Graphite furnace detection limit

Graphite furnace AA detection limits were determined using 50-/il sample volumes, a L vov platform, and full STPF conditions (Model 5100 PC with 5100 ZL Zeeman Furnace Module or Model 4110 ZL). SIMAA 6000 detection limits are similar in its multisource and are typically two to five times better in its single-source mode. Graphite furnace detection limits can be enhanced further by the use of replicate injections. [Pg.3371]

Hi) Methods Based on the Formation of Volatile Compounds. A specific method for the determination of fluoride is based on the formation of gaseous SiF4- The atomic absorption signal of silicon is measured at 251.6 nm using either a dinitrogen oxide-acetylene flame (detection limit 30/ug of F ) or a graphite furnace (detection limit 0.17 fig of F ). [Pg.143]

Atomic absorption spectroscopy is more suited to samples where the number of metals is small, because it is essentially a single-element technique. The conventional air—acetylene flame is used for most metals however, elements that form refractory compounds, eg, Al, Si, V, etc, require the hotter nitrous oxide—acetylene flame. The use of a graphite furnace provides detection limits much lower than either of the flames. A cold-vapor-generation technique combined with atomic absorption is considered the most suitable method for mercury analysis (34). [Pg.232]

Atomic absorption spectroscopy is an alternative to the colorimetric method. Arsine is stiU generated but is purged into a heated open-end tube furnace or an argon—hydrogen flame for atomi2ation of the arsenic and measurement. Arsenic can also be measured by direct sample injection into the graphite furnace. The detection limit with the air—acetylene flame is too high to be useful for most water analysis. [Pg.232]

Electrothermal vaporization can be used for 5-100 )iL sample solution volumes or for small amounts of some solids. A graphite furnace similar to those used for graphite-furnace atomic absorption spectrometry can be used to vaporize the sample. Other devices including boats, ribbons, rods, and filaments, also can be used. The chosen device is heated in a series of steps to temperatures as high as 3000 K to produce a dry vapor and an aerosol, which are transported into the center of the plasma. A transient signal is produced due to matrix and element-dependent volatilization, so the detection system must be capable of time resolution better than 0.25 s. Concentration detection limits are typically 1-2 orders of magnitude better than those obtained via nebulization. Mass detection limits are typically in the range of tens of pg to ng, with a precision of 10% to 15%. [Pg.638]

GC-AAS has found late acceptance because of the relatively low sensitivity of the flame graphite furnaces have also been proposed as detectors. The quartz tube atomiser (QTA) [186], in particular the version heated with a hydrogen-oxygen flame (QF), is particularly effective [187] and is used nowadays almost exclusively for GC-AAS. The major problem associated with coupling of GC with AAS is the limited volume of measurement solution that can be injected on to the column (about 100 xL). Virtually no GC-AAS applications have been reported. As for GC-plasma source techniques for element-selective detection, GC-ICP-MS and GC-MIP-AES dominate for organometallic analysis and are complementary to PDA, FTIR and MS analysis for structural elucidation of unknowns. Only a few industrial laboratories are active in this field for the purpose of polymer/additive analysis. GC-AES is generally the most helpful for the identification of additives on the basis of elemental detection, but applications are limited mainly to tin compounds as PVC stabilisers. [Pg.456]

Bishop [75] determined barium in seawater by direct injection Zeeman-modulated graphite furnace atomic absorption spectrometry. The V203/Si modifier added to undiluted seawater samples promotes injection, sample drying, graphite tube life, and the elimination of most seawater components in a slow char at 1150-1200 °C. Atomisation is at 2600 °C. Detection is at 553.6 nm and calibration is by peak area. Sensitivity is 0.8 absorbance s/ng (Mo = 5.6 pg 0.0044 absorbance s) at an internal argon flow of 60 ml/min. The detection limit is 2.5 pg barium in a 25 ml sample or 0.5 pg using a 135 ml sample. Precision is 1.2% and accuracy is 23% for natural seawater (5.6-28 xg/l). The method works well in organic-rich seawater matrices and sediment porewaters. [Pg.141]

Guevremont et al. [117] studied the use of various matrix modifiers in the graphite furnace gas method of determination of cadmium in seawater. These included citric acid, lactic acid, aspartic acid, histidine, and EDTA. The addition of less than 1 mg of any of the compounds to 1 ml seawater significantly decreased matrix interference. Citric acid achieved the highest sensitivity and reduction of interference, with a detection limit of 0.01 pg cadmium per litre. [Pg.148]

Pruszkowska et al. [135] described a simple and direct method for the determination of cadmium in coastal water utilizing a platform graphite furnace and Zeeman background correction. The furnace conditions are summarised in Table 5.1. These workers obtained a detection limit of 0.013 pg/1 in 12 pi samples, or about 0.16 pg cadmium in the coastal seawater sample. The characteristic integrated amount was 0.35 pg cadmium per 0.0044 A s. A matrix modifier containing di-ammonium hydrogen phosphate and nitric acid was used. Concentrations of cadmium in coastal seawater were calculated directly from a calibration curve. Standards contained sodium chloride and the same matrix modifier as the samples. No interference from the matrix was observed. [Pg.148]

The chemiluminescence technique has been used to determine trivalent chromium in seawater. Chang et al. [187] showed Luminol techniques for determination of chromium (III) were hampered by a salt interference, mainly due to magnesium ions. Elimination of this interference is achieved by seawater dilution and utilising bromide ion chemiluminescence signal enhancement (Fig. 5.7). The chemiluminescence results were comparable with those obtained by a graphite furnace flameless atomic absorption analysis for the total chromium present in samples. The detection limit is 3.3 x 10 9 mol/1 (0.2 ppb) for seawater with a salinity of 35%, with 0.5 M bromide enhancement. [Pg.159]

A Cis column loaded with sodium diethyldithiocarbamate has been used to extract copper and cadmium from seawater. Detection limits for analysis by graphite furnace atomic absorption spectrometry were 0.024 pg/1 and 0.004 xg/l, respectively [283]. [Pg.172]

In this method, inorganic lead in seawater samples are converted to tetra-ethylead using sodium tetraethylboron (NaB(C2H5)4) which is then trapped in a graphite furnace at 400 °C. Quantitation is achieved by using a simple calibration graph prepared from aqueous standards. An absolute detection limit of (3relative standard deviation. [Pg.189]

Andreae [564] coprecipitated tellurium (V) and tellurium (VI) from seawater and other natural waters with magnesium hydroxide. After dissolution of the precipitate with hydrochloric acid, the tellurium (IV) was reduced to tellurium hydride in 3 M hydrochloric acid. The hydride was trapped inside the graphite tube of a graphite furnace atomic absorption spectrometer, heated to 300 °C, and tellurium (IV) determined. Tellurium (VI) was reduced to tellurium (IV) by boiling with hydrochloric acid and total tellurium determined. Tellurium (VI) was then calculated. The limit of detection was 0.5 pmol per litre and precision 10-20%. [Pg.223]

Graphite furnace atomic absorption spectrometry has also been used to determine zinc [610,611] in seawater with a detection limit of 2 ig [611]. Guevre-mont [610] has discussed the use of organic matrix modifiers for the direct determination of zinc. [Pg.233]

Cadmium, copper, and silver have been determined by an ammonium pyrrolidine dithiocarbamate chelation, followed by a methyl isobutyl ketone extraction of the metal chelate from the aqueous phase [677], and finally followed by graphite furnace atomic absorption spectrometry. The detection limits of this technique for 1% absorption were 0.03 pmol/1 (copper), 2 nmol/1 (cadmium), and 2 nmol/1 (silver). [Pg.242]

High-performance liquid chromatography coupled with fluorescence detection [106, 107] or ion-exchange high-performance liquid chromatography with detection by graphite furnace atomic absorption spectroscopy [108] proved to be sensitive methods, but may lack from limitations in separation power and ease of identification of unknown products. [Pg.420]

Stephenson and Smith [84] used graphite furnace atomic absorption spectrometry to determine tributyltin in sediments. Recoveries from spiked samples ranged to from 72-111%. The detection limit was 2.5mg kgp1 of sample. [Pg.420]


See other pages where Graphite furnace detection limit is mentioned: [Pg.238]    [Pg.113]    [Pg.238]    [Pg.261]    [Pg.43]    [Pg.238]    [Pg.113]    [Pg.238]    [Pg.261]    [Pg.43]    [Pg.14]    [Pg.48]    [Pg.232]    [Pg.410]    [Pg.69]    [Pg.625]    [Pg.634]    [Pg.259]    [Pg.489]    [Pg.611]    [Pg.443]    [Pg.195]    [Pg.231]    [Pg.242]    [Pg.251]    [Pg.251]    [Pg.402]    [Pg.457]    [Pg.340]    [Pg.844]    [Pg.83]    [Pg.32]    [Pg.55]    [Pg.60]    [Pg.363]    [Pg.365]    [Pg.261]    [Pg.262]   