Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sensor current measurement

The decrease of oxygen reduction current measured with the Clark oxygen sensor indicates the concentration of glucose. [Pg.443]

Dusts, Mists, Aerosols and Fumes. The P-5 Digital Dust Indicator is another sensor currently available for use as a component of the Chronotox System. Suitable for the measurement of silica, lead fumes, pharmaceutical powders as well as many other types of particulates found in manufacturing or laboratory situations, the battery-operated P-5 uses the light scattering technique to measure dusts over a range of either 0.01-100 mg/m or 0.001-10 mg/m (Figure 6). [Pg.529]

The circuitry used for the breadboard testing of NO and NOp sensor cells was very similar to that shown in Figure 2 only the applied potential was changed. An applied potential of +1.30 V versus the SHE reference electrode was used for NO oxidation while a potential of 0.75 V versus the same reference electrode was used for N02 reduction. Current measurements were again made by measuring the voltage drop across resistor RA. Three electrode systems were used for both gases. [Pg.557]

Flow rates were varied from 60 to 5 cc/min. The N0 concentrations which were used varied from 2 3 ppm at low flow rates to 11.9 ppm at higher flow rates, Values of n at higher (k0 - 60 cc/min) flow rates were found to be independent of N0 concentration. Due to the low sensor cell currents measured at low flow rates, N0 concentrations were held constant. [Pg.562]

The physiological state and concentration of microorganisms in a batch or a fed-batch process change as a fermentation progresses therefore, it is desirable to determine current control strategy (set-point) in real time based on current process variables. However, the limited availability of online sensors to measure... [Pg.231]

An electronic torque balance transmitter incorporating an LVTD (Fig. 6.13) has a sensitivity of 1 mA output current per 1 kN/m2 change in measured differential pressure, where the current is measured across an output impedance of 100 kO. This transmitter is connected to a recorder which has an input impedance of 10 kf) and a sensitivity of 1 kN/m2 per mA change in current from the transmitter (Fig. 6.73). If the pressure sensor is measuring a true pressure differential of 5 kN/m2, what will be the corresponding reading on the recorder ... [Pg.545]

In MIP chronoamperometric sensors, current due to diffusion of the analyte towards the electrode surface, coated with the recognition MIP material, is measured at a constant applied potential. The MIP film used for that purpose must be sufficiently permeable in order to allow for unhindered diffusion of the analyte. The measured current is directly proportional to the concentration of the analyte and depends upon the rate of the analyte diffusion towards and, in some cases, away from the electrode surface. The analyte determination involves a preliminary step of removal of the template from the MIP film. [Pg.247]

Chronoamperometric transduction can be applied to electroinactive analytes as well as electroactive, which are sorbed by the MIP film and then undergo an electrochemical reaction [25]. In the latter case, the analyte should be able to diffuse freely both towards and away from the electrode surface for the current to flow. The primary requirement of chronoamperometric sensing is a linear relationship between the current measured at the constant potential and the concentration of the analyte. Moreover, the electrochemically generated species should readily diffuse away from the electrode surface coated by the sensing film. By way of example, a few representative chronoamperometric sensors based on MIPs are presented below. [Pg.247]

These chapters divide the discussion of electrochemical sensors by the mode of measurement. This chapter is an introduction to the general parameters and characteristics of electrochemical sensors. Chapter 6 focuses on potentiometric sensors, which measure voltage. Chapter 7 describes amperometric sensors, which measure current. Chapter 8 examines conductometric sensors, which measure conductivity. [Pg.99]

Fig. 6.26 Individual and differential response of the NaCI ion-sensitive field-effect transistor (ISFET) sensor (a) response of Na-ISFET and (b) Cl ISFET both measured against regular reference electrode (c) differential current measurement of concentration of NaClin 0.01MMgSO4 solution (adapted from Bezegh et al., 1987)... Fig. 6.26 Individual and differential response of the NaCI ion-sensitive field-effect transistor (ISFET) sensor (a) response of Na-ISFET and (b) Cl ISFET both measured against regular reference electrode (c) differential current measurement of concentration of NaClin 0.01MMgSO4 solution (adapted from Bezegh et al., 1987)...
Studies of atmospheric properties using IR spectroscopy techniques have been reported in the literature for nearly 100 years. This paper presents a brief historical review of the development of this area of science and discusses the common features of spectrographic instruments. Two state of the art instruments on opposite ends of the measurement spectrum are described. The first is a fast response iri situ sensor for the measurement of the exchange of CO2 between the atmosphere and the earth s surface. The second is a rocketborne field-widened spectrometer for upper atmosphere composition studies. The thesis is presented that most improvements in current measurement systems are due to painstakingly small performance enhancements of well understood system components. The source, optical and thermal control components that allow these sensors to expand the state of the art are detailed. Examples of their application to remote canopy photosynthesis measurement and upper atmosphere emission studies are presented. [Pg.217]

Linear sweep or preset potential and current measurement providing instrument can be used with immunosensor and enzyme-free sensor. Potential measured instrument can be used with antioxidant activity sensor. [Pg.644]

These experiments are described in detail in reference [101]. The glucose/ lactate device was first one point calibrated with a protein-free buffer solution. The measured sensitivities were applied to the sensor currents for calculation of glucose and lactate levels and the time course of glucose and lactate levels of a combined intravenous and oral glucose tolerance test together with reference values was determined (Figs. 9 and 10). [Pg.203]

By means of amperometric sensors a simultaneous determination of oxygen and hydrocarbons or nitric oxides is possible. For such sensors tube [vi,vii] or planar [viii] designs were described. The current measured is proportional to the oxygen concentration and the concentration of hydrocarbons and nitric oxides, respectively. [Pg.296]

Even before the introduction of miniaturized biosensor arrays, however, some systems able to simultaneously measure glucose and lactate had been reported in the literature. One example is provided by Osborne et al. [157], who described plastic film carbon electrodes fabricated in a split-disk configuration and then modified to obtain a dual biosensor. They achieved a continuous monitoring of these metabolites by placing the dual electrode in a thin-layer radial flow cell coupled to a microdialysis probe. The stability of the sensors was sufficient for short-term in vivo experiments in which the crosstalk, i.e., the percentage of current measured by one biosensor but due to product generated by the partner biosensor, was acceptable for an in vivo application. [Pg.253]

With regard to in vivo gas-sensing devices, the majority of the work reported to date has involved oxygen-sensitive devices which operate as an electrolytic, not galvanic, type of electrochemical cell (i.e., current measured, not potential). Since such oxygen-sensing catheters are not based on ISEs, they will not be considered in this review. There has been, however, some limited work concerning the development of potentiometric sensors, particularly for in vivo COg measurements. One approach has been to devise... [Pg.24]


See other pages where Sensor current measurement is mentioned: [Pg.68]    [Pg.68]    [Pg.36]    [Pg.651]    [Pg.83]    [Pg.161]    [Pg.122]    [Pg.234]    [Pg.678]    [Pg.148]    [Pg.14]    [Pg.518]    [Pg.154]    [Pg.562]    [Pg.632]    [Pg.245]    [Pg.518]    [Pg.98]    [Pg.117]    [Pg.440]    [Pg.1]    [Pg.46]    [Pg.259]    [Pg.132]    [Pg.198]    [Pg.254]    [Pg.123]    [Pg.289]    [Pg.83]    [Pg.482]    [Pg.867]    [Pg.24]    [Pg.148]    [Pg.430]    [Pg.679]    [Pg.146]   
See also in sourсe #XX -- [ Pg.67 ]




SEARCH



Current Measuring

Current measurement

Measured current

Measurement sensors

Measuring sensor

© 2024 chempedia.info