Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample addition

This expression includes the use of detector arrays of detectors with additive signals and sample addition of samples to improve sensitivity. Typical sensor parameters are = 1 mW/cm, NEP = 30 pW, = 1.5E — 5 cm, = 60, = 1 for imaging and ca 600 for nonimaging gas detection. [Pg.293]

Methods for data collection vary from lysimeters to streamwater samples. Additional examples of inputs and losses may be found in Jordan (80), Melillo and Gosz (85) and Vitousek and Sanford (26). [Pg.431]

According to test protocols described above, RPA-FT test were performed at 100°C, 1 Hz on all samples additional tests at 100°C, 0.5 Hz were performed on IMA TR, IMA FM, and IMA-AG samples. Essentially three types of data will be discussed hereafter The complex modulus G (as derived from the main torque component in the FT torque spectrum), the corrected total torque harmonic component, i.e., cTTHC, and the Q1/Q2 ratio. [Pg.831]

Determination of Hydroxy-Nitrosamines. The column extraction procedure has proven flexible and convenient for isolating NDELA and BHP from a variety of matrices. No artifactual formation of these nitrosamines has been observed when sulfamic acid was incorporated with the sample. Addition of excess acid prevents elution of amines from the Celite column, minimizing nitrosation reactions at later stages. The triisopro-panolamine sample examined contained approximately 250 mg/kg BHP (Table IV). [Pg.340]

Each commodity required a specifically customized workbook, containing a worksheet for each analyte determined in the commodity. Each laboratory received electronic copies of either three or four workbooks, which served as templates for the three or four commodities assigned to the laboratory. Each set of up to 10 commodity samples scheduled for colleetion and analysis required the creation of a copy of the appropriate template. Each workbook template contained one primary worksheet for each analyte, in which analytical data were recorded and residue levels were calculated, as described below. Eor example, the template for green beans contained 17 primary worksheets, one for each of the 17 analytes determined in each green bean sample. Additional worksheets were inserted into copies of the template as needed, to describe results of further analyses, such as confirmation of analytes present above the limit of quantitation (LOQ) or dilutions to bring the concentration of the analyte into the calibration range. [Pg.244]

Place the Cig SPE column on the vacuum manifold, and prepare for extraction by washing the column sequentially three times with approximately 3-mL volumes of methanol and Optima water, respectively. Following the final water wash, allow a volume of liquid to remain on top of the resin bed, and avoid allowing the column to become dry before sample addition. [Pg.373]

In Table 7, a comparison of actual measurements, and also two well-known pedo-transfer functions, can be found by depth. It is important to note that there is a large difference in water content between the disturbed soil core samples and the undisturbed samples. Additionally, the two pedo-transfer functions also exhibit a large difference in predicted water content. Therefore, when doing calculations or trying... [Pg.887]

HPLC/MS and HPLC/MS/MS analyses are susceptible to matrix effects, either signal enhancement or suppression, and are often encountered when the cleanup process is not sufficient. To assess whether matrix effects influence the recovery of analytes, a post-extraction fortified sample (fortified extract of control sample that is purified and prepared in the same manner as with the other samples) should be included in each analytical set. The response of the post-extraction fortified sample is assessed against that of standards and samples. Matrix effects can be reduced or corrected for by dilution of samples, additional cleanup, or using calibration standards in the sample matrix for quantitation. [Pg.1152]

Besides EPR, in many cases one can make use of spectroscopic, electronic, and kinetic methods. These techniques require tedious procedures on obtaining representative samples. Additionally, applying the methods mentioned one comes across numerous experimental problems, particularly if the experiment should be performed in situ. [Pg.171]

To measure the response of the biotinylated protein sample, add 3 ml of the (strept)avidin solution plus 75 pi of the HABA dye to a cuvette. Mix well and measure the absorbance of the solution at 500 nm. Next, add a small amount of sample to this solution and mix. Record the absorbance at 500 nm. If the change in absorbance due to sample addition was not sufficient to obtain a significant difference from the initial (strept)avidin-HABA solution, add another portion of sample and measure again. Determine the amount of biotin present in the protein sample by using the standard curve. The number of moles of biotin divided by the moles of protein present gives the number of biotin modifications on each protein molecule. [Pg.923]

The first and very simple solid contact polymeric sensors were proposed in the early 1970s by Cattrall and Freiser and comprised of a metal wire coated with an ion-selective polymeric membrane [94], These coated wire electrodes (CWEs) had similar sensitivity and selectivity and even somewhat better DLs than conventional ISEs, but suffered from severe potential drifts, resulting in poor reproducibility. The origin of the CWE potential instabilities is now believed to be the formation of a thin aqueous layer between membrane and metal [95], The dominating redox process in the layer is likely the reduction of dissolved oxygen, and the potential drift is mainly caused by pH and p02 changes in a sample. Additionally, the ionic composition of this layer may vary as a function of the sample composition, leading to additional potential instabilities. [Pg.125]

The TGA system was a Perkin-Elmer TGS-2 thermobalance with System 4 controller. Sample mass was 2 to 4 mgs with a N2 flow of 30 cc/min. Samples were initially held at 110°C for 10 minutes to remove moisture and residual air, then heated at a rate of 150°C/min to the desired temperature set by the controller. TGA data from the initial four minutes once the target pyrolysis temperature was reached was not used to calculate rate constants in order to avoid temperature lag complications. Reaction temperature remained steady and was within 2°C of the desired temperature. The actual observed pyrolysis temperature was used to calculate activation parameters. The dimensionless "weight/mass" Me was calculated using Equation 1. Instead of calculating Mr by extrapolation of the isothermal plot to infinity, Mr was determined by heating each sample/additive to 550°C under N2. This method was used because cellulose TGA rates have been shown to follow Arrhenius plots (4,8,10-12,15,16,19,23,26,31). Thus, Mr at infinity should be the same regardless of the isothermal pyrolysis temperature. A few duplicate runs were made to insure that the results were reproducible and not affected by sample size and/or mass. The Me values were calculated at 4-minute intervals to give 14 data points per run. These values were then used to... [Pg.337]

PCR or PLS establish a mathematical relationship (calibration) between the matrix that is formed by the spectra taken of a collection of samples and the vector of properties or qualities for these same samples. Additionally, both methods allow the prediction of the quality for new samples, just based on their spectra. In contrast to most methods discussed so far, PCR and PLS do not require any order in the data set. [Pg.295]

Sample Additive Amount of additive (wt.% of sample) Temperature (K) ... [Pg.184]

Electrochemical detection places a number of restrictions on the mobile phase used in the chromatographic separation but is well suited to reversed-phase separations. Even so, the mobile phase must be oxygen-free, which requires bubbling an inert gas such as helium through the mobile phase reservoir as well as the sample. Additionally, both the mobile phase and the sample must be free of metals to ensure baseline stability during the readings. Despite these constraints, electrochemical... [Pg.222]

The iirfrastracture of national security and homeland defense research needs improvement. Though many scientists are interested in assisting national security efforts, researchers who work with dangerous or iirfectious materials often have difficulties obtaining authorization for their work and even for receiving their samples. Additionally, research is not as efficient as possible due to lack of industry collaboration of access to intellectual property (IP) and classified iirfor-mation and of peer review of research proposals, engineered applications, and... [Pg.21]

Since the of the backup section is 2.2 times the of the front section, the backup section should sample additional material at 46 percent of the sampling rate of the front section. (See Figure 3.)... [Pg.580]

The kinetics of this type of reaction are slow, so this can be avoided by starting the reaction immediately after sample addition. [Pg.227]

Customarily, semiconductor surfaces are chemically or physically prepared to optimize their chemical and/or electro-optical properties. For chemical sensing applications, a freshly etched surface often provides greater chemical sensitivity. A Br2/MeOH etch of n-CdSe, for example, has typically yielded larger luminescence responses to analytes than have polished samples. Additionally, transducing films have been used to modify semiconductor surfaces to enhance the selectivity of CdSe for particular analytes [2]. [Pg.346]

D. M. Santos, M. M Pedroso, L. M. Costa, A. R. A. Nogueira and J. A. Nobrega, A new procedure for bovine milk digestion in a focused microwave oven gradual sample addition to pre-heated acid, Talanta, 65(2), 2005, 505-510. [Pg.143]

Propene on HY was, therefore, selected for the first in situ variable-temperature study using the CAVERN method. These experiments were carried out in early 1988 and published in 1989 (93). The central features of the CAVERN experiments were that the propene was introduced into the zeolite at cryogenic temperature and the sample was manipulated so that spectral acquisition could commence with an unreacted sample. Additional spectra were then acquired as the sample was slowly raised to room temperature. Detailed experiments of this sort were carried out for propene-2-l3C and propene-7-13C and less extensive experiments were performed for propene-3-13C. These experiments showed, among other things, that the 250 ppm peak was formed coincident with a second peak at ca. 156 ppm and the relative intensities of these peaks were 2 1. A careful study of the literature of carbenium ion chemistry in sulfuric acid and superacid solution media suggested the assignment of these resonances (250 and 156 ppm) to alkyl-substituted cyclopentenyl cations similar to 4. [Pg.141]

Evacuated bulbs (Figure 4.7C) are generally used for trapping volatile components. Since this technique does not concentrate the sample, additional sample preparation may be required. For substances with high infrared absorptivity, the sample may be trapped directly in an evacuated infrared gas cell and analyzed directly. For nonvolatile samples that may condense on the inside walls, the cell must be heated before analysis. [Pg.166]


See other pages where Sample addition is mentioned: [Pg.261]    [Pg.72]    [Pg.989]    [Pg.142]    [Pg.186]    [Pg.418]    [Pg.844]    [Pg.344]    [Pg.158]    [Pg.59]    [Pg.391]    [Pg.176]    [Pg.271]    [Pg.367]    [Pg.407]    [Pg.150]    [Pg.201]    [Pg.152]    [Pg.213]    [Pg.362]    [Pg.405]    [Pg.458]    [Pg.32]    [Pg.578]    [Pg.87]    [Pg.302]    [Pg.584]    [Pg.70]    [Pg.595]   
See also in sourсe #XX -- [ Pg.140 ]




SEARCH



Addition of Sample

Additional Methods of Sample Supply

Electrodeposition sample addition

How to analyse a sample using the method of standard additions in FAAS

Problems sample addition

Sample matrix additives

© 2024 chempedia.info