Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sensitivity, mass

Another aspect of cost reduction would be solvent economy. The need to preferentially select inexpensive solvents and employ the minimum amount of solvent per analysis would be the third performance criteria. Finally, to conserve sample and to have the capability of determining trace contaminants, the fourth criterion would be that the combination of column and detector should provide the maximum possible mass sensitivity and, thus, the minimum amount of sample. The performance criteria are summarized in Table 1. Certain operating limits are inherent in any analytical instrument and these limits will vary with the purpose for which the instrument was designed. For example, the preparative chromatograph will have very different operating characteristics from those of the analytical chromatograph. [Pg.362]

The apparatus must provide the maximum mass sensitivity. [Pg.363]

The maximum and minimum flow rate available from the solvent pump may also, under certain circumstances, determine the minimum or maximum column diameter that can be employed. As a consequence, limits will be placed on the mass sensitivity of the chromatographic system as well as the solvent consumption. Almost all commercially available LC solvent pumps, however, have a flow rate range that will include all optimum flow rates that are likely to be required in analytical chromatography... [Pg.363]

Another critical instrument specification is the total extra-column dispersion. The subject of extra-column dispersion has already been discussed in chapter 9. It has been shown that the extra-column dispersion determines the minimum column radius and, thus, both the solvent consumption per analysis and the mass sensitivity of the overall chromatographic system. The overall extra-column variance, therefore, must be known and quantitatively specified. [Pg.363]

The optimum mobile phase velocity will also be determined in the above calculations together with the minimum radius to achieve minimum solvent consumption and maximum mass sensitivity. The column specifications and operating conditions are summarized in Table 4. [Pg.366]

It is seen that with 1 ml samples the peak for a concentration of 1 ppm is well off scale and a clearly defined peak was observed for 10 ppb. When a 10 ml sample was used, acetophenone at a concentration of 1 ppb could just be detected. Under the conditions used, the ultimate mass sensitivity of the system was about 10 ng. This level of sensitivity was achieved, on the one hand, by the sample concentrating process, and on the other, as a result of the high mass sensitivity of small bore columns. [Pg.208]

It is seen that there are a large number of compounds capable of dispersive interactions with the reverse phase, contained in the serum that have been extracted and separated. Again the results have been obtained, partly as a result of the extraction and concentration properties of the sampling system, and partly as a result of the high mass sensitivity of the small bore columns. [Pg.210]

It will be assumed for the moment that the non-bonded atoms will pass each other at the distance Tg (equal to that found in a Westheimer-Mayer calculation) if the carbon-hydrogen oscillator happens to be in its average position and otherwise at the distance r = Vg + where is a mass-sensitive displacement governed by the probability distribution function (1). The potential-energy threshold felt is assumed to have the value E 0) when = 0 and otherwise to be a function E(Xja) which depends on the variation of the non-bonded potential V with... [Pg.11]

Under many experimental conditions, the mass spectrometer functions as a mass-sensitive detector, while in others, with LC-MS using electrospray ionization being a good example, it can behave as a concentration-sensitive detector. The reasons for this behaviour are beyond the scope of this present book (interested readers should consult the text by Cole [8]) but reinforce the need to ensure that adequate calibration and standardization procedures are incorporated into any quantitative methodology to ensure the validity of any results obtained. [Pg.34]

The spectrometer is behaving as a concentration-sensitive detector as the signal intensity remains constant as the flow rate increases. If it were mass-sensitive, the detector response would increase. [Pg.296]

Mass-sensitive detector see Mass-flow-sensitive detector)... [Pg.307]

The immobilization of the hyperbranched spherical structures onto physical transducers greatly increases the binding capacity of the surface and leads to enhanced sensitivity and extended linearity of biosensors. Nucleic acid dendrimers were prepared and their amplification properties for the detection of DNA were examined using mass-sensitive transducers [45, 46]. Antibodies... [Pg.254]

Column and detector properties determine the minimum amount of a component that can be reliably distinguished from the background noise. If we arbitrarily select a signal to noise ratio of 4 as the minimum value for the confident determination of a peak in a chromatogram then for a mass sensitive detector the minimum detectable amount is given by... [Pg.29]

RELATIOMSHIP BETWEEN COLDMN DIAMETER, SOLVENT OMSOMPTKNI, MASS SENSITIVITY AND PEAK STANDRAD DEVIATION... [Pg.560]

Fig. 2.6.5 Hardware for high field NMR remote probe in (c) contains a relatively large saddle-detection. Photographs (a) and (b) show la- coil and is used for (flow) imaging. The detec-boratory-built remote detection probes with tor probe in (d) contains a microsolenoid coil both rf coils built into the same body (c), (d) for optimized mass sensitivity, which is parti-and (e) are detector-only remote probes that cularly useful for microfluidic NMR applica-can be inserted from the top or bottom into the tions. The same probe is shown in (e) with a NMR imaging assembly, so that the well mounted holder for a microfluidic chip that is... Fig. 2.6.5 Hardware for high field NMR remote probe in (c) contains a relatively large saddle-detection. Photographs (a) and (b) show la- coil and is used for (flow) imaging. The detec-boratory-built remote detection probes with tor probe in (d) contains a microsolenoid coil both rf coils built into the same body (c), (d) for optimized mass sensitivity, which is parti-and (e) are detector-only remote probes that cularly useful for microfluidic NMR applica-can be inserted from the top or bottom into the tions. The same probe is shown in (e) with a NMR imaging assembly, so that the well mounted holder for a microfluidic chip that is...

See other pages where Sensitivity, mass is mentioned: [Pg.548]    [Pg.396]    [Pg.391]    [Pg.392]    [Pg.158]    [Pg.366]    [Pg.367]    [Pg.383]    [Pg.404]    [Pg.407]    [Pg.407]    [Pg.82]    [Pg.198]    [Pg.53]    [Pg.117]    [Pg.206]    [Pg.6]    [Pg.34]    [Pg.163]    [Pg.416]    [Pg.14]    [Pg.518]    [Pg.42]    [Pg.48]    [Pg.560]    [Pg.791]    [Pg.140]    [Pg.177]    [Pg.179]    [Pg.179]   
See also in sourсe #XX -- [ Pg.25 ]

See also in sourсe #XX -- [ Pg.172 ]

See also in sourсe #XX -- [ Pg.316 ]

See also in sourсe #XX -- [ Pg.152 , Pg.225 , Pg.235 , Pg.243 ]

See also in sourсe #XX -- [ Pg.172 ]

See also in sourсe #XX -- [ Pg.91 ]

See also in sourсe #XX -- [ Pg.531 , Pg.532 ]

See also in sourсe #XX -- [ Pg.83 ]

See also in sourсe #XX -- [ Pg.23 ]

See also in sourсe #XX -- [ Pg.280 ]




SEARCH



Accelerator mass spectrometry sensitivity

Biosensors mass sensitive sensors

Chemical-ionization mass spectrometry sensitivity

Coupling with Molar-Mass-Sensitive Detectors

Detection mass sensitivity

Detector, atomic spectrometer mass sensitivity

Electrochemical detectors mass sensitivity

Electronic fundamentals, mass sensitive

High mass sensitive

Liquid chromatography/ultraviolet/mass sensitivity

Liquid mass-sensitive

Mass Resolution and Abundance Sensitivity

Mass Sensitivity of the APM

Mass Spectrometric Sensitivity

Mass abundance sensitivity

Mass analysis high-sensitivity

Mass flow sensitive detectors

Mass sensitive

Mass sensitive

Mass sensitivity Table

Mass sensitivity acoustic plate mode device

Mass sensitivity overload

Mass sensitivity, absolute

Mass sensitivity, trace analysis

Mass sensitivity,optimization

Mass spectrometer sensitivity

Mass spectrometry differential sensitivity

Mass spectrometry sensitivity

Mass spectroscopy sensitivity

Mass-Sensitive Approaches

Mass-change sensitivity

Mass-sensitive applications

Mass-sensitive detection

Mass-sensitive detectors

Mass-sensitive devices

Mass-sensitive gas sensors

Mass-sensitive magnetoelastic immunosensors

Mass-sensitive sensor arrays

Mass-sensitive transducers

Mass-sensitive transducers crystal microbalances Surface

Molar mass sensitive detector

Sensitivity acoustic mass sensors

Sensitivity analysis accelerator mass spectrometry

Sensitivity analysis high-resolution mass spectrometry

Sensitivity analysis liquid chromatography-mass

Sensitivity analysis mass spectrometric imaging

Sensitivity analysis mass spectrometry

Sensitivity electrospray ionization mass spectroscopy

Sensitivity in mass spectrometry

Sensitivity mass analyzers

Sensitivity of mass spectrometers

Sensitivity tandem mass spectrometry

Sensors mass sensitive

Spark source mass relative sensitivity

Spray mass spectrometric sensitivity

TSM Resonator Mass Sensitivity

The Mass Sensitivity of a Chromatographic System

Triple quadrupole mass spectrometers sensitivity

© 2024 chempedia.info