Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrument radiation

Cary 14 diagram (ca. 1953) The arrows on the optical diagram trace the path of the UV and vis radiation through the instrument. Radiation from the D2 or W lamp is directed to the monochromator entrance slit D by appropriate lenses and mirrors. From mirror E it travels to prism F where it is refracted, then to mirror G which reflects it to variable-width intermediate slit H. Mirror I reflects the radiation to grating J and from there the monochromatic beam is directed to mirror K and exits the monochromator through slit L. Semicircular mirror O, driven by motor Q, chops the beam at 30 Hz and alternately sends half the beam to the reference and half to the sample. Elements V, V1, W, and W1 pass the separated beams to the phototube. The light pulses of the two beams are out of phase with each other so that the phototube receives light from only one beam at a time. The photomultiplier for UV-vis work is shown at X and the NIR detector for 700-2600 nm is shown at Y. [Pg.666]

In optical null instruments, radiation is passed through the sample and reference paths simultaneously or the instrument may be designed to pass the radiation through each path alternately (pre-sample chopping). A wedge or comb attenuator is moved in or out of the reference beam until absorption in both beams is equal. The movement of the... [Pg.237]

The transition from radiochemical separation described in Chapter 6 to instrumental radiation detection in Chapter 8 is source preparation for counting. The analyst wants to prepare a source that represents the radionuclide in the collected sample, can be measured reliably by its radiation, and is stable. The analyst selects a detector that is sensitive to the radiation that characterizes the radionuclide, stable as defined by its QA program, and calibrated for efficiency and—if needed—energy. Source preparation concerns are addressed here for the four types of detectors that are described in Chapters 2 and 8, but these considerations can apply to sources prepared for measurement by other detectors. [Pg.121]

Modern alarming radiation detection instrument (radiation pager). [Pg.108]

In modem instruments, radiation of 2 or 3 wavelengths is passed through the solution before reaching the detector. This reduces interferences (Anal.Chem.44 972) 93A). [Pg.25]

The suggested method is appropriately implemented at the practice. The cost and working hours of unit measurement of it is less than of any alternative method of destructive test and with respect to the authenticity inspection of Stress-Deformation the given method is inferior only to destructive testing. The method was successfully implemented while evaluation of service life of main pipe-lines sections and pressure vessels as well. Data of method and instrument are used as official data equally with ultrasonic, radiation, magnetic particles methods, adding them by the previously non available information about " fatigue " metalwork structure. [Pg.29]

The radiation and temperature dependent mechanical properties of viscoelastic materials (modulus and loss) are of great interest throughout the plastics, polymer, and rubber from initial design to routine production. There are a number of laboratory research instruments are available to determine these properties. All these hardness tests conducted on polymeric materials involve the penetration of the sample under consideration by loaded spheres or other geometric shapes [1]. Most of these tests are to some extent arbitrary because the penetration of an indenter into viscoelastic material increases with time. For example, standard durometer test (the "Shore A") is widely used to measure the static "hardness" or resistance to indentation. However, it does not measure basic material properties, and its results depend on the specimen geometry (it is difficult to make available the identity of the initial position of the devices on cylinder or spherical surfaces while measuring) and test conditions, and some arbitrary time must be selected to compare different materials. [Pg.239]

Flowever, in order to deliver on its promise and maximize its impact on the broader field of chemistry, the methodology of reaction dynamics must be extended toward more complex reactions involving polyatomic molecules and radicals for which even the primary products may not be known. There certainly have been examples of this notably the crossed molecular beams work by Lee [59] on the reactions of O atoms with a series of hydrocarbons. In such cases the spectroscopy of the products is often too complicated to investigate using laser-based techniques, but the recent marriage of intense syncluotron radiation light sources with state-of-the-art scattering instruments holds considerable promise for the elucidation of the bimolecular and photodissociation dynamics of these more complex species. [Pg.881]

A number of mixing experiments have therefore been used to generate both pulses and CW THz radiation. Among these, diode-based mixers used as upconvertors (that is, heterodyne spectroscopy m reverse ) have been the workliorse FIR instruments. Two such teclmiques have produced the bulk of the spectroscopic results ... [Pg.1247]

The isotope boron-10 is used as a control for nuclear reactors, as a shield for nuclear radiation, and in instruments used for detecting neutrons. Boron nitride has remarkable properties and can be used to make a material as hard as diamond. The nitride also behaves like an electrical insulator but conducts heat like a metal. [Pg.14]

Structure determination m modern day organic chemistry relies heavily on instrumental methods Several of the most widely used ones depend on the absorption of electromagnetic radiation... [Pg.575]

Stray radiation is the second contribution to instrumental deviations from Beer s law. Stray radiation arises from imperfections within the wavelength selector... [Pg.387]

Infrared instruments using a monochromator for wavelength selection are constructed using double-beam optics similar to that shown in Figure 10.26. Doublebeam optics are preferred over single-beam optics because the sources and detectors for infrared radiation are less stable than that for UV/Vis radiation. In addition, it is easier to correct for the absorption of infrared radiation by atmospheric CO2 and 1420 vapor when using double-beam optics. Resolutions of 1-3 cm are typical for most instruments. [Pg.393]

The determination of an analyte s concentration based on its absorption of ultraviolet or visible radiation is one of the most frequently encountered quantitative analytical methods. One reason for its popularity is that many organic and inorganic compounds have strong absorption bands in the UV/Vis region of the electromagnetic spectrum. In addition, analytes that do not absorb UV/Vis radiation, or that absorb such radiation only weakly, frequently can be chemically coupled to a species that does. For example, nonabsorbing solutions of Pb + can be reacted with dithizone to form the red Pb-dithizonate complex. An additional advantage to UV/Vis absorption is that in most cases it is relatively easy to adjust experimental and instrumental conditions so that Beer s law is obeyed. [Pg.394]

The emission spectrum from a hollow cathode lamp includes, besides emission lines for the analyte, additional emission lines for impurities present in the metallic cathode and the filler gas. These additional lines serve as a potential source of stray radiation that may lead to an instrumental deviation from Beer s law. Normally the monochromator s slit width is set as wide as possible, improving the throughput of radiation, while being narrow enough to eliminate this source of stray radiation. [Pg.418]

In principle, emission spectroscopy can be applied to both atoms and molecules. Molecular infrared emission, or blackbody radiation played an important role in the early development of quantum mechanics and has been used for the analysis of hot gases generated by flames and rocket exhausts. Although the availability of FT-IR instrumentation extended the application of IR emission spectroscopy to a wider array of samples, its applications remain limited. For this reason IR emission is not considered further in this text. Molecular UV/Vis emission spectroscopy is of little importance since the thermal energies needed for excitation generally result in the sample s decomposition. [Pg.434]

One instrumental limitation to Beer s law is the use of polychromatic radiation instead of monochromatic radiation. Consider a radiation source that emits two wavelengths of... [Pg.450]

A second instrumental limitation to Beer s law is stray radiation. The following data were obtained using a cell with a pathlength of 1.00 cm, when stray light is insignificant... [Pg.450]


See other pages where Instrument radiation is mentioned: [Pg.134]    [Pg.135]    [Pg.135]    [Pg.862]    [Pg.319]    [Pg.41]    [Pg.695]    [Pg.134]    [Pg.135]    [Pg.135]    [Pg.862]    [Pg.319]    [Pg.41]    [Pg.695]    [Pg.368]    [Pg.469]    [Pg.496]    [Pg.912]    [Pg.1059]    [Pg.873]    [Pg.1200]    [Pg.1236]    [Pg.1633]    [Pg.1143]    [Pg.524]    [Pg.387]    [Pg.388]    [Pg.388]    [Pg.388]    [Pg.390]    [Pg.390]    [Pg.391]    [Pg.394]    [Pg.398]    [Pg.420]    [Pg.435]    [Pg.442]    [Pg.446]    [Pg.235]    [Pg.59]   


SEARCH



Instrumentation 303RADIATION CHEMISTRY

Instrumentation radiation level detection

Instrumentation radiation sources

Instruments to measure the radiation field

Ionizing radiation instruments

Levelness instrumentation radiation

Portable radiation surveying instruments

Process radiation instrumentation

Radiation Surveying instruments

Radiation Thermometers (Point-Sensing Instruments)

Radiation instrumentation

Radiation instrumentation

Synchrotron radiation instrumentation

© 2024 chempedia.info