Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Optimization HPLC separations

Optimized HPLC separation allows most betaxanthins to be separated on a Cl8 reversed phase stationary phase according to their respective polarities. - Considerable progress was achieved by the introduction of a highly polar silica-based column, which allowed major improvement of peak resolution, especially at early... [Pg.512]

K. M. Kirkland, K. L. Neilson, D. A. Me Combs, and J. J. DeStef-ano, Optimized HPLC separations of racemic drugs using ovomucoid protein-based chiral column, LC-GC, 10 322 (1992). [Pg.106]

The optimal HPLC separation conditions for the nemadectins precluded the use of "buffer only" ionization. The Finnigan TSP-46 thermospray source is equipped with a filament and a discharge electrode as choices for an external ionization method. Both positive and negative ion detection are available. The best choice for an ionization method and for the mode of ion detection would have to provide adequate sensitivity for minor component analysis, unambiguous molecular weight information, and sufficient fragmentation to differentiate between components with similar retention times and identical molecular weights. [Pg.146]

Column temperature is very important in optimizing HPLC separations, especially in RP-HPLC. Early RP-HPLC protein separations were often done at ambient temperature. Fluctuations in room temperature, however, significantly decreased run-to-mn reproducibility of elution times [53]. Thus, maintaining constant column temperature (and, ideally, preheating solvents to this temperature) is essential for good reproducibility. [Pg.557]

The above optimized HPLC separation conditions were next interfaced with optimized DCP operating conditions, realized in the FIA-DCP studies. Figure 9.15 illustrates a typical HPLC-DCP chromatogram for the separation-detection of both standard Se species at the levels and conditions indicated. It was apparent that the desired separation and detection of only Se species could be accomplished. [Pg.273]

Solvent triangle for optimizing reverse-phase HPLC separations. Binary and ternary mixtures contain equal volumes of each of the aqueous mobile phases making up the vertices of the triangle. [Pg.582]

In this experiment a theoretical model is used to optimize the HPLC separation of substituted benzoic acids by adjusting the pH of the mobile phase. An empirical model is then used... [Pg.699]

In the context of chemometrics, optimization refers to the use of estimated parameters to control and optimize the outcome of experiments. Given a model that relates input variables to the output of a system, it is possible to find the set of inputs that optimizes the output. The system to be optimized may pertain to any type of analytical process, such as increasing resolution in hplc separations, increasing sensitivity in atomic emission spectrometry by controlling fuel and oxidant flow rates (14), or even in industrial processes, to optimize yield of a reaction as a function of input variables, temperature, pressure, and reactant concentration. The outputs ate the dependent variables, usually quantities such as instmment response, yield of a reaction, and resolution, and the input, or independent, variables are typically quantities like instmment settings, reaction conditions, or experimental media. [Pg.430]

Though we and others (27-29) have demonstrated the utility and the improved sensitivity of the peroxyoxalate chemiluminescence method for analyte detection in RP-HPLC separations for appropriate substrates, a substantial area for Improvement and refinement of the technique remains. We have shown that the reactions of hydrogen peroxide and oxalate esters yield a very complex array of reactive intermediates, some of which activate the fluorophor to its fluorescent state. The mechanism for the ester reaction as well as the process for conversion of the chemical potential energy into electronic (excited state) energy remain to be detailed. Finally, the refinement of the technique for routine application of this sensitive method, including the optimization of the effi-ciencies for each of the contributing factors, is currently a major effort in the Center for Bioanalytical Research. [Pg.153]

The Window diagram method for the optimization of separation was developed by Laub and Purnell [73], and it has been used both for gas chromatography and HPLC. Recently it is applied in TLC and HPTLC [19,74—76]. [Pg.92]

Fixed pathlength transmission flow-cells for aqueous solution analysis are easily clogged. Attenuated total reflectance (ATR) provides an alternative method for aqueous solution analysis that avoids this problem. Sabo et al. [493] have reported the first application of an ATR flow-cell for both NPLC and RPLC-FUR. In micro-ATR-IR spectroscopy coupled to HPLC, the trapped effluent of the HPLC separation is added dropwise to the ATR crystal, where the chromatographic solvent is evaporated and the sample is enriched relative to the solution [494], Detection limits are not optimal. The ATR flow-cell is clearly inferior to other interfaces. [Pg.491]

Berridge, J.C. (1985). Techniques forthe Automated Optimization of HPLC Separations. John Wiley Sons, Inc., New York. [Pg.144]

System B is the most widely used for CL detection after an HPLC separation. In this system, two pumps are required for delivering the reagent solutions in the following cases (1) the solutions for CL reaction are first combined and then mixed with an eluent (2) CL reaction conditions (e.g., pH, water and organic solvent contents, and salt concentration) need to be optimized before mixing with the CL reagent. [Pg.397]

Optimizing an HPLC separation using five ternary mobile phases. Peaks ... [Pg.141]

A number of papers have appeared reporting the HPLC separation of CTC from its isomers and/or other tetracyclines. There is not a consensus of opinion as to the most satisfactory approach thus, it appears that at this time one must still verify the optimal system for a particular instrument. Methods found in the literature for CTC are described in Table 6. EDTA is added to prevent the formation of complexes of the tetracyclines with metallic surfaces. [Pg.127]

Eigure 3 shows the comparison of the optimized HPLC and CE methods for a stressed sample of the basic drug substance LAS 35917. While the main degradants (II, 12, 13) coelute in the HPLC method, they could be baseline separated in the optimized CE system. [Pg.113]

A.K. Smilde, A. Knevelman and P.M.J. Coenegracht, Introduction of multi-criteria decision making in optimization procedures for HPLC separations. Journal of Chromatography, 369 (1986) 1-10. [Pg.305]

J.W. Weyland, C.H.P. Bruins, D.A. Doombos, Use of 3-dimensional a-plots for optimization of mobile phase composition for RP-HPLC separation of sulfonamides. Journal of Chromatographic Science, 22 (1984) 31-39. [Pg.306]

Fig. 13. A Chromatographic System. From J. Berridge TK hmques for the automated optimization of HPLC separations , 1984, J. Wiley Sons, England, page 20. Reproduced by permission of Wiley Sons, England... Fig. 13. A Chromatographic System. From J. Berridge TK hmques for the automated optimization of HPLC separations , 1984, J. Wiley Sons, England, page 20. Reproduced by permission of Wiley Sons, England...
The first results of optimization in chromatography were published in 1975 Since then a growing number of optimization experiments in HPLC using the Simplex procedure has been reported (table 9). The examples are mainly reversed-phase separations, in which the composition of the ternary or binary mobile phase composition is optimized. The factors optimized are usually a selection from flow rate, column temperature and length, the eluents constitution (e.g. organic modifier content, buffer concentration and pH), the gradient shape. Seven years after the first applications of Simplex optimization had appeared, the first fully automated optimization of HPLC separations was published by Berridge in 1982. This development coincid-... [Pg.23]


See other pages where Optimization HPLC separations is mentioned: [Pg.12]    [Pg.217]    [Pg.676]    [Pg.2121]    [Pg.391]    [Pg.12]    [Pg.217]    [Pg.676]    [Pg.2121]    [Pg.391]    [Pg.203]    [Pg.29]    [Pg.204]    [Pg.244]    [Pg.291]    [Pg.120]    [Pg.259]    [Pg.346]    [Pg.542]    [Pg.146]    [Pg.318]    [Pg.323]    [Pg.51]    [Pg.230]    [Pg.445]    [Pg.5]    [Pg.220]    [Pg.353]    [Pg.278]    [Pg.456]    [Pg.16]   
See also in sourсe #XX -- [ Pg.118 ]




SEARCH



HPLC OPTIMIZATION

HPLC separation

Optimization of Enantiomer Separations in HPLC

Principles for Optimization of Enantioselective HPLC Separations

Separator optimized

© 2024 chempedia.info