Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Peak-zero method

The peak-zero (PZ) method of evaluation is used only in special cases. The vertical distance z from the zero line is measured (Fig. 2-25), which is proportional to the absolute value of the derivative [52]. It is suitable for higher derivatives which have nearly symmetrical signals with respect to the abscissa. This method is also recommended if individual curves overlap in an undistorted state, and one of the signals passes through zero at this A position (Fig. 2-26). [Pg.34]

Distorted main peaks lead to nonlinearity of standard lines or give incorrect results. In this case, it can be helpful to use a satellite-zero (SZ) distance for evaluation [Pg.34]


Figure 2-25. Peak-zero method (PZ method), where Zn is proportional to the concentrations of peak P . Figure 2-25. Peak-zero method (PZ method), where Zn is proportional to the concentrations of peak P .
For quantitative measurement, all methods are suited which evaluate the signal directly, such as the peak-peak, peak-tangent, and peak-zero methods (Sec. 2.6.1.1-2.6.1.3). However, if derivatives need to be compared, the signals must be normalized, which means that the concentration must be identical or it must be eliminated. [Pg.132]

Background from previous runs vary from instrument to instrument, from laboratory to laboratory, and from day to day. Memory has an adverse elfeet on the aeeuracy and preeision of isotopie eompositions in the 0.1 %o range and the so-eaUed on-peak zero (OPZ) methods. [Pg.140]

Zero Coverage. The peaks at Infinite dilution were slightly skewed (skew ratio 0.8), with virtually no dependence of retention volume on Injection size. Instead of the peak maximum method, retention volumes were measured by the method proposed by Conder and Young (32). To ensure that the adsorption of n-alkanes on carbon fibers was taking place under equilibrium conditions, the flow rate was varied In the range 20 to 32 cm3 min-1. The net retention volumes were essentially Independent of flow rate. [Pg.173]

Figure 2-27. Evaluation of satellite peaks. SZ satellite-zero methods S satellite method. Figure 2-27. Evaluation of satellite peaks. SZ satellite-zero methods S satellite method.
The absorbance can be measured before the analyte has been converted quantitatively into the detectable compound if the conditions for the reaction (time and temperature) are kept constant and the system is zeroed by washing with a blank or pure water before the next sample is introduced. In this case, the reacted portion of the analyte may be related to its concentration (peak-detecting method). The time required to analyse one sample including washing time is about 1-2 min. Allowing for quantitative reaction (steady-state method) takes about twice as long. However, no intermediate washing is required, because the spec-... [Pg.209]

Random position samplers are also available, which can randomly select, under programme control, sample containers in racks or on trays. The sizes of the sample plate and containers depend on the amount of sample required for the determination of each individual compound. Using the peak-recording principle 2-3 mL of sample usually is needed per analysis of each compound while the steady-state method uses twice the amount. A pick-up device controlled by a mechanical or electrical timer directs a thin stainless steel tube into the sample vessel and the sample is pumped into the analytical system for a pre-determined time interval. The pick-up is then lifted and moved into a rinsing solution reservoir filled with the appropriate zero water . Using the peak-detecting method, the rinsing time equals... [Pg.211]

The same method has been applied to the spherically symmetrical chloro-substituent through equilibration of (16) and (17) in chloroform in the presence of aluminium chloride at temperatures between 210 and 334 K- Chloride ratios were estimated by g.l.c. after hydrolysis to the alcohols. The derived thermodynamic parameters are AH° = 0.68 0.03 kcal mol and A5° = 2.2 + 0.1 cal deg" mol" and the value of AS may be associated with the value indicated by symmetry considerations (see above) and suggests that for the isomerization (16) (17) AS° should approximate to zero. The present value for AH bears comparison with earlier studies with cyclohexyl chloride thus Reisse obtained —AH equal to 0.52kcalmol" with AS ss 0, and Jensen found — A.ff = 0.53 kcalmol" at — 80°C, both by the reliable peak area method. Calculations indicate —AH° to be 0.56 kcal mol for cyclohexyl chloride. [Pg.182]

Another related issue is the computation of the intensities of the peaks in the spectrum. Peak intensities depend on the probability that a particular wavelength photon will be absorbed or Raman-scattered. These probabilities can be computed from the wave function by computing the transition dipole moments. This gives relative peak intensities since the calculation does not include the density of the substance. Some types of transitions turn out to have a zero probability due to the molecules symmetry or the spin of the electrons. This is where spectroscopic selection rules come from. Ah initio methods are the preferred way of computing intensities. Although intensities can be computed using semiempirical methods, they tend to give rather poor accuracy results for many chemical systems. [Pg.95]

These reactances are measured by creating a fault, similar to the method discussed in Section 14.3.6. The only difference now is that the fault is created in any of the phases at an instant, when the applied voltage in that phase is at its peak, i.e. at Vni- so that the d.c. component of the short-circuit current is zero and the waveform is symmetrical about its axis, as shown in Figure 13.19,... [Pg.354]

Because of peak overlappings in the first- and second-derivative spectra, conventional spectrophotometry cannot be applied satisfactorily for quantitative analysis, and the interpretation cannot be resolved by the zero-crossing technique. A chemometric approach improves precision and predictability, e.g., by the application of classical least sqnares (CLS), principal component regression (PCR), partial least squares (PLS), and iterative target transformation factor analysis (ITTFA), appropriate interpretations were found from the direct and first- and second-derivative absorption spectra. When five colorant combinations of sixteen mixtures of colorants from commercial food products were evaluated, the results were compared by the application of different chemometric approaches. The ITTFA analysis offered better precision than CLS, PCR, and PLS, and calibrations based on first-derivative data provided some advantages for all four methods. ... [Pg.541]

It is still possible to enhance the resolution also when the point-spread function is unknown. For instance, the resolution is improved by subtracting the second-derivative g x) from the measured signal g x). Thus the signal is restored by ag x) - (7 - a)g Xx) with 0 < a < 1. This llgorithm is called pseudo-deconvolution. Because the second-derivative of any bell-shaped peak is negative between the two inflection points (second-derivative is zero) and positive elsewhere, the subtraction makes the top higher and narrows the wings, which results in a better resolution (see Fig. 40.30). Pseudo-deconvolution methods can correct for sym-... [Pg.555]

The absorbance ratio AnlAxi for the solute peak should be close to zero. If it is not, then this suggests that the peak is not what we think it is. For example, there may be another component that elutes at the same time, so the ratio method is a simple way of indicating the purity of the peaks. [Pg.58]

In general, if all (n = l,. .., A7e) are distinct, then A will be full rank, and thus a = A 1 /3 as shown in (B.32). However, if any two (or more) (< />) are the same, then two (or more) columns of Ai, A2, and A3 will be linearly dependent. In this case, the rank of A and the rank of W will usually not be the same and the linear system has no consistent solutions. This case occurs most often due to initial conditions (e.g., binary mixing with initially only two non-zero probability peaks in composition space). The example given above, (B.31), illustrates what can happen for Ne = 2. When ((f)) = ()2, the right-hand sides of the ODEs in (B.33) will be singular nevertheless, the ODEs yield well defined solutions, (B.34). This example also points to a simple method to overcome the problem of the singularity of A due to repeated (< />) it suffices simply to add small perturbations to the non-distinct perturbed values need only be used in the definition of A, and that the perturbations should leave the scalar mean (4>) unchanged. [Pg.398]

Increasing standard amounts of analyte are added to the sample and the resulting peak areas, which should show an increase with concentration added, are measured. This method is not as useful in GC as it would be in atomic absorption (see Chapter 9), since the sample matrix is not an issue in GC as it is in atomic absorption, due to the fact that matrix components become separated. However, standard additions may be useful for convenience s sake, particularly when the sample to be analyzed already contains a component capable of serving as an internal standard. Thus, standard additions could be used in conjunction with the internal standard method (see Experiment 45), and the internal standard would not have to be independently added to the sample and to the series of standards — it is already present, a convenient circumstance. Area ratio would then be plotted vs. concentration added and the unknown concentration determined by extrapolation to zero area ratio. Please refer to Chapter 9 for other details of the method of standard additions. [Pg.355]

The internal standard method uses an internal standard substance added in a constant amount to all standards and the sample. Area ratio of analyte peak to internal standard peak is plotted vs. concentration of analyte. The standard additions method uses the addition of the analyte in increasing amounts to the sample. Peak area is plotted vs. concentration added and the line is extrapolated to zero peak area to get the sample concentration. [Pg.535]


See other pages where Peak-zero method is mentioned: [Pg.1]    [Pg.34]    [Pg.1]    [Pg.34]    [Pg.73]    [Pg.1761]    [Pg.57]    [Pg.633]    [Pg.739]    [Pg.170]    [Pg.217]    [Pg.134]    [Pg.229]    [Pg.308]    [Pg.217]    [Pg.212]    [Pg.302]    [Pg.225]    [Pg.611]    [Pg.128]    [Pg.27]    [Pg.76]    [Pg.167]    [Pg.227]    [Pg.197]    [Pg.138]    [Pg.24]    [Pg.17]    [Pg.139]    [Pg.174]    [Pg.187]    [Pg.59]    [Pg.192]    [Pg.33]   
See also in sourсe #XX -- [ Pg.34 ]




SEARCH



Peaks zero

© 2024 chempedia.info