Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Impedance measurements reference electrode

Fig. 9.9 Effect of the low-frequency (LF) impedance of reference electrode, reflecting the sum of contact and charge-transfer resistances, on the error in polarization resistance of the working electrode measured using three-point LO-pin configuration [46]. The error, ctdculated... Fig. 9.9 Effect of the low-frequency (LF) impedance of reference electrode, reflecting the sum of contact and charge-transfer resistances, on the error in polarization resistance of the working electrode measured using three-point LO-pin configuration [46]. The error, ctdculated...
As Figure 8-10 illustrates, two voltage-measuring reference electrodes are connected through a high-impedance input device that in principle should reject all current flow through these electrodes. This input impedance consists of the capacitance with a typical value of 10 pF in parallel with... [Pg.180]

There are two procedures for doing this. The first makes use of a metal probe coated with an emitter such as polonium or Am (around 1 mCi) and placed above the surface. The resulting air ionization makes the gap between the probe and the liquid sufficiently conducting that the potential difference can be measured by means of a high-impedance dc voltmeter that serves as a null indicator in a standard potentiometer circuit. A submerged reference electrode may be a silver-silver chloride electrode. One generally compares the potential of the film-covered surface with that of the film-free one [83, 84]. [Pg.116]

By means of a resistance in the circuit the spontaneous corrosion reaction can be made to proceed at a predetermined rate, and the rate can be measured by means of an ammeter A. At the same time the potentials of the individual electrodes can be measured by means of a suitable reference electrode, a Luggin capillary and high-impedance voltmeters and Kj. At equilibrium there is no net transfer of charge (/ = A = 0). the e.m.f. of the cell is a maximum and equals the difference between the reversible potentials of the two electrodes... [Pg.85]

Logarithmic scale for expressing acidity or alkalinity of water (7.0 to 0 indicates increasing acidity 7.0 to 14 indicates increasing alkalinity). Measured by means of a glass electrode/reference electrode pair immersed in the water sample under test. The potential difference depends upon the pH which is then displayed on a pH meter (high input impedance, millivoltmeter). [Pg.749]

Metal/molten salt interfaces have been studied mainly by electrocapillary833-838 and differential capacitance839-841 methods. Sometimes the estance method has been used.842 Electrocapillary and impedance measurements in molten salts are complicated by nonideal polarizability of metals, as well as wetting of the glass capillary by liquid metals. The capacitance data for liquid and solid electrodes in contact with molten salt show a well-defined minimum in C,E curves and usually have a symmetrical parabolic form.8 10,839-841 Sometimes inflections or steps associated with adsorption processes arise, whose nature, however, is unclear.8,10 A minimum in the C,E curve lies at potentials close to the electrocapillary maximum, but some difference is observed, which is associated with errors in comparing reference electrode (usually Pb/2.5% PbCl2 + LiCl + KC1)840 potential values used in different studies.8,10 It should be noted that any comparison of experimental data in aqueous electrolytes and in molten salts is somewhat questionable. [Pg.147]

Nonfaradaic components associated with the uncompensated resistance between reference electrodes (7 ) and the double layer capacitance (Qi) can be accurately determined by AC impedance measurements. In this technique, a small AC potential perturbation is superimposed to the DC bias, and the resulting AC current is measured as a function of the frequency of modulation. The simplest circuit considered for a polarizable... [Pg.203]

FIGURE 10.1 A schematic diagram for a typical electrode system for potentiometric pH measurements. A potential is established on the pH sensitive membrane-solution interface of a pH electrode that responds to the activity or concentration of hydrogen ions in the solution. The reference electrode has a very stable half-cell potential. The cell potential, which is proportional to the pH in the test solution, is measured using a high input impedance voltmeter between the pH electrode and the reference electrode. [Pg.289]

The measured impedance, Zr, generally includes the electrolyte resistance between the working electrode and the reference electrode, Rx, and hence... [Pg.81]

Problems with reference electrode/electrolyte. The sample solution can react with the electrolyte. The greatest source of error in biological solutions is through the formation of insoluble silver sulfide, often at the ceramic frit. A blackened spot is usually observed in a pH electrode that has been in service for a few weeks. This precipitate can impede the free flow of electrolyte and cause the probe response to become sluggish and cause large errors in the measured pH. [Pg.422]

The impedance of the system described above will generally consist of in parallel with Cg, i.e. only the bulk electrolyte between the metal electrode and the reference electrode is likely to make a contribution to the impedance. Although Qi in parallel with R t is theoretically present at sufficiently low frequencies it is only measurable in special circumstances - for example when the metal cations have an unusually high mobility so that Rq is comparatively small, enabling R i to be observed. [Pg.283]

As an example. Figure 54a shows the zero-bias impedance of LSC electrodes on rare-earth-doped ceria in air at 750 °C measured using a symmetric cell incorporating a traditional reference electrode. Although the two screen-printed electrodes (1 and 2) were processed identically and aligned to an accuracy of 0.1 mm, the cell response is highly asymmetric... [Pg.597]

Since the resistance in the circuit containing the reference electrode is approximately 10,000 ohms, an accurate measure of the cathode-reference electrode potential can only be obtained by the use of a potential measuring device with an imput impedance of at least 100,000 ohms. Although a vacuum-tube voltmeter (VTVM, typical imput impedance 11 x 10 ohms) is suitable for this purpose, a common multimeter (VOM, typical imput impedance 20,000 ohms per volt) is not a satisfactory alternative. [Pg.16]

Data from electrochemical impedance diagrams yield a simplified quantitative analysis for an appropriate interpretation of the linear sweep voltammetry (LSV) experiments. In fact, the Si electrode potential measured with respect to the reference electrode represents the value within the bulk of the material. The direct current flow for the electrochemical reaction has to overcome the resistance of the space charge layer, which can reach extremely high values when a depletion layer is formed. For p-type Si in the potential range for the HER onset, this excess surface resistance is over 10 f2 cm. Thus, even with a bias of —1 V, the DC... [Pg.316]

The pH meter is a specialized voltmeter that measures the potential difference (in mV) between the sensing and reference electrode and converts it to a display of pH. To provide an accurate measurement of the voltage of an extremely high resistance electrode (108 Q) [5], this specialized voltmeter must be designed with high input resistance or impedance characteristics (100 times that of the electrode used). Since the measurement potential difference per pH change is very small (59.16 mV/pH unit at 25°C), a reliable amplifier in the pH meter is also essential. It should be sufficiently sensitive to detect changes of at least 0.05 pH unit (or 3 mV). [Pg.234]

Potentiometry is a method of obtaining chemical information by measuring the potential of an indicator electrode under zero current flow. It is based on the Nernst equation, which expresses the electrode potential as a function of the activity (or activities) of the chemical species in solution. The information obtained varies with indicator electrode, from the activity (concentration) of a chemical species to the redox potential in the solution. The potential of the indicator electrode is measured against a reference electrode using a high inptit-impedance mV/pH me-... [Pg.148]

In potentiometry, we measure the emf of a cell consisting of an indicator electrode and a reference electrode. For emf measurements, we generally use a pH/ mV meter of high input impedance. The potential of the reference electrode must be stable and reproducible. If there is a liquid junction between the indicator electrode and the reference electrode, we should take the liquid junction potential into account. [Pg.167]


See other pages where Impedance measurements reference electrode is mentioned: [Pg.387]    [Pg.1686]    [Pg.942]    [Pg.465]    [Pg.466]    [Pg.55]    [Pg.765]    [Pg.96]    [Pg.246]    [Pg.226]    [Pg.227]    [Pg.431]    [Pg.432]    [Pg.228]    [Pg.216]    [Pg.268]    [Pg.297]    [Pg.305]    [Pg.1211]    [Pg.125]    [Pg.669]    [Pg.13]    [Pg.233]    [Pg.274]    [Pg.597]    [Pg.598]    [Pg.63]    [Pg.164]    [Pg.166]    [Pg.172]    [Pg.13]    [Pg.27]    [Pg.160]    [Pg.161]   
See also in sourсe #XX -- [ Pg.109 ]




SEARCH



Electrode measurements

Electrodes impedance measurement

Impedance references

Measuring electrode

Reference electrode impedance

Reference electrodes

Reference measurements

© 2024 chempedia.info