a liquid are the pycnometer and the vibrating tube densimeter. The pycnometer method involves measuring the mass of a liquid in a vessel of known volume. The volume of the pycnometer, either at the temperature of measurement or at some reference temperature, is determined using a density standard, usually water or mercury. Using considerable care and a precision analytical balance accurate to 10 5 g, it is possible to achieve densities accurate to a few parts in 10s with a pycnometer having a volume of 25 cm3 to 50 cm3. [Pg.8]

Also, it seems that most of these properties are interdependent. For example, deaeration and permeability (Mainwaring and Reed, 1987) and perhaps the bulk density ratio (Jones and Mills, 1989) seem to provide an adequate mechanism to detect changes in material performance due to different particle size distribution, density and/or shape. However, possibly the greatest disadvantage or limitation of these empirical techniques is the need to standardize the experimental apparatus and techniques. For exam-... [Pg.729]

Surface roughness was measured according to ANSI standard 1346.1-(1978) using a Tencor Instruments Alpha-Step 200 stylus profilometer located at the National Nanofabrication Facility of Cornell. Five surface roughness measurements were made for each sample and their average values recorded. Details of the experimental apparatus set-up and its operation are given elsewhere (Dems, B. C. et. al. Inti. Polvm. Proc.. in press.). [Pg.242]

Future experimentalists will come to regard deconvolution as an essential tool in the standard repertoire. Through deconvolution they will make otherwise-expensive observations with inexpensive instruments. The limits of state-of-the-art experimental apparatus will be extended. This progress, in turn, will produce data that cannot fail to reveal phenomena that would otherwise lie hidden. The new experimental results will stimulate advances in physical theory, thereby providing incentives for better and better apparatus. Because of its applicability across disciplinary lines, deconvolution will play a key role in the advancement of all branches of science that yield to the tools of measurement. [Pg.132]

The experimental apparatuses for hydrate phase equilibria underwent considerable evolution during the nineteenth century. During the last half century the standard methods for measuring macroscopic equilibria have not changed considerably. Table 6.1 summarizes the different macroscopic experimental methods used to study hydrate properties. [Pg.320]

One of the principal qualities of operant behavior is the stability of the behavioral baselines between different executions of the procedure. Even with the present learning task, where the behavior is by definition labile, the possibility of obtaining comparable results on different occasions represents a major advantage of the procedure. The stable behavioral baselines undoubtedly result from the highly standardized experimental environment, where the behavior is measured automatically and there is a minimum of experimenter influence. Automation also permits numerous animals to be tested simultaneously by the same technician, thereby offsetting the cost of the apparatus and the time-consuming nature of the training procedures themselves. [Pg.41]

The experimental apparatus, as shown in Figure 11-1, was a standard molecular beam machine with a heated pulsed valve for vaporization of the non-volatile species and for supersonic cooling. Samples of 1-methyluracil, 1,3-dimethyluracil and thymine were purchased from Aldrich Co. and used without further purification. The sample 1,3-dimethylthymine was synthesized from thymine following a literature procedure [33], and its purity was checked by nuclear magnetic resonance (NMR) and infrared absorption (IR) spectroscopy. The heating temperatures varied for different samples 130°C for DMU, 150°C for MU, 180°C for DMT, and 220°C for thymine. No indication of thermal decomposition was observed at these... [Pg.303]

It would be helpful if various workers would adopt a few standard experimental arrangements, since the value of 6 and its variation with D and v could quickly be determined for each technique. An alternative would be to run a known reaction with each new method of stirring, thus establishing an "apparatus constant." Sometimes a knowledge of 6 is useful in determining the mechanism of a reaction. For example, in the dissolution of Cu in 02 aturated NH3 solutions, studied by Halpern (28), the overall reaction is... [Pg.369]

The experimental apparatus required to perform UK experiments is relatively simple (Figure 1). The two isotopic feeds are routed via a four-way valve to either a vent or the reactor. Their flow rates are set as closely as possible using standard mass flow controllers. The pressures of the two gas streams are then equalised using a differential pressure transducer and fine needle valve at the vent outlet. The reactor effluent is monitored continuously by a quadmpole mass spectrometer (Fisons Instruments GasLab 300) via a capillary sampling system. [Pg.381]

The details of the experimental apparatus and procedures used can be found in reference 8. In brief, the five catalysts were made by aqueous impregnation of RhClj.SV O onto cabosil silica. The catalysts were calcined at 450°C for two hours in air. Hydrogen uptake measurements were made using standard volumetric methods [9]. One variation was that the catalysts were saturated with hydrogen at 180°C instead of room temperature before measuring their isotherms. [Pg.258]

The reactions were carried out in a standard glass apparatus. The experimental procedure was as follows. A solution of thiophene (10- mol) and acid chloride (2. lO mol) in 50 ml chlorobenzene, was heated to 100°C. The freshly calcinated zeolite (HY Si/Al =15,150 mg, supplied by Zeocat, activation temperature 400°C during 6h, in air) was added to the stirred solution. Samples were withdrawn periodically and analysed by GLC using a OVl capillary column (30 m x 0.22 mm). [Pg.596]

The experimental apparatus for a potentiometric titration can be quite simple only a pH or millivolt meter, a beaker and magnetic stirrer, reference and indicator electrodes, and a burette for titrant delivery are really needed for manual titrations and point-by-point plotting. Automatic titrators are available that can deliver the titrant at a constant rate or in small incremental steps and stop delivery at a preset endpoint. The instrument delivers titrant until the potential difierence between the reference and indicator electrodes reaches a value predetermined by the analyst to be at, or very near, the equivalence point of the reaction. Alternatively, titrant can be delivered beyond the endpoint and the entire titration curve traced. Another approach to automatic potentiometric titration is to measure the amount of titrant required to maintain the indicator electrode at a constant potential. The titration curve is then a plot of volume of standard titrant added versus time, and is very useful, for example, for kinetic studies. The most extensive use of this approach has been in the biochemical area with so-called pH-stats—a combination of pH meter, electrodes, and automatic titrating equipment designed to maintain a constant pH. Many enzymes consume or release protons during an enzymatic reaction therefore, a plot of the volume of standard base (or acid) required to maintain a constant pH is a measure of the enzyme activity, the amount of enzyme present. [Pg.42]


See other pages where Standard Experimental Apparatus is mentioned: [Pg.18]    [Pg.21]    [Pg.21]    [Pg.23]    [Pg.25]    [Pg.458]    [Pg.461]    [Pg.676]    [Pg.679]    [Pg.451]    [Pg.18]    [Pg.21]    [Pg.21]    [Pg.23]    [Pg.25]    [Pg.458]    [Pg.461]    [Pg.676]    [Pg.679]    [Pg.451]    [Pg.104]    [Pg.138]    [Pg.595]    [Pg.252]    [Pg.29]    [Pg.23]    [Pg.96]    [Pg.225]    [Pg.297]    [Pg.204]    [Pg.574]    [Pg.75]    [Pg.468]    [Pg.7]    [Pg.353]    [Pg.131]    [Pg.33]    [Pg.217]    [Pg.270]    [Pg.294]   


SEARCH



Experimental apparatus

Standardized apparatus

© 2024 chempedia.info