Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Concentration determination calibration methods

All instrumental analytical methods except coulometry (Chapter 15) require calibration standards, which have known concentrations of the analyte present in them. These calibration standards are used to establish the relationship between the analytical signal being measured by the instrument and the concentration of the analyte. Once this relationship is established, unknown samples can be measured and the analyte concentrations determined. Analytical methods should require some sort of reference standard or check standard. This is also a standard of known composition with a known concentration of the analyte. This check standard is not one of the calibration standards and should be from a different lot of material than the calibration standards. It is run as a sample to confirm that the calibration is correct and to assess the accuracy and precision of the analysis. Reference standard materials are available from government and private sources in many countries. Government sources include the National Institute of Standards and Technology (NIST) in the US, the National Research Council of Canada (NRCC), and the Laboratory of the Government Chemist in the UK. [Pg.15]

It is known that Selenium catalyzes reaction of some dye reduction by Sulphide. On this basis spectrophotometric and test-techniques for Selenium determination are developed. Inefficient reproducibility and low sensitivity are their deficiencies. In the present work, solid-phase reagent on silica gel modified first with quaternary ammonium salt and then by Indigocarmine was proposed for Selenium(IV) test-determination. Optimal conditions for the Selenium determination by method of fixed concentration were found. The detection limit of Se(IV) is 10 ftg/L = 2 ng/sample). Calibration curve is linear in the range 50-400 ftg/L of Se(IV). The proposed method is successfully applied to the Selenium determination in multivitamins and bioadditions. [Pg.397]

The IR spectra are finally analyzed to determine the effluent concentration from each reactor channel. The quantification of species concentration is performed using either univariate or multivariate calibration methods. For non-overlapping peaks, like CO, C02, and N20, we can use univariate calibration. This is simply performed by baseline correction, the peak areas and/or peak heights and then converting these values... [Pg.329]

This method assumes that the response factor is constant over a range of concentrations and it is often more acceptable to determine the response factor for a range of test concentrations. In this method, a calibration curve is produced by incorporating a fixed amount of the internal standard in samples that contain known amounts of the test compound. For each concentration the ratio of peak heights is determined and plotted against concentration (Procedure 3.2). For quantitation of a test sample, the same amount of the internal standard is introduced in its usual way and the ratio of peak heights for the standard and unknown is used to determine the concentration of the unknown from the calibration curve. [Pg.112]

Minimum reportable concentration. The lower concentration limit for a method is usually measured by determining the detection limit. This is basically an instrument signal to noise ratio, and it does not include calibration effects. At low concentrations the calibration process often has a major adverse effect on precision. Detection limits are useful for comparing the inherent sensitivity of methods, but they are not realistic indexes of measurable concentrations in routine analysis. [Pg.126]

These iodometric calibration methods are based on the assumption that there is a stoichiometric reaction between ozone and the iodine in the various potassium iodide procedures. Three essentially independent methods have been used to test the accuracy of this assumption measuring the absorption of ultraviolet radiation at 254 nm by ozone in air, measuring the absorption of infrared radiation at 9,480 nm by ozone in air, and determining the ozone concentration in air by titration with nitric oxide. [Pg.253]

The simplest but also the least reliable calibration method is the use of a single standard solution. The electrode response is assumed to be Nernstian. The slope of the potential versus concentration dependence can also be determined experimentally, by using two standard solutions with different concentrations. To avoid large errors, the standard concentration should be as close as possible to the sample concentration in calibration with a single standard solution. [Pg.102]

The use of drying as a method of determining water content has been described in unit au. In the current discussion, indirect methods which depend upon the estimation of solute concentration will be considered. This estimation will utilize some form of calibration curve which correlates a measured property with the concentration. Many traditional methods exist, since water content—or solute (solids) content—is an important measure to describe the state of a system. It is a particularly important measure in commerce, where either the solute content or the water content may be an important indicator of quality or value. [Pg.29]

The RF should be determined using standards at various concentrations, and an average RF value should be used in the calculation. Both the external standard and internal standard calibration methods for GC analysis are fully discussed in Chapter 1.3. [Pg.164]

Air drawn through Palmes tube with three triethanolamine (TEA)-treated screens analyte converted into nitrite ion (NO, ) NO, treated with an aqueous solution of a reagent mixture containing sulfanilamide, H3P04, and V-1 -naphthy 1-ethylenediamine dihydrochloride color develops absorbance measured at 540 nm by a spectrophotometer, concentration determined from a standard calibration curve made from NaN02 (NIOSH Method 6700, 1984). [Pg.363]

For most spectroscopic applications, the goal of multivariate calibration is to predict the concentration of a given analyte(s) in a future (prospective) sample using only its measured spectrum and a previously determined model. To do this, the inverse calibration method is used in which equation (12.2) is rewritten as... [Pg.335]

The most common calibration method is to prepare standards of known concentrations, covering the concentration range expected in the sample. The matrix of the standard should be as close to the samples as possible. For instance, if the sample is to be extracted into a certain organic solvent, the standards should be prepared in the same solvent. The calibration curve is a plot of detector response as a function of concentration. A typical calibration curve is shown in Figure 1.3. It is used to determine the amount of analyte in the unknown samples. The calibration can be done in two ways, best illustrated by an example. Let us say that the amount of lead in soil is being measured. The analytical method includes sample preparation by acid extraction followed by analysis using atomic absorption (AA). The stan-... [Pg.5]

Of the methods for determining lignin in solution based on a specific chemical reaction, that involving nitrosation, the so-called Pearl-Benson method, has found the widest application. In this procedure, reaction of the phenolic units in lignin with acidified sodium nitrite leads to the formation of a nitrosophenol which, upon addition of alkali, is tautomerized to an intensely colored quinone mono-oxime. The absorbance of the latter structure is measured at 430 nm and related to lignin concentration by calibration with a standard lignin. The procedure described below is essentially that developed by Barnes et al. (1963), who modified the original Pearl-Benson method (Pearl and Benson 1940) to improve its sensitivity. [Pg.55]

Most of the static headspace methods determine the partition coefficient by quantifying volatile concentration above a sample by gas-chromatography. The vapour phase calibration method (VPC) uses an external vapour standard for calibration. One must assure that the pure component is completely vaporized before injection. A widely employed alternative is the Liquid calibration static headspace (LC-SH) method (YoiWey et al. 1991 Nedjma 1997). A third approach uses HS-SPME. SPME may be used to determine partition coefficients if short sampling times are applied the process must only sample the headspace and not disrupt the equilibrium (Jung and Ebeler 2003). This method has become very popular to study the effect of wine macromolecules on the liquid-vapor equilibrium, (Whiton and Zoecklein 2000 Escalona et al. 2002 Hartmann et al. 2002 Aronson and Ebeler 2004). [Pg.419]

An example of use of the above rules for calibration purposes is a procedure called the method of gradient ratios (gradient ratio calibration method, GRCM) [8], which is one way of implementing the dilution method in flow analysis. In this approach, sample and standard solutions are injected one after the other into a flow system. The obtained peaks overlap each other (as shown in Fig. 3.12) and are considered point-by-point from the maximum points along the softer sides of both peaks. On the basis of the values of the two measurement points obtained for standard and sample after the same time, analyte concentrations can be determined in various parts of the sample segment. For the final analytical result, the average of these concentrations is taken (only if their values differ from each other randomly). [Pg.41]

At typical flow rates, the concentration in the dialysate, Cout, is less than the actual concentration in the extracellular fluid, Cext (23). The ratio of Cout/Cext is defined as relative recovery, R, and must be considered for probe calibration and sampling optimization. In vitro, R is easily calculated because the dialysate and the extracellular fluid are homogenous therefore, probe calibration is easily obtained. However, in in vivo studies, calculation of R is difficult because of the active removal of neurotransmitters by uptake and tortuosity. Movement of analytes is impeded by tissue that surrounds the probe, and this movement cannot be easily accounted for with in vitro calibrations. Therefore, the most common method to determine concentrations in vivo is the zero-net flux method, in which known analyte concentrations are added to the perfusate (Cin), and then the analyte concentration is measured at the probe outlet (Cout)- The difference between analyte concentration at the inlet and outlet is used to establish the actual analyte concentration in the tissue, and the relative recovery rate can be calculated. This calibration method can be used to estimate basal levels of neurotransmitters. For example, the zero-net flux method has been used to determine that basal concentrations of dopamine are approximately 1-3.5 nM (24, 25). Although basal level concentrations... [Pg.1242]

For comparison with the flame simulations, the relative CH and CN concentration profiles along the centerline of the burner were recorded followed by a calibration of the relative concentration profiles using a N2 Rayleigh calibration method. Linear LIF was used to determine the CH and CN signal intensities as a function of height above the burner. In this case the relationship between the detected LIF intensity /lip and, e.g. the CH number density Nqh is given by... [Pg.220]

In many cases, however, if only the monitoring of changes in concentration of a given analyte is of interest, determination of the absolute concentration may not be necessary and simpler calibration methods can be applied. For example, when microdialysis is used for monitoring the concentration of a substance in blood, the dialysate concentration may be calibrated against blood concentrations, assuming that there is a linear relationship between the dialysate and blood concentration. This approach has been used for subcutaneous glucose sensors [64,65]. [Pg.229]

The external standard calibration method is a simple but less precise method and should only be used when the sample preparation is simple and small or no instrumental variations are observed. The method is not suitable for use with complicated matrices but is often used in pharmaceutical product analysis characterized by simple matrices and easy sample preparation. To construct a standard curve, standard solutions containing known concentrations of the analyte must be prepared and a fixed volume injected into the column. The resulting areas or heights of the peaks in the chromatogram are measured and plotted versus the amount injected. Unknown samples are then prepared, injected and analyzed in exactly the same manner, and their concentrations are determined from the calibration plot. The term external standard calibration implies that the standards are analyzed in chromatographic runs that are separate from those of the unknown sample. [Pg.26]


See other pages where Concentration determination calibration methods is mentioned: [Pg.235]    [Pg.605]    [Pg.417]    [Pg.339]    [Pg.215]    [Pg.224]    [Pg.74]    [Pg.134]    [Pg.309]    [Pg.296]    [Pg.294]    [Pg.103]    [Pg.342]    [Pg.359]    [Pg.429]    [Pg.236]    [Pg.477]    [Pg.482]    [Pg.39]    [Pg.308]    [Pg.455]    [Pg.562]    [Pg.567]    [Pg.214]    [Pg.83]    [Pg.84]    [Pg.300]    [Pg.341]    [Pg.41]    [Pg.135]   
See also in sourсe #XX -- [ Pg.115 ]




SEARCH



Calibration concentrations

Calibration determination

Concentration determine

Concentration methods

Concentration, determination

Determining concentration

Method calibration

© 2024 chempedia.info