Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Filter sets

The simplest sound level meter consists of a microphone, an amplifier and a meter of some type. Sound level meters are graded according to British and international standards, and the most common type used for accurate measurement purposes it known as the Precision Grade or Type 1 meter. In practice, a basic sound level meter will incorporate weighting networks with either in-built octave filters or provision for connecting an external filter set (Figure 42.6). [Pg.654]

The position of these fixed filters should be set to assure capture of information that is needed. The filter settings are determined from the speed range (i.e., no-load to full-load) of the primary driver. In addition, the lower and upper limits of each filter should be adjusted by 10 to 15 per cent to allow for slight variations in speed. [Pg.715]

Expertise required to operate One of the objectives for using microprocessor-based predictive maintenance systems is to reduce the expertise required to acquire error-free, useful vibration and process data from a large population of machinery and systems within a plant. The system should not require user input to establish maximum amplitude, measurement bandwidths, filter settings, or allow free-form data input. All of these functions force the user to be a trained analyst and will increase both the cost and time required to routinely acquire data from plant equipment. Many of the microprocessors on the market provide easy, menu-driven measurement routes that lead the user through the process of acquiring accurate data. The ideal system should require a single key input to automatically acquire, analyze, alarm and store all pertinent data from plant equipment. This type of system would enable an unskilled user to quickly and accurately acquire all of the data required for predictive maintenance. [Pg.806]

Accuracy of data The microprocessor should be capable of automatically acquiring accurate, repeatable data from equipment included in the program. The elimination of user input on filter settings, bandwidths and other measurement parameters would greatly improve the accuracy of acquired data. The specific requirements that determine data accuracy will vary depending on the type of data. For example, a vibration instrument should be able to average... [Pg.806]

The main problem has been a methodological one. The patch clamp analysis of single channels views the world of channels through a very small analytical window [10]. A single channel event (opening) needs to be sufficiently long-lived and sufficiently large to be picked up within the current noise band under optimized conditions, and with the low-pass filter set to say 2 kHz. The open time needs to be close to a millisecond and the current amplitude close to 0.5 pA to permit detection. [Pg.277]

Fig. 41.7. Concentrations of the reactant A (reaction A B) as a function of time (dotted line) (ca = 1, cb = 0) state updates (after a new measurement), O state extrapolations to the next measurement (see Table 41.11 for Kalman filter settings). Fig. 41.7. Concentrations of the reactant A (reaction A B) as a function of time (dotted line) (ca = 1, cb = 0) state updates (after a new measurement), O state extrapolations to the next measurement (see Table 41.11 for Kalman filter settings).
Recently, SETA BioMedicals has developed a new near-infrared squaraine-based label Seta-633, which can be used to study the interaction between low-molecular-weight analytes and proteins using fluorescence lifetime as the readout parameter [19]. This label exhibits lower quantum yields and shorter fluorescence lifetimes when free in solution, but these values substantially increase upon interaction with proteins, which is contrary to tracers like Cy5 or Alexa 647. It was demonstrated in a model assay that a biotinylated Seta-633 binds to anti-biotin with high specificity. Importantly, the lifetime of Seta-633-biotin increases about 2.76 fold upon binding to a specific antibody (anti-biotin, MW =160 kDa), while the titration with BSA or nonspecific antibody does not result in a noticeable change in lifetime (Fig. 13). The label is compatible with readily available light sources (635 nm or 640 nm lasers) and filter sets (as for Cy5 or Alexa 647) and its... [Pg.95]

Note that G as derived here relates the FRET-induced sensitized emission in the S channel to the loss of donor emission in the D channel and that it is identical to the correction factor y/ [2] or G [6, 14]. Note however, that if the correction factors or 5 change, G and

correction factor C [3] is a constant that depends only on fluorophore properties and filter settings, and therefore it does not change with excitation intensity or detector gain. This is a clear advantage for confocal filterFRET. C (Eq. (7.14)) and G (Eq. (7.19)) are related as ... [Pg.321]

A final distinction is that on confocal microscopes S and A images are commonly acquired with the exact same emission filter settings whereas for CCD microscopes they typically involve physically separate- and therefore slightly different—filter cubes.6 This simplifies the calculation of leak-through terms [3], In Appendix of this chapter, we rather generalized the treatment of filterFRET by not making assumptions on the filter settings for S and A. [Pg.327]

Measure the absorbance of all wells using a microplate reader with a filter set at 410nm. [Pg.773]

The simulation discussed above was based on full color produced by individual red, green, and blue emitters. Other full-color reproduction approaches have been proposed for OLED displays including color from blue emitter by means of energy down conversion fluorescent filter [177], and color from white emitters by means of transmission color filter sets similar to that used in LCD industry [178,179]. Table 1.5 compares the EL efficiency of equivalent white... [Pg.32]

The standard fluorescence microscope can be equipped with filter sets for observation of the four types of fluorescent label that are in common use ... [Pg.136]

Filter set for visualization of AMCA, Alexa Fluor 350 or nuclear dye DAPI. [Pg.136]

Fluorophore absorbs ultraviolet light (or violet, blue or green) and emits light of longer wavelength. Fluorophores are used in immunohistochemistry for labeling primary or secondary antibodies in direct and indirect immunostain-ing methods, respectively. They can be visualized in fluorescence microscopy using special filter sets. [Pg.145]

Fig. 18.8 Similarity profile for filtered set of commercially available compounds. 5000 randomly selected compounds from the Available Chemicals Directory that pass the REOS filter were ranked according to their Tanimoto similarity scores (vertical axis) using Daylight fingerprints. 2886 compounds (58%) had similarity scores below 0.85. Fig. 18.8 Similarity profile for filtered set of commercially available compounds. 5000 randomly selected compounds from the Available Chemicals Directory that pass the REOS filter were ranked according to their Tanimoto similarity scores (vertical axis) using Daylight fingerprints. 2886 compounds (58%) had similarity scores below 0.85.
Typically, the distribution of each fluorochrome is viewed or photographed separately hy selection of the appropriate optical filter set. Alternatively, filter... [Pg.103]

Fluorescent microscope with standard fluorescein filter set. [Pg.144]

The spectral irradiance of a typical mercury xenon lamp is shown in Figure 45. There is a characteristic high intensity mercury emission at 313 nm and a low intensity emission at 334 nm The MUV projection printing tools isolate these two lines through insertion of band-pass filter sets into the optical path, such that little or no light to the blue of 300 nm or to the red of 350 nm is transmitted to the resist surface. The transmission of a filter set of this sort for the Perkin Elmer Micralign 500 is shown in Figure 46. [Pg.145]

A filter set is formed by two sheets of pure metal of thicknesses such that the absorption curves differ only in the range of the peak to be measured. Each element possesses a pronounced absorption maximum, the K corner, that is close to the K/J transition of that element (Fig. 13.10). For example, copper can be quantified by its Kq line using filters made out of l li and Co. [Pg.247]


See other pages where Filter sets is mentioned: [Pg.189]    [Pg.717]    [Pg.62]    [Pg.75]    [Pg.108]    [Pg.192]    [Pg.211]    [Pg.300]    [Pg.305]    [Pg.318]    [Pg.320]    [Pg.326]    [Pg.352]    [Pg.435]    [Pg.492]    [Pg.18]    [Pg.96]    [Pg.136]    [Pg.136]    [Pg.136]    [Pg.136]    [Pg.136]    [Pg.138]    [Pg.171]    [Pg.407]    [Pg.104]    [Pg.147]    [Pg.167]    [Pg.177]    [Pg.133]    [Pg.68]    [Pg.307]   


SEARCH



Band pass filter set

Filter Sets in Fluorescence Microscopy

Filter sets function

Filter sets parts

Filters settings

Filters settings

Long pass filter set

© 2024 chempedia.info