Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sensitivity, concentration enhancement

The main aims in environmental analysis are sensitivity (due to the low concentration of microcontaminants to be determined), selectivity (due to the complexity of the sample) and automation of analysis (to increase the throughput in control analysis). These three aims are achieved by multidimensional chromatography sensitivity is enhanced by large-volume injection techniques combined with peak compression, selectivity is obviously enhanced if one uses two separations with different selectivi-ties instead of one, while on-line techniques reduce the number of manual operations in the analytical procedure. [Pg.334]

Higher salt concentrations (enhanced water conductivity) from urban and industrial waste waters result in lower dilution. Salinity limits the distribution of sensitive plant and animal taxa, and triggers the abundance of others. The presence of the shrub Tamarix canariensis has increased in the saline soils of the low flow affected Tablas de Daimiel (Spain), while the once dominant RopM/Ms alba retreated. [Pg.28]

The only drawback to NMR is its low sensitivity. Concentrations in the millimolar range are sometimes required, although with computer enhancement techniques (such as Fourier transform) signals at 10 -10 M concentrations can be detected. This is especially important for nuclei that have a low natural abundance, such as (1.1%) or deuterium, (0.015%). [Pg.59]

Adsorption of dye to the surface increased the local concentration, enhancing the TIRF sensitivity. [Pg.50]

Nitro compounds may add to carbon-centered radicals and thus also with the majority of the DNA radicals (Chap. 6.3 only the very strongly reducing radicals such as eaq and C02 reduce the nitro sensitizers to (unstable) hydroxyl-amines McClelland et al. 1984). Originally, the nitro compounds and 02 have just been taken as oxidants irrespective of their mode of action, especially as the efficiency of the sensitizers correlates with their reduction potential (Adams and Cooke 1969 Tallentire et al. 1972 Simic and Powers 1974 Adams et al. 1976a, 1981). This concept is expressed in relationship (88), where C is the sensitizer concentration required to achieve a constant sensitizing response (e.g. an enhancement ratio of 1.6) and E7 the one-electron reduction potential of the sensitizer at pH 7. [Pg.441]

Since the TCD Is a flow sensitive (concentration) detector, the low flow rates generally associated with capillary systems should enhance the observed response of a low volume (30) TCD. The observed response, l.e. peak height, or peak area, is Inversely proportional to the gas flow rate at the detector (assuming the same carrier gas flow rate through the column, with any additional gas being Introduced as a make-up flow just prior to the detector). Optimum detector response should be obtained at those flow rates Just sufficient to efficiently sweep-out the dead volumes associated with the detector and connecting tubing. [Pg.60]

One of the advantages of hydrodynamic electrodes over those employed in stationary solution is that both steady-state and time-dependent measurements can be made. Whilst, for the majority of applications, channel electrodes are operated under steady-state conditions, switching to the "time-dependent mode is often useful. Firstly, the sensitivity is enhanced (this is of particular benefit in analytical applications) and secondly, the concentration of species adsorbed or deposited on the electrode can be controlled. In systems in which the electrode is prone to fouling, for example, the time... [Pg.193]

Note The analytical concentration range may be adjusted by varying the buffer concentration in the acceptor stream. Thus, when the buffer is decreased to 4 ml/500 ml acceptor solution, and 1 ml/500 ml acceptor solution, the sensitivity is enhanced and the analytical ranges are lowered to 3-30 and 2-15 mg 1 respectively. [Pg.213]

Figure 20 Schematic of the flow-through competitive MIA [29,56-58,60], (a) There is a constant concentration of chromophore/fluorophore in the mobile phase, and a constant amount in the binding sites of the MIP. (b) When analyte is injected, it displaces the (intensely colored/fluorescent) probe from the MIP binding sites so the concentration in the mobile phase transiently increases, (c) The presence of the analyte is detected as a peak, followed by a trough as the analyte itself is eluted from the MIP and probe is readsorbed from the mobile phase. Interferents that do not bind to the MIP do not displace probe, so no signal is detected. Sensitivity is enhanced over conventional chromatography because the probe is much more readily detected via absorbance or fluorescence than is the analyte itself. Figure 20 Schematic of the flow-through competitive MIA [29,56-58,60], (a) There is a constant concentration of chromophore/fluorophore in the mobile phase, and a constant amount in the binding sites of the MIP. (b) When analyte is injected, it displaces the (intensely colored/fluorescent) probe from the MIP binding sites so the concentration in the mobile phase transiently increases, (c) The presence of the analyte is detected as a peak, followed by a trough as the analyte itself is eluted from the MIP and probe is readsorbed from the mobile phase. Interferents that do not bind to the MIP do not displace probe, so no signal is detected. Sensitivity is enhanced over conventional chromatography because the probe is much more readily detected via absorbance or fluorescence than is the analyte itself.
In the present research work, serum and urine samples were selected for metabolite analysis. Only sample dilution and concentration enhancement were integrated into the microchip-CE devices developed. However, for variable and complex biomedical samples, which require additional sample pretreatment and present a bigger challenge than normal sample. Although the use of UV detection is universal and applicable for most analytes, its sensitivity is insufficient for the detection of trace analytes. Other detectors such as chemiluminescence and electrochemical detector can be developed coupled with microchip-CE. [Pg.279]

Therefore these results compared to the others already mentioned confirm that the A6 desaturation is sensitive to insulin concentration enhancing polyunsaturated fatty acid biosynthesis. [Pg.94]


See other pages where Sensitivity, concentration enhancement is mentioned: [Pg.441]    [Pg.766]    [Pg.229]    [Pg.182]    [Pg.204]    [Pg.9]    [Pg.461]    [Pg.259]    [Pg.63]    [Pg.96]    [Pg.327]    [Pg.63]    [Pg.590]    [Pg.754]    [Pg.754]    [Pg.668]    [Pg.938]    [Pg.318]    [Pg.217]    [Pg.214]    [Pg.188]    [Pg.23]    [Pg.25]    [Pg.943]    [Pg.770]    [Pg.197]    [Pg.177]    [Pg.152]    [Pg.273]    [Pg.568]    [Pg.569]    [Pg.588]    [Pg.801]    [Pg.1015]    [Pg.1034]    [Pg.513]    [Pg.362]   
See also in sourсe #XX -- [ Pg.160 ]




SEARCH



Concentration enhancement

Sensitivity enhancing

Sensitizer concentrations

© 2024 chempedia.info