Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instrument Operation Modes

The objective in SIMS is to record secondary ion intensities over time (this is relayed as secondary ion current over time) and/or to record secondary ion intensities over space (this is relayed as the secondary ion current emanating from some specific location on the samples surface) as this allows for insight into the relative concentration and/or concentration gradients of the respective isotopes, elements, or molecules of interest on or within the solid of interest. [Pg.225]

Detection limit (the ability to detect the signal of interest if present at low concentrations) [Pg.225]

Mass resolution (the ability to separate signals in close proximity to each other in mass) [Pg.225]


According to the distance from probe to the sample, three operation modes can be classified for the AFM. The first and foremost mode of operation is referred to as contact mode or repulsive mode. The instrument lightly touches the sample with the tip at the end of the cantilever and the detected laser deflection measures the weak repulsion forces between the tip and the surface. Because the tip is in hard contact with the surface, the stiffness of the lever needs to be less than the effective spring constant holding atoms together, which is on the order of 1 — 10 nN/nm. Most contact mode levers have a spring constant of <1 N/m. The defection of the lever can be measured to within 0.02 nm, so for a typical lever force constant at 1 N/m, a force as low as 0.02 nN could be detected [50]. [Pg.20]

In this chapter, we present the principles of conventional Mossbauer spectrometers with radioactive isotopes as the light source Mossbauer experiments with synchrotron radiation are discussed in Chap. 9 including technical principles. Since complete spectrometers, suitable for virtually all the common isotopes, have been commercially available for many years, we refrain from presenting technical details like electronic circuits. We are concerned here with the functional components of a spectrometer, their interaction and synchronization, the different operation modes and proper tuning of the instrument. We discuss the properties of radioactive y-sources to understand the requirements of an efficient y-counting system, and finally we deal with sample preparation and the optimization of Mossbauer absorbers. For further reading on spectrometers and their technical details, we refer to the review articles [1-3]. [Pg.25]

Specificity is unsurpassed. Traditionally, MS was performed on very large and expensive high-resolution sector instruments operated by experienced specialists. The introduction of low-resolution (1 amu), low-cost, bench-top mass spectrometers in the early 1980s provided analysts with a robust analytical tool with a more universal range of application. Two types of bench-top mass spectrometers have predominated the quadrupole or mass-selective detector (MSD) and the ion-trap detector (ITD). These instruments do not have to be operated by specialists and can be utilized routinely by residue analysts after limited training. The MSD is normally operated in the SIM mode to increase detection sensitivity, whereas the ITD is more suited to operate in the full-scan mode, as little or no increase in sensitivity is gained by using SIM. Both MSDs and ITDs are widely used in many laboratories for pesticide residue analyses, and the preferred choice of instrument can only be made after assessment of the performance for a particular application. [Pg.740]

The cellular reaction product consisted of alternating plates of C03W lamellae in a solute-depleted Co matrix (eCo). Ffomogenized ingots were cut into slices 1 mm thick and then heat-treated. Spark erosion was used to trepan 3 mm diameter discs which were then jet electropolished to form thin foils for TEM examination in a Philips EM 430 STEM instrument operating at 300 kV in the nanoprobe mode with a probe size of 5 to 10 nm. [Pg.159]

A mass accuracy better than 1 ppm is routinely obtained by the better instruments when they are operated in a peak matching mode. An interlaboratory study of mass accuracy of different instruments and operating modes can be found in Reference 231. [Pg.48]

Figure 8.1 Schematic representation of NIR chemical imaging instrument operating in diffuse reflectance mode. Radiation from the illumination source interacts with the sample. Light reflected off of the sample is focused onto a NIR sensitive 2D detector after passing through a solid-state tunable wavelength selection filter. Figure 8.1 Schematic representation of NIR chemical imaging instrument operating in diffuse reflectance mode. Radiation from the illumination source interacts with the sample. Light reflected off of the sample is focused onto a NIR sensitive 2D detector after passing through a solid-state tunable wavelength selection filter.
Detailed Instrumentation and Separation Operation Modes 12.3.2.1 Cross-Flow FFF... [Pg.340]

Progress in the application of sensor arrays to gas analysis will be made through increasingly independent data channels using novel combinations of sensors and operating modes. Computational approaches will be modified to suit specific types of sensor arrays and to make economical use of computational space for portable instrument applications. The primary challenges of the near future will be to solve the "needle-in-the-haystack" problem and to proceed to complex mixture analysis using a plurality of sensor responses. [Pg.307]

This experimental design has the big advantage that it can be easily accomplished. The operation mode of the NMR instrument from routine NMR data acquisition to the LC-NMR mode can be easily changed by removing the routine probe from the room-temperature bore of the cryomagnet and inserting... [Pg.11]

Another measurement principle is the DSC, after Boersma [8]. In this case, no compensation heating is used and a temperature difference is allowed between sample crucible and reference crucible (Figure 4.5). This temperature difference is recorded and plotted as a function of time or temperature. The instrument must be calibrated in order to identify the relation between heat release rate and temperature difference. Usually this calibration is by using the melting enthalpy of reference substances. This allows both a temperature calibration and a calorimetric calibration. In fact, the DSC after Boersma works following the isoperibolic operating mode (see Section 4.2.2). Nevertheless, the sample size is so small (3 to 20 mg) that it is close to ideal flux. [Pg.91]

MAT 311A instrument operating at 70 eV [electron impact (El) mode] and reported as m/z and relative intensity (%). Field desorption (FD) mass measurements were carried out on a ZAB 2-SE-FDP instrument. Microwave-assisted synthesis was performed using a CEM-Discovery monomode microwave system utilizing an IR temperature sensor and magnetic stirrer in sealed 10 -mL glass vials with aluminum caps and a septum. All reactions were monitored and controlled using a personal computer. [Pg.114]

Problems related to sample introduction, high voltage isolation, or other typical of other gradient methods are avoided. Samples can be injected in both electro-kinetic and hydrodynamic mode without affecting other aspects of the instrument operation or causing sample waste. [Pg.71]


See other pages where Instrument Operation Modes is mentioned: [Pg.325]    [Pg.24]    [Pg.225]    [Pg.50]    [Pg.325]    [Pg.24]    [Pg.225]    [Pg.50]    [Pg.2032]    [Pg.59]    [Pg.144]    [Pg.33]    [Pg.680]    [Pg.69]    [Pg.444]    [Pg.1185]    [Pg.306]    [Pg.334]    [Pg.71]    [Pg.789]    [Pg.253]    [Pg.14]    [Pg.246]    [Pg.70]    [Pg.228]    [Pg.329]    [Pg.53]    [Pg.113]    [Pg.560]    [Pg.565]    [Pg.179]    [Pg.212]    [Pg.273]    [Pg.373]    [Pg.197]    [Pg.144]    [Pg.16]    [Pg.273]    [Pg.206]    [Pg.6]   


SEARCH



Operating mode

Operation mode

Operation modes mode

© 2024 chempedia.info