Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Entrance slit, monochromator

The construction of a typical monochromator is shown in Figure 10.12. Radiation from the source enters the monochromator through an entrance slit. The radiation is collected by a collimating mirror, which reflects a parallel beam of radiation to a diffraction grating. The diffraction grating is an optically reflecting surface with... [Pg.376]

Radiation exits the monochromator and passes to the detector. As shown in Figure 10.12, a polychromatic source of radiation at the entrance slit is converted at the exit slit to a monochromatic source of finite effective bandwidth. The choice of... [Pg.377]

A simple spectrometer that we have used successfully is shown in Figure 2. Electrons from an electron microscope hairpin tungsten filament are focused with an Einzel lens onto the monochromator entrance slit, pass through the monochromator and exit slit, and are focused on the sample s surface by additional electrostatic... [Pg.447]

A modern laser Raman spectrometer consists of four fundamental components a laser source, an optical system for focusing the laser beam on to the sample and for directing the Raman scattered light to the monochromator entrance slit, a double or triple monochromator to disperse the scattered light, and a photoelectric detection system to measure the intensity of the light passing through the monochromator exit slit (Fig. 7). [Pg.306]

SI(220) double crystal monochromator was used with entrance slit (1 mm high 20 m from the source point) chosen to give a bandpass of 2 eV at the Pt edge, 11,563.7 eV.( ) The operation of the catalyst... [Pg.282]

The Compton profile measurements on Cu and Cu 953AI0047 were performed at ID 15b of the ESRF. Figure 1 shows the setup of the scanning-type Compton spectrometer used. It consists of a Si (311) monochromator (M), a Ge (440) analyzer (A) and a Nal detector (D). The signal of an additional Ge solid state detector (SSD) was used for normalization. ES, CS and DS denote the entrance slit, the collimator slit and the detector slit, respectively. For each sample 10 different directions were measured with approximately 1.5-2 x 103 7 total counts per direction. The incident energy was 57.68 keV for the Cu and 55.95 keV for the Cuo.953Alo.047 measurement. [Pg.315]

Figure 2. Experimental set-up for Raman spectroscopy. The desired laser line is isolated from other plasma lines by a narrow bandpass filter or broadband prism monochromator, then focused onto a sample in a capillary tube. A collecting lens placed at a 90° angle to the incident beam focuses the scattered light onto the entrance slit of a monochromator with output to a photomultiplier tube (in the case of a scanning instrument) or a diode array detector. Figure 2. Experimental set-up for Raman spectroscopy. The desired laser line is isolated from other plasma lines by a narrow bandpass filter or broadband prism monochromator, then focused onto a sample in a capillary tube. A collecting lens placed at a 90° angle to the incident beam focuses the scattered light onto the entrance slit of a monochromator with output to a photomultiplier tube (in the case of a scanning instrument) or a diode array detector.
Atomic emission from the plasma is focused on to the entrance slit of the monochromator using a combination of convex or plano-convex lenses or a concave mirror. The combination of focusing optics, monochromator and detector is generally referred to as a spectrometer, although the heart of the device is the monochromator. A monochromator is an instrument that... [Pg.93]

Electrothermal atomizers are also suitable for AFS as, when an inert gas atmosphere is used, quenching will be minimized. In the nuclear, electronic, semiconductor and biomedical industries where detection limits have to be pushed as low as 1 part in lO (or 0.1 pg g- in the original sample), electrothermal atomization with a laser as excitation source (LIF-ETA) may be used. Figure 6.5 shows schematically a common way of observing the fluorescence in LIF-ETA. The fluorescence signal can be efficiently collected by the combination of a plane mirror, with a hole at its centre to allow excitation by the laser, positioned at 45° with respect to the longitudinal axis of the tube and a lens chosen to focus the central part of the tube into the entrance slit of the fluorescence monochromator. [Pg.142]

As the beam leaves the prism predisperser, it is focused on the entrance slit of the grating monochromator. The slit is curved, has variable width, and opens symmetrically about the chief ray (optical center line of system). The monochromator itself is of the off-axis Littrow variety (James and Sternberg, 1969 Stewart, 1970 Jennings, 1974) and uses a double-pass system described by McCubbin (1961). The double-pass aspect of the system doubles the optical retardation of the incident wave front and theoretically doubles the resolution of the instrument. The principal collimating mirror is a 5-m-focal-length, 102-cm-diam parabola. [Pg.158]

The simplest method of determining the function S, in the visible region of the spectrum is to take photomultiplier readings when the entrance slit of the monochromator is illuminated by a tungsten lamp giving light of known spectral distribution. If RSL represents the values so obtained, the spectral sensitivity is then calculated from... [Pg.315]

The light beam emitted by the source, which must be at the wavelength required for measurement, passes through the flame (or graphite furnace) in which the element is located in its atomic state. The beam is then focused on the entrance slit of the monochromator, located after the sample. The monochromator s role is to select a very narrow band of wavelengths. The optical path ends at the entrance slit of the photomultiplier tube. [Pg.258]

The monochromator can be replaced by a polychromator. In this instance, a plate with multiple slits is placed at the focal point of the entrance slit to collect the selected lines of radiation. A detector is placed behind every slit (Fig. 15.5). This optical set-up is fixed since there is no rotation of the grating. [Pg.277]


See other pages where Entrance slit, monochromator is mentioned: [Pg.162]    [Pg.519]    [Pg.162]    [Pg.519]    [Pg.382]    [Pg.438]    [Pg.627]    [Pg.258]    [Pg.428]    [Pg.663]    [Pg.132]    [Pg.353]    [Pg.278]    [Pg.396]    [Pg.445]    [Pg.54]    [Pg.74]    [Pg.207]    [Pg.305]    [Pg.42]    [Pg.78]    [Pg.78]    [Pg.110]    [Pg.122]    [Pg.114]    [Pg.120]    [Pg.57]    [Pg.34]    [Pg.316]    [Pg.429]    [Pg.71]    [Pg.222]    [Pg.385]    [Pg.157]    [Pg.373]    [Pg.41]    [Pg.62]    [Pg.278]    [Pg.169]    [Pg.429]   
See also in sourсe #XX -- [ Pg.180 , Pg.181 , Pg.185 , Pg.187 , Pg.188 , Pg.197 , Pg.261 ]




SEARCH



Entrance slit

Monochromate

Monochromator

Monochromators

Monochromic

Slits

© 2024 chempedia.info