Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Monitors/monitoring emissions

The second phase, becoming effective January 1, 2000, will require approximately 2000 utilities to reduce their emissions to a level equivalent to the product of an emissions rate of (1.2 lbs of S02/mm Btu) X (the average mm Btu of their 1985-1987 fuel use). In both phases, affected sources will be required to install systems that continuously monitor emissions in order to track progress and assure compliance. [Pg.401]

Air-Quality Monitoring, Emission Monitoring, Control Methods, Studies... [Pg.433]

Thus, as illustrated by Table 17.13 monitoring emissions of hazardous ehemieals into die environment may be required for a variety of reasons sueh as ... [Pg.307]

Figure 5.7 Room-temperature electronic absorption, excitation and emission spectra for 2 in aqueous solution. The excitation spectrum of 2 was recorded by monitoring emission at 400 nm. Reproduced with permission from [31]. Copyright (2004) Royal Society of Chemistry. Figure 5.7 Room-temperature electronic absorption, excitation and emission spectra for 2 in aqueous solution. The excitation spectrum of 2 was recorded by monitoring emission at 400 nm. Reproduced with permission from [31]. Copyright (2004) Royal Society of Chemistry.
Monitor emission control system every 15 min Daily inspection Daily operations test... [Pg.966]

Some commercially available detectors have a number of detection modes built into a single unit. Fig. 2.4o is a diagram of the detector used in the Perkin Elmer 3D system, which combines uv absorption, fluorescence and conductivity detection. The uv function is a fixed wavelength (254 nm) detector, and the fluorescence function can monitor emission above 280 nm, based on excitation at 254 nm. The metal inlet and outlet tubes act as the electrodes in the conductance cell. The detection modes can be operated independently or simultaneously, using a multichannel recorder. In the conductivity mode, using NaCl, a linear range of 103 and a noise equivalent concentration of 5 x 10 8 g cm-3 have been obtained. [Pg.74]

Fluorescence Function It can monitor emission above 280 nm, based on excitation at 254 nm,... [Pg.464]

Figure 1.9 The energy-level and transition schemes and possible luminescence spectra of a three-level ideal phosphor (a) the absorption spectrum (b, c) emission spectra under excitation with light of photon energies hvi and /iV2, respectively (d, e) Excitation spectra monitoring emission energies at /i( V2 — vi) and at /i vi, respectively. Arrows mark the absorption/emission transitions involved in each spectrum. Eixed indicates that the excitation or emission monochromator is fixed at the energy (wavelength) corresponding to this transition. Figure 1.9 The energy-level and transition schemes and possible luminescence spectra of a three-level ideal phosphor (a) the absorption spectrum (b, c) emission spectra under excitation with light of photon energies hvi and /iV2, respectively (d, e) Excitation spectra monitoring emission energies at /i( V2 — vi) and at /i vi, respectively. Arrows mark the absorption/emission transitions involved in each spectrum. Eixed indicates that the excitation or emission monochromator is fixed at the energy (wavelength) corresponding to this transition.
The Texas Air Control Board was responsible for monitoring emissions of hydrogen sulfide and sulfur dioxide during all phases of the Sulphlex construction The measurements indicated that neither gas was present at levels near those which would pose a safety or environmental hazard during any phase of the operation. [Pg.221]

Prior to monitoring of the Cu+ and Zn2+ photoluminescence spectra, the Cu2+ and Zn2+ molecular sieves were calcined in an oxygen stream at 620 K for 3 hrs with a heating rate of 1 K/min. For monitoring emission spectra of Zn2+- and absorption spectra of Co2+-(A1)MCM-41, samples were then dehydrated at 750 K under vacuum of 7xl0 2 Pa in a silica flask connected with a optical cell. Dehydration was carried out with a heating rate of 5 K/min in two steps 370 K for 30 min and 750 K for 3 h. For monitoring spectra of Cu+-(A1)MCM-41, samples were dehydrated at 750 K for 1 h, subsequently reduced in a stream of carbon... [Pg.236]

Recently quantum yields of O( S) from N20 have been determined as a function of incident wavelength. The yield is near unity at 1290 A [McEwan et al. (680), Black et al. (113)]. Because the photolysis of N20 is a convenient source for the production of O( S), N(2D), N2(A3E), and N2(B3n), their quenching rates by many gases have been measured by monitoring emissions produced by the photolysis of N20 and quenching gas mixtures. Chamber-lain and Simons (203) believe that in the region 1400 to 1550 A NO is produced mostly from two reactions... [Pg.52]

A very profound temperature effect was observed for the emission intensity. Figure 1 presents an emission-temperature profile at open circuit in sulfide electrolyte the relative invariance of the sample s spectral distribution with temperature allowed us to monitor emission intensity at the band maximum. Emission intensity was matched for 501.7 and 514.5 nm excitation at 20°C using 17 times as much 501.7 nm intensity. Over the 20-100°C excursion emission intensity is seen to drop by factors of 8 and 30 for 501.7 and 514.5 nm excitation, respectively. [Pg.297]

Another interesting example of the effect of non-radiative relaxation on linewidths is visible in the two-photon absorption spectrum of Eu2+ in CaF2 (Downer et al., 1983). These experiments involve the excitation of 4f7 states within the 4f65d band, followed by non-radiative decay to 4T55d states. The excitation was measured by monitoring emission from 41fl5d to 4f7. It is apparent from the spectra that the linewidths of the 4f7 states vary significantly. [Pg.90]

The quantitation of BaP was accomplished by reversed-phase HPLC (Li-Chrosorb RP18) with a mobile phase of 10% water in acetonitrile. The column was eluted isocratically, and the BoP was determined with a fluorometer using a 366 nm excitation and monitoring emission at 385 nm. [Pg.351]

Equations have recently been derived (87) which extend the MIDP method to systems whose monitored emission originates from two zf levels. [Pg.352]

Figure 4. Excitation spectrum of Mo in Ar at 14 K. The---------is found by monitoring emission at 7561 A and scanning the laser wavelength as shown. The---------... Figure 4. Excitation spectrum of Mo in Ar at 14 K. The---------is found by monitoring emission at 7561 A and scanning the laser wavelength as shown. The---------...
U.v-visible MP fragmentation of Cr(CO)g studied by 542 monitoring emission of excited Cr atoms. Photodissociation mechanisms discussed... [Pg.110]

On the other hand, in the excitation spectrum, the emission intensity 4, at the monitored emission band, is plotted as a function of the wavelength A of the excitation light, w hich varies as the extinction coefficient of the absorbing molecules. Therefore, the excitation spectrum exhibits the same spectral appearance as that of the absorption spectrum. The advantage of measuring the excitation spectrum in addition to the emission spectrum is the greater sensitivity even for low concentrations of photoluminescent material compared to standard absorption measurements. [Pg.134]

Out-of-laboratory measurements are undertaken across a broad range of industrial and analytical sectors for a variety of reasons in clinical and medical diagnostics for the control of chemical and petrochemical production processes and to monitor emissions and discharges to the environment. The validity of data derived from such measurements is clearly of vital importance, for example to demonstrate compliance with environmental legislation. However, it is particularly difficult to obtain valid and reliable measurements outside the laboratory. The inability to control the environment in which the measurements are made and the use of untrained operators both have potential to impact significantly on the reliability of data. The situation is made worse because of the lack of adequate QA and QC procedures, the shortage of reference materials and calibration standards, and... [Pg.144]


See other pages where Monitors/monitoring emissions is mentioned: [Pg.427]    [Pg.384]    [Pg.88]    [Pg.372]    [Pg.6]    [Pg.1314]    [Pg.444]    [Pg.50]    [Pg.75]    [Pg.104]    [Pg.267]    [Pg.277]    [Pg.187]    [Pg.345]    [Pg.1317]    [Pg.425]    [Pg.206]    [Pg.223]    [Pg.29]    [Pg.262]    [Pg.384]    [Pg.1011]    [Pg.426]    [Pg.45]    [Pg.263]    [Pg.24]    [Pg.255]    [Pg.125]   
See also in sourсe #XX -- [ Pg.156 ]




SEARCH



Acid Gas Monitoring in Stack Emissions

Acoustic emission monitoring

Continuous emission monitoring

Continuous emissions monitoring instruments

Continuous emissions monitoring systems

Emissions monitoring

Emissions monitoring INDEX

Formaldehyde surface emission monitor

Fugitive emissions monitoring

Hydrogen sulfide emissions monitoring

Inductively coupled plasma/atomic emission monitoring

Monitor particle emissions

Monitoring optical emission

Monitors/monitoring continuous emissions

Monitors/monitoring fugitive emissions

Pollution continuous emissions monitoring

Process emissions monitoring

Solvent emission lines monitored

© 2024 chempedia.info