Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Ejection resolution

For low ion populations, a first estimate of achievable ejection resolution might be obtained from the cyclotron frequency spread that occurs over the range of cyclotron orbit radii through which the ion must pass to be ejected. This is based on the notion that an ejection waveform that is just adequate to eject one ion must have a frequency spectral peak that is at least as wide as the above spread of frequencies. Such a waveform would then excite, at least to some extent, all ions with frequencies falling within the width of the peak, thus limiting the ejection resolution. For ions with low z-mode amplitudes, we can use Dunbar s (46) approximate expression for the average radial field strength,... [Pg.52]

From the expression for Aueff, the ejection resolution becomes... [Pg.52]

Tool RF burst waveforms serve as a convenient means for determining the ejection resolution limits that are due to effects intrinsic in the operation of the cubic trap. More general computed waveforms (59) are obtained via inverse discrete Fourier transformation. These computed waveforms are just a linear superposition of a finite number of RF bursts. As a result, it is proposed that the best performance obtained with RF bursts anticipate the best performance obtained with computed waveforms,... [Pg.52]

The MS-1 resolution is determined through the use of ejection spectra. These spectra are produced in an experiment using two rf bursts, the first is the test ejection waveform and the second quantitates the number of unejected ions. The spectra are plots of ion signal intensity versus the frequency of the test waveform as the frequency is stepped through the resonance of the ion. At the ion resonance, the plot shows a dip having a width that indicates the ejection resolution of the test waveform. [Pg.53]

The first step in the measurement of the intrinsic ejection resolution limits of the trap is to find the minimum ejection waveform time - amplitude product for which the dip minimum shows complete loss of ion signal. This is done by varying the amplitude of the test waveform and fixing the time at -1 msec to assure the peak width in the frequency spectrum of the test waveform is wide enough to cover the range of cyclotron frequencies of the ejected ion. [Pg.53]

By using this technique, the ejection resolution measured for the example given in the "Model" section is certainly no better than 2.H K. This compares with a normal mass resolution of 30 K under the same trap conditions. [Pg.53]

Sputtered Neutral Mass Spectrometry (SNMS) is the mass spectrometric analysis of sputtered atoms ejected from a solid surface by energetic ion bombardment. The sputtered atoms are ionized for mass spectrometric analysis by a mechanism separate from the sputtering atomization. As such, SNMS is complementary to Secondary Ion Mass Spectrometry (SIMS), which is the mass spectrometric analysis of sputtered ions, as distinct from sputtered atoms. The forte of SNMS analysis, compared to SIMS, is the accurate measurement of concentration depth profiles through chemically complex thin-film structures, including interfaces, with excellent depth resolution and to trace concentration levels. Genetically both SALI and GDMS are specific examples of SNMS. In this article we concentrate on post ionization only by electron impact. [Pg.43]

The kinetics study [38] utilized a Fourier transform-ion cyclotron resonance (FT-ICR) mass spectrometer to measure the pathway branching ratios. The ability to eject selected masses and the extremely high mass resolution of this technique ensured that the observed CD3CH2 was in fact a primary product of the reaction. Temporal profiles from this reaction are shown in Fig. 1. Noticeably absent from the mass spectrum are the cations C2D2H3 and... [Pg.229]

The second and third energy terms in equation (1) could be interchanged without any effect (i.e. it is impossible to say which electron fills the initial core hole and which is ejected as an Auger electron they are indistinguishable. The existence of different electronic states within the final doubly ionized atom may furthermore lead to fine structure in high-resolution spectra. [Pg.172]

XPS or ESCA (electron spectroscopy for chemical analysis) is a surface sensitive technique that only probes the outer atomic layers of a sample. It is very useful tool to study polymer surfaces [91]. An XPS spectrum is created by focusing a monochromatic beam of soft (low-energy) X-rays onto a surface. The X-rays cause electrons (photoelectrons) with characteristic energies to be ejected from an electronic core level. XPS, which may have a lateral resolution of ca. 1-10 pm, probes about the top 50 A of a surface. [Pg.433]

Modem instrumentation has improved substantially in recent years, which has enabled the measurement of XPS spectra of superior resolution necessary to reveal the small BE shifts present in highly covalent compounds such as those studied here. In a laboratory-based photoelectron spectrometer, a radiation source generates photons that bombard the sample, ejecting photoelectrons from the surface that are transported within a vacuum chamber to a detector (Fig. 2). The vacuum chamber is required to minimize the loss of electrons by absorption in air and, if a very high quality vacuum environment is provided (as is the case with modem instruments), the surface contamination is minimized so that the properties of the bulk material are more readily determined. [Pg.95]

TOF analyzer it is critical for the mass resolution that the secondary ions are ejected at a precisely defined time. This means that the primary ion pulse should be as narrow in time as possible, preferably < 1 ns. At the same time maximum lateral resolution is desired. Unfortunately, there is a trade-off between these two parameters if the primary ion intensity is not to be sacrificed [122], Therefore, TOF-SIMS instruments have two modes of operation, high mass resolution and high lateral resolution. An advantage with the pulsed source is that an electron flood gun can be allowed to operate when the primary ion gun is inoperative. Thus, charge-compensation is effectively applied when analyzing insulating materials. [Pg.33]

The orbitrap is the most recently invented mass analyzer. Like with the QIT, ions are trapped and stored in a potential well. However, instead of ejecting the ions for external detection the frequency of the trapped oscillationg ions is measured. This method provides substantially better resolution and mass accuracy in normal operation. [Pg.55]

The ink-jet process relies on using a piezoelectric printhead that can create deformation on a closed cavity through the application of an electric field. This causes the fluid in the cavity to be ejected through the nozzle whose volume is determined by the applied voltage, nozzle diameter, and ink viscosity. The final width of the drop of the substrate is a result of the volume of fluid expelled and the thickness of the droplet on the surface. In addition, the drop placement is critical to the ultimate resolution of the display. Typical volumes expelled from a printhead are 10 to 40 pi, resulting in a subpixel width between 65 and 100 pm. Drop accuracies of +15 pm have been reported such that resolutions better than 130 ppi are achievable however, because the solvent to polymer ratio is so high, the drops must be contained during the evaporation process to obtain the desired resolution and film thickness. This containment can be a patterned photoresist layer that has been chemically modified so that the EL polymer ink does not stick to it. [Pg.574]

Ion trapping devices are sensitive to overload because of the detrimental effects of coulombic repulsion on ion trajectories. The maximum number of ions that can be stored in a QTT is about 10 -10, but it reduces to about 10 -10 if unit mass resolution in an RF scan is desired. Axial modulation, a sub-type of resonant ejection, allows to increase the number of ions stored in the QIT by one order of magnitude while maintaining unit mass resolution. [160,161] During the RF scan, the modulation voltage with a fixed amplitude and frequency is applied between the end caps. Its frequency is chosen slightly below V2 of the fundamental RF frequency, because for Pz < 1, e.g., = 0.98, we have z = (0 + 0.98/2) = 0.49 x... [Pg.160]

AES is similar to XPS in its function, but it has unparalleled high sensitivity and spatial resolution (of approximately 30-50 nm). Both AES and XPS involve the identification of elements by measurement of ejected electron energies. Fig. 2.12... [Pg.26]

Although not capable of the micrometer-sized lateral resolutions available with the aforementioned techniques, the surface spectroscopy, electron spectroscopy for chemical analysis (ESCA), also deserves mention. The ESCA experiment involves the use of X-rays rather than electrons to eject core electrons (photoelectrons), and it has comparable surface specificity and sensitivity to that of Auger electron spectroscopy (AES) (25, 26, 29). The principal advantage of ESCA relative to AES is that small... [Pg.140]


See other pages where Ejection resolution is mentioned: [Pg.51]    [Pg.52]    [Pg.52]    [Pg.51]    [Pg.52]    [Pg.52]    [Pg.1756]    [Pg.1842]    [Pg.378]    [Pg.269]    [Pg.333]    [Pg.73]    [Pg.280]    [Pg.280]    [Pg.293]    [Pg.308]    [Pg.475]    [Pg.32]    [Pg.66]    [Pg.275]    [Pg.435]    [Pg.469]    [Pg.143]    [Pg.382]    [Pg.96]    [Pg.31]    [Pg.300]    [Pg.161]    [Pg.79]    [Pg.27]    [Pg.28]    [Pg.140]    [Pg.258]    [Pg.187]    [Pg.56]    [Pg.313]   


SEARCH



Ejection

© 2024 chempedia.info