Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Lock parameters

Figure 21. One-dimensional elTective potential i (R) for ion-dipole interactions for dipole locking parameter C = 0,2,4, and 6. Distance Rf, is (oui /kT), where q is ion charge and a is molecule polarizability. Figure 21. One-dimensional elTective potential i (R) for ion-dipole interactions for dipole locking parameter C = 0,2,4, and 6. Distance Rf, is (oui /kT), where q is ion charge and a is molecule polarizability.
In the case of polar molecules an analytical expression for Rj cannot be found. As shown in Fig. 21, the transition state in is well defined for all values of the dipole-locking parameter C = Ho/2ctkT). The transition state rate of the stochastic theory is plotted against C in Fig. 22, where a comparison is made with other theories. The important point is that the result of this theory is identical to that of CVTST and, of free-energy VTST. When corrected for access to bound states, the result is close to that of i-VTST. This shows that use of the free energy surface for systems with internal degrees of freedom does not introduce significant errors in the evaluation of the low-friction limit. [Pg.442]

Figure 2.2 Variation of locking parameter C with the ratio These data were taken... Figure 2.2 Variation of locking parameter C with the ratio These data were taken...
Next up is the ADO rate coefficient, which takes into account the permanent dipole moment of acetone. The locking parameter, C, is obtainable from Figure 2.2 or from Reference 24, but this requires the correct units for In units ofD/A po/a =... [Pg.37]

Also, using dyes as laser media or passive mode-locked compounds requires numerous special parameters, the most important of which ate the band position and bandwidth of absorption and fluorescence, the luminiscence quantum efficiency, the Stokes shift, the possibiHty of photoisomerization, chemical stabiHty, and photostabiHty. AppHcations of PMDs in other technical or scientific areas have additional special requirements. [Pg.499]

As we further change the parameter R, the hysteresis interval ends (the invariant circle stops existing) and the only attractor is the stable periodic frequency locked solution N. Both sides of the unstable manifold of the sad e-type frequency locked solution are attracted to N (Point G, inset 2e). [Pg.289]

Historically, this has been the most constrained parameter, particularly for confocal laser scanning microscopes that require spatially coherent sources and so have been typically limited to a few discrete excitation wavelengths, traditionally obtained from gas lasers. Convenient tunable continuous wave (c.w.) excitation for wide-held microscopy was widely available from filtered lamp sources but, for time domain FLIM, the only ultrafast light sources covering the visible spectrum were c.w. mode-locked dye lasers before the advent of ultrafast Ti Sapphire lasers. [Pg.158]

Under MAS the quadrupole splitting becomes time dependent, Qg = Qg (f) (see Sect. 2.3.4). This influences both the spin-locking behavior [223] and the polarization transfer [224], with the latter being further affected by the periodic modulation of the IS dipolar interaction. The effect of MAS on spin-locking of the S magnetization depends on the magnitude of the so-called adiabaticity parameter ... [Pg.167]

Fig. 2 The example BATSEQ has been run with the Colors, Data Points and Grid selected. From left Lock, Overlay, Table, Legend, Parameters, Colors, Dashed Lines, Data Points, Grid, Value Output and Zoom. Fig. 2 The example BATSEQ has been run with the Colors, Data Points and Grid selected. From left Lock, Overlay, Table, Legend, Parameters, Colors, Dashed Lines, Data Points, Grid, Value Output and Zoom.
Analytical change control is the monitoring of any changes to analytical methodology, which has to be justified and evaluated for its performance and impact on the quality of the product. For process analytical methods, the change control approach requires significant documentation and can impede the efficient optimization of the method in the early phases of implementation if parameters have to be locked down based on an insufficient data set. [Pg.35]

In tandem MS mode, because the product ions are recorded with the same TOF mass analyzers as in full scan mode, the same high resolution and mass accuracy is obtained. Isolation of the precursor ion can be performed either at unit mass resolution or at 2-3 m/z units for multiply charged ions. Accurate mass measurements of the elemental composition of product ions greatly facilitate spectra interpretation and the main applications are peptide analysis and metabolite identification using electrospray iomzation [68]. In TOF mass analyzers accurate mass determination can be affected by various parameters such as (i) ion intensities, (ii) room temperature or (iii) detector dead time. Interestingly, the mass spectrum can be recalibrated post-acquisition using the mass of a known ion (lock mass). The lock mass can be a cluster ion in full scan mode or the residual precursor ion in the product ion mode. For LC-MS analysis a dual spray (LockSpray) source has been described, which allows the continuous introduction of a reference analyte into the mass spectrometer for improved accurate mass measurements [69]. The versatile precursor ion scan, another specific feature of the triple quadrupole, is maintained in the QqTOF instrument. However, in pre-... [Pg.35]


See other pages where Lock parameters is mentioned: [Pg.80]    [Pg.409]    [Pg.443]    [Pg.185]    [Pg.35]    [Pg.80]    [Pg.409]    [Pg.443]    [Pg.185]    [Pg.35]    [Pg.2105]    [Pg.617]    [Pg.87]    [Pg.182]    [Pg.408]    [Pg.108]    [Pg.33]    [Pg.397]    [Pg.33]    [Pg.249]    [Pg.289]    [Pg.151]    [Pg.302]    [Pg.20]    [Pg.12]    [Pg.467]    [Pg.225]    [Pg.302]    [Pg.261]    [Pg.319]    [Pg.317]    [Pg.163]    [Pg.44]    [Pg.264]    [Pg.78]    [Pg.72]    [Pg.349]    [Pg.17]    [Pg.247]    [Pg.238]    [Pg.84]    [Pg.248]    [Pg.253]   
See also in sourсe #XX -- [ Pg.80 ]




SEARCH



© 2024 chempedia.info