Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Gradient elution pressure

Schmitz et al. Polystyrene Packed column LiChrosorb Si 100 Butane-1, 4-dioxane UV Gradient elution, pressure program... [Pg.233]

Schmitz and Hilgers Vinyl naphthalenes Packed column Lichrosorb Si Pentane-dioxane UV Gradient elution pressure program... [Pg.235]

In the pneumatic pumping system, the pressure (and not the flow rate) is maintained constant as variations in chromatographic conditions occur. Thus, a change in mobile phase viscosity (e.g. gradient elution) or column back pressure will result in a change in flow rate for these types of pumps. The gas displacement pump in which a solvent is delivered to the column by gas pressure is an example of such a pneumatic pump. The gas displacement system is among the least expensive pumps available and is found in several low cost instruments. While the pump is nonpulsating and hence, produces low noise levels with the detectors in current use, its flow stability and reproducibility are only adequate. In addition, its upper pressure limit is only 2000 psi which may be too low in certain applications. [Pg.232]

Mixture-design, aobile phase optiaization (LC) 480 Mobile phase (GC) flow control 232 aodifiers 43 pressure control 232 purification 232 selection 4i velocity 44 viscosity 43 Mobile phase (LC) classification 460 degassing 553 gradient elution 485 ideal properties 458 aixed solvents 465 reservoirs 553... [Pg.514]

MacNair, J.E., Patel, K.D., and Jorgenson, J.W., Ultrahigh-pressure reversed-phase capillary liquid chromatography isocratic and gradient elution using columns packed with 1.0 pm particles, Anal. Chem., 71, 700, 1999. [Pg.70]

Bulk property detectors function by measuring some bulk physical property of the mobile phase, e.g., thermal conductivity or refractive index. As a bulk property is being measured, the detector responses are very susceptible to changes in the mobile phase composition or temperature these devices cannot be used for gradient elution in LC. They are also very sensitive to the operating conditions of the chromatograph (pressure, flow-rate) [31]. Detectors such as TCD, while approaching universality in detection, suffer from limited sensitivity and inability to characterise eluate species. [Pg.178]

Implementation of SFC has initially been hampered by instrumental problems, such as back-pressure regulation, need for syringe pumps, consistent flow-rates, pressure and density gradient control, modifier gradient elution, small volume injection (nL), poor reproducibility of injection, and miniaturised detection. These difficulties, which limited sensitivity, precision or reproducibility in industrial applications, were eventually overcome. Because instrumentation for SFC is quite complex and expensive, the technique is still not widely accepted. At the present time few SFC instrument manufacturers are active. Berger and Wilson [239] have described packed SFC instrumentation equipped with FID, UV/VIS and NPD, which can also be employed for open-tubular SFC in a pressure-control mode. Column technology has been largely borrowed from GC (for the open-tubular format) or from HPLC (for the packed format). Open-tubular coated capillaries (50-100 irn i.d.), packed capillaries (100-500 p,m i.d.), and packed columns (1 -4.6 mm i.d.) have been used for SFC (Table 4.27). [Pg.206]

LC-MS-MS was also the method of choice for the analysis of UV filters in solid matrices. Both LC and UPLC have been applied in three out of the four methods available for the determination of UV filters in sludge. Separation was performed on C8 and C18 LC-chromatographic columns (Zorbax, Eclipse, Vydac, and Purosphere) using binary gradient elution of mobile phases consisting of water/ methanol or water/acetonitrile. MS-MS detection was performed in SRM with ESI and atmospheric pressure photoionization (APPI) in both positive and negative modes. For each compound, two characteristics transitions were monitored. In addition to MS and MS-MS, diode array detection (DAD) was occasionally applied to the determination of OT. Spectra were recorded between 240 and 360 nm and discrete channels at 310 nm. [Pg.55]

The stationary phase is selected to provide the maximum selectivity. Where possible, the retention factor is adjusted (by varying the mobile phase composition, temperature, or pressure) to an optimum value that generally falls between 2 and 10. Resolution is adversely affected when k 2, while product dilution and separation time increase greatly when k 10. When this is not possible for all feed components and large differences exist among the -values of the different solutes, gradient elution should be considered. [Pg.48]

Where binary, ternary or quaternary gradient elution (p. 91) is required, a microprocessor controlled low-pressure gradient former is the most suitable (Figure 4.31(c)). The solvents from separate reservoirs are fed to a mixing chamber via a multiport valve, the operation of which is preprogrammed via the microprocessor, and the mixed solvent is then pumped to the column. For the best reproducibility of solvent gradients small volume pumps (< 100 gl) are essential. [Pg.121]

D. (3S, 4E)-Methyl 3-(dimethylphenylsilyl)-4-hexenoate (4a). A stirred solution of 6.0 g (29.11 mmol) of alcohol 3a and 60 mL of dry toluene in a 200-mL flask equipped with a reflux condenser is treated with 15 mL (116.89 mmol) of trimethyl orthoacetate and 0.15 mL (2.01 mmol, 0.07 equiv) of propionic acid (Note 15). The reaction mixture is heated under reflux for 48 hr and then allowed to cool to room temperature. Solvents and volatile material are removed under reduced pressure to leave the crude (E)-crotylsilane 4a as a yellow oil. The crude residue is chromatographed on silica gel (gradient elution 5% EtOAc/hexanes). The eluant (Note 6) affords 6.58 g (25.1 mmol, 86%) of pure 4a as a faint yellow, viscous oil (Notes 6, 16 and 17). [Pg.41]

A. ( )-1-(Dimethylphenylsilyl)-1-buten-3-ol (2a). A solution of 10.0 g (0.143 mol) of racemic 3-butyn-2-ol (Note 1) dissolved in 255 mL of tetrahydrofuran (THF, Note 2) in a 1-L, round-bottomed flask equipped with a reflux condenser and nitrogen atmosphere is prepared. Dimethylphenylsilane (21.4 g, 0.157 mol) (Note 3) and a small piece of sodium metal (ca. 5 mg) (Note 4) are placed in the reaction mixture. The solution is stirred for 15 min and 12 mg (2.05 x 10 5 mol) of bis(q-divinyltetramethyldisiloxane)tri-tert-butylphosphineplatinum(O) (Note 5) is added. The reaction mixture is then heated under reflux for 12 hr. The orange solution is cooled to ambient temperature, and the solvent is removed under reduced pressure to yield a crude orange residue containing 2a. The oil is subjected to column chromatography on silica gel (Note 6) (gradient elution 5, 10, 20, 35% EtOAc/hexanes) providing 25.4 g (123.23 mmol, 86%) of pure 2a as a yellow oil (Note 7). [Pg.187]


See other pages where Gradient elution pressure is mentioned: [Pg.597]    [Pg.360]    [Pg.110]    [Pg.110]    [Pg.21]    [Pg.24]    [Pg.163]    [Pg.222]    [Pg.77]    [Pg.184]    [Pg.290]    [Pg.302]    [Pg.566]    [Pg.790]    [Pg.795]    [Pg.799]    [Pg.901]    [Pg.206]    [Pg.208]    [Pg.210]    [Pg.175]    [Pg.120]    [Pg.151]    [Pg.374]    [Pg.64]    [Pg.576]    [Pg.146]    [Pg.115]    [Pg.116]    [Pg.117]    [Pg.573]    [Pg.574]    [Pg.12]    [Pg.13]   
See also in sourсe #XX -- [ Pg.312 ]




SEARCH



Gradient elution

Gradient elution high-pressure mixing

Gradient elution high-pressure systems

© 2024 chempedia.info