Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Constant-force mode

Images can be made in variable or constant force mode. In the latter case the difference signal from the photo detectors is used to adjust the distance between tip and surface, such that the force between the two, and thus the deflection of the cantilever, remains constant. An important advantage of working in constant force mode is that the overall orientation of the surface with respect to the z direction is not so critical, because the z piezo compensates for any inclination of the sample. [Pg.200]

Fig. 2.55 (a) Atomic force microscopy image (constant force mode) of islands at the surface of a PS PUMA diblock (M = 82 kg mol1) copolymer film (Maaloum et al. 1992). The height of the islands is 310A. (b) Section of one domain with a diameter of Afim. (c) Assumed structure at the domain edge. [Pg.110]

Images can be made in either variable or constant force mode. In the latter case, the difference signal from the photo detectors is used to compensate the dis-... [Pg.201]

Examples of some common forces that may exist between a surface and an AFM tip are Van der Waal, electrostatic, covalent bonding, capillary, and magnetic. In addition to providing information regarding the topography of the surface (constant force mode), forces may be applied to understand the morphology of a surface — for example, to determine the frictional force between the tip and surface, or the elasticity/hardness of a surface feature. For instance, see Tranchida, D. Piccarolo, S. Soliman, M. Macromolecules 2006,39,4547, and references therein. [Pg.430]

Bottino et al. [1994] used silicon nitride tips and silicon nitride gold-coated cantilevers to obtain the images in the height mode (or constant force mode). The AFM technique has been compared quite well with SEM using alpha-alumina membrane samples. The found that AFM could determine the sizes of individual constituent gamma-alumin.. particles in the membranes and detect subtle differences in their dimension ratios. Base. [Pg.100]

Fig. 3.40 Egg PC bilayer patches on glass imaged in constant force mode CM-AFM in Milli-Q water. The imaging conditions for the image on the right hand side changed during the scan, hence the imaging force was close the transition from the regime of electrostatic repulsion to had wall contact (compare Fig. 3.39) [89]... Fig. 3.40 Egg PC bilayer patches on glass imaged in constant force mode CM-AFM in Milli-Q water. The imaging conditions for the image on the right hand side changed during the scan, hence the imaging force was close the transition from the regime of electrostatic repulsion to had wall contact (compare Fig. 3.39) [89]...
The engagement procedure is carried out with minimized forces as described in Sect. 3.2.2. After engagement, the gains and the scan rate must be optimized. It is advisable to image the films in constant force mode with minimized forces in order... [Pg.168]

The scanner is moved up and down in z-direction during the scan, in order to keep the resonance frequency constant. Similarly to the constant force mode in contact mode a surface of constant force gradient is measured. The computer generates image as a function of the motion of the scanner. [Pg.367]

Constant Height Mode AFM Constant Force Mode AFM Constant Error Mode AFM Lateral Force Microscopy (LFM) Spreading Resistance Imaging... [Pg.358]

The cantilever position is held fixed in the AFM laser head while the sample is scanned imder the tip. See Fig. 6.7. The feedback system relies on an optical detection system that monitors the cantilever vertical deflection, dz, as the tip follows the contours of the sample surface. In the constant force mode of operation, as in the STM constant current case, dz is translated into a cor-... [Pg.140]

AFM images of surface morphology and particle size were obtained with FemtoScan 001 (Advanced Technologies Center, Moscow) operated in the constant force mode (1-10 nN) using nanoprobe cantilevers (spring constant 0.32 N/m) with oxide-sharpened Si3N4 integral tips. [Pg.382]

An automated Langmuir trough was used for measurements of surface pressure-area per molecule (n-A) isotherms and film deposition. The isotherms were recorded at a compression speed of 0.2-0.3 A /(moleculexmin). AFM images were obtained with a Nanoscope Ilia (Digital Instruments, USA) operated in the constant force mode (1-10 nN). [Pg.416]

Instrumentation. A cantilever with a sharp tip interacting with the surface under investigation is used. The actual bending of the cantilever is measured with a laser beam deflected from a mirror-like surface spot on the back of the cantilever towards a position-sensitive photodetector. The measured signal is used to control the piezo actuators. A constant force mode in which the cantilever-surface distance is kept at a preset interaction force and a constant height mode of scanning operation are possible. The principle of operation is schematically outlined in Fig. 7.9. [Pg.261]

Lateral foree microscopy (LEM) is a variant of the dc-contact mode, specifieally the constant-force mode, in which the laser-beam detector is arranged so as to allow monitoring not only of the vertical component of the tip deflection (topography), but also the... [Pg.381]

In C-AFM, the tip makes physical contact with the sample. As the tip is moved across the sample, the contact force causes the cantilever to bend according to changes in topography. In constant force mode, the tip is constantly adjusted to maintain a constant deflection, and therefore constant height above the surface. It is this adjustment that is displayed as data. However, the ability to track the surface in this manner is limited by the feedback circuit. Sometimes the tip is allowed to scan without this adjustment, and one measures only the deflection. This is useful for small, high-speed atomic resolution scans, and is known as variable-deflection mode. [Pg.33]


See other pages where Constant-force mode is mentioned: [Pg.279]    [Pg.455]    [Pg.201]    [Pg.84]    [Pg.74]    [Pg.273]    [Pg.186]    [Pg.202]    [Pg.203]    [Pg.447]    [Pg.216]    [Pg.447]    [Pg.35]    [Pg.105]    [Pg.133]    [Pg.195]    [Pg.559]    [Pg.658]    [Pg.171]    [Pg.262]    [Pg.378]    [Pg.379]    [Pg.34]    [Pg.2955]    [Pg.121]    [Pg.122]    [Pg.53]    [Pg.430]    [Pg.7447]    [Pg.7456]   
See also in sourсe #XX -- [ Pg.480 ]

See also in sourсe #XX -- [ Pg.329 ]




SEARCH



Constant modes

Force constant

© 2024 chempedia.info