Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Phase presetting

Defects in the transmitter path may cause total signal loss, increased I, noise or diamond patterns. Problems with the frequency control unit or the synthesizer will lead to no signal or amplitude or phase instabilities. Random variations in amplitude or phase will increase q noise. A more subtle point arises when the phase presetting time is too short such that the... [Pg.70]

Operation of the relay may not necessarily start at the preset value due to certain allowable tolerances. As in lEC 60947-4-1, the relay must not trip within two hours at 105% of FLC but it must trip within the next two hours when the current rises to 120% of FLC. Also, it should trip ill two hours in the event of single phasing when the line current in the healthy phases is 115%, but it should not trip in less than two hours during a healthy condition. [Pg.284]

AMSCO International Inc. [2.14] offers complete installations for VHP sterilization of freeze drying plants, which are programmable and automatic 4 to 400 g H202 per injection can be preset, as well as the parameter of the drying-, sterilization- and venting phases. The process is documented. Figure 2.47 shows an AMSCO-VHP generator. [Pg.182]

Automatic headspace samplers are available from manufacturers of gas chromatographs. These devices are based on the technique of sampling an amount of vapor above the sample itself. Samples are sealed, neat or in a suitable solvent, in containers, and hold at a preset temperature in a thermostatted liquid bath. The headspace vapor results as a partition equilibrium is established between the liquid or solid and the gaseous phase of the volatiles. As each sample is presented to the analyzer, the vessel is punctured and a portion of the headspace gas is withdrawn by a pneumatic injection technique and forced into the column. The main application for those samplers is in the routine analysis of low-boiling fractions in samples containing nonvolatile solids or high-boiling components. Some of the more popular applications today are ... [Pg.321]

In principle, to study the local stability of a stationary point from a linear approximation is not difficult. Some difficulties are met only in those cases where the real parts of characteristic roots are equal to zero. More complicated is the study of its global stability (in the large) either in a particular preset region or throughout the whole phase space. In most cases the global stability can be proved by using the properly selected Lyapunov function (a so-called second Lyapunov method). Let us consider the function V(c) having first-order partial derivatives dY/dCf. The expression... [Pg.37]

In phase I, the product moisture increases. In this phase, the amount of moisture that is removed by drying gas is smaller compared to the liquid that is sprayed on product bed. After preblending, when the product is heated up to a preset granulation temperature, the granulation liquid spray starts. The product bed temperature decreases slightly after liquid spray starts and then reaches a relatively stable temperature throughout the whole stage. [Pg.82]

The progress in the development of improved monitoring techniques for the heterogeneous synthetic reactions involved in the solid phase method provided the feedback control which was necessary for the full automation of the peptide synthesizers. A monitoring system, which is based on the titration of the unreacted polymer-peptide chains with picric acid 34,35), on coupling with a peptide synthesizer provided an automatic feedback 36). This feedback is to implement the next step in the synthesis if the level of the unreacted peptide chains is below the acceptable preset value, or otherwise to repeat the last step. [Pg.126]

Crystalline phases were identified by X-ray diffraction (XRD) on a SIEMENS D5005 diffractometer with Ni-filtered CuKa radiation at room temperature. KCI powder was used as an internal standard for the determination of peak positions. XRD data were collected over 10-70 with a step interval of 0.02 and a preset time of... [Pg.40]

Most of the USALLE approaches reported are of the discrete type and use an ultrasonic bath. Usuaiiy a vessei containing the sampie and the immiscibie, acceptor phase is immersed in the transmitting iiquid heid in a bath and the process involves application of US for a preset time, phase separation and repetition of the extraction cycle, if required. With siight differences, this procedure has been used to extract aroma compounds from grape must, wine [3-5], aged brandies and aqueous aicohoiic wood extracts [6], or specific compound famiiies such as monoterpenoids [7] as weii as voiatiles [8] both from wine, pesticides from honey [9] and methyimercury from bioiogicai materials [10]. [Pg.194]

Apb is the scattering length density difference, Q is Porod s invariant, and Y the mean chord length. For the calculation of Yo(r) we approximated I(q) hy a cubic spline. The equations used for the calculation of " pore and " soUd are to be found in [8,30,39-41,47]. Analytical expressions for the descriptors of RES were published in [10,11,13,42,43]. In its most simple variant, the stochastic optimization procedure evolves the two-point probability S2 (r) of a binary representation of the sample towards S2(r) by randomly excWiging binary ceUs of different phases, starting from a random configuration which meets the preset volume fractions. After each exchange the objective function... [Pg.357]

Once a "learning set" (i.e., database) satisfying preset criteria for acceptance (3)has been developed, the model building phase can begin. In general, this is a straightforward process that is specific for the SAR method employed. [Pg.829]

When the concentrations of the target analytes are relatively low, a low precision is obtained that can be ascribed to the fact that equilibrium conditions in the gas phase held in the donor chamber are not readily reached. This shortcoming can be circumvented by previously saturating the air gap with the volatile species (i.e. by injecting the required volume of a standard solution containing an appropriate volatile concentration and driving it to the lower part of the separation module). The flow of both the donor and the acceptor stream is then stopped for a preset time and, once resumed, the experiment can be continued with fairly good results (the precision thus obtained in the determination of sulphide was 5.46% as RSD) [145]. [Pg.142]


See other pages where Phase presetting is mentioned: [Pg.120]    [Pg.78]    [Pg.139]    [Pg.799]    [Pg.221]    [Pg.161]    [Pg.171]    [Pg.547]    [Pg.72]    [Pg.123]    [Pg.129]    [Pg.136]    [Pg.137]    [Pg.142]    [Pg.258]    [Pg.120]    [Pg.388]    [Pg.99]    [Pg.253]    [Pg.451]    [Pg.71]    [Pg.20]    [Pg.321]    [Pg.271]    [Pg.322]    [Pg.230]    [Pg.114]    [Pg.93]    [Pg.247]    [Pg.248]    [Pg.1430]    [Pg.1430]    [Pg.1430]    [Pg.1431]    [Pg.1435]    [Pg.1436]    [Pg.1779]   
See also in sourсe #XX -- [ Pg.70 ]




SEARCH



Preset

Presets

© 2024 chempedia.info