Column packings monoliths

Correlation was found between domain size and attainable column efficiency. Column efficiency increases with the decrease in domain size, just like the efficiency of a particle-packed column is determined by particle size. Chromolith columns having ca. 2 pm through-pores and ca. 1pm skeletons show H= 10 (N= 10,000 for 10 cm column) at around optimum linear velocity of 1 mm/s, whereas a 15-cm column packed with 5 pm particles commonly shows 10,GOO-15,000 theoretical plates (7 = 10—15) (Ikegami et al., 2004). The pressure drop of a Chromolith column is typically half of the column packed with 5 pm particles. The performance of a Chromolith column was described to be similar to 7-15 pm particles in terms of pressure drop and to 3.5 1 pm particles in terms of column efficiency (Leinweber and Tallarek, 2003 Miyabe et al., 2003). Figure 7.4 shows the pressure drop and column efficiency of monolithic silica columns. A short column produces 500 (1cm column) to 2500 plates (5 cm) at high linear velocity of 10 mm/s. Small columns, especially capillary type, are sensitive to extra-column band... [Pg.156]

Current commercial silica-based columns have two important characteristics (1) they can produce efficiency similar to that of columns packed with 3.5 /tm particles and (2) they typically produce a pressure drop of half that caused by a column packed with 5 /tm particles.35 Monolithic columns have been shown to exhibit flat van Deemter curves, resulting in little loss of efficiency at high flow rates.36 As a result, high-throughput separations on conventional HPLC instruments can be achieved by increasing flow rate up to nine times (up to 9 ml/min) the usual rate in a conventional packed column. Cycle times for HPLC analysis as short as 1 min (injection-to-injection) have been reported by users of monolithic columns. Additional benefits of monolithic columns cited include... [Pg.257]

In more demanding separations that require higher plate counts, specially designed rapid analysis columns packed with very high efficiency 2 to 3 /.an porous particles are available from several manufacturers. In addition, monolithic columns with improved flow-through characteristics are also commercially available. Figure 13.4 depicts a comparison of inlet pressure and flow rate for 4.6 mm inner diameter x 50, 100, and 150 mm columns packed with 5 /an particles. [Pg.343]

Horvath et al. sintered the contents of a capillary column packed with 6 pm oc-tadecylsilica by heating to 360 °C in the presence of a sodium bicarbonate solution [101]. These conditions also strip the alkyl ligands from the silica support, thus significantly deteriorating the chromatographic properties. However, the performance was partly recovered after resilanization of the monolithic material with dimethyloctadecylchlorosilane allowing the separation of aromatic hydrocarbons and protected aminoacids with an efficiency of up to 160,000 plates/m. [Pg.28]

Recent chromatographic data indicate that the interactions between the hydrophobic surface of a molded poly(styrene-co-divinylbenzene) monolith and solutes such as alkylbenzenes do not differ from those observed with beads under similar chromatographic conditions [67]. The average retention increase, which reflects the contribution of one methylene group to the overall retention of a particular solute, has a value of 1.42. This value is close to that published in the literature for typical polystyrene-based beads [115]. However, the efficiency of the monolithic polymer column is only about 13,000 plates/m for the isocratic separation of three alkylbenzenes. This value is much lower than the efficiencies of typical columns packed with small beads. [Pg.108]

The effect that the quality of the bed structure has on the chromatographic properties of columns packed with particles has been well known for a long time [1]. Similarly, the efficiency of capillary electrophoretic separations reaches its maximum for a specific capillary diameter, and then decreases steeply for both larger and smaller size [ 117]. Therefore, any improvement in the efficiency of the polymeric monolithic columns for the isocratic separations of small molecules is likely to be achieved through the optimization of their porous structure rather than their chemistry. [Pg.108]

Excellent performance for the elution of another peptide, insulin (molecular weight 5800 g/mol), was also observed using silica-based monoliths. The efficiency of the monolithic column was much better than that of a column packed with beads, and did not change much even at high flow rates. [Pg.114]

Commercially available monolithic columns are based either on silica or organic polymer and are generally characterized as a polymeric skeleton with macropores, with a diameter of approximately 2 pm, and mesopores, with a diameter of approximately 13 nm. The role of the macropores (through-pores) is to provide channels with high compounds permeability, which permits the use of higher flow rates with respect to columns based on conventional particle size, and an extended surface area, which is comparable to conventional columns packed with 3 pm particles. [Pg.53]

Development of short monolithic silica columns with high efficiencies (up to 200.000 plates per meter) and high-speed separations, performance parameters that can be obtained with column packed with sub-2 pm particles, but achieved with monolithic silica columns at very lower system back-pressure. [Pg.53]

Unger K.K. Skudas R. Schulte M.M. Particle Packed Columns and Monolithic Columns in High Performance Liquid Chromatography-Comparison and Critical Appraisal. Journal of Chromatography A, 2008, 1184,393-415. [Pg.68]

CEC is generally performed in one of three types of columns, namely open-tubular (coated columns), packed columns, or continuous-bed (monolithic) columns. ... [Pg.451]

Due to the fact that the polymer can chemically be attached to the column wall during polymerization, monolithic stationary phases do not necessitate frits to retain the column packing. [Pg.16]

The absolute permeability (8 x 10 ° cm ) of currently available silica monoliths (Chromolith, Merck) is similar to that of columns packed with 9 jm particles [213], To increase efficiency, columns are usually packed with particles of 2, 3, and 5 jm, which possess absolute permeabilities of approximately 4x10 ", 9x10 ", and 2.5xl0" °cm, respectively. If we try to compare columns packed with 5 pm particles and monolithic columns with through-pores of an average size of 1.5 pm, we observe that the permeability of the monolithic columns is around three to four times larger than that of packed columns. [Pg.30]

The experimental of peptides or proteins depend more on the type of the column than on the gradient program and are constant within % at various gradient times. The variance in the (p is lower as the size of molecules increases. The values of (p are slightly higher with totally porous particles than with columns packed with superficially porous, nonporous particles and monolithic columns [97]. [Pg.135]


See other pages where Column packings monoliths is mentioned: [Pg.57]    [Pg.296]    [Pg.236]    [Pg.237]    [Pg.240]    [Pg.148]    [Pg.154]    [Pg.155]    [Pg.157]    [Pg.160]    [Pg.171]    [Pg.216]    [Pg.76]    [Pg.259]    [Pg.325]    [Pg.27]    [Pg.51]    [Pg.56]    [Pg.65]    [Pg.79]    [Pg.114]    [Pg.118]    [Pg.25]    [Pg.59]    [Pg.54]    [Pg.391]    [Pg.392]    [Pg.411]    [Pg.95]    [Pg.618]    [Pg.470]    [Pg.13]    [Pg.30]    [Pg.31]    [Pg.82]    [Pg.84]    [Pg.488]   