Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Laboratories types

In laboratory-type experiments, eertain aspeets of die proeess are investigated by handling relatively small amounts of raw materials to reduee die material eonstraints to a minimum. In laboratory experiments, a series of measurements are taken eoneerning all die meehanisms that are independent of size (thermodynamies and ehemieal kineties). A number of physieal properties, sueh as densities, viseosities, speeifie heats, and phase equilibria, involved in the model must be aseertained throughout the operating eonditions of the proeess. [Pg.1036]

Cold-Wall Reactors. In a cold-wall reactor, the substrate to be coated is heated directly either by induction or by radiant heating whi 1 e th e rest of the reactor remains cool, or at least cooler. Most CVD reactions are endothermic, i.e., they absorb heat and deposition takes place preferentially on the surfaces where the temperature is the highest, in this case the substrate. The walls of the reactor, which are cooler, remain uncoated. A simple laboratory-type reactor is shown... [Pg.118]

Fig. 3.8 Various types of reference electrode vessels (A) laboratory-type calomel electrode (B), (C) portable calomel electrodes, (D) a Ag/AgCl micropipette with a KC1 electrolyte immobilized with agar in the tip of the capillary... [Pg.188]

The in vitro battery would ideally include measures of opacity, cytotoxicity, and inflammation. The actual test method(s) will vary depending upon the experience of the laboratory, types of compounds to be tested, and so on. If the measured endpoint(s) indicates that the test material is approximately equipotent with known irritants, one would presume the unknown to be an irritant and further testing would not generally be required. One should keep in mind, however, that in many cases in vitro assays are more sensitive than whole-animal testing, so a positive response in vitro may not always indicate an in vivo irritant. If the assays give equivocal results or responses similar to those seen with non- or mild irritants, some type of animal testing may be indicated as confirmation. [Pg.667]

Workers at the Water Research Centre, UK [52], have drawn up a list of organic substances present in the environment from data contributed by laboratories participating in the COST Project 64b Micropollutants , and from the literature, mainly from 1960 onwards. Where possible, the concentration of each substance is given, the contributing laboratory, type of sample, sampling date, method of analysis, and references. [Pg.306]

If the bubble tube is connected to the inlet side of the pump, no corrections are required. This requires the sampler bubble tube shown in Figure 15. If a standard laboratory type bubble tube is connected to the exhaust side of the pump, the volume of air which has gone through the sampler pump becomes additionally humidified by the soap solution in the bubble tube causing an increase in the volume. Connecting a bubble tube to the exhaust side of the pump can cause errors in the range of 1 to 3% if the proper correction is omitted(9). [Pg.501]

The notebook used should be a laboratory-type notebook with bound pages. The notebook should be clearly labeled as to the audit type, date, and auditor(s). Many auditors prefer to use a notebook for a single audit so it may be filed with the checklist and the final report. [Pg.225]

Selection of the chemical, and the proper concentration, requires special knowledge of surface chemistry and laboratory-type testing in bench scale units... [Pg.190]

There are four basic system types. Type I are basic isocratic systems used for simple, routine analysis in a QA/QC environment often for fingerprinting mixtures or final product for impurity/yield checking. Type II systems are flexible research gradient systems used for methods development, complex gradients, and dial-mix isocratics for routine analysis and standards preparation. They fit the most common need for an HPLC system. Type III systems are fully automated, dedicated systems used for cost-per-test, round-the-clock analysis of a variety of gradient and isocratic samples typical of clinical and environmental analysis laboratories. Type TV systems are fully automated gra-... [Pg.17]

Copper Gold-cobalt 0.0005-0.025 -450 0 Reasonable above 60°K Good output at very low temperature Expensive laboratory-type TC... [Pg.515]

With idealized assumptions inherent losses due to interconnection are between about 5 and 8% depending on the front TCO conductivity and the width of the interconnection structure. Additional losses to consider in module production compared to laboratory type small area test cells appear due to production type processes, nonuniformity, and peripheral film removal [8,70]. [Pg.375]

Fig. 8.26. Haze as function of RMS roughness <5rms for different textured TCO front contacts according to Lechner et al. [145]. The TCO films were prepared under different conditions and consist of surface features with differing distribution, size, and shape. The TCOs labeled series A, B, and M are texture etched ZnO Al films, while Asahi U is a high quality laboratory type Sn02 F covered glass [71], More details of the corresponding TCO films can be found in the original papers... Fig. 8.26. Haze as function of RMS roughness <5rms for different textured TCO front contacts according to Lechner et al. [145]. The TCO films were prepared under different conditions and consist of surface features with differing distribution, size, and shape. The TCOs labeled series A, B, and M are texture etched ZnO Al films, while Asahi U is a high quality laboratory type Sn02 F covered glass [71], More details of the corresponding TCO films can be found in the original papers...
It is instructive to project this relationship to die silica/alumina range for type Z faujasite. Granting the uncertainty in extrapolation, the predicted value of pH = 11 for Si/Al = 5 (Si02/Al203 = 10) is entirely reasonable. Current laboratory type Y synthesis is near pH = 12.3 certainly reduction of die pH of die synthesis batch should produce a more silica-rich product The size of die pH reduction required to produce 10 Si02/Al203 is in doubt, but die direction is unmistakable. [Pg.442]

Where small-scale bioseparations have been developed, particularly in the biopharmaceutical industry, there has been a tendency to retain laboratory type equipment even if this results in more labour and capital intensive processing. The reason for this is often to avoid the need for extended periods of process development work with new equipment designs, which might delay the launch of a product where competitors are not far behind. Manufacturers are also wary of adopting new bioseparation techniques for processes if there is any risk that regulators such as the U.S. Food and Drug Administration (FDA) will require more evidence that the equipment is fit for the purpose. This conservative tendency is understandable and may influence the choice of bioseparation equipment for pharmaceutical manufacturing in particular. [Pg.638]

Once a plant is built, the conditions of agitation, aeration, mass (oxygen) transfer, and heat transfer become set. Therefore, those environmental conditions achievable in plant-scale equipment should be scaled down to the pilot plant and laboratory-type equipment (shake flask, micro-reactor) to ensure that the earlier studies are carried out under conditions that can be duplicated. [Pg.1325]

Shortly after the condensation of the first minerals to form within the Solar System, thermal processing of refractory inclusions began. The overall processing of these materials into igneous rocks was short-lived, perhaps only 100000 years (Bizzarro et al. 2007). The critical constraints on the thermal histories of these objects is essentially limited to Type-B CAIs (Connolly Desch 2004), which have been generated by the experimental reproduction of these objects in the laboratory. Type-B CAIs are composed mostly of melilite, anorthite, aluminous spinel, and a titanium-rich pyroxene known in the parlance as fassaite (for a summary of minerals common to the Solar System see Appendix 1). The general consensus is... [Pg.245]

Two typical calomel electrode designs (a) laboratory type, shown unsaturated ... [Pg.609]

Model VG201, Spectroderm International, Inc., Fairfax, Virginia 22030. If volatile chemical carcinogens are used, a separate vacuum pump or vacuum system should be used in conjunction with an appropriate laboratory-type containment cabinet. [Pg.202]

Multiple Phases Reactions between gas/liquid, liquid/liquid, and fluid/solid phases are often tested in CSTRs. Other laboratory types are suggested by the commercial units depicted in appropriate sketches in Sec. 23. Liquids can be reacted with gases of low solubili-... [Pg.533]

In addition to the discussion of the elements for evaluation in the individual studies section of the TSD, a section entitled "Data Adequacy and Research Needs" is included in the text of the TSD. A summary of the data-adequacy discussion is also included in the derivation summary tables in the appendix of the TSD and in the summary section of the TSD. The text of the TSD relates the studies used to derive or support the derivation of the AEGL values to the discussion of the adequacy of the available data. Brief summaries of this discussion are included in the summary and derivation summary tables. The data-adequacy section also presents and integrates the weight of evidence by considering all information as a whole for each AEGL developed. In addition to considering the elements for evaluation as relevant in the discussion, a number of other factors must be considered. They include repeatability of experiments between laboratories, consistency of data between experiments and laboratories, types and number of species tested, variability of results between species, and comparison of AEGL values with the valid... [Pg.76]

The above table was compiled for use at the Bureau of Standards, and is only applicable to thermometers of the laboratory type used under total-immersion conditions, i.e.y with bulb and stem containing the mercury column at the same temperature. [Pg.413]

Industrial Thermometers.—The use of the chemical or laboratory type of mercurial thermometer is restricted as the name implies to the laboratory since unprotected glass thermometers are much too frail to withstand the rough usage of the plant and shop, and are comparatively difficult to read as well. [Pg.413]

Multiphase Reactors Reactions between gas-liquid, liquid-liquid, and gas-liquid-solid phases are often tested in CSTRs. Other laboratory types are suggested by the commercial units depicted in appropriate sketches in Sec. 19 and in Fig. 7-17 [Charpentier, Mass Transfer Rates in Gas-Liquid Absorbers and Reactors, in Drew et al. (eds.), Advances in Chemical Engineering, vol. 11, Academic Press, 1981]. Liquids can be reacted with gases of low solubilities in stirred vessels, with the liquid charged first and the gas fed continuously at the rate of reaction or dissolution. Some of these reactors are designed to have known interfacial areas. Most equipment for gas absorption without reaction is adaptable to absorption with reaction. The many types of equipment for liquid-liquid extraction also are adaptable to reactions of immiscible liquid phases. [Pg.35]

Plebani M, Carraro P. Mistakes in a stat laboratory types and frequency. Clin Chem 1997 43 1348-51. (http //www.clinchem.Org/cgi/content/fuU/43/8/1348). [Pg.526]

Grain-to-grain contact is variable and such a phenomenon influences attempts to repack mined sand, as may be the case in studies involving bitumen removal from the sand in laboratory-type in situ studies. This phenomenon also plays a major role in the expansion of the sand during processing where the sand to be returned to the mine site might occupy 120-150% of the volume of the original as-mined material. [Pg.2949]

Choosing a good thermometer to monitor the temperature of the batches is an area which cannot be ignored. A laboratory type thermometer, of course, will work just fine, but many kitchen thermometers will not. Any portion of the thermometer which... [Pg.16]

Face-centered cubic crystal lattice. Burns when heated with a hot enough flame (over 800, oxygen torch), df 3.513. rt 2 4173. Hardness — 10 (Mohs scale), Sp heat at 100°K 0.606 cal/g-atom/ K. Entropy at 298.I6 K 0,5684 cal/g-atom/ K. Band gap energy 6.7 ev. Dielectric constant 5.7. Electron mobility —1800 cm1/v-sec. Hole mobility 1200 cmz/v-sec. Can be pulverized in a steel mortar. Attacked by laboratory -type cleaning soln (potassi um dichromate + coned HiSO ), In the jewelry trade the unit of weight for diamonds is one carat — 200 mg. Ref Wall Street J. 164, no. 36, p 10 (Aug 19, 1964),... [Pg.470]

Fig. 6 depicts the type of relationship that might be found between downstream gas flow rate and upstream gas pressure in a typical in-process automated bubble point test. The transition pressure is not clearly defined. Actual bubble points (transition pressures) obtained with this type of equipment differ from theoretical bubble points calculated for the same membrane from direct measurement of pore size, and from laboratory-type bubble point testing. [Pg.169]


See other pages where Laboratories types is mentioned: [Pg.206]    [Pg.239]    [Pg.451]    [Pg.812]    [Pg.1530]    [Pg.514]    [Pg.398]    [Pg.611]    [Pg.820]    [Pg.394]    [Pg.566]    [Pg.1586]    [Pg.713]    [Pg.186]    [Pg.294]    [Pg.79]    [Pg.47]    [Pg.89]    [Pg.79]   
See also in sourсe #XX -- [ Pg.2 , Pg.3 , Pg.4 , Pg.5 ]




SEARCH



Dispersion Laboratory reactor types

Laboratory clothing, types

Laboratory glass types

Laboratory reactor types

Types of Management Standards for Laboratories

© 2024 chempedia.info